Open access peer-reviewed chapter

Toric Intraocular Lenses

By Zequan Xu

Submitted: May 26th 2019Reviewed: October 15th 2019Published: November 25th 2019

DOI: 10.5772/intechopen.90153

Downloaded: 102

Abstract

This chapter described a short history about the toric intraocular lenses (IOLs) and then discussed some interesting topics such as the measurement of (front and posterior) corneal astigmatism and surgically induced astigmatism; the manual marking techniques and image-guided systems and intraoperative aberrometry-based methods; the new toric lens calculation calculators and toric IOLs formulas; the post operation care of toric IOLs and re-rotation of misaligned toric IOLs; and some relevant issues on multifocal toric intraocular lens. Meanwhile, this chapter also discussed toric IOL in some special cases like keratoconus corneal ectatic disorders, post-refractive surgery and post-keratoplasty, etc.

Keywords

  • astigmatism
  • cataract
  • toric IOL

1. Introduction

Up to more than one-third of cataract patients have preoperative corneal astigmatism of more than 1.0 diopter (D) [1], while 26.2% have more than 1.5 D [2, 3], 8–14.9% have more than 2.0 [1, 3], and 2.6–7.4% have more than 3.0 D [1, 3]. Astigmatism is one of the most important factors that affect postoperative vision quality. More than 0.5 D of residual astigmatism can reduce visual performance and patient satisfaction [4, 5, 6]. Currently, implanting a toric lens is recognized as the most accurate form of astigmatic correction during cataract surgery, especially astigmatism of more than 1 D [7]. Actually, toric IOLs correct preexisting regular corneal astigmatism usually ranging from 0.75 to 4.75 D [8]. However, the outcomes after toric IOL implantation are still influenced by many factors including accurate preoperative measurement of corneal astigmatism, IOL selection, marking techniques, intraoperative alignment and postoperative care, etc.

2. A short history and clinical outcomes of toric IOLs

The first article reporting a toric IOL (Nidek NT -98B) was published in 1994 [9], which had a cylinder power of 2.00 or 3.00 D. In the study, Shimizu et al. had relatively positive results, although some negative results still occurred in some eyes of which the lens axis rotated more than 30° [9]. Ever since then, with the predictability increasing and the safety enhancing, toric IOLs have definitely become a considerable option to correct significant astigmatism when undergoing cataract surgery [10, 11]. At present, standard toric IOLs are available in cylinder powers of 1.0 to 6.0 D, while higher cylinder powers are also available (see Table 1).

IOLMaterialDesignAsphericSpherical power (D)Cylinder power (D) at IOL planeIncision size (mm)
Acri. comfort (Carl Zeiss Meditec) [12]Hydrophilic acrylic with hydrophobic surfacePlate haptic, 11.0-mm dialectY−10.0 to +32.01.0–12.0 (0.50 steps)<2.0
T-flex (Rayner) [13]Hydrophilic acrylicC-loop haptic with AVH technology, 12.0–12.5-mm dialectY−10.0 to +35.01.0–11.0 (0.25 steps)<2.0
AF-1 Toric (Hoya) [7]Hydrophobic acrylic with PMMA haptic tipsPMMA-modified C-loop haptic, 12.5-mm dialectY+6.0 to +30.01.5–6.0 (0.75 steps)2.0
AcrySof (Alcon) [14, 15, 16, 17, 18, 19]Hydrophobic acrylicC-loop haptic,13.0-mm dialectY+6.0 to +34.01.0–6.0 (0.75 steps)2.2
TECNIS Toric IOL (Abbott Medical Optics) [20]Hydrophobic acrylic“Tri-Fix” modified C haptic integral with optic, 13.0-mm dialectY+5.0 to +34.01.5–6 (0.5–1.0 steps)2.2
Precizon toric IOL (OPHTEC) [21, 22]Hydrophilic acrylicBiconvex transitional conic toric design offset-shaped hapticY+1.0 to +34.01.0–10.0 (0.5 steps)2.2
Morcher 89A, 92S (Morcher GmbH) [23, 24]Hydrophilic acrylicBag-in-the-lens, 7.5-mm dialectN+10.0 to +30.0 D0.5–8.0 (0.25 steps)2.5
LENTIS Tplus (Oculentis) [7]Hydrophilic acrylic with hydrophobic surfaceC/Plate haptic, 12.0–11.0-mm dialectY−10.0 to +35.00.25–12.0 (0.75–1.0 steps)2.6
STAAR (STAAR Surgical Company) [25]SiliconePlate haptic, 10.8–11.2-mm dialectN+9.5 to +28.52.0 or 3.52.8
Light-adjustable lens (Calhoun Vision) [26]Silicone with PMMA hapticsModified C-loop PMMA haptics, 13.0-mm dialectY+17.0 to +24.00.75–2.03.0
Microsil (HumanOptics) [27]Silicone with PMMA hapticsC-loop haptic, 11.6-mm dialectN−10.0 to +35.01.0–15.0 (1.0 steps)3.4

Table 1.

Summary of commercially available toric IOLs.

Toric IOL had achieved increasingly great visual outcomes. An uncorrected distance visual acuity (UDVA) of 20/40 or better is achieved in more than 70% of the cases, and spectacle independence has been reported in more than 60% of the patients in previous studies [12, 13, 15, 16, 17, 18, 19, 20, 21, 22, 23, 25, 26, 27, 28, 29, 30], which is significantly increased compared with nontoric monofocal IOLs [31, 32]. A randomized controlled trial (RCT) compared the outcomes of AcrySof toric IOLs with conventional spherical IOLs and observed a UDVA of 20/40 or better in 92.2% of cases undergoing toric IOL implantation, with 63.4% having a UDVA of 20/25 or better. In contrast, only 81.4% of cases undergoing nontoric IOL implantation had a UDVA of 20/40 or better and 41.4% had a UDVA of 20/25 or better [9]. Similar results were found in another high-quality RCT [29].

Compared with incisional astigmatic keratotomy, toric IOLs offered better predictability and stability of correction [17], especially in moderate to high astigmatism [30]. In a recent meta-analysis (including 13 RCTs with 707 eyes), toric IOLs provided better distance visual acuity and lower amounts of residual astigmatism, combined with greater spectacle independence, than nontoric IOLs even when relaxing incisions were used [33].

From a social cost-effectiveness perspective, toric IOLs were inferior to monofocal IOLs in a recent prospective study [34], which should be noted in health-care decision-making.

3. The measurement of astigmatism

For a toric IOL, the keratometric astigmatism (both axis and magnitude) of the cornea must be accurately measured.

3.1 Anterior corneal curvature

Traditionally, keratometry and topography take into account only the anterior corneal curvature [35]. However, nomograms predict total corneal astigmatism based on the power and axis of the anterior corneal astigmatism, assuming a fixed ratio between the anterior and posterior curvature [36]. These methods obviously cannot take outliers and irregularities into account (e.g., post-refractive surgery eyes) [35], thus leading to significant postoperative and/or overcorrection. However, if the agreement of measurement of astigmatism between instruments of different kinds is poor (more than 10°), the selection of toric IOLs requires extra care.

3.2 Posterior corneal curvature

The astigmatism of posterior cornea is generally minus lens of against-the-rule. As mentioned above, ignoring effects of actual posterior corneal curvature may lead to inaccuracies in total astigmatism estimation in some eyes. In a recent study [36], for those eyes who received IOLs with 2 diopters of cylinder or less, a coefficient of adjustment of 0.75 for with-the-rule astigmatism and 1.41 for against-the-rule astigmatism can be applied to the corneal astigmatism power value to calculate a more appropriate IOL cylinder power than that be calculated by using unadjusted anterior corneal curvature measurements.

Since minimizing the residual refractive error is especially critical in toric multifocal IOLs [37], imaging systems that measure posterior corneal curvature, as well as the new algorithm that incorporates the effect of posterior corneal astigmatism, are increasingly being invented. For example, the Scheimpflug imaging systems, slit scanning systems, and OCT systems could measure posterior corneal curvature, besides the anterior curvature. In a comparative study [35] including a Scheimpflug tomography (OCULUS Pentacam), a Placido topographer (Tomey TMS-5 in Placido mode), a swept source/Fourier domain OCT (CASIA SS-1000), an autokeratometer (Haag-Streit Lenstar), and a hybrid topographer (Tomey TMS-5), the OCULUS Pentacam has the disadvantage of high measuring noise on posterior corneal curvature. Meanwhile, the highest precision for planning toric IOL power and axis was achieved by combining the keratometry and OCT data. In a recent study, Lu et al. found that a novel multicolored spot reflection topographer system could provide high repeatable measurements in (both anterior and posterior) corneal power and astigmatism [38].

3.3 Surgically induced astigmatism

Besides naturally occurring astigmatism, the surgically induced astigmatism (SIA) is also an important factor for the appropriate option of a toric IOL. The SIA could be influenced by position and length of incisions [39]. Meanwhile, to achieve minimum residual refractive astigmatism for specific patients, the incisions could be determined by the magnitude and axis of preoperative keratometric astigmatism [4]. The application of femtosecond laser-assisted cataract surgery (FLACS) could minimize SIA.

4. IOL power calculation

An accurate biometry is a precondition not only for toric IOLs but also for regular IOL power calculation. The axial length may be measured by either ultrasonic biometry or optical systems, and SRK/T, Holladay 2, Hoffer Q, and Barrett formula are recommended to be used to calculate sphere power. Nguyen et al. adjusted the power of an existing hydrophobic acrylic IOL by a femtosecond laser [40], which is definitely a promising idea.

There are several toric calculators available for surgical planning that have been developed to predict postoperative cylinder power, such as Barrett toric calculator [41], Holladay toric calculator, and Alcon toric calculator(the revised Alcon toric calculator is a derivation of the Barrett calculator). In general, an ideal IOL power calculation formula should take into account the posterior corneal curvature, the effective lens position (ELP), as well as the SIA. And there are a few formulas available such as Abulafia-Koch linear regression formula [42], Baylor nomogram (a method from Koch) [43], Barrett formula, Abulafia-Koch formula, etc.

4.1 IOL power calculation considering posterior cornea

A few online toric IOL calculators have been revised to take into account the contribution of the posterior cornea in IOL power calculation, but it proved itself valuable. The Baylor nomogram which incorporates the posterior corneal curvature has been observed to be more precise than traditional Alcon and Holladay toric calculator without posterior corneal astigmatism compensation [44]. However, the revised AcrySof toric calculator incorporates the Barrett toric algorithm, which takes into account both the ELP and the posterior corneal astigmatism, and had better predictability than the Baylor nomogram as well as Holladay and traditional Alcon toric calculator [44]. Other toric IOL calculators such as TECNIS calculator also incorporate posterior corneal astigmatism compensation.

4.2 IOL power calculation considering ELP

Failing to consider the anterior chamber depth and cornea thickness may result in inaccurate calculations, especially in eyes with extremes of axial lengths [45]. As mentioned above, the revised AcrySof online toric calculator and iTrace toric planner takes into account the ELP [14, 46]. The TECNIS calculator incorporates the anterior chamber depth based on the axial length and keratometry values [46], and the Holladay formula incorporates the ELP in its calculations.

4.3 Intraoperative wavefront aberrometry

Intraoperative wavefront aberrometry is increasingly being used to estimate the toric IOL power and axis of placement based on the aphakic refraction, especially in post-refractive surgery cases. A recent study reported only a mean error of 0.43 ± 0.33 D with Optiwave Refractive Analysis (ORA; WaveTec Vision Systems Inc., CA, USA) in post laser-assisted in situ keratomileusis (LASIK) cases undergoing toric IOL implantation, which were more accurate than those obtained by the standard SRK/T formula and the online ASCRS calculator.

5. Surgery techniques

Many issues, such as accurate marking technique, clear corneal incisions, intraoperative alignment of the toric IOL, capsulorhexis, and IOL centration, play a significant role in achieving optimal outcomes.

5.1 Marking techniques

Preoperative reference and axis marking techniques could be broadly categorized as manual methods, image-guided systems, and intraoperative aberrometry-based methods.

The three-step manual technique is at present most commonly used [47], which is fairly accurate [48]. The first step is preoperative marking of the reference axis, which is commonly placed in the horizontal 3’o and 9’o clock positions. The second step is intraoperative alignment of the reference mark. The marking may be performed with a skin marking pen or needle. The patient should be sitting erect in a straight-ahead gaze while marking the reference axis. A change in patient position from sitting to supine may induce significant cyclotorsion; studies reported up to 28° of cyclotorsion in 68% of cases [49]. The manual marking methods have been limited by smudging of the dye, irregular, and broad marks.

Image-guided systems and intraoperative aberrometry have advantages compared with manual marking. The image-guided system based on the concept of landmarks to place the axis marks [50], which could be iris crypts, nevi, brush fields, etc. The systems capture a preoperative reference image and calculated the location of these marks and their distance in degrees from the target IOL axis. Then the system generated a final plan which provides simple angular directions from each reference mark to the planned axis of IOL placement.

There are a few image-guided systems at present such as CALLISTO Eye and Z Align (Carl Zeiss Meditec, Jena, Germany), VERION (Alcon, Fort Worth, Texas), TrueGuide (TrueVision 3D Surgical System, Santa Barbara, Calif), Osher Toric Alignment System (OTAS, Haag-Streit, Koeniz, Switzerland), and iTrace System (Tracey Technologies, Houston, Tx). Besides alignment, image-guided systems also contribute to planning the incisions, capsulorhexis size, and optimal IOL centration.

5.2 Intraoperative toric IOL alignment

Intraoperative IOL positioning is the key procedure to sustain rotation stability. During IOL alignment, the IOL should be left about 3–5° anticlockwise of the final desired lens position, followed by complete OVD removal and hydration of the wounds. Most open-loop IOLs can be rotated only clockwise, and a complete re-rotation will be needed if the IOL rotates further clockwise of the target axis.

The image-guided systems and intraoperative aberrometry could be definitely more useful than manual alignment. As mentioned above, the image-guided systems capture a preoperative reference image and an intraoperative image and then match the two images with respect to each other using landmarks. During the operation, a graphic overlay is then superimposed on the surgical field along the target axis, which provides a guide for toric IOL alignment. The image-guided systems and intraoperative aberrometry have improved the precision of toric IOL alignment, with <5° of deviation from the intended axis in the majority of cases.

Compared with manual marking, Elhofi et al. had observed more precise alignment with VERION image-guided system [51], which offers comprehensive astigmatism management, the incision location optimization, toric IOL power calculation, as well as decreasing SIA.

However, Solomon et al. claimed that, compared with the surgeon’s standard of care, the use of the VERION combined with intraoperative aberrometry (Optiwave Refractive Analysis system with VerifEye) did not significantly optimize the outcomes [52]. The accuracy of CALLISTO Eye is also very effective [53], and it also assists in planning the position of limbal relaxing incisions.

6. Complications

Postoperative toric IOL misalignment is the major complication after toric IOL implantation. Toric IOL misalignment could harm visual quality. In a recent experimental study, 5° IOL axis rotation from the intended position determined a decay in the image quality of 7.03%, 10° of IOL rotation caused 11.09% decay, and 30° rotation caused 45.85% decay [54].

Toric IOL misalignment may be attributed to three factors: (1) inaccurate preoperative prediction of the axis of IOL alignment; (2) inaccurate intraoperative alignment; and (3) postoperative IOL rotation. IOL rotation may be observed as early as 1 hour after surgery, and a majority of rotations occur within the initial 10 days [18]. Early IOL rotation likely results from incomplete OVD removal, whereas late postoperative rotation is influenced by the IOL architecture, design, and axial length. In a recently published case report, the toric IOL was rotated more than 115° shortly after a neodymium: YAG (Nd:YAG) laser posterior capsulotomy [55].

Rotational stability of the IOL varies with design and material and strength of IOL capsular bag adhesions. Maximum rotational stability has been observed with hydrophobic acrylic lenses, followed by Hydrophobic acrylic, hydrophilic acrylic, PMMA and silicone. Loop haptic IOLs are better than plate-haptic IOLs on postoperative rotation stability when using silicone IOL, but they are similar when using acrylic IOL. A study of AT TORBI 709 M, which had one-piece hydrophilic acrylic with hydrophobic surface and a supporting four-haptic design, had rotation of more than 5° in 10% cases in 6 months [56]. Another study of AT TORBI 709 M reported 13% eyes had rotation of more than 10° [57], while another study reported 100% rotation of more than 10° [58]. Scialdone et al. found similar results in rotation stability between AT TORBI 709 M and AcrySof toric IOLs [59]. A long-term of 2-year study of AcrySof toric IOLs (hydrophobic acrylic IOL with Flexible loop haptic) reported postoperative rotation of more than 10° in 1.68% eyes, more than 5° in 23.3% eyes [18]. A recent cohort study [60] of 1273 eyes showed that AcrySof toric IOL was less likely to rotate, with 91.9% of eyes rotated 5° in AcrySof toric IOL eyes compared with 81.8% in TECNIS Toric IOL eyes (P < 0.0001); rotation 10° (97.8% Acrysof vs. 93.2% TECNIS, P = 0.0002) and 15° (98.6% Acrysof vs. 96.4% TECNIS, P = 0.02). Furthermore, a hydrophilic IOL with C-flex design (Rayner 600S IOL) was reported to have excellent rotational stability: average 1.83° ± 1.44° at 6 months and no lens rotated more than 5° [61].

In cases with more than 10° of rotation, realignment of the toric IOL is needed [62]. In a study by Oshika et al., 6431 eyes are implanted with toric IOLs, and realignment was performed in 0.653% of cases [63]. An early repositioning performed after 1 week of primary cataract surgery had optical outcomes.

IOL tilt could also induce astigmatism: tilting toric IOLs aligned at 180° would decrease with-the-rule astigmatism, bringing in undercorrection, while aligned at 90° increased against-the-rule astigmatism, bringing in overcorrection [64].

Meanwhile, LASIK, customized surface ablation, or femtosecond laser-assisted intrastromal keratotomies could also be used to correct residual astigmatism [65]. Some toric rotation check, such ashttps://www.astigmatismfix.com/, could help determine the amount of IOL rotation, and the expected residual refraction. When the large residual cylinder not amenable to correction by rotation alone or refractive surgery, an IOL exchange, piggyback IOLs procedures may be considered.

7. Multifocal toric IOLs

Toric designs are even more required in multifocal IOLs [66] because patients undergoing multifocal IOLs may not tolerate residual astigmatism of <1 D, and multifocal IOLs without toric design perform best with less than 0.75 D of cylinder [67].

In previous studies [68, 69, 70, 71, 72], toric multifocal IOLs achieved good visual performance, with UDVA better than 20/40 in more than 97% of patients, uncorrected near visual acuity better than 20/40 in 100% of patients, spectacle independence in more than 80% of patients, and residual refractive astigmatism lower than 0.50 D in 38–79% of patients. Toric trifocal IOLs such as a trifocal spherical hydrophilic IOL (FineVision POD F) [73] also showed great performance.

But on the other hand, the selection of multifocal toric IOLs should be more restricted than monofocal toric IOLs, especially for the following candidates: (1) patients who had unrealistic expectations of visual quality when having related ocular comorbidities; (2) patients who may not tolerate dysphoric symptoms such as glare and halos; and (3) patients who had specific contraindications for multifocal IOLs, such as abnormal κ or α angle, etc. Thus, a comprehensive ocular examination should be undertaken to rule out any ocular comorbidities that may interfere with the postoperative outcomes.

8. Special cases

Normally, cases with irregular astigmatism, corneal ectatic disorders, post-refractive surgery, post-keratoplasty, and high myopia are not ideal candidates for toric IOL implantation, partly because they are unlikely to achieve complete refractive correction with toric IOLs. However, the amount of astigmatism may be partly reduced, decreasing spectacle dependence. And such cases may be considered for surgery after adequate counseling. As a consequence, the applications of toric IOLs are expanding to include special cases such as pellucid marginal degeneration [74, 75], mild keratoconus with cataract [76], astigmatism after keratoplasty [77, 78, 79, 80], and high astigmatism [81]; even toric trifocal IOLs were used in high astigmatism cases [82]. In general, the indications of toric IOL are still controversial and expanding.

9. Conclusions

The outcomes after toric IOL implantation are influenced by a few factors: accurate astigmatism measurement, marking techniques, intraoperative alignment, and postoperative care. The importance of posterior corneal curvature is increasingly being recognized, and advanced toric calculators and fumulars that account for both the anterior and posterior corneal power are becoming the standard of care. The image-guided systems and intraoperative aberrometry could provide a markless IOL alignment and optimize incisions, capsulorhexis size, and optimal IOL centration. New toric IOLs with superior design are still being looked forward although they have already achieved great performance.

Acknowledgments

Thanks to my wife, Wenzhe Li, who dedicated her precious time to help when I finish this manuscript. Also, this research was supported by the Chinese Capital Clinical Features Key Project—Clinical Application on Chinese Keratoprosthesis (Project No: Z161100000516012), National Natural Science Foundation of China (Grant No. 81770887), and National Natural Science Foundation of China (Grant No. 81670830).

Conflict of interest

The authors have declared that no competing interests exist.

Other declarations

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

Ddiopter
IOLintraocular lens
RCTrandomized control trial
UDVAuncorrected distance visual acuity
SIAsurgically induced astigmatism
FLACSfemtosecond laser-assisted cataract surgery
ELPeffective lens position
LASIKlaser-assisted in situ keratomileusis
ORAOptiwave Refractive Analysis
UNVAuncorrected near visual acuity

© 2019 The Author(s). Licensee IntechOpen. This chapter is distributed under the terms of the Creative Commons Attribution 3.0 License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

How to cite and reference

Link to this chapter Copy to clipboard

Cite this chapter Copy to clipboard

Zequan Xu (November 25th 2019). Toric Intraocular Lenses, Intraocular Lens, Xiaogang Wang and Felicia M. Ferreri, IntechOpen, DOI: 10.5772/intechopen.90153. Available from:

chapter statistics

102total chapter downloads

More statistics for editors and authors

Login to your personal dashboard for more detailed statistics on your publications.

Access personal reporting

Related Content

This Book

Next chapter

Pseudophakic Dysphotopsia

By Emely Zoraida Karam Aguilar

Related Book

First chapter

Current Applications of Optical Coherence Tomography in Ophthalmology

By Nadia Al Kharousi, Upender K. Wali and Sitara Azeem

We are IntechOpen, the world's leading publisher of Open Access books. Built by scientists, for scientists. Our readership spans scientists, professors, researchers, librarians, and students, as well as business professionals. We share our knowledge and peer-reveiwed research papers with libraries, scientific and engineering societies, and also work with corporate R&D departments and government entities.

More About Us