Abstract
In this chapter, we describe several configurations of digital holographic microscopes operating with partially coherent illumination, including the implementation of color holography. The different ways of implementing partially coherent illumination are described and discussed, as well as the respective improvements they provide depending on the properties of the observed objects. Several significant applications in the biomedical and environmental fields are presented. In particular, we carry out researches on correlative quantitative phase-contrast-fluorescence imaging on blood with a special focus on the behavior of platelets in relation to relevant pathologies. In the environmental domain, we demonstrate the use of high-throughput digital holographic microscopy to monitor and analyze plankton and microalgal cultures.
Keywords
- digital holographic microscopy
- partially coherent illumination
- coherent noise reduction
- quantitative phase-contrast imaging
- digital refocusing
- environmental applications
- biomedical applications
1. Introduction
Classical optical microscopy is highly limited by the small depth of focus due to the high magnification coupled with the high numerical apertures. With such a constraint, the analysis of a three-dimensional (3D) sample is complex and even impossible in the case of dynamic phenomena, which makes the extension of the depth of focus an important challenge. With the impressive recent developments of digital cameras, digital holography has provided an elegant and natural tool to counteract this limitation by enabling numerical refocusing at different depths within an experimental volume [1, 2]. In this context, digital holographic microscopy (DHM), which drastically increases the depth of focus (about 100×), became a very powerful tool to investigate numerous applications, in particular in the biomedical and environmental fields. Moreover, thanks to the wave nature of holography, DHM may provide the quantitative phase-contrast imaging (QPCI) that considerably extends the potential of this technology, especially in biological and environmental applications [3, 4, 5, 6]. Several ways exist to implement DHM. The conceptually simplest implementation consists of using an interferometric configuration: it allows to record the interference pattern between the object beam, resulting from the transmission through—or the refection by—the object, and a separated reference beam [2, 7, 8]. Hereinafter, those configurations are called two-channel interferometers. Several other ways to proceed may be considered. Firstly, common-path interferometers [5, 9] allow to simplify the configuration or to improve mechanical robustness. However, other limitations may arise from the common-path configurations as, in some cases, the apparition of halo effects in the phase images or the difficulties to implement an illumination of partial coherence. Other configurations were also developed that aim to reach the same goals as, for instance, the implementation of the transport of the intensity equation in a usual microscope [10], the in-line holography [11] and the ptychography [12]. Those configurations are interesting but request more complex computations to get the results, which can make reconstruction ability or the quantitative phase recovery more difficult. However, with two-paths interferometers, the full complex amplitude, required for the refocusing process and the QPCI, is obtained with relatively simple and straightforward computation. Regarding the mechanical robustness of the two-channel configuration, our experience has shown that it is not actually an issue. The design of the compact optical system with usual optomechanical components indeed allows to build systems that are able to operate even in the harsh environment [13, 14, 15].
The most usual way to implement holography in a two-channel configuration is to use a laser source; it is then easy to have mutually coherent object and reference beams, giving interference patterns on the camera. However, coherent beams are very sensitive to any disturbance in their propagation, which leads to noisy results. This noise that alters the images, which is related to the speckle effect, is multiplicative and hence difficult to remove by post-processing. Additionally, the speckle disturbance is created everywhere in the optical setup and experimental cells from any small defect. For this reason, we developed several DHM configurations taking advantage of partially coherent illuminations. The use of such illuminations drastically reduces the noise in both the amplitude modulus and phase images.
It has to be noted that a lot of researches were also performed to develop digital holography to get holographic information with completely incoherent light, and in particular from fluorescent objects. Pioneer articles in this field are given by the references [16, 17, 18].
In Section 2, we present the DHM configurations, developed with partially coherent illumination, and the processing to obtain the amplitude modulus and the phase from the recorded hologram. Section 2 also describes how the fluorescence can be implemented to achieve multimodal DHM (see details in Section 4.1) and the extension of the technology to color DHM with partially coherent illumination. The benefit of using partially coherent illumination to drastically reduce the speckle noise is demonstrated in Section 3. In Section 4, we describe several valuable applications we implemented.
2. DHM configurations with partially coherent sources
When implemented in a DHM configuration, it is requested that the partial spatial coherence of the beam illuminating the sample is uniform over the complete field of view. Therefore, the mutual coherence function
Such a mutual coherence function can be obtained from a spatially incoherent source or from a spatially highly coherent source according to the schemes in Figure 1 [19].
On the left configuration in Figure 1, a fully spatially incoherent beam is illuminating an aperture
where
It is also possible to obtain an equivalent partial spatial coherence illumination by the right configuration in Figure 1, in which a laser source is focused close to a rapidly moving ground glass placed at the front focal plane of a collimating lens. The ground glass generates a speckle field, whose width is adjusted by changing the position of the focused laser beam with respect to this ground glass and by the choice of the focal length of the collimating lens. It can be observed that, with time averaging, this source can be seen as a partially spatially coherent source, whose typical coherence width is equal to the average speckle size [20].
In the following, we describe two particular implementations of DHM with partially coherent illumination that we developed. The first one uses monochromatic illumination and can be coupled with a fluorescence module to achieve a multimodal instrument. It is shown in Figure 2a. A coherent beam emerging from a mono-mode laser diode is focalized by the lens ML1 close to the rough surface of the rotating ground glass (GG). The light transmitted by the GG is scattered and collimated by the lens L1. A first beam splitter BS1 divides the collimated beam into an object beam and a reference beam. The object beam passes through the sample and then through the microscope lens ML3. In a symmetric way, the reference beam passes through the microscope lens ML2, identical to ML3. Reference and object beams are then recombined by a second beam splitter (BS2) and transmitted by lens L2. Lenses ML3–L2 image one plane of the sample onto the camera sensor. The magnification Gs is then given by the ratio between the focal length of L2 and the focal length of ML3. Depending on the applications, the magnification is adjusted by changing the couple of lenses ML2 and ML3 (typically we used microscope lenses ×10, ×20, ×40, ×63, and ×100 (immersion oil)).
In most of the applications, the interferometer is adjusted to have a slanted reference beam with respect to the object beam in such a way that a thin fringe pattern is recorded on the camera sensor. The computation of the full complex amplitude from every recorded hologram, involving the amplitude modulus and the phase, is obtained by using the Fourier method [21]. In the resulting phases, it happens that unwanted phase modulations appear, which can be originated from the experimental cell or from small misalignments of the interferometer. To compensate them, powerful correction techniques can be applied and are given in more detail in Section 4.2 [22]. An example of a quantitative phase-contrast image obtained, of an alga
The fluorescence is implemented in the epifluorescence configuration, in such a way that the same sample is sequentially imaged by DHM and fluorescence modes on the same camera sensor. Applications of the combination of the DHM and fluorescence modes are given in Section 4.1.
In some applications, the color information is relevant. However, the configuration of Figure 2a is complex to adapt to several optical sources of different wavelengths. Indeed, in this case, the slant between the object and reference beams is identical for all the wavelengths, giving rise to fringe patterns with different inter-fringe spaces. This would make more difficult the processing with the Fourier method. In order to counter this limitation, we developed the color DHM instrument shown in Figure 3 [23].
This color DHM is also based on a Mach–Zehnder configuration in which we added two afocal systems, L2–L3 and L4–L5, in the object and reference channels, respectively, between the microscope lenses and the lens in front of the camera sensor. In the object channel, a neutral density filter is placed in the back focal plane of L2. In the reference channel, a Ronchi grating is placed in the back focal plane of L4 and is imaged on the camera sensor by L5–L6. In the back focal plane of L5, an optical stop is placed to keep only one of the first diffraction orders. In this way to proceed, the slant between the object and reference beams on the camera sensor is changing with the wavelength, in such a way that the same spatial frequency of the fringes is kept on the sensor. Therefore, this configuration is operating for any wavelength in the visible range without having to adjust the fringe spacing to implement the Fourier method. In order to operate in the red (R), green (G), and blue (B) wavelengths, the partially spatially coherent illumination is created by the combination of three laser beams into the system using a ground glass as described hereinabove. It has to be emphasized that the grating also enables the use of reduced temporal coherence sources, such as LEDs [23]. Indeed, if a reduced temporal coherence source is used in the configuration of Figure 2a, the slant between the object and reference beams makes the fringe contrast not uniform over the field of view. We can even expect to have no fringe at all in some areas of the field of view. With the configuration of Figure 3, however, due to the properties of the diffraction by a grating, this issue is solved. This configuration indeed provides a uniform fringe contrast with reduced temporal coherence illumination, accordingly, the optical paths between the object and reference beams are finely equalized. This equalization is adjusted, thanks to the rotating mount on which the mirror M2 and the beam splitter BS2 are placed.
An example where the color DHM can provide relevant information is the routine analysis of clinical cervical samples for the prevention and the checking of the development of cancerous cells. Analyses are usually performed on classical ThinPrep slides with the Papanicolaou stain. In this way, we performed the analysis of samples with the color DHM and recorded multispectral holograms (Figure 4a). From the recorded hologram, we extracted the intensity and the phase images in three wavelengths separately: 639 nm (R), 532 nm (G), and 488 nm (B). The combination of the RGB intensity images of stained slides gives color intensity images comparable to those obtained with classical microscopy used by histopathologists (Figure 4b). In addition to this RGB intensity image, DHM gives, simultaneously on the same sample, quantitative phase-contrast imaging. In Figure 4c, it can be seen that typical dysplastic cells (big nucleus and small cytoplasm) have the highest optical thickness.
One of the targets with the DHM is the refocusing of objects that have been recorded out of focus. For that purpose, the implementation in Fourier optics, of the convolutional form of the free space propagation is used [24]. The complex amplitude distribution
where
where
The phase
where
This holographic reconstruction does not indicate for which distance
3. Noise reduction by the use of partially coherent illumination
To study the illumination by a partially spatially coherent illumination, the left image of Figure 1 may be considered. In this configuration, the illumination is performed by the set of mutually incoherent point sources located within the aperture. As the locations of those points are different, each of them gives rise, after the collimation lens, to plane waves with different propagation angles. Therefore, each object is illuminated by a set of mutually incoherent plane waves having different propagation directions. For an unfocused object, it results that the propagation up to the input focus plane of lens ML3 (in Figure 2) gives rise to a set of mutually incoherent diffraction patterns that are specifically shifted, due to the different propagation directions. Those shifts increase with the defocus distance
where
It results in an increasing loss of resolution when the defocus distance increases. The way to manage it consists to adjust the partial coherence by adapting
The comparison of the results in Figures 5 and 6 shows that the reduction of the spatial coherence has a major impact on the reduction of the noise created by out-of focus disturbances. It is also to be noted that the reduction of the temporal coherence by replacing the lasers with LEDs (coherence length of 30 nm) gives, in this example, a weak reduction of the noise. This is due to the fact that the considered DHM configuration is in transmission and the noise-generating slide is relatively thin.
4. Applications
4.1 Multimodal DHM
The fluorescence mode in microscopy is of central importance in biology and biomedical applications, as specific fluorescent labeling allows us to tag specific structures and molecules in cells, tissues, and to perform cell analysis.
We were pioneers in the development of a multimodal DHM to combine the quantitative phase-contrast images provided by the recorded holograms and the fluorescence images of samples, labeled with fluorophores or that present an auto-fluorescence [19, 30, 31].
The quantitative phase-contrast imaging provides morphology and structural information on the sample. Fluorescence imaging provides information on molecules, thanks to the use of specific fluorophores. Therefore, the combination of both quantitative phase-contrast imaging and fluorescence imaging is of great interest, in particular in biomedical applications and cell biology. For that reason, this combination was afterward also performed by several authors [9, 32, 33].
One of the configurations we developed is presented in Figure 2a. In this setup, the fluorescence mode is obtained in reflection as in classical microscopy (epifluorescence mode), with the use of a classical fluorescence filter set selected for the fluorophore(s) to be studied. This filter set is inserted just below the microscope objective ML3 in the sample path (Figure 2a). A specific light source (LED) for microscope fluorescence mode is implemented for the excitation of the fluorophore(s) or the autofluorescence. The holograms and the fluorescence images are recorded with the same camera in a sequential mode. As the hologram is recorded in one shot, it allows to record the fluorescence signals with a very short delay.
The multimodal DHM working with partially coherent illumination allows to directly obtain the optical phase of the sample, the bright field image with reduced noise, and the fluorescence image. The combination of those images allows us to analyze the specimen in detail and brings precious information. An example obtained on a phytoplankton organism is presented in Figure 7.
Thanks to the overlay of the quantitative phase-contrast images and the fluorescent images, this DHM is a powerful tool to perform correlative quantitative phase-contrast-fluorescence microscopy. An example is presented in Figure 8, which shows the validation of the visualization of nucleoli inside the nucleus of HeLa cells in the phase image. The nucleoli are clearly visible on the phase image with their highest optical height. It is, therefore, possible to study the nucleoli thanks to the phase information, without fluorescence and in a noninvasive way.
Moreover, the recording of the quantitative phase information with low noise also enables the emulation of other classical microscopy modes. In particular, we emulated the differential interference contrast (DIC) mode, a classical tool to observe living cells [34].
4.2 In-flow analysis using DHM
The DHM configurations described in Section 2 are based on Mach–Zehnder off-axis interferometers. As explained above, this allows the one-shot acquisition of the hologram, that is, the extraction of the complex amplitude of the optical field,
We developed in-flow DHM in partially coherent illumination, with specific algorithms for processing the holograms, resulting in high-quality refocused images of the objects of interest, regarding both optical phase and intensity [4, 35]. We used it in many applications, mainly in the environmental field, as illustrated in Figure 9, and in the biomedical field, as shown in Figure 10. Monochromatic ([4], Figure 10) or color RGB DHM ([35], Figure 9) were used in the different studies. The illumination was partially coherent, either both spatially and temporally (Figure 9) or only spatially while keeping a high temporal coherence (Figure 10).
The DHM configurations are those described in Section 2 above and in [4, 35]. The microscope objectives are Leica, magnification 40×, and numerical aperture 0.6. A flow chamber is placed in the object arm of the interferometer (position SA in Figure 3). The objects of interest are biological cells: either algal cells in Figure 9, or red blood cells in Figure 10. The cells are present in suspension in the sample, which is pulled into the flow chamber in such a way that the cells appear in the field of view of the microscope. A sequence of holograms is recorded in order to analyze a large number of cells.
In interferometric DHM, the knowledge of the complex amplitude,
where
However, in the case of the in-flow imaging in DHM, the objects of interest are moving in the field of view, whereas the permanent defects and aberrations are static. As we showed in [4, 35], the blank image,
where
The blank,
The computation of the blank is efficient from about 100 averaged images. If the sequence experiment is very long, the blank
where 2
Dividing every image of the sequence by the computed blank provides the automatic correction of the defects and aberrations, which are permanent in the field of view. For the color in-flow DHM, the described method also automatically corrects the color balance in the images as illustrated in Figure 9. The method is very powerful, giving efficiently corrected complex amplitudes, resulting in high-quality images of both the optical phase and intensity, for every image of the sequence.
Once the images are corrected, every object of interest is automatically refocused as explained in Section 2. This allows to observe each object present in the image, even if it was recorded out of its focal plane. The whole depth of the flow chamber is thus analyzed.
4.2.1 An environmental application: the automatic analysis of plankton
We used the color in-flow DHM with LED illumination for the analysis of the microorganisms of plankton in pond and river water samples [35]. The algorithm described above was applied for the automatic correction of the permanent defects and aberrations, including the color balance, and for the automatic refocusing. Figure 9 illustrates the study and shows an example of observed microorganisms.
4.2.2 A biomedical application: the analysis of human red blood cells through a layer of endothelial cells
We also present here another interesting application of in-flow DHM with partially coherent illumination: the
4.3 Other blood applications
DHM is well suited for the analysis of blood cells, especially human blood cells [6, 37, 38]. Firstly, the quantitative phase analysis yields a powerful measurement tool for the accurate characterization of blood cells. The phase is indeed related to the optical thickness of the cells. Secondly, the holographic reconstruction in depth allows the extension of the depth of investigation, in comparison with the usual depth of field in classical optical microscopy. This allows to study the motion of blood cells in thick samples, of the same order of magnitude as real human blood vessels. Thirdly, the knowledge of both the quantitative phase and intensity makes easier the automatic processing of a large number of images. Therefore, statistically significant samples can be efficiently studied.
In addition to the in-flow analysis described in 4.2.2, we present here two other applications we performed, illustrating the potential of DHM in blood cell analysis.
4.3.1 Distribution of size and shape of red blood cells within one individual
All the red blood cells within one individual are not identical. Their size and shape are distributed over some range. Studying this distribution is important in biomedical research since it impacts platelet diffusion in blood flows. We used DHM for characterizing this distribution.
An example is shown in Figure 11, for one healthy individual (male, 27 years old). Microscope lenses Olympus, magnification 63
where
with
with
The normalized correlation product ranges from 0 to 1, exhibiting local maxima, close to 1, when a red blood cell is detected. Therefore, a fixed threshold applied to the image of the normalized correlation product allows to identify the position of each red blood cell. The segmentation is operated by applying a threshold to the phase image. Four parameters are then measured in each detected red blood cell: the mean radius, the minimal phase in the central part, the mean radius of the crest, and the mean phase value on the crest. From these four parameters, the parameters
with
4.3.2 Statistical analysis of spread platelets from healthy and dialyzed donors
We also used DHM for an accurate analysis of the spreading of platelets, while comparing healthy volunteers and dialyzed patients [6]. Results are illustrated in Figure 12. The spreading occurred in glass slides surmounted by wells. For each of the 14 individuals of the study, seven healthy and seven dialyzed, six wells were used: two coated with laminin, two coated with collagen IV, and two without any coating, denoted as control. Platelet-rich plasma from each individual was poured into each well. Subsequent sedimentation allows the platelets to spread onto the slide, before washing and fixation. Using this experimental procedure, with the sedimentation step for the spreading of platelets, enhances the reproducibility. Each well was then analyzed using a color DHM working in partially spatially coherent illumination, made by three laser sources followed by a rotating ground glass. Forty holograms were recorded for each well. The processing of holograms, including the compensation of aberrations, the refocusing, the detection, and the segmentation of platelets, was performed as described in 4.2 and 4.3.1. For every detected platelet, several parameters were measured, whose surface area
Whereas the surface area
5. Conclusions
In this chapter, we have described the advantages of digital holographic microscopy (DHM) working with partially coherent illumination. This illumination drastically reduces the speckle noise that alters the images in both the amplitude modulus and the phase images. This opens the investigation of numerous applications with instruments that are robust and easy to use. The instrumentation is accompanied by software allowing the hologram processing in an automated way.
We have depicted several DHM configurations that we developed as a color DHM and a multimodal fluorescence DHM. We have presented different valuable applications that we implemented and that show the high-quality image obtained, which demonstrate the high capabilities of the technology in the biomedical and the environmental fields.
Acknowledgments
The research was supported by the Walloon Region in the frame of the Decisiv and the HoloCancer projects and L’Institut d’encouragement de la Recherche scientifique et de l’Innovation de la Région de Bruxelles-Capitale (IRSIB-INNOVIRIS) in the frame of the Holoflow Impulse project.
Dr. Jérôme Dohet-Eraly was Aspirant and then Chargé de Recherches du Fonds de la Recherche Scientifique – FNRS (F.R.S.-FNRS, Belgium).
References
- 1.
Schnars U, Jüptner WPO. Digital recording and numerical reconstruction of holograms. Measurement Science and Technology. 2002; 13 :R85-R101. DOI: 10.1088/0957-0233/13/9/201 - 2.
Dubois F, Joannes L, Legros JC. Improved three-dimensional imaging with a digital holography microscope with a source of partial spatial coherence. Applied Optics. 1999; 38 :7085-7094. DOI: 10.1364/AO.38.007085 - 3.
Marquet P, Rappaz B, Magistretti PJ, Cuche E, Emery Y, Colomb T, et al. Digital holographic microscopy: a noninvasive contrast imaging technique allowing quantitative visualization of living cells with subwavelength axial accuracy. Optics Letters. 2005; 30 :468-470. DOI: 10.1364/OL.30.000468 - 4.
Yourassowsky C, Dubois F. High throughput holographic imaging-in-flow for the analysis of a wide plankton size range. Optics Express. 2014; 22 :6661-6673. DOI: 10.1364/OE.22.006661 - 5.
O’Connor T, Shen J-B, Liang BT, Javidi B. Digital holographic deep learning of red blood cells for field-portable, rapid COVID-19 screening. Optics Letters. 2021; 46 :2344-2348. DOI: 10.1364/OL.426152 - 6.
Dohet-Eraly J, Zouaoui Boudjeltia K, Rousseau A, Queeckers P, Lelubre C, Desmet JM, et al. Three-dimensional analysis of blood platelet spreading using digital holographic microscopy: a statistical study of the differential effect of coatings in healthy volunteers and dialyzed patients. Biomedical Optics Express. 2022; 13 :502-513. DOI: 10.1364/BOE.448817 - 7.
Zhang T, Yamaguchi I. Three-dimensional microscopy with phase-shifting digital holography. Optics Letters. 1998; 23 :1221-1223. DOI: 10.1364/OL.23.001221 - 8.
Cuche E, Bevilacqua F, Depeursinge C. Digital holography for quantitative phase-contrast imaging. Optics Letters. 1999; 24 :291-293. DOI: 10.1364/OL.24.000291 - 9.
Park Y, Popescu G, Badizadegan K, Dasari RR, Feld MS. Diffraction phase and fluorescence microscopy. Optics Express. 2006; 14 :8263-8268. DOI: 10.1364/OE.14.008263 - 10.
Barty A, Nugent KA, Paganin D, Roberts A. Quantitative optical phase microscopy. Optics Letters. 1998; 23 :817-819. DOI: 10.1364/OL.23.000817 - 11.
Luo W, Zhang Y, Feizi A, Göröcs Z, Ozcan A. Pixel super-resolution using wavelength scanning scanning. Light: Science and Applications. 2016; 5 :e16060. DOI: 10.1038/lsa.2016.60 - 12.
Tian L, Liu Z, Yeh L-H, Chen M, Zhong J, Waller L. Computational illumination for high-speed in vitro Fourier ptychographic microscopy. Optica. 2015;2 :904-911. DOI: 10.1364/OPTICA.2.000904 - 13.
Podgorski T, Callens N, Minetti C, Coupier G, Dubois F, Misbah C. Dynamics of vesicle suspensions in shear flow between walls. Microgravity Science and Technology. 2011; 23 :263-270. DOI: 10.1007/s12217-010-9212-y - 14.
Callens N, Minetti C, Coupier G, Mader M-A, Dubois F, Misbah C, et al. Hydrodynamic lift of vesicles under shear flow in microgravity. Europhysics Letters. 2008; 83 :24002. DOI: 10.1209/0295-5075/83/24002 - 15.
Zegers I, Carotenuto L, Evrard C, Garcia-Ruiz J, De Gieter P, Gonzales-Ramires L, et al. Counterdiffusion protein crystallisation in microgravity and its observation with PromISS (protein microscope for the international space station). Microgravity Science and Technology. 2006; 18 :165-169. DOI: 10.1007/BF02870402 - 16.
Schilling BW, Poon T-C, Indebetouw G, Storrie B, Shinoda K, Suzuki Y, et al. Three-dimensional holographic fluorescence microscopy. Optics Letters. 1997; 22 :1506-1508. DOI: 10.1364/OL.22.001506 - 17.
Rosen J, Brooker G. Digital spatially incoherent fresnel holography. Optics Letters. 2007; 32 :912-914. DOI: 10.1364/OL.32.000912 - 18.
Rosen J, Brooker G. Non-scanning motionless fluorescence three-dimensional holographic microscopy. Nature Photonics. 2008; 2 :190-195. DOI: 10.1038/nphoton.2007.300 - 19.
Dubois F, Yourassowsky C. Method and device for obtaining a sample with three-dimensional microscopy. Patents. EP 01870147 (2001), EP1399730 B1, US 7009700 B2 (2006) - 20.
Dubois F, Requena ML, Minetti C, Monnom O, Istasse E. Partial spatial coherence effects in digital holographic microscopy with a laser source. Applied Optics. 2004; 43 :1131-1139. DOI: 10.1364/AO.43.001131 - 21.
Kreis T. Digital holographic interference-phase measurement using the Fourier-transform method. Journal of the Optical Society of America. A. 1986; 3 :847-855. DOI: 10.1364/JOSAA.3.000847 - 22.
Zouaoui Boudjeltia K, Ribeiro de Sousa D, Uzureau P, Yourassowsky C, Perez-Morga D, Courbebaisse G, et al. Quantitative analysis of platelets aggregates in 3D by digital holographic microscopy. Biomedical Optics Express. 2015; 6 :3556-3563. DOI: 10.1364/BOE.6.003556 - 23.
Dubois F, Yourassowsky C. Full off-axis red-green-blue digital holographic microscope with LED illumination. Optics Letters. 2012; 37 :2190-2192. DOI: 10.1364/OL.37.002190 - 24.
Nazarathy M, Shamir J. Fourier optics described by operator algebra. Journal of the Optical Society of America. 1980; 70 :150-159. DOI: 10.1364/JOSA.70.000150 - 25.
Dubois F, Schockaert C, Callens N, Yourassowsky C. Focus plane detection criteria in digital holography microscopy by amplitude analysis. Optics Express. 2006; 14 :5895-5908. DOI: 10.1364/OE.14.005895 - 26.
Dohet-Eraly J, Yourassowsky C, Dubois F. Refocusing based on amplitude analysis in color digital holographic microscopy. Optics Letters. 2014; 39 :1109-1112. DOI: 10.1364/OL.39.001109 - 27.
Dubois F, El Mallahi A, Dohet-Eraly J, Yourassowsky C. Refocus criterion for both phase and amplitude objects in digital holographic microscopy. Optics Letters. 2014; 39 :4286-4289. DOI: 10.1364/OL.39.004286 - 28.
Dohet-Eraly J, Yourassowsky C, Dubois F. Fast numerical autofocus of multispectral complex fields in digital holographic microscopy with a criterion based on the phase in the Fourier domain. Optics Letters. 2016; 41 :4071-4074. DOI: 10.1364/OL.41.004071 - 29.
Dohet-Eraly J, Yourassowsky C, El Mallahi A, Dubois F. Quantitative assessment of noise reduction with partial spatial coherence illumination in digital holographic microscopy. Optics Letters. 2016; 41 :111-114. DOI: 10.1364/OL.41.000111 - 30.
Dubois F, Yourassowsky C, and Monnom O. Microscopie en holographie digitale avec une source partiellement cohérente. In: Faupel M, Smigielski P, Grzymala R, editors. Imagerie et Photonique pour les Sciences du Vivant et de la Médecine. Fontis Media; 2004.p. 287-302 - 31.
Dubois F, Yourassowsky C, Callens N, Minetti C, Queeckers P, Podgorski T, et al. Digital holographic microscopy working with a partially spatial coherent source. In: Ferraro P, Wax A, Zalevsky Z, editors. Coherent Light Microscopy, Imaging and Quantitative Phase Analysis. Berlin: Springer; 2011. p. 31-59 - 32.
Quan X, Nitta K, Matoba O, Xia P, Awatsuji Y. Phase and fluorescence imaging by combination of digital holographic microscopy and fluorescence microscopy. Optical Review. 2015; 22 :349-353. DOI: 10.1007/s10043-015-0042-4 - 33.
Nygate YN, Singh G, Barnea I, Shaked NT. Simultaneous off-axis multiplexed holography and regular fluorescence microscopy of biological cells. Optics Letters. 2018; 43 :2587-2590. DOI: 10.1364/OL.43.002587 - 34.
Dubois F, Yourassowsky C, Monnom O, Legros JC, Debeir O, Van Ham P, et al. Digital holographic microscopy for the three-dimensional dynamic analysis of in vitro cancer cell migration. Journal of Biomedical Optics. 2006; 11 :054032. DOI: 10.1117/1.2357174 - 35.
Dohet-Eraly J, Yourassowsky C, Dubois F. Color imaging-in-flow by digital holographic microscopy with permanent defect and aberration corrections. Optics Letters. 2014; 39 :6070-6073. DOI: 10.1364/OL.39.006070 - 36.
Colomb T, Cuche E, Charrière F, Kühn J, Aspert N, Montfort F, et al. Automatic procedure for aberration compensation in digital holographic microscopy and applications to specimen shape compensation. Applied Optics. 2006; 45 :851-863. DOI: 10.1364/AO.45.000851 - 37.
O’Connor T, Santaniello S, Javidi B. COVID-19 detection from red blood cells using highly comparative time-series analysis (HCTSA) in digital holographic microscopy. Optics Express. 2022; 30 :1723-1736. DOI: 10.1364/OE.442321 - 38.
Bianco V, Merola F, Miccio L, Memmolo P, Gennari O, Paturzo M, et al. Imaging adherent cells in the microfluidic channel hidden by flowing RBCs as occluding objects by a holographic method. Lab on a Chip. 2014; 14 :2499-2504. DOI: 10.1039/c4lc00290c