Open access peer-reviewed chapter

Development of An Impedimetric Nanoplatform for Cu(II) Detection Based on Carbon Dots and Chitosan Nanocomposite

Written By

Mosaab Echabaane and Chérif Dridi

Submitted: 28 May 2022 Reviewed: 17 June 2022 Published: 15 February 2023

DOI: 10.5772/intechopen.105924

From the Edited Volume

Frontiers in Voltammetry

Edited by Shashanka Rajendrachari, Kiran Kenchappa Somashekharappa, Sharath Peramenahalli Chikkegouda and Shamanth Vasanth

Chapter metrics overview

69 Chapter Downloads

View Full Metrics

Abstract

In this chapter, an impedemitric sensor based on carbon dots (CDs) and chitosan (CS) nanocomposite for Cu (II) detection has been studied. The synthesized carbon dots were characterized by TEM, FTIR, XRD, UV-visible, and PL. The obtained carbon dot/chitosan/GCE structure was characterized by scanning electron microscopy (SEM) and electrochemical impedance spectroscopy (EIS). The behavior, interface processes of GCE/electrolyte, and sensor properties of the studied structures were investigated. In particular, the performance of the proposed impedimetric sensor to Cu ions, such as sensitivity, dynamic range, detection limit, and selectivity, was studied. Subsequently, the proposed sensor was applied for the determination of copper ions in real water samples.

Keywords

  • carbon dots
  • chitosan
  • nanocomposite
  • copper(II)
  • Impedimetric sensors
  • introduction

1. Introduction

Essential transition metals such as copper, zinc, iron, and manganese play vital roles in several biological processes [1]. One of the essential transition metal ions, copper is the most abundant metal ion in the human body. It contributes to blood formation, connective tissue development, transcriptional events, and the functioning of several enzymes [2, 3]. However, exposure to an excess level of copper ion can cause gastrointestinal disturbance, neurodegenerative diseases, and liver and kidney damage [4]. The maximum content of Cu2+ ions in drinking water is limited to 30 nM by the World Health Organization (WHO) [5].

Hence, it is very important to develop a simple, fast, economically advantageous, selective, and sensitive analytical technique for the determination of copper in aquatic environments to avoid side effects. Notably, numerous techniques have been used to determine the copper ion, including atomic absorption spectroscopy, chromatography spectrofluorimetry, and spectrophotometry [6, 7, 8, 9]. These techniques can provide good accuracy and sensitivity, but they have various disadvantages, such as expensive equipment, time-consuming operation, and an inability to do onsite analysis. Thus, their use is restricted. Therefore, to circumvent these limitations, electrochemical sensors have attracted great attention for detecting trace levels of copper ions due to their inherent advantages, such as simple operation, low-cost, rapidity, excellent sensitivity, and selectivity [10]. Notably, the presence of numerous advanced electrode modified nanomaterials improved sensing performance significantly [11, 12, 13, 14, 15]. Amongst such nanomaterials, nanocomposites have been employed in electrochemical sensors for environmental analysis [16, 17, 18, 19, 20, 21]. Nowadays, analytical measurements are altered by nanocomposites of various sizes, shapes, and compositions. Chitosan as well as carbonaceous nanoparticles have an interest in the construction of electrochemical sensors [22, 23, 24, 25, 26, 27, 28]. Chitosan exhibits interesting properties due to its nontoxicity, film-forming ability, high permeability, adsorption ability, and biocompatibility. It has been extensively used as a good biomaterial in designing sensors [29, 30]. As compared to other carbonaceous nanoparticles, carbon dots have considerable importance because of their simple synthetic routes, high surface area, high biocompatibility, and high conductivity [31, 32]. The aim of this chapter is to develop an impedimetric sensor for copper ion detection based on carbon dots and chitosan nanocomposite. The synthesized carbon dots were characterized by transmission electron microscope (TEM), X-ray diffraction (XRD), FTIR spectroscopy, UV–Vis absorption, and photoluminescence spectroscopy. The morphological and electrochemical properties of carbon dot/chitosan/GCE structure were investigated by scanning electronic microscopy (SEM) and electrochemical impedance spectroscopy (EIS). The weather sensors’ performance was discussed. The proposed sensor can be used to determine copper ions in real water samples.

Advertisement

2. Experimental details

2.1 Reagents

Copper sulfate (CuSO4, ≥ 99.9%), mercury nitrate monohydrate (Hg(NO3)2.·H2O, ≥ 99.9%), cadmium nitrate tetrahydrate (Cd(NO3)2,·4H2O, ≥ 99%), lead nitrate (Pb(NO3)2, ≥ 99%), nickel nitrate hexahydrate (Ni(NO3)2·6H2O, ≥ 98.5%), zinc nitrate hexahydrate (Zn(NO3)2·6H2O, ≥ 98%), sodium chloride (NaCl, ≥ 99%), calcium chloride (CaCl2, ≥ 96%), potassium chloride (KCl, ≥ 99%), glucose (≥ 99.5%), sulfuric acid (H2SO4, ≥ 99.9%), Deionized (DI) water, sodium hydroxide (NaOH, ≥ 97%), ammonium acetate (CH3COONH4, ≥ 98%), and chitosan (CS, ≥ 85%). All chemicals used in this work were purchased from of Sigma- Aldrich.

2.2 Instrumentations

For further investigation of the synthesized carbon dots (CDs), an X-ray diffractometer (Bruker AXS D8 Advance), Fourier Transform Infrared Spectroscopy (FTIR) (Perkin Elmer, Spectrum Two), UV–vis spectrophotometry (Specord 210 Plus), and a spectrophotometer (Perkin Elmer) were employed.

Scanning Electron Microscopy (SEM) (Jeol JSM-5400 emission) was used to characterize the CS-CDs film.

The electrochemical impedance measurements were carried out using an impedance analyzer (EC-LAB BIOLOGIC) with a sinusoidal excitation signal amplitude of 10 mV and a 0.01 Hz-100 kHz frequency range. The polarization potential used for these measurements was optimized at −600 mv. An aqueous ammonium acetate (solution of 0.1 M, with pH = 7, was used as an electrolyte in the electrochemical cell formed by three electrodes: a working electrode (GCE), a platinum counter electrode, and an Ag/AgCl electrode (SCE).

2.3 Synthesis of CDs

Glucose (3 g) and H₂SO₄ (1 ml) were dissolved in DI water (10 ml). Then, the mixed solution was heated for 30 min at a temperature of 100°C (Figure 1). The solution changed from colorless to yellow-brownish, signaling the formation of CDs. Then, this latter was neutralized using NaOH. In order to remove impurities, the prepared solution was centrifuged (14,000 rpm), filtered, and dried for 6 hours at a temperature of 80°C [33].

Figure 1.

Schematic representation of CDs formation from glucose.

2.4 Preparation of modified glassy carbon electrode (GCE)

The surface of the GCE electrode was polished with 0.05 μm and 0.3 μm alumina powder. Then, the electrode was washed successively with distilled water, ethanol, and distilled water in an ultrasonic bath, and dried at room temperature. For the preparation of the CDs-CS nanocomposite solution: 0.5 mL of 1.0% CS solution was added to 1.5 mL of CDs solution by ultrasonication. Then, 2 μL of nanocomposite was dropped onto the surface of GCE and dried in the oven at 60°C for 30 min.

Advertisement

3. Characterization of CDs and CS-CDs/GCE

The TEM image of the synthesized CDs is shown in Figure 2(a). It is noticed that the CDs are spherical in shape and their mean size is approximately 3.2 nm. The X-ray diffraction pattern of CDs is shown in Figure 2(b). It exhibited a broad peak at 2θ=22.35°, corresponding to the (002) hkl plane of the graphitic structure (JCPDS card no. 26–1076). The FTIR transmittance spectrum of the CDs is presented in Figure 2(c). The peaks at 3319 cm−1, 1727 cm−1, 1642 cm−1, and 1092 cm−1 were assigned to O-H, C = O, C = C, and C-O. These indicate that carboxyl and hydroxyl groups may be found on the surface of CDs. The UV–Vis absorption spectrum of CDs exhibited two absorption bands located at 280 nm and 375 nm (Figure 2(d)), which are ascribed to π-π* and n-π* respectively. Figure 2(e) presents the PL spectra of the CDs. The fluorescence emission peak of CDs located at 470 nm is a blue light luminescent spectral region.

Figure 2.

Characterization of CDs: (a) TEM images, (b) XRD pattern, (c) FTIR spectra, (d) UV–visible spectra, (e) PL spectra.

SEM images of the bare GCE and the CS-CDs/GCE film are shown in Figure 3. It exhibits that the bare GCE electrode has a flat and smooth surface (Figure 3(a)), whereas the surface of the CS-CDs nanocomposite is rough with a few apparent pores (Figure 3(b)). This result shows that the CS-CDs film was deposited on the GCE surface.

Figure 3.

Images SEM of (a) bare GCE electrode, (b) CS-CDs nanocomposite/GCE electrode.

Advertisement

4. EIS characterization of the CS-CDs/GCE electrode

Electrochemical impedance spectroscopy is a sensitive tool for investigating the characteristics of the electrode-solution interface. The Nyquist diagrams of the bare GCE electrodes and functionalized CS-CDs film are shown in Figure 4. The Nyquist of CS-CDs/GCE electrodes is higher than that of bare CGE electrodes. This is due to the immobilization of the film. Besides, Figure 4 shows that the Nyquist plot of the CS-CDs/GCE structure is a single semicircle with a large diameter. This might be explained by the fact that it is an overlapping of two semicircles.

Figure 4.

Impedance spectra of bare GCE, and CS-CDs/GCE structures.

The EIS spectrum of the CS-CDs/GCE electrode was simulated with the equivalent electrical circuit presented in Figure 5. This circuit is an association of three components.

Figure 5.

Equivalent circuit used to fit the impedance spectra.

This circuit is an association of three components.

  • The first one, in the low-frequency range, is used to represent the GCE/electrolyte interface, including in a parallel combination the constant phase element CPEdl that substitutes the electrical double layer capacitance (Cdl) with the charge transfer resistance Rct.

  • The second component at higher frequencies is attributed to the film/electrolyte interface formed by a parallel combination of the resistance Rm and the capacitance CPEm. Where Rm refers to the resistance of the film and CPEm donates the constant phase element capacitance, is assigned to the electric capacitor consisting of the metal and the electrolyte, with the film as the dielectric.

  • The last component is the series resistance of the electrolyte solution.

The two constant phase elements (CPE) are non-ideal capacitances and can be expressed by (Eq. (1)):

ZCPE=1QnE1

where Q is a constant, j is the imaginary number, ω is the angular frequency, and 0 < n < 1. CPE becomes more capacitive, when the value n tends to be 1.

Finally, the experimental EIS measurements were fitted using the ZView software. The different fits were made with a total error value of χ2 ≈ 10−4.

Advertisement

5. Impedance characterization for the detection of Cu ion

The electrochemical impedance response of copper at the CS-CDs/GCE is characterized using EIS. The Nyquist spectra of the CS-CDs film obtained after the addition of different concentrations of copper ions is shown in Figure 6.

Figure 6.

Impedance spectra for different copper concentrations of CS-CDs/GCE.

It is noted that the Nyquist plot of the CS-CDs film clearly decreases as a function of Cu2+ concentration from 10−9 M to 10−5 M. The response of the CS-CDs modified electrode to copper ions is due to a complexation process. This is induced by coordination between the donor groups of nanocomposites and the copper ions. In fact, the interaction of the copper ions (Lewis acid, acceptor electrons) with the amide and hydroxyl groups of CS (Lewis base, donor electrons) is possible by the acid–base reaction. Furthermore, the negatively charged carboxyl and hydroxyl groups existing on the surface of CDs enable interactions with the positively charged copper ions. Thus, the interaction mechanism is evolved by an electrostatic interaction that occurs between the copper ions and the nanocomposite film.

Table 1 summarizes the circuit parameters of the CS-CDs film as determined by the fit to the previous equivalent circuit.

[Cu2+] (M)Rs (kΩ)CPEm (μF)n1Rm (kΩ)CPEdl (μF)n2Rct (kΩ)χ2 (10−4)
10−90.111.700.9017.409.200.5967.242.7
10−80.101.730.8916.2410.010.6056.566.8
10−70.111.600.8815.019.220.6045.057.3
10−60.121.560.9013.409.520.6136.162.1
10−50.131.660.8812.5010.100.6227.724.2

Table 1.

Fitting parameters of CS-CDs/GCE versus copper concentrations.

Advertisement

6. Physical meaning of parameters obtained from the Nyquist plots

Figure 7 shows the plot of the film resistance versus p[Cu2+]. The value of film resistance (Rm) decreases when increasing the concentration of copper. This is may be explained by the increase in the ionic conductivity of the CS-CDs film.

Figure 7.

Variation of the Rm of CS-CDs/GCE.

Figure 8 represents the variation of the charge transfer resistance (Rct) parameters upon the addition of Cu2+. It was observed that Rct decreased with increasing Cu2+ concentration. This can be due to improvement in the charge transfer at the GCE/CS-CDs electrode electrolyte interface.

Figure 8.

Variation of the Rct of CS-CDs/GCE.

Table 2 presents the Rm and Rct slope values. It is noted that the Rct slope value is higher than the Rm slope value. This can indicate that the recognition process was at the electrolyte/GCE interface. Therefore, the Rct is provided as the primary factor determining the sensors’ response.

Electrode/p [Cu2+]Rm (KΩ)Rct (KΩ)
CS-CDs/GCE1.39.9

Table 2.

Rm and Rct slope values of CS-CDs/GCE.

Advertisement

7. Analytical performance of impedimetric sensors

Figure 9 shows the variation of -log(Rtc/Rtc0) as a function of copper concentration, where Rtc0 is the charge transfer resistance value without copper ions in the solution. Here, a linear behavior with a large detection range between 10−9 and 10−5 M was observed. The equation of linear regression is -log (Rtc/Rtc0) = 0.1 × log[copper] + 0.9, where the slope is 0.1 of the sensor sensitivity. The detection limit of our copper sensor was ≈ 5 × 10−10 M, according to the formula LOD = (3 × standard deviation)/slope.

Figure 9.

Calibration curve of -log (Rtc/Rtc0) versus log ([copper]).

Advertisement

8. Specificity

The specificity of the GCE/CS-CDs sensor has been studied towards some interfering cations: Hg2+, Cd2+, Pb2+, Ni2+, Zn2+, Na+, Ca+, and K+. Figure 10 shows the relative signal change RtcRtc0Rtc0 as a function of the interfering cations. The results noted that GCE/CS-CDs displayed good selectivity for Cu2+ in the presence of other ions.

Figure 10.

Influence of interferences (10−4 M) on the response copper (10−6 M).

Advertisement

9. Determination of copper ion in water samples

To evaluate the practical feasibility of the GCE/CS-CDs sensor, it is applied for the detection of copper in water samples (tap water). 1 mL of sample solution (tap water) is added to 1 mL of electrolyte (acetate ammonium at pH 7). The impaled copper amounts are of 10−6 M and 10−8 M. The recovery percentage is calculated as FoundMAddedM×100. It can be seen in Table 3 that the recoveries are less than 5%. Therefore, the proposed GCE/CS-CDs impedimetric sensor is efficient for the detection of Cu2+ ions in real water samples.

SampleAdded (M)Found (M)Recovery (%)RSD (%)
Tap water10−89.70 × 10−997.04.12
10−61.02× 10−6102. 01.47

Table 3.

Determination of copper ion in tap water samples (n = 3).

Advertisement

10. Conclusion

In this chapter, we fabricate a highly sensitive, selective, impedimetric sensor for the determination of copper-based on carbon dots/chitosan nanocomposite modified glassy carbon electrode. A simple, fast, low-cost method to prepare carbon dots using glucose by heating has been demonstrated. The electrochemical performance of the copper sensor was examined by electrochemical impedance spectroscopy. Based on the excellent properties of CS-CDs/GCE, the impedimetric sensors had high sensitivity, a low limit of detection, and good selectivity. The newly developed impedimetric sensor for the determination of copper ions was applied in water for real samples with satisfactory results. Thus, this indicates that these sensors offer promising potential for water monitoring.

References

  1. 1. Cowan JA. Handbook of Inorganic Biochemistry: An Introduction. 2nd ed. New York: Wiley; 1997. 456 p. DOI: 978-0-471-18895-7
  2. 2. Nasiri MM, Taher MA, Fazelirad H. Preparation and application of a simple electrochemical sensor for the determination of copper in some real and standard samples. Ionics. 2016;22:289-296. DOI: 10.1007/s11581-015-1533-9
  3. 3. Rouis A, Echabaane M, Sakly N, Bonnamour I, Ouada HB. Characterization of a sensitive and selective copper optode based on β-ketoimine modified calix[4]arene derivative. Materials Science and Engineering: C. 2015;46:125-131. DOI: 10.1016/j.msec.2014.10.026
  4. 4. Ansari R, Delavar AF, Aliakbar A, Mohammad-khah A. Solid-state Cu (II) ion-selective electrode based on polyaniline-conducting polymer film doped with copper carmoisine dye complex. Journal of Solid State Electrochemistry. 2012;16:869-875. DOI: 10.1007/s10008-011-1436-2
  5. 5. Wu T, Xu T, Ma Z. Sensitive electrochemical detection of copper ions based on the copper (II) ion assisted etching of Au@ Ag nanoparticles. Analyst. 2015;140:8041-8047. DOI: 10.1039/C5AN01888A
  6. 6. Soylak M, Narin I, Dogan M. Trace enrichment and atomic absorption spectrometric determination of lead, copper, cadmium and nickel in drinking water samples by use of an activated carbon column. Analytical Letters. 1997;30:2801-2810. DOI: 10.1080/00032719708001823
  7. 7. Ali A, Shen H, Yin X. Simultaneous determination of trace amounts of nickel, copper and mercury by liquid chromatography coupled with flow-injection on-line derivatization and preconcentration. Analytica Chimica Acta. 1998;369:215-223. DOI: 10.1016/S0003-2670(98)00252-9
  8. 8. Amjadi M, Fakhri ZA. Gold nanostar-enhanced chemiluminescence probe for highly sensitive detection of Cu (II) ions. Sensors and Actuators B: Chemical. 2018;257:629-634. DOI: 10.1016/j.snb.2017.11.013
  9. 9. Agnihotri NK, Singh VK, Singh HB. Derivative spectrophotometric determination of copper (II) in non-ionic micellar medium. Talanta. 1997;45:331-341. DOI: 10.1016/S0039-9140(97)00137-9
  10. 10. Kaur I, Sharma M, Kaur S, Kaur A. Chemical sensors based on self-assembled chelating dithiol on gold electrode for trace level detection of copper (II) ions. Sensors and Actuators B: Chemical. 2020;312:127935-127946. DOI: 10.1016/j.snb.2020.127935
  11. 11. Mahmoud A, Echabaane M, Omri K, ElMir L, Chaabane RB. Development of an impedimetric non enzymatic sensor based on ZnO and Cu doped ZnO nanoparticles for the detection of glucose. Journal of Alloys and Compounds. 2019;786:960-968. DOI: 10.1016/j.jallcom.2019.02.060
  12. 12. Somashekharappa KK, Rajendrachari S. Handbook of Carbon Nanomaterial-Based Sensors. 1st ed. Elsevier. 2022. p. 3. DOI: 10.1016/B978-0-323-91174-0.00001-9
  13. 13. Rajendrachari S, Ramakrishna D. Handbook of Functionalized Nanomaterial-Based Electrochemical Sensors. 1st ed. Elsevier; 2022. p. 3. DOI: 10.1016/B978-0-12-823788-5.00010-7
  14. 14. Shashanka R, Jayaprakash GK, Prakashaiah BG, Kumar M, Swamy BEK. Electrocatalytic determination of ascorbic acid using a green synthesised magnetite nano-flake modified carbon paste electrode by cyclic voltammetric method. Materials Research Innovations. 2022;26:229-239. DOI: 10.1080/14328917.2021.1945795
  15. 15. Shashanka R, Swamy BEK. Simultaneous electro-generation and electro-deposition of copper oxide nanoparticles on glassy carbon electrode and its sensor application. SN Applied Sciences. 2020;2:956. DOI: 10.1007/s42452-020-2785-1
  16. 16. Jebril S, Aguilera LC, Santander JMP, Dridi C. A novel electrochemical sensor modified with green gold sononanoparticles and carbon black nanocomposite for bisphenol a detection. Materials Science and Engineering B. 2021;264:114951-114961. DOI: 10.1016/j.mseb.2020.114951
  17. 17. Raghu GK, Sampath S, Pandurangappa M. Chemically functionalized glassy carbon spheres: A new covalent bulk modified composite electrode for the simultaneous determination of lead and cadmium. Journal of Solid State Electrochemistry. 2012;16:1953-1963. DOI: 10.1007/s10008-011-1595-1
  18. 18. Yin T, Pan D, Qin W. A solid-contact Pb2+ −selective polymeric membrane electrode with nafion-doped poly (pyrrole) as ionto-electron transducer. Journal of Solid State Electrochemistry. 2012;16:499-504. DOI: 10.1007/s10008-011-1358-z
  19. 19. Lisak G, Bobacka J, Lewenstam A. Recovery of nanomolar detection limit of solid-contact lead (II)-selective electrodes by electrode conditioning. Journal of Solid State Electrochemistry. 2012;16:2983-2299. DOI: 10.1007/s10008-012-1725-4
  20. 20. Intarakamhang S, Schuhmann W, Schulte A. Robotic heavy metal anodic stripping voltammetry: Ease and efficacy for trace lead and cadmium electroanalysis. Journal of Solid State Electrochemistry. 2013;17:1535-1542. DOI: 10.1007/s10008-013-2018-2
  21. 21. Özcan A, Hamid F, Özcan AA. Synthesizing of a nanocomposite based on the formation of silver nanoparticles on fumed silica to develop an electrochemical sensor for carbendazim detection. Talanta. 2021;222:121591-112160. DOI: 10.1016/j.talanta.2020.121591
  22. 22. Rizwan M, Elma S, Lim SA, Ahmed MU. AuNPs/CNOs/SWCNTs/chitosan-nanocomposite modified electrochemical sensor for the label-free detection of carcinoembryonic antigen. Biosensors & Bioelectronics. 2018;107:211217. DOI: 10.1016/j.bios.2018.02.037
  23. 23. Jiang H, Du C, Zou Z, Li X, Akins DL, Yang H. A biosensing platform based on horseradish peroxidase immobilized onto chitosan-wrapped single-walled carbon nanotubes. Journal of Solid State Electrochemistry. 2009;13:791-798. DOI: 10.1007/s10008-008-0612-5
  24. 24. Li Y, Li Y, Yang Y. A new amperometric H2O2 biosensor based on nanocomposite films of chitosan6MWNTs, hemoglobin, and silver nanoparticles. Journal of Solid State Electrochemistry. 2012;16:1133-1140. DOI: 10.1007/s10008-011-1503-8
  25. 25. Wang L, Zhu H, Hou H, Zhang Z, Xiao X, Song Y. A novel hydrogen peroxide sensor based on Ag nanoparticles electrodeposited on chitosan-graphene oxide/cysteamine-modified gold electrode. Journal of Solid State Electrochemistry. 2012;16:693-1700. DOI: 10.1007/s10008-011-1503-8
  26. 26. Huang J, Xie Z, Huang Y, Xie L, Luo S, Fan Q, et al. Electrochemical immunosensor with Cu (I)/Cu (II)-chitosan-graphene nanocomposite-based signal amplification for the detection of Newcastle disease virus. Scientific Reports. 2020;10:1-12. DOI: 10.1038/s41598-020-70877-3
  27. 27. Comino CS, Rassas I, Minot S, Bessueille F, Arab M, Chevallier V, et al. Voltammetric sensor based on molecularly imprinted chitosan-carbon nanotubes decorated with gold nanoparticles nanocomposite deposited on boron-doped diamond electrodes for catechol detection. Materials. 2020;13:688-699. DOI: 10.3390/ma13030688
  28. 28. Jalali F, Hassanvand Z, Barati A. Electrochemical sensor based on a nanocomposite of carbon dots, Hexadecyltrimethylammonium bromide and chitosan for Mesalazine determination. Journal of Analytical Chemistry. 2020;75:544-552. DOI: 10.1134/S1061934820040061
  29. 29. Nazari F, Ghoreishi SM, Khoobi A. Bio-based Fe3O4/chitosan nanocomposite sensor for response surface methodology and sensitive determination of gallic acid. International Journal of Biological Macromolecules. 2020;160:456-469. DOI: 10.1016/j.ijbiomac.2020.05.205
  30. 30. Sheng Q, Liu R, Zheng J, Lin W, Li Y. Direct electrochemistry of glucose oxidase immobilized on porous carbon nanofiber/room temperature ionic liquid/chitosan composite film and its biosensing application. Journal of Solid State Electrochemistry. 2012;6:739-745. DOI: 10.1007/s10008-011-1420-x
  31. 31. Wang X, Feng Y, Dong P. Huang J a mini review on carbon quantum dots: Preparation, properties and electrocatalytic application. Frontiers in Chemistry. 2019;7:671-680. DOI: 10.3389/fchem.2019.00671
  32. 32. Campuzano S, Sedeño PY, Pingarrón JM. Carbon dots and graphene quantum dots in electrochemical biosensing. Nanomaterials. 2019;9:634-652. DOI: 10.3390/nano9040634
  33. 33. Echabaane M, Hfaiedh S, Smiri B, Saidi F, Dridi C. Development of an impedimetric sensor based on carbon dots and chitosan nanocomposite modified electrode for Cu(II) detection in water. Journal of Solid State Electrochemistry. 2021;25:1797-1806. DOI: 10.1007/s10008-021-04949-3

Written By

Mosaab Echabaane and Chérif Dridi

Submitted: 28 May 2022 Reviewed: 17 June 2022 Published: 15 February 2023