Open access peer-reviewed chapter

Ceramic Biomaterials for Dental Implants: Current Use and Future Perspectives

By Federico Mussano, Tullio Genova, Luca Munaron, Maria Giulia Faga and Stefano Carossa

Submitted: October 17th 2015Reviewed: February 24th 2016Published: August 17th 2016

DOI: 10.5772/62701

Downloaded: 1083

Abstract

Although titanium implants have the longest traceable record of predictable clinical performance and by far the widest diffusion in the market, some drawbacks have been recently pointed out. Titanium is not a completely bioinert material, since it may elicit allergenic reactions and is capable to diffuse not only within the adjacent tissues, which is proven by the elevated concentrations found in peri-implant bone and regional lymph nodes, but also systemically. Ceramic materials for oral application have been used for 40 years. Presently, the material of choice is yttria-stabilized tetragonal zirconia, which presents excellent mechanical and tribological properties together with biocompatibility. Concerns remain about the long-term durability of the material, owing to the report of in vivo failures that were caused by the low-temperature degradation of zirconia. To address this issue, research has developed improved oxide-based materials such as alumina–zirconia composites along with non-oxidic ceramics such as silicon nitride.

Keywords

  • surface roughness
  • dental implants
  • ceramic materials
  • surface modifications
  • interface

1. Introduction

Modern oral implantology has been based on titanium since the research line originated by Brånemark’s first discovery and subsequent experiments [1]. Titanium implants have the longest traceable record of predictable clinical performance with a cumulative success rate of 98.8% for 15 years [2]. High biocompatibility, favorable tissue response and adequate strength and corrosion resistance rendered titanium implants widely diffused in the market. The number of dental implant brands grew from 45 systems in 1988 [3] to 600 systems produced by 146 manufacturers in 2008 [4]. Currently, worldwide, there are more than 350 dental implant manufacturers producing an estimate of 1600 different systems, 98% of which are titanium implants. Titanium, however, is no longer considered a completely bioinert material, instead it might be an allergen as reported by several studies [58]. Elevated titanium concentrations have been found in the vicinity of oral implants [8], in regional lymph nodes [9], serum and urine [10], which is potentially hazardous to human body. Besides these issues, some dental patients are metal-phobic and demand to be treated solely with metal-free dental implants [11].

Only recently, truly viable alternative materials were proposed to titanium, although the first ceramics for oral applications dated back to the 1970s. Historically, indeed, high-density, high-purity aluminum oxide (alumina) was chosen for dental implant manufacturing, as it combined excellent corrosion resistance, good bio-compatibility, high wear resistance, and high strength. Despite these promising features, the material was brittle and prone to fracture under unfavorable load. Thus, the positive preclinical and clinical outcomes of the first studies could not prevent alumina implant systems to be withdrawn from the market [12]. Research and manufacturing technology have greatly improved the offer of bio-ceramics, thanks to the introduction of yttria-partially stabilized tetragonal zirconia polycrystals (Y-TZP), whilst a possible future use of alumina zirconia composites and silicon nitride–titanium nitride composites may further expand the offer of reliable devices on the market. Three distinct sections of the present chapter are dedicated to each of these materials. Specifically, the literature regarding zirconia was thoroughly revised in Section 2, whilst some novel data of our group are exposed and discussed in light of and along with previous work as for alumina zirconia composites (Section 3) and silicon nitride–titanium nitride non-oxidic ceramics (Section 4).

2. Yttria-partially stabilized tetragonal zirconia polycrystals (Y-TZP)

2.1. Material features

Zirconia (ZrO2) is a crystalline dioxide of zirconium: as thoroughly reviewed elsewhere [12], unalloyed zirconia can assume three crystallographic forms depending on the temperature, at ambient pressure. At room temperature and upon heating up to 1170°C, the symmetry is monoclinic (P21/c). The structure becomes tetragonal (P42/nmc) between 1170 and 2370°C and cubic (Fm3m) above 2370°C and up to the melting point [13]. Upon cooling at ~950°C, during the transformation from the tetragonal (t) phase to the monoclinic (m) phase, a substantial increase in volume (~4.5%) occurs, which is sufficient to lead to catastrophic failure. By alloying pure zirconia with stabilizers such as calcium oxide (CaO), magnesium oxide (MgO), yttrium oxide (Y2O3), or cerium oxide (CeO2), the tetragonal structure is maintained, even at room temperature, and the stress-induced t → m transformation is controlled, efficiently arresting crack propagation [14, 15]. Indeed, when a crack develops, tetragonal grains convert immediately to monoclinic form. The propagation of the crack develops sufficient stress within the tetragonal structure to transform also the grains around the crack to stable monoclinic form. Thus, the expansion volume of zirconium dioxide crystals produces compressive stress around the crack and prevents further propagation of crack [1618]. This mechanism is known as transformation toughening and is influenced by temperature, vapor, particle size, micro- and macrostructure, and concentration of stabilizing oxides [19].

Yttria-stabilized zirconia (Y-TZP) [20] is endowed with excellent mechanical, and tribological properties together with biocompatibility and rightly represent a good choice for preparing dental implants. As yttria decreases the driving force of the t-m transformation [21, 22], biomedical grade zirconia are usually stabilized with 3 mol% yttria (Y2O3) (hence 3Y-TZP) [16]. The salient mechanical properties of Y-TZP are reported in Table 1, but it is noteworthy that the Weibull modulus is strongly dependent on the type of surface finish and the processing conditions [23].

Notwithstanding the excellent mechanical properties of Y-TZP [19, 24], recent reports of in vivo failures [2527] have questioned the long-term stability of the material. The low-temperature degradation (LTD) of zirconia [22, 2831], also known as aging process, plays here a fundamental role. Involving the t→m transformation, LTD can be favored, even at room temperature, by the penetration of water radicals into zirconia lattice, thus leading to the formation of tensile stresses in zirconia surfaces. The activation barrier for the transformation is lowered, and the phase transition is promoted. The main consequences of this aging process include surface degradation with grain pullout, microcracking, and strength degradation. As reported by Cattani-Lorente et al. [32] also, Young’s modulus and hardness of Y-TZP bars were reduced by 30%, when they were subjected to hydrothermal cycling. The increase of monoclinic–tetragonal phase ratio was associated with microcracking and resulted responsible for the decline in mechanical parameters [32].

To control the aging phenomenon, several factors can be taken into account: from the obvious use of stabilizers to the modulation of residual stress [33]. Likewise, adjusting crystal size and removing impurities during manufacturing was proposed with the same anti-aging scope [34]. Interestingly, surface finishing could affect the aging kinetics of 3Y-TZP, according to Deville et al. [35]. More precisely, rough polishing produced a compressive surface stress layer beneficial for the aging resistance, whilst smooth polishing lead preferential transformation nucleation around scratches, due to elastic/plastic damage tensile residual stresses.

In an extensive review of his, Jerome Chevalier concluded that “although in the 1990s, 3Y-TZP ceramics were considered very promising materials for biomedical applications, long-term follow-up is needed to address the critical problem of aging in vivo. Moreover, most zirconia implants were processed at a time when aging was not yet fully understood. Methods to assess a priori the aging sensitivity of a given zirconia ceramic have been developed and should lead to safer implants. In the meantime, new zirconia or zirconia-based materials that overcome the major drawback of the standard 3Y-TZP are now available” [22].

2.2. Manufacturing methods

Hot isostatic press (HIP) is the most common method used for preparing zirconia dental implant. By subjecting encapsulated powder, or sintered yet porous parts, to inert gas at isostatic pressure at a high temperature, HIPing is deemed an excellent method to obtain high-density homogenous products [36]. HIPing enables the application of an equally distributed pressure in all directions resulting in greater material uniformity and higher strength [2]. HIPing of Y-TZP enhances the strength, eliminates fracture sources such as pores, and reduces the aging phenomenon [37]. The preparation entails many steps as summarized in Figure 1. Briefly, Y-TZP blocks are presintered at temperatures below 1500°C to reach a density of at least 95% of the theoretical density. Hot isostatic pressing (HIP) is applied to the blocks at temperatures between 1400 and 1500°C under high pressure. A HIP cycle after sintering is recommended to achieve a full density close to the theoretical values (d = 6.1 g/cm3 = 100% dense). Since HIPing changes the color of Y-TZP into dark-grey, a heat treatment in air is usually performed to restore the material whiteness by oxidation, prior to be machined using a specially designed milling system. Because of the high hardness of fully sintered Y-TZP, the milling system is to be particularly robust [3840].

Figure 1.

Manufacturing process for Y-ZPT.

The relatively recent and yet pervasive introduction of computer-aided design/computer-aided manufacturing (CAD/CAM) technology has provided dentistry with an alternative to HIPing [42]. Usually, dental CAD/CAM systems recur to partly sintered yttria-stabilized tetragonal zirconia polycrystal (Y-TZP) blanks. The use of this partly sintered state of the Y-TZP ceramic renders the milling process faster and reduces the tools wear, compared to systems employing densely sintered blanks (HIP process). Of course, the final sintering shrinkage must be taken into account during the CAD phase by enlarging the shapes before milling, whilst this compensation is not necessary with the HIPed Y-TZP blanks that are directly ground to the desired dimensions [5]. The salient mechanical properties of Y-TZP subjected to the two manufacturing work-flow described above are compared in the following [43] table (Table 1), along with pressed and sintered polycrystalline α-alumina.

HIPed Y-TZPPressure-less sintered Y-TZPPressed and sintered polycrystalline α-alumina
Density (g/cm 3 )6.164
Microhardness (Vickers)1000–13001000–12002300
Young’s modulus (GPa)200200420
Bending strength (MPa)1200800500
Toughness KIC (MPa m1/2 )9–109–104

Table 1.

Values refer to Duraccio et al. [12].

2.3. Biological properties

In vitro experiments on different cell lines, in vivo studies on animals and clinical studies on humans supported the safety and the high level of biocompatibility of zirconia. In a preliminary in vitro investigation [44], one-piece zirconia implants were proven to possibly fulfill the biomechanical requirements for anterior teeth restoration. In addition, the mean fracture strength of zirconia implants was investigated after chewing simulation and it was found to be within the limits of clinical acceptance. However, the preparation of a one-piece zirconia implant prior to prosthetic finalization may significantly compromise fracture strength. Therefore, long-term clinical data were deemed necessary before one-piece zirconia implants could be recommended for clinical practice [45]. For the same reason, two-piece zirconia implants were considered clinically inadequate due to the increased risk of fracture at the implant head level [46].

Evidence from in vitro studies on osteoblasts supported the possible favorable response of zirconia ceramics in vivo [47, 48]. When implanted in bone or soft tissues, these materials could elicit no inflammatory reactions, nor fibrous encapsulation, according to Hisbergues et al. [41]. Interestingly, Scarano et al. [49] reported the osseointegration of unloaded zirconia implants inserted in rabbit bones without any signs of inflammation or mobility. The possible role of surface roughness was investigated by comparing the removal torque of machined zirconia implants to roughened ones [50]. Notably, the roughened implants performed better than the smooth ones and behaved similarly to the oxidized titanium implants used as control. Loaded zirconia implants were studied and compared to titanium implants by Kohal and co-workers [51], who could find no significant difference in the osseointegration level between the two groups. Akagwa et al. [52] reported a similar bone to implant interlock in loaded and unloaded zirconia implants, but a crestal bone loss higher around the former group. In favor of the clinical use of Zirconia, it must be cited its maintenance of bending strength of over 700 MPa after immersion in 95°C saline solution for over 3 years [53]. Furthermore, zirconia blanks did not show any significant mechanical detriment even after being embedded in the medullary cavity of the tibia of rabbits for 30 months.

Scarce are the clinical studies dedicated to the long-term performance of zirconia implants. The short follow-up period and the often small sample size hinder their quality of evidence, so that Andreiotelli and coauthors [29] could only include three retrospective cohort studies on one-piece zirconia dental implants in their systematic review, reaching in total 231 patients and 416 implants. The studies by Mellinghoff et al. [54] and Oliva et al. [55] investigated, respectively, 189 and 100 zirconia implants and estimated 1-year survival rates of 93 and 98%. Almost all of the failures occurred during the healing phase, as only one implant failed after prosthetic reconstruction due to fracture. Lambrich and Iglhaut [56] observed 127 zirconia and 234 titanium implants for a mean period of 21.4 months. Notably, in this study, the survival rate of zirconia implants was similar to that of titanium in the mandible (Y-TZP = 98.4% vs. Ti = 97.2%), whilst differed considerably in the maxilla (Ti = 98.4% vs. Y-TZP = 84.4%). Again, all failures occurred during the healing phase owing to increased implant mobility. These findings are consistent with the paper by Depprich et al. [57], where the survival rate of zirconia implants obtained from 17 clinical studies was between 74 and 98% after 12–56 months. Payer et al. [58] followed up for 2 years 19 immediately loaded zirconia implants, reporting a 95% survival rate, as determined clinically and radiographically. These results are in accordance with Oliva et al. [59] who determined the same survival rate at 5 years in 371 patients who received 831 one-piece zirconia implants. Kohal et al. [60] found that immediately restored one-piece zirconia implants have 1-year cumulative survival rate comparable to titanium counterparts. In conclusion, the clinical data currently available for Y-TZP implants may not be sufficient to recommend their routine clinical use. Zirconia, however, may have the potential to be a successful implant material, although this is as yet not fully supported by present investigations and further good-quality research is needed.

3. Zirconia-toughened alumina (ZTA) and alumina-toughened zirconia (AZT)

3.1. Background

The demand of structural ceramics has led to an increased interest in Alumina–Zirconia composites for biomedical [21, 61] and dental implant application [62, 63]. Two composite materials can be prepared: ZrO2 reinforced with alumina particles, which is denominated alumina toughened zirconia (ATZ), and Al2O3 reinforced with zirconia particles, which is known as zirconia-toughened alumina (ZTA). Thus, higher fracture values can be reached if compared with the monophase ceramics [64] (Table 2).

ATZZTA
Hardness (GPa)15.3 ± 0.921.3 ± 1.5
Young’s modulus (GPa)245 ± 9363 ± 5
Bending strength (MPa)633 ± 127441 ± 24
Toughness KIC (MPa m1/2)7.1 ± 0.13.9 ± 0.05

Table 2.

Physical and mechanical properties of ATZ and ZTA data are extracted from Faga et al. [85].

These composites benefit from combining the characteristics of Alumina, namely the high hardness and stiffness, with the superior strength and toughness of Zirconia, which improves remarkably the resistance to crack growth [65]. In addition, alumina increases the hydrothermal stability of tetragonal Zirconia phase [65, 66], owing mainly to the formation of a stiff matrix capable to keep the Zirconia particles in a metastable tetragonal state [67], thus acting as mechanical stabilizer. The only commercially used ATZ oral implant was tested both statically and dynamically for its fracture resistance in different simulated oral conditions with satisfying results [68].

The main features of the aforementioned implant are reported in the following table (Table 3).

ATZ
Density (g/cm 3 )5.5
Average grain size (μm)<0.5
Microhardness (Vickers)1000–1200
Young’s modulus (GPa)220
Bending strength (MPa)2000
Toughness KIC (MPa m 1/2)8

Table 3.

Physical and mechanical properties of ATZ values refer to Spies et al. (2015) – [68]

Very recently, a complete powder injection molding process was developed to fabricate cylindrical ZTA parts recurring to a binder system made of high-density polyethylene, paraffin wax, and stearic acid. The effects of sintering temperature on shrinkage, relative density, and hardness of the sintered part were taken into account and proved the technology suitable for the production of ZTA parts with sufficient mechanical properties [69]. However, ATZ and ZTA are usually produced through the classic workflows described above for Y-TZP (see Section 2.2).

The favorable mechanical features and the biological safety of different ZTA and ATZ composites have been the object of several studies in the last years [64, 7078]. Whilst ATZ materials show increased mechanical stability [79] and improved aging resistance versus Y-TZP, still they exhibit a certain degree of aging [79]. ZTA materials display much better aging resistance than both monolithic Y-TZP and ATZ [21, 79, 80]. In a recent work by our research group, both ATZ and ZTA were functionalized with two laminins as a preliminary investigation for improving soft tissue healing around implants. The simple adsorption of these two different isoforms was sufficient to induce some of the most important cell kinases in the epithelial cells grown on the surface of the two Alumina–Zirconia composites, supporting the possible advantages of these materials in dental implantology [81]. On this basis, we further studied the behavior of ATZ dental implants treated with a patented hydrothermal process, comparing them to a clinical use titanium surface in a minipig model. Bone healing was assessed through histology and mRNA expression at different time points (8, 14, 28, and 56 days). The most interesting outcome was a statistically significant higher percentage of newly formed bone along ATZ implants, at 56 days, suggesting that the tested material proved to be a promising candidate among the possible ceramic dental implants [82]. Interestingly, by comparing the bone-to-implant contact of moderately roughened ATZ implants (Sa = 1.51 μm) to an anodized titanium standard (Sa = 1.31 μm) in Sprague–Dawley rats, Kohal et al. [83] found that titanium greatly (58%/75%) outperformed the ceramic implant (24%/41%) after a healing period of 14 and 28 days. In addition, at the same time points, the mechanical interlock measured as push-in values increased from 20 to 39 N for titanium and from 10 to 25 N for ATZ. Although the moderately roughened ATZ implants were well accepted in rat bone, their osseointegration process seemed to proceed more slowly than that of anodized titanium.

However, the concerns raised in light of the in vivo data reported by Kohal et al. [51] seemed not to be completely consistent with the promising outcomes of the clinical study conducted by the same research group. Indeed, the cumulative survival rate (94.2%) of one-piece ATZ implants immediately restored with partial fixed prostheses was comparable to that of the loaded titanium implants, in a human clinical trial involving 40 patients after 3 years of observation. In addition to the marginal bone loss (0.79 mm), several soft tissue parameters and patient-reported outcome measures were evaluated suggesting the potential of ATZ for clinical utilization [84]. Notwithstanding the clinical use, little information is still available about the ideal surface treatment that a ceramic dental implant should receive. To better understand whether roughness or hydroxyapatite precipitation capability were more likely to be efficient in terms of surface modifications, we designed a simple in vitro pilot study.

3.2. Material and methods

3.2.1. Sample preparation

Two high purity, ready-to-press powders were used to produce the ATZ (ZrO2-20 wt% Al2O3, TZ-3Y20AB, Tosoh, Japan) and ZTA (Al2O3-16 wt% ZrO2, Taimicron, Taimei, Japan) samples. As reported elsewhere [85], specimens were prepared through linear pressuring at 80 MPa followed by cold isostatic pressing at 200 MPa. The process parameters for sintering were as follows: heating up to 700°C at a rate of 50°C/h, followed by a 2-h dwell; heating up to 1500°C at a rate of 100°C/h, followed by a 2-h dwell. The resulting fully dense materials were 12-mm disks with thickness ranging between 4 and 5 mm.

As reported in the diagram below, both ATZ and ZTA discs were either mirror polished with diamond suspension in ethanol with decreasing granulometry to the final surface roughness of <1 micron. Also as-fired samples were used to evaluate the influence of the surface roughness on the biological response. Subsequently, the samples were either bioactivated with phosphoric acid under hydrothermal conditions (patent numbers: TO2012A000029 and PCT/IB2013/050425) or left untreated (Figure 2).

Figure 2.

Schematic representation of samples treatments.

3.2.2. Surface characterization

Microstructure was studied by means of a scanning electron microscope Zeiss EVO 50 with energy dispersion spectroscopy analyzer for elemental composition detection. Surface roughness was measured with a non-contact profilometer, Talysurf CCI 3000A. The tests were performed in an air-conditioned laboratory, where temperature is kept at 20°C, on a representative surface of 90 μm2. To quantify the amount of protein adsorbed, fetal bovine serum (FBS) was diluted in phosphate-buffered saline (PBS) at a concentration of 2% and was used to incubate the samples at 37°C for 30 min. After two wash in PBS, the adsorbed protein was eluted from the disks using Tris Triton buffer (10 mM Tris (pH 7.4), 100 mM NaCl, 1 mM EDTA, 1 mM EGTA, 1% Triton X-100, 10% Glycerol, and 0.1% SDS) for 10 min. Finally, the total protein amount was quantified using Pierce™ BCA Protein Assay Kit (Life Technologies, Milan, Italy) following the manufacturer’s instructions.

3.2.3. Biological response

Pre-osteoblastic murine cells MC3T3-E1 (ECACC, Salisbury, UK) were used to characterize the biological response in vitro. Cells were maintained in alpha MEM supplemented with 10% FBS (Life Technologies, Milan, Italy), 100 U/ml penicillin, 100 mg/ml streptomycin, under a humidified atmosphere of 5% CO2 in air, at 37°C. To prevent contact inhibition, cells were always passaged at subconfluency. When required, to differentiate MC3T3 cells, the culture medium was supplemented with 10 mM ß-glycerophosphate and 50 ug/ml ascorbic acid.

To examine cell morphology, MC3T3 cells were seeded at a concentration of 5000 cells/well in a 24-well plate. After 1 day, cells were fixed in 4% paraphormaldheyde in PBS. Rodamine–Phalloidin and Dapi (Life Technologies, Milan, Italy) were, respectively, used to stain cytoskeleton and cell nuclei, thus evaluating cell adhesion and morphology.

Alkaline phosphatase activity was quantified using the Alkaline Phosphatase Assay Kit (Abcam, Cambridge, UK). Following the manufacturer’s instruction, the OD was measured at a wavelength of 405 nm. The calcium deposed within the extracellular matrix was quantified colorimetrically through the Calcium Assay Kit (Cayman Chemical, Michigan, USA). Absorbance of the lysates was measured at 570 nm.

Data were analyzed recurring to GraphPad Prism6 (GraphPad Software, Inc., La Jolla, CA, USA). Each experiment was repeated at least three times. Statistical analysis was performed using the Student t-test. A p value of <0.05 was considered significant.

3.3. Results and discussion

The success of dental implants is directly related to the bone implant interlock, which can be experimentally evaluated in animal living bone, by histomorphometry and/or biomechanical testing [50]. A moderately rough surface topography is known to positively affect the interfacial tissue reaction [86]. Surface modification of zirconia and its composites is, however, challenging. Among the roughening techniques used to attain proper bone–implant interfaces, it is convenient to remember the apposition of sintering particles, nano-technology, sandblasting and acid etching, and laser technology [50, 8790]. In recent animal studies, in vivo evidence was found that alumina-toughened-zirconia is a suitable candidate for dental implantology [82], which was further supported by very recent clinical data at the University of Freiburg [84, 91]. Following Dohan Ehrenfest’s classification [92], surface roughness was moderate (Sa = 1.51 μm) in case of Ziraldent implants or very high (Sa = 5.4 μm) for our research group. Here, the microstructure of ATZ and ZTA was determined by SEM (Figure 3).

The materials show an almost defect-free surface, with a homogeneous distribution of both zirconia and alumina. Submicrometric grains are present in both composites, the darkest representing alumina phase. It is noteworthy that similar dimensions can be observed for alumina and zirconia grains only in ATZ material, whilst ZrO2 growth is inhibited by the predominant alumina content in ZTA composite.

Figure 3.

Scanning electron micrographs of ATZ (A) and ZTA (B).

Surface roughness was measured via profilometry (Table 4). As it can clearly be seen, polished and as-fired samples were, respectively, endowed with a very smooth and a highly rough surface, according to the expected values.

ATZZTA
Mirror polishedAs firedMirror polishedAs fired
Sa (μm)0.0439900.3764920.0511740.485283
SD0.0036040.0189700.0055540.079454

Table 4.

Surface roughness measured via profilometry.

Interestingly, from Figure 4, it can be inferred that the only condition capable to affect significantly the protein adsorption was surface roughness. No statistically significant difference was found among materials (ATZ vs. ZTA) or chemical treatment (hydrothermal cycle present + vs. absent -), although a trend in facilitating protein adsorption could be noted in roughened-treated surfaces.

Figure 4.

Quantification of adsorbed bovine serum albumin (BSA) by samples, as measured trough BCA assay (see Methods). The rough surfaces significantly increase the amount of adsorbed proteins.

The same trend described for protein adsorption was observed in the other cell-based assays whether they were focusing on the early cell response as in the focal adhesion density (Figure 5), or they were dealing with intermediate and late stages of osteogenic differentiation in vitro such as alkaline phosphatase activity (Figure 6A) and calcium deposition within the extracellular matrix (Figure 6B).

Figure 5.

Morphology of MC3T3 cells seeded on different surfaces and stained with phalloidin–rhodamine and DAPI to visualize, respectively, the cytoskeleton and nucleus (see Methods). (A) Quantification of focal adhesion density measured by normalizing the number of focal adhesions on cell area (see Methods). (B) The rough surfaces significantly increase the density of focal adhesion.

Figure 6.

Colorimetric quantification of ALP activity (A) and calcium deposition (B) (see Methods). The rough surfaces significantly increase the level of either ALP activity (A) and calcium deposition (B).

Although surface chemistry is known to play a role in cueing the biological systems [81], the present experimental data showed that roughened surfaces were more efficient in inducing an osteogenic response in vitro independently of the application of the chemical treatment. In other terms, roughness per se seemed to overpower the effect of the chemical treatment which was deemed bioactive on the ground of the Kokubo tests previously performed (i.e., the capacity to induce hydroxyapatite precipitation) [85]. Within the obvious limits of this experimental setting, our results support the importance of roughening modifications over the chemical treatment.

4. Silicon nitride–titanium nitride

4.1. Background

Silicon nitride (Si3N4) is a high-strength and tough ceramic used as a viable implant material [9395]. Since the first clinical trial in 1986 [96], over two decades have passed before the introduction of Si3N4 to the biomedical market of the US and EU. Since 2008, it has been used as a fusion cage for arthrodesis of the cervical and thoracolumbar spine [97], with few adverse reported events [98]. Silicon nitride has been shown to possess favorable cell interaction characteristics [94, 95, 99104], along with bacteriostatic properties [105, 106]. Also, porous or unpolished Si3N4 osseointegrates with adjacent bone [104, 105, 107109].

Silicon nitride derives its strength and toughness through its microstructure, which is composed of asymmetric needle-like interlocking grains surrounded by a thin (<2 nm) refractory grain-boundary glass [110]. Unlike other ceramics, no phase transformation is involved. Thus, similar to alumina, Si3N4 exists as an irreversibly stable phase at room temperature, but an advancing crack must navigate a high energy path through the ceramic, and bridging grains within the crack wake restrict its continued propagation [111113].

Industrial standards have been adopted for Si3N4 composition, processing, and properties [114, 115]. However, sintered Si3N4 is usually machined by hard grinding with diamond tools and the high hardness of Si3N4 makes the production of complex shapes through conventional mechanical machining difficult and expensive. To address this issue, electrically conductive reinforcements, such as TiN, TiC, TiB2, ZrB2, were added to the Si3N4 matrix, generating composites suitable to be wrought by electrical discharge machining (EDM) [116]. The EDM has been introduced with encouraging results, achieving complex shapes from dense electroconductive bulks with high densification [94]. Accurate semi-finished Si3N4–TiN surfaces may be either used as they are, or further finished through diamond polishing [116]. Some preliminary data comparing in vitro the osteogenic behavior of two different surface modifications of a silicon nitride–titanium nitride (Si3N4–TiN) composite are here presented. The two surfaces were, respectively, the very product of the EDM process (henceforth Si3N4–TiN_A) and the result of partial polishing with diamond suspensions (henceforth called Si3N4–TiN_B). For material and methods please refer to Sections 3.2.2 and 3.2.3.

4.2. Results and discussion

A detail of the two silicon nitride–titanium nitride surfaces is reported in Figure 7.

Figure 7.

Surface electron micrographs of Si3N4–TiN_A (A) and Si3N4–TiN_B (B) at 10.000 magnifications.

Si3N4–TiN_A showed an interesting coalesced structure derived from the melting generated during the manufacturing process, whilst, in Si3N4–TiN_B the microstructure of silicon nitride–titanium nitride is clearly appreciable along with the remnants of the peaks after polishing. The tridimensional analysis of Si3N4–TiN_A and Si3N4–TiN_B is graphically depicted in Figure 8, whilst Sa values were, respectively, 2.92 ± 0.07 and 0.88 ± 0.06 μm. Thus, Si3N4–TiN_A resulted rougher than Si3N4–TiN_B.

Figure 8.

Tridimensional graphical representation of Si3N4–TiN_A (A) and Si3N4–TiN_B (B).

MC3T3 cells grew well on both samples. Notably, fluorescent images of adherent cells at 24 h (Figure 9A) clearly show that Si3N4–TiN_A induced a more complex morphology with more tapered shape cells than Si3N4–TiN S, as expected for rougher surfaces. Consistently, a higher density of focal adhesions was quantified on the Si3N4–TiN R surface [117] (Figure 9B).

Figure 9.

Morphology of MC3T3 cells seeded on Si3N4–TiN_A and Si3N4–TiN_B and stained with phalloidin–rhodamine and DAPI to visualize, respectively, the cytoskeleton and nucleus (see Methods). MC3T3 cells seeded on Si3N4–TiN_B display a more complex shape with a lower spreading level than Si3N4–TiN_A (A). Quantification of focal adhesion density measured by normalizing the number of focal adhesions on cell area (see Methods) (B). Si3N4–TiN_B significantly increase the density of focal adhesion.

The osteogenic differentiation was evaluated based on the alkaline phosphatase activity as well as the deposition of bone matrix on the specimens. A statistically significant difference between Si3N4–TiN_A and Si3N4–TiN_B was determined in favor the former, when ALP activity was determined (Figure 10A).

Figure 10.

Colorimetric quantification of ALP activity (A) and calcium deposition (B) (see Methods). Si3N4–TiN_B surface significantly increase the level of either ALP activity (A) and calcium deposition (B).

The rougher surface promoted a greater osteogenic response than the smooth surface in terms of calcium deposition (Figure 10B).

The biological responses induced in MC3T3 cells, a widely diffused osteoblast model, were correlated with the surface roughness, even in this case. The effect of roughness on osteoblast adhesion has been mainly attributed to an increased surface-to-volume ratio that may provide more sites for cell attachment [118]. Consistently, the rougher surface tested (Si3N4–TiN_A) could promote better cell viability, higher density of focal adhesions and more pronounced calcium deposition than the smoother one (Si3N4–TiN_B). Taken together, these data confirmed the biocompatibility of silicon nitride–titanium nitride composites in accordance with the literature, which has indeed so far explored preferably the pristine Si3N4 material [93, 94, 99, 119]. The possible application of surfaces directly obtained by EDM to Si3N4–TiN is therefore noteworthy. Further research should be oriented at investigating the in vivo effects of such surface finishing, as well as the importance of the texture in the pattern recognition operated by cells.

5. Concluding remarks

As stated in previous sections, even though titanium and titanium alloys are the material of choice for dental implants, they are not without drawbacks. Among the possible issues, for instance, the hypersensitivity in allergic patients and some aesthetic concerns deserve attention. To address these problems, ceramics have been introduced to the market in the last decades. Y-TZP was first proposed owing to its biocompatibility, white root-like color and low plaque affinity. More recently, oxidic composites containing variable amounts of zirconium oxide such as zirconia-toughened alumina (ZTA) and especially alumina-toughened zirconia (AZT) were recently considered an improved alternative to Y-TZP. These implants seem very suitable to replace the anterior teeth to avoid the formation of dark shimmer in the presence of thin gingival biotype. However, one-piece ceramic implants may be more difficult to place than two-pieces titanium implants if angulated abutments are required.

Nevertheless, the demand of non-metallic materials endowed with high mechanical features is prompting research and industry to explore also ceramics such as silicon nitride. This non-oxidic material, whose use is almost completely limited to orthopedics in the biomedical field, possesses really promising quality even for dental application. The possibility to dope silicon nitride with titanium nitride, thus rendering it electroconductive, enables a range of manufacturing processes like the electro discharge machining. This opens compelling perspectives in the future as biomaterials are supposed to be increasingly customizable, maneuverable, and adaptable to the particular necessity of the single case, possibly entering the digital work-flow.

© 2016 The Author(s). Licensee IntechOpen. This chapter is distributed under the terms of the Creative Commons Attribution 3.0 License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

How to cite and reference

Link to this chapter Copy to clipboard

Cite this chapter Copy to clipboard

Federico Mussano, Tullio Genova, Luca Munaron, Maria Giulia Faga and Stefano Carossa (August 17th 2016). Ceramic Biomaterials for Dental Implants: Current Use and Future Perspectives, Dental Implantology and Biomaterial, Mazen Ahmad Jawad Amin Almasri, IntechOpen, DOI: 10.5772/62701. Available from:

chapter statistics

1083total chapter downloads

2Crossref citations

More statistics for editors and authors

Login to your personal dashboard for more detailed statistics on your publications.

Access personal reporting

Related Content

This Book

Next chapter

Ceramic Materials as an Alternative to Titanium for Dental Implant Fabrication

By Mobilio Nicola, Mollica Francesco and Catapano Santo

Related Book

First chapter

The Growth of Dental Implant Literature from 1966 to 2016: A Bibliometric Analysis

By Andy Wai Kan Yeung and Wai Keung Leung

We are IntechOpen, the world's leading publisher of Open Access books. Built by scientists, for scientists. Our readership spans scientists, professors, researchers, librarians, and students, as well as business professionals. We share our knowledge and peer-reveiwed research papers with libraries, scientific and engineering societies, and also work with corporate R&D departments and government entities.

More About Us