Rapid Qualitative and Quantitative HPLC/MS Analysis of an Antioxidant Couple Consisted of Glutathione and Ascorbic Acid in a Pharmaceutical Product

Written By

Stanislav V. Yefimov

Submitted: 06 December 2021 Reviewed: 01 February 2022 Published: 02 November 2022

DOI: 10.5772/intechopen.102975

From the Edited Volume

Medicinal Plants

Edited by Sanjeet Kumar

Chapter metrics overview

65 Chapter Downloads

View Full Metrics

Abstract

Vitamin C and glutathione are mostly found together in pharmaceutical products. These two components protect each other from oxidation by forming an antioxidant couple and mutually reinforcing each other’s actions. This paper describes a method for determining the activity of the antioxidant couple in pharmaceutical products using HPLC/MS. An Agilent 6125 HPLC equipped with MSD and DAD detectors was used. The first detector is for the detection of glutathione, and the second is for the detection of ascorbic acid. The MS spectrum of glutathione (Glut) was dominated by the signal m / z + = 308.2, which corresponds to the Glut-H + cation. The MS spectrum of ascorbic acid (AA) was dominated by signals m/z + = 177 and 375, which corresponds to the cations AK-H + and 2AK-Na +. The use of HPLC with two detectors (MS and DAD) made it possible to simultaneously determine both components of the antioxidant couple in pharmaceutical products without derivatization and any preliminary sample preparation. The method meets the FDA criteria for accuracy, selectivity, robustness, and reproducibility, and has a low detection limit of both components of the antioxidant couple.

Keywords

  • antioxidant couple
  • glutathione
  • ascorbic acid; potency
  • HPLC/MS; Validation

1. Introduction

In living cells, glutathione protects cells from oxidative damage and maintains redox balance [1]. Vitamin C (ascorbic acid) is one of the powerful reducing agents, it works as a neutralizer of oxidizing free radicals in a living cell [2]. Glutathione and vitamin C often coexist in pharmaceutical products. Both components enhance each other’s actions and protect each other from oxidation. This combination is called the antioxidant couple [3].

Determination of the activity of an antioxidant couple in pharmaceutical products is carried out using modern instruments and specially developed methods. In this work, we used HPLC/MS from Agilent [4].

When developing methods of analysis, it is customary to follow the recommendations of the International Conference on Harmonization (ICH), which contains the necessary criteria for validating methods [5]. It is also important for developers to achieve simplicity (fewer stages) and speed of analysis. Both factors affect its cost.

The complexity of the simultaneous analysis of glutathione and ascorbic acid is that Glutathione (Figure 1) is relatively poorly visible to UV-VIS detectors. To make it highly visible, chemists are forced to derivatize glutathione by adding a radical containing a benzoic or furan ring through the sulfhydryl group.

Figure 1.

Structures of the molecules forming an antioxidant couple.

Ellman’s reagent (5,5'-dithiobis (2-nitrobenzoic acid) [6, 7], O-phthalic aldehyde [8], 4-vinyl pyridine, 4-fluoro-7-sulfobenzofurazan ammonium salt, 4- (aminosulfonyl)-7-fluorobenzofurazan and N- (9-acridinyl) maleimide [9] are used as derivatizing agent. However, the glutathione molecule contains two carboxyl groups and can be ionized at (ESI) and detected by MSD [10, 11], or by HPLC/ MS/MS [12] without derivatization.

The ascorbic acid molecule (Figure 1) contains a furan ring and is well detected by the UV-Vis detector. There are some literature data about methods for the determination of ascorbic acid and ascorbates by HPLC in fruits, blood serum, and pharmaceuticals [13, 14, 15, 16, 17].

Usually, the UV detector is used [13, 18, 19, 20], some authors prefer an electrochemical detector [16]. The tandem of MS and UV-VIS detectors is a useful and effective tool [14, 17, 21, 22].

As a mobile phase for HPLC/MS, methanol and acetonitrile volatile solutions are most popular [14, 17, 21, 22]. The not volatile sodium dihydrogen orthophosphate solutions may be used as a mobile phase for HPLC-UV [13, 16, 17, 18, 19].

To simultaneously analyze several components in a pharmaceutical product, chemists select the appropriate detector and suitable chromatography conditions. In the case when the components differ in nature and their concentration is very different, a good result has given the use of two detectors, each of which is focused on the detection of its group of components. In the present study, we used two detectors, one for glutathione and the other for ascorbic acid. This approach allowed us to obtain a satisfactory result of the analysis with a tenfold difference in the concentration of the components and to exclude derivatization.

The antioxidant couple analysis method presented in this article was developed for the analysis of nasal spray and injection solution. The specificity, precision, reproducibility, reliability, and selectivity of the method have been validated. The recommendations of the European Medicines Agency [5], as well as the FDA guidelines for the validation of analytical procedures [23, 24, 25], were used.

Advertisement

2. Materials and methods

Chemicals. L-ascorbic acid analytical standard from Sigma-Aldrich; formic acid 98-100% analytical grade from Merck; glutathione from European Pharmacopoeia Reference Standard, water HPLC grade purchased from Agilent. All the solvents used were of HPLC grade. Branded pharmaceutical formulation, in form of nasal spray and injection solution, was obtained from commercial sources and used as received, without any further purification.

Samples. All the samples were from a freshly prepared product. The products were tested: Glutathione injection solution, containing 20% of glutathione and 2% of ascorbic acid and, glutathione nasal spray, containing 20% of Glutathione and 3.5% ascorbic acid.

The instruments. The Agilent instrument (Agilent Single Quadrupole LC/MS instrument, 2019) includes the following components: quaternary pump 600 bar maximal pressure; single quadrupole (SQ) mass selective detector (MSD) with electrospray ionization (ESI), 150 V fragmentor, gas flow: 7 L/min, gas temperature 300°C, capillary 4000 V and nebulizer 15 psi; OpenLAB CDS Version 2.2. software. Reversed-phase (RP) column Poroshell 120 EC-C18, 50 × 4.6 mm, particles size 2.7 μm, with guard precolumn. Isocratic elution was performed with mobile phase: 0.1% formic acid water solution and flow rate of 1.0 mL/min.

A qualitative analysis was made using MDS spectrum. Glutathione shows the predominant signal of m/z+=308 which corresponds to the cation Glut-H+. Ascorbic acid (AA) shows m/z+=177, 199, and 375 which corresponds to the AA-H+ cation, AA-Na+ cation, and to the 2AA-Na+ cation respectively.

A quantitative analysis was made based on the standard calibration curve.

Preparation of standard stock solution. An accurately weighed of ascorbic acid and glutathione was dissolved in water. The solution was filtered through a 0.45 μm cellulose acetate membrane filter. The stock solution was stepwise diluted to make a set of standard solutions.

System suitability was tested and validated according to the Centre for Drug Evaluation and Research [24] and Agilent recommendations [25].

Calibration curve. Five standard concentration solutions were tested in the range of glutathione concentrations from 0.1 g/l to 1.0 g/l and AA concentrations from 0.04 g/l to 2.1 g/l. Each point of the calibration curve was the mean of five measurements. Slope (a), intercept (b), and correlation coefficient (r) were calculated by least squares using OpenLAB CDS software. The range of concentrations for which the correlation coefficient was equal to or greater than 0.999 was taken as the working range [24]. For MSD, the type of calibration curve (linear or parabolic) was chosen so that the correlation coefficient was maximum.

Accuracy was expressed as mean absolute recovery and percent relative standard deviation (RSD), for AA and glutathione samples in five copies for each concentration.

The precision of the method was determined by comparing the measurement results of five samples under the same experimental conditions. Intra- and inter-day tests were carried out.

Limits of detection (LOD) and quantification (LOQ). LOD is the minimum measurable amount of a substance, LOQ is the minimum measurable amount with acceptable linearity, accuracy, and precision. For a linear calibration curve, LOD is calculated using the formula (1) [5] where (σ) is the standard deviation of the response, and (a) is the slope of the line.

LOD=3.3σ/aE1

If the calibration curve is a parabola (y=ax2 +bx + c), then the LOD is the root of the quadratic equation, which is calculated by the formula (2).

LOD=b+SQRTbb4ac3.3σ/2aE2

The standard deviation of the response (σ) is determined based on the calibration curve as the residual standard deviation of the regression line [5]. The LOQ may be estimated as 3 times LOD.

Repeatability was evaluated in the same tests as for the accuracy and recovery. The measure of repeatability is RSD.

To assess the specificity of the method, samples of the pharmaceutical compound and a standard solution of approximately the same concentration of the tested components were compared in terms of tail coefficient, retention time, and the number of theoretical plates. The difference was expressed in RSD.

The robustness of the method was evaluated by the effect of small changes in flow rate, column temperature, and mobile phase composition on the measurement result. The system suitability parameters (T and N) were determined, and the results were compared with acceptable limits.

Statistical analysis. Each experimental point was the average of five measurements. The results p < 0.05 were considered statistically significant. Data variation was expressed as standard deviation (SD) and relative standard deviation. A regression analysis based on the Least Squares Method was used to construct a calibration curve. The correlation coefficient (r) and the coefficient of determination (r2) were calculated.

Advertisement

3. Results

System suitability. The solution of a standard sample with a volume of 1 μl was injected five times. The relative standard deviation of the peak area, retention time, the number of theoretical plates, and tail factor were determined and compared with the acceptable limits (Table 1) according to the recommendations [26, 27].

Test parametersMeanS.D.% RSDAcceptable limit
AA Peak area (mAU⋅min)1,023.9018.951.9RSD ≤2
Glut. Peak area (counts⋅min)2,543,221.4422,859.330.9RSD ≤2
AA Retention time (min)0.780.000.0RSD ≤2
Glut. Retention time (min)1.020.000.2RSD ≤2
AA Theoretical plates (N)*30,128.504,273.051.6>2,000
Glut. Theoretical plates (N)*47,000.004,242.649.0>2,000
AA Tailing factor (T)*1.290.010.6≤2
Glut. Tailing factor (T)*1.600.010.4≤2

Table 1.

System suitability.

Ascorbic acid (AA) is detected by DAD and glutathione (Glut.) is detected by MSD.

Ascorbic acid standard solution 2.7 μg/μL, injection volume 0.1 μL. Values are presented as mean ±S.D., n = 5, *p < 0.05 Glutathione standard solution 0.5 μg/μL, injection volume 1.0 μL. Values are presented as mean ±S.D., n = 5, *p < 0.05.

The MS spectra and chromatograms are shown in Figures 2 and 3. As we can see in Figure 2B, the ascorbic acid DAD signal is clearly visible, while the glutathione peak is barely visible. The situation is the reverse of Figure 2A, the glutathione peak is high, and the ascorbic acid peak is very small. As a result of electrospray ionization, glutathione is converted into a single cation m/z+=308 (Figure 3A), and ascorbic acid into several cations, including those with m/z+= 177 and 375 (Figure 3B).

Figure 2.

(a) Chromatogram detected by MSD, hight peak corresponds to glutathione. (b) Chromatogram detected by DAD (272 and 380 nm), hight peak corresponds to ascorbic acid. Concentration of glutathione is 0.62 μg/μL, concentration of ascorbic acid is 0.122 μg/μL; mobile phase is aqua solution of 0.1% formic acid.

Figure 3.

Extracted spectrum (m/z+ value) of glutathione (A) and ascorbic acid (B). Concentration of glutathione is 0.62 μg/μL, concentration of ascorbic acid is 0.122 μg/μL; mobile phase is aqua solution of 0.1% formic acid.

Linearity, range, and limit of detection. The working range for glutathione was from 0.2 to 1.0 μg, and for AA from 0.5 to 1.0 μg. The limit of detection was calculated based on the standard deviation of the response (σ) and slope (a) according to Eq. (1) for ascorbic acid, and Eq. (2) for glutathione. The results of the LOD calculation are shown in Table 2. Linear regression in the case of MSD does not have a good correlation coefficient, so parabolic regression was used for MSD, it gives a satisfactory correlation coefficient (r ≥ 0.999).

Table 2.

Linearity, range and LOD.

Ascorbic acid (AA) is detected by DAD and glutathione (Glut.) is detected by MSD. Calibration curves are in the table. Standard errors of the calibration points are represented by the size of the circles (p < 0.05).

“x”—the content of AA or Glut. in the sample; “Y”—the peak area; “Y calc.”—the calculated peak area; “ΔY”—the residues; “a”—the slope of the regression line; “b”—the intercept; “r”—the correlation coefficient; “S.D. ΔY”—the residual standard deviation of the regression line (σ).

Accuracy/recovery and precision. The accuracy of the method was validated for recovery at 3 different concentrations in five replicate tests. Recovery was determined based on the calibration curve. The results are shown in Tables 3 and 4. Interday analysis shows the stability of the AA and Glut samples during the day.

AA (μg)Mean recovery (μg)±SDRSD (%)Recovery (%)
1.3351.3160.0776.0*99
1.7801.7760.0540.0*100
2.6702.6770.0920.0*100
♥1.7801.700.0104.0*96

Table 3.

Glutathione. Accuracy, recovery, repeatability.

Ascorbic acid (AA) is detected by DAD.

Recovery data presents an average value of five independent determinations (n = 5). ♥The bottom row corresponds to the inter-day analysis. Samples were kept overnight in closed vials at 19°C. *p < 0.05

Glut. (μg)Mean recovery (μg)SDRSD (%)Recovery (%)
0.5340.540.035.0*101
0.840.880.023.0*105
1.3351.230.000.0*92
♥0.840.840.034.0*100

Table 4.

Ascorbic acid. Accuracy, recovery, repeatability.

Glutathione (Glut.) is detected by MSD.

Recovery data presents an average value of five independent determinations (n = 5). ♥The bottom row corresponds to the inter-day analysis. Samples were kept overnight in closed vials at 19°C. *p < 0.05.

Selectivity analysis (Table 5). Two peaks of both components were compared, one for the standard solution and the other for the drug. It has been shown that the presence of other ingredients in the composition of the drug does not affect the recovery of glutathione and ascorbic acid. In particular, the relative standard deviation of recovery does not exceed 0.72%. Thus, the method is specific to the antioxidant couple.

Active component, detectorConcentration of active components (μg/μL)Mean peak area. (Standard) (n = 3)Mean peak area (Drug) (n = 3)RSD (%)
Glut.0.200175584917590240.13
AA0.0542182200.72

Table 5.

Specificity.

Ascorbic acid (AA) is detected by DAD and glutathione (Glut.) is detected by MSD.

Two peaks are compared: one for a standard solution, and the other for a dilute solution for injection containing the following components: glutathione 0.2 g/L, ascorbic acid 0.054 g/L, disodium EDTA 0.001 g/L, NaOH 0.02 g/L, benzyl alcohol 0.015 g/L.

To test for robustness, the flow rate, column temperature, and formic acid concentration in the mobile phase were varied. The system suitability parameters (T and N) were determined, they were within acceptable values for all changes in the analysis conditions (Table 6). Thus, the method is robust.

ParameterAscorbic acid (0.1 μg)Glutathione (0.5 μg)
T%RSDN%RSDT%RSDN%RSD
Flow rate 1.0 ml/min1.241.530,7091.81.591.644,5041.9
Flow rate 1.05 ml/min1.301.528,3201.71.601.750,6931.9
Temperature 22°C1.241.530,7091.81.591.644,5041.9
Temperature 24°C1.421.729,5521.51.811.843,3321.7
Mobile phase
Formic acid 0.10%1.241.530,7091.81.591.644,5041.9
Formic acid 0.13%1.211.430,8891.61.561.544,6511.8

Table 6.

Robustness.

Ascorbic acid is detected by DAD and glutathione Is detected by MSD.

T = Tailing factor (mean); N = Theoretical plates (mean); n = 5.

Advertisement

4. Discussion

The use of two detectors (MSD and DAD) for the analysis of the antioxidant couple is efficient and extends the range of HPLC analysis. In the tandem of detectors, DAD comes first because it is a non-destructive detection method. MSD not only provides information on the qualitative chemical composition of the sample but also in certain situations allows us to determine the quantitative content of the component (glutathione) that is not visible to DAD. The calibration curve for the MSD is not linear, but parabolic, which probably reflects the mutual repulsion of like-charged particles in the gas phase. The use of MSD for quantitative analysis requires further study and justification.

Advertisement

5. Conclusion

The HPLC/MS method has been developed for the determination of the antioxidant couple consisting of glutathione and ascorbic acid in pharmaceutical products. The use of a tandem of DAD and MSD detectors is substantiated. The method has been validated for accuracy, stability, and precision. The method has a low detection limit. The presence of foreign components in samples including sodium hydroxide, disodium EDTA, and benzyl alcohol does not impair the accuracy of the analysis. The method provides a fast, sensitive, accurate, and reproducible means of determining the antioxidant couple in pharmaceuticals. Preliminary special preparation of samples is not required.

Advertisement

Conflict of interest

The authors claim that there is no conflict of interest.

Advertisement

Abbreviations

ACNacetonitrile
AAAscorbic acid
DADDiode Array Detector
ESIelectrospray ionization
FDAFood and drug administration
Glut.Glutathione
ICHInternational Conference on Harmonization
LODlimit of detection
m/zmass per charge unit
MSDmass selective detector
MWMolecular weight
RPreversed-phase
SQsingle quadrupole
UV-VISUltraviolet-Visible

References

  1. 1. Forman HJ, Zhang H, Rinna A. Glutathione: Overview of its protective roles, measurement, and biosynthesis. Molecular Aspects of Medicine. 2009;30(1-2):1-12
  2. 2. Pehlivan FE. Vitamin C: An Antioxidant Agent. InertechOpen. 2017. https://www.intechopen.com/books/vitamin-c/vitamin-c-an-antioxidant-agent [Consulted January 19, 2021]
  3. 3. Meister A. The antioxidant effects of glutathione and ascorbic acid. In: Pasquier C, Olivier RY, Auclair C, Packer L, editors. Oxidative Stress, Cell Activation and Viral Infection. Birkhäuser Basel: Molecular and Cell Biology Updates; 1994. pp. 101-111
  4. 4. Agilent Single Quadrupole LC/MS Instrument. 2019. https://www.agilent.com/en/products/liquid-chromatography-mass-spectrometry-lc-ms/lc-ms-instruments/single-quadrupole-lc-ms/lc-msd [Consulted January 19, 2021]
  5. 5. European Medicines Agency. ICH. Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. 2006. https://www.ema.europa.eu/en/documents/scientific-guideline/ich-q-2-r1-validation-analytical-procedures-text-methodology-step-5_en.pdf [Consulted January 19, 2021]
  6. 6. Raju N, Appala, Sridevi Chigurupati, Raju VV, Appala SS, Kesavanarayanan Krishnan Selvarajan, and Jahidul Islam Mohammad. A simple HPLC-UV Method for the Determination of Glutathione in PC-12 Cells. 2016. https://www.hindawi.com/journals/scientifica/2016/6897890/ [Consulted January 19, 2021]
  7. 7. Florholmen-Kjær Å, Lyså RA, Fuskevåg O-M, Goll R, Revhaug A, Mortensen KE. A sensitive method for the analysis of glutathione in porcine hepatocytes. Scandinavian Journal of Gastroenterology. 2014;49(11):1359-1366
  8. 8. Hou Y, Li X, Dai Z, Wu Z, Bazer FW, Wu G. Analysis of glutathione in biological samples by HPLC involving pre-column derivatization with o-phthalaldehyde. Methods in Molecular Biology. 2018;1694:105-115
  9. 9. A Derivatizing Agent for Glutathione Assays. The Protein Man’s Blog. 2016. https://info.gbiosciences.com/blog/a-derivatizing-agent-for-glutathione-assays [Consulted January 19, 2021]
  10. 10. Mika A, Skorkowski E, Stepnowski P. The use of different MS techniques to determine glutathione levels in marine tissues. Food Analytical Methods. 2013;6:789-802
  11. 11. Rellán-Álvarez R, Hernández LE, Abadía J, Álvarez-Fernández A. Direct and simultaneous determination of reduced and oxidized glutathione and homoglutathione by liquid chromatography–electrospray/mass spectrometry in plant tissue extracts. Analytical Biochemistry. 2006;356:254-264
  12. 12. Herzog K, Ijlst L, van Cruchten AG, van Roermund CWT, Kulik W, Wanders RJA, et al. An UPLC-MS/MS assay to measure glutathione as marker for oxidative stress in cultured cells. Metabolites. 2019;9(3):45
  13. 13. Sawant L, Prabhakar B, Pandita N. Quantitative HPLC analysis of ascorbic acid and gallic acid in Phyllanthus Emblica. Journal of Analytical and Bioanalytical Techniques. 2010;1:111
  14. 14. Frenich AG, Torres ME, Vega AB, Vidal JL, Bolaños PP. Determination of ascorbic acid and carotenoids in food commodities by liquid chromatography with mass spectrometry detection. Journal of Agricultural and Food Chemistry. 2005;53:197371-197376
  15. 15. Somchai B, Sasiwimon L, Yaneenart S, Supaluk P, Virapong P. Analysis of ascorbic acid and isoascorbic acid in orange and guava fruit juices distributed in Thailand by LC-IT-MS/MS. Food Analytical Methods. 2016;9(6):1616-1626
  16. 16. Gazdik Z, Zitka O, Petrlova J, Adam V, Zehnalek J, Horna A, et al. Determination of Vitamin C (ascorbic acid) using high performance liquid chromatography coupled with electrochemical detection. Sensors. 2008;8:7097-7112
  17. 17. Mitić SS, Kostić DA, Nasković-Dokić DC, Mitic MN. Rapid and reliable HPLC method for the determination of Vitamin C in pharmaceutical samples. Tropical Journal of Pharmaceutical Research. 2011;10(1):105
  18. 18. Racz E, Parlagh-Huszar K, Kecskes T. HPLC Method for Determination of Ascorbic Acid in fruits and Vegetables. Budapest: Technical University; 1990
  19. 19. Watada AE. A high-performance liquid chromatography method for determining ascorbic acid content of fresh fruits and vegetables. Horticulture. Science. 1982;17(3):334-335
  20. 20. Lloyd LL, Warner FP, White CA, Kennedy JF. Quantitative reversed phase HPLC analysis of L-Ascorbic acid (Vitamin-C) and identification of its degradation products. Chromatographia. 1987;24:371-374
  21. 21. Yefimov SV. Express qualitative and quantitative HPLS/MS analysis of the ascorbic acid in pharmaceutical product. International Journal of Chemical and Pharmaceutical Sciences. 2019;10(4):40-45
  22. 22. Szultka M, Buszewska-Forajta M, Kaliszan R, Buszewski B. Determination of ascorbic acid and its degradation products by high performance liquid chromatography-triple quadrupole mass spectrometry. Electrophoresis. 2014;35(4):585-592
  23. 23. FDA Guidance for Analytical Procedures and Methods Validation for Drugs and Biologics Guidance for Industry. 2015. https://www.fda.gov/media/87801/download [Consulted January 19, 2021]
  24. 24. Centre for Drug Evaluation and Research (CDER), FDA. Reviewer Guidance, Validation of Chromatographic Methods. Rockville: FDA. 1994. https://www.fda.gov/media/75643/download [Consulted January 19, 2021]
  25. 25. Evaluating System Suitability CE, GC, LC and A/D ChemStation. Revisions: A.03.0x-A.08.0x. 2019. https://www.agilent.com/cs/library/Support/Documents/a10424.pdf [Consulted January 19, 2021]
  26. 26. Deepak. How to calculate System Suitability in Chromatography. 2013. https://lab-training.com/2013/02/27/how-to-calculate-system-suitability-in-chromatography/ [Consulted January 19, 2021]
  27. 27. Bose A. HPLC Calibration Process Parameters in Terms of System Suitability Test. 2014. https://pdfs.semanticscholar.org/c378/3cd6c88c294be0aecee3b77cfd1eb35789fc.pdf [Consulted January 19, 2021]