Abstract

Wireless power transmission (WPT) can provide an alternative for wireless power in implantable medical devices (IMDs). The WPT in implantable medical devices will involve many emerging biomedical topics, such as implantable pacemakers, optogenetic devices, and bio-impedance sensors. To this end, this chapter comprehensively reviews the recent WPT studies for those mentioned above emerging biomedical applications. The specific key components are carried out for those applications. Besides, the operation principle and system design are presented. In conclusion, this chapter’s significance can help evolve reliable implantable device development in the future.

Keywords

  • bio-impedance sensor
  • electrical field coupling
  • implantable medical devices
  • magnetic field coupling
  • optogenetic device
  • pacemaker
  • resonance circuit

1. Introduction

Wireless power transmission (WPT) is an emerging technology to transmit power without wires [1, 2]. It is widely studied in various applications such as consumer electronics charging [3], radio-frequency identification (RFID) system [4], electric vehicle charging [5], biomedical applications [6], etc. Especially in implantable devices, the market benefit of global implantable medical devices achieved a value of US$ 120.5 billion in 2021 by IMARC Group. The market can be expected to reach US$ 168.3 billion by 2027, exhibiting at a compound annual growth rate (CAGR) of 5.54% during 2022–2027. In practice, implantable devices solve today’s problems using advanced medical technologies such as a well-known pacemaker and capsule endoscopy [7, 8]. The pacemaker and capsule endoscopy invention brings new perspectives for cardiovascular disease therapy and direct health diagnosis of the whole small bowel. Although several miniatured battery systems have been developed to provide several years of power supply for implantable devices, a battery replacement would induce patients’ inconvenient and unexpected surgery risks [9, 10]. Recently, the WPT technique has been widely applied in implantable medical devices that are expected to power supply issues in the future [11, 12].

WPT technique can be mainly divided into the coupling and radiative approaches [13, 14], as shown in Figure 1. The WPT coupling can be categorized into magnetic and electric field methods. The radiative approach can be categorized into electromagnetic waves and optical power transmission. In the radiative method (far-field), power is transferred by electromagnetic radiation in terms of microwave, laser, and light. Although radiative techniques can transfer power over longer distances, the large environmental loss-induced transfer efficiency is considerable due to poor air medium. The nonradiative method (near field) utilizes magnetic field and electric field coupling to approach wireless power transfer. The main advantage of nonradiative techniques is high transfer efficiency; however, its technical limitation is that the transfer distance is limited.

Figure 1.

WPT classification, including nonradiative and radiative methods.

The nonradiative method is also called coupling-based WPT, which can be described as the power-transmitting side (primary side) and power-receiving side (secondary side). The coupling type can be classified as a magnetic field (Figure 2a) and electrical field coupling (Figure 2b). Both magnetic and electric field coupling methods can be further categorized as resonance mode, based on LC resonance (Figure 2c and d).

Figure 2.

Magnetic and electric field coupling with and without resonant circuit.

Advertisement

2. Magnetic field coupling without resonance

The magnetic field coupling without resonance type utilizes the principle of electromagnetic induction, as shown in Figure 3. According to Ampere’s law as Eq. (1), the magnetic pulsing (dB/dt) of the primary coil can be generated by time-varying driving current (dI/dt). Where B, μ0, J are magnetic field, permeability, and current density. The electric field E can be induced on the secondary coil, according to Faraday’s law as Eq. (2) [9, 15]. Thus, the charge can be stored via a secondary coil.

Figure 3.

Circuit topology of electromagnetic induction.

×B=μ0JE1
E=B∂tE2

To understand the principle of magnetic field coupling without resonance, Figure 3 shows the topology of the circuit analysis. The mutual inductance, Lm=kL1L2, depends on the distance x between the two coils, where k is the coupling factor (0 < k < 1) between the primary and the secondary coil. Based on Kirchhoff’s voltage law, Eqs. (3) and (4) can be obtained as below [16]:

L1I1+I1R1+LmI2=U1E3
L2I2+I2R2+I2RL+LmI1=0E4

I2 can be expressed as Eq. (5), according to Eq. (4).

I2=I1ωLmωL2jR2+RLE5

The relation between I1 and U1 can be obtained as Eq. (6), according to Eqs. (3) and (5).

I1=R2+RL+L2R1+L1L2+R2+RL+ω2Lm2U1E6

The relation between I1 and I2 can be expressed as Eq. (7), according to Eqs. (5) and (6).

I2=LmR1+L1L2+R2+RL+ω2Lm2U1E7

The ratio of I1 and I2 can be expressed as Eq. (8), according to Eqs. (6) and (7).

I1I2=R2+RL+L2LmE8

To better understand the input impedance and power efficiency of the circuit, a T-type equivalent circuit is used in Figure 4.

Figure 4.

T-type equivalent circuit for electromagnetic induction.

VLm2=LmI1E9
VLm2=LmR2+RL+L2R1+L1L2+R2+RL+ω2Lm2U1E10

According to Eq. (11), Zin2 can be obtained from Eqs. (8) and (9).

Zin2=VLm2I2=LmI1I2=R2+RL+L2E11

Then, Z2 can be further obtained, according to Eq. (12).

Z2=VLm1I1=LmI2I1=ω2Lm2R2+RL+L2E12

Finally, the Zin1 can be acquired, according to Eq. (13).

Zin1=R1+L1+Z2=R1+L1+ω2Lm2R2+RL+L2E13

To understand the power efficiency, the energy loss ratio for the primary-side internal resistance, secondary resistance, and load can be denoted by PR1, PR2, and PRL as Eq. (14).

PR1:PR2:PL=I12R1:I22R2:I22RLE14

The square ratio of I2 and I1 can be expressed as Eq. (15), according to Eq. (8)

I2I12=ω2Lm2R2+RL2+ω2Lm2E15

The power ratio can be rewritten as Eq. (16), according to Eqs. (14) and (15).

PR1:PR2:PL=R2+RL2+ω2Lm2R1:ω2Lm2R2:ω2Lm2RLE16

Finally, the power efficiency can be obtained, according to Eq. (17).

η=PLPin=PLPR1+PR2+PL=ω2Lm2RLR2+RL2+ω2Lm2R1+ω2Lm2R2+ω2Lm2RLE17
Advertisement

3. Magnetic field coupling with resonance

Compared with electromagnetic induction, magnetic resonance coupling utilizes resonance with the series capacitor in primary and secondary sides to approach high efficiency and high power (Figure 5). Based on Kirchhoff’s voltage law, Eqs. (18) and (19) can be obtained as below [16]:

Figure 5.

Circuit topology of magnetic field coupling with resonance.

1C1I1+R1I1+L1I1+LmI2=U1E18
L2I2+R2I2+1C2I2+RLI2+LmI1=0E19

I2 can be expressed as Eq. (20), according to Eq. (18).

I2=LmL2+1C2+R2+RLI1E20

The relation between I1 and U1 can be expressed as Eq. (21), according to Eqs. (18) and (20).

I1=R2+RL+jωL21ωC2(R1+jωL11ωC1R2+RL+jωL21ωC2+ω2Lm2U1E21

The relation between I2 and U1 can be expressed as Eq. (22), according to Eqs. (20) and (21).

I2=Lm(R1+jωL11ωC1R2+RL+jωL21ωC2+ω2Lm2U1E22

The relation between I1 and U1 can be expressed as Eq. (23), according to Eqs. (21) and (22).

I1I2=R2+RL+ωL21ωC2LmE23

When the operating and resonant frequencies are equal on primary and secondary sides, I1 and I2 can be expressed as (25) and (26), according to Eqs. (21), (22), and (24).

ω0=ω1=ω2=1L1C1=1L2C2E24
I1=R2+RLR1R2+RL+ω2Lm2U1E25
I2=LmR1R2+RL+ω2Lm2U1E26

The ratio of I1 and I2 can be rewritten as Eq. (27), according to Eqs. (25) and (26).

I1I2=R2+RLLmE27

To further understand the input impedance and power efficiency of the magnetic field coupling with the resonance circuit, a T-type equivalent circuit is used as Figure 6.

Figure 6.

T-type equivalent circuit for magnetic field coupling with resonance.

VLm2=LmI1E28
VLm2=LmR2+RLR1R2+RL+ω2Lm2U1E29

Zin2 can be obtained as Eq. (30), according to Eqs. (26) and (29).

Zin2=VLm2I2=LmI1I2=R2+RLE30

Then, Z2 can be further obtained as Eq. (31).

Z2=VLm1I1=LmI2I1=ω2Lm2R2+RLE31

Finally, the Zin1 can be acquired, according to Eq. (32).

Zin1=R1+1C1+L1+Z2=R1+1C1+L1+ω2Lm2R2+RLE32

To understand the power efficiency, the ratio of energy loss for the primary-side internal resistance, secondary resistance, and load can be denoted by PR1, PR2, and PRL as Eq. (14).

PR1:PR2:PL=I12R1:I22R2:I22RLE33

The square ratio of I2 and I1 can be expressed as Eq. (34), according to Eq. (27).

I2I12=ω2Lm2R2+RL2E34

The power ratio can be rewritten as Eq. (35), according to Eqs. (33) and (34).

PR1:PR2:PL=R2+RL2R1:ω2Lm2R2:ω2Lm2RLE35

Finally, the power efficiency can be obtained, according to Eq. (36).

η=PLPin=PLPR1+PR2+PL=ω2Lm2RLR2+RL2R1+ω2Lm2R2+ω2Lm2RLE36

Table 1 summarizes the results from the previous derivation in terms of the efficiency and power of the two types of magnetic field coupling circuits. With C1 and C2 inserted on the primary and secondary sides, the power efficiency and power at Rload will increase, which is why resonance is important.

EfficiencyPower at Rload
Magnetic field coupling without resonanceω2Lm2RLR2+RL2+ω2Lm2R1+ω2Lm2R2+ω2Lm2RLLmR1+L1L2+R2+RL+ω2Lm22U12RL
Magnetic field coupling With resonanceω2Lm2RLR2+RL2R1+ω2Lm2R2+ω2Lm2RLLmR1R2+RL+ω2Lm22U12RL

Table 1.

Comparison between magnetic field coupling under either no or with resonance.

Advertisement

4. Electric field coupling

Compared with magnetic field coupling, the electric field coupling without resonance uses a displacement current instead of an electromagnetic induction (Figure 7), as Eq. (37). Where ε, E, and D are permittivity, electric field intensity, electric flux density, respectively [17, 18, 19].

Figure 7.

Diagram of displacement current.

I=D∂tS=εE∂tSE37

In Figure 8, the mutual inductance of Cm between two capacitances on the distance x between the two coils can be written as Cm=kC1C2.

Figure 8.

Circuit topology of electric field coupling with resonance.

The electrical field coupling schematic diagram can be equivalent to π-type coupling, as shown in Figure 9. Based on the mesh-analysis method, the basic mathematical expression can be described as a matrix, according to Eq. (38).

Figure 9.

π-type equivalent circuit for electric field coupling with resonance.

U00=R1+jωL1+1C1Cm1C1Cm01C1Cm1C1Cm+1Cm+1C2Cm1C2Cm01C2CmR2+RL+jωL2+1C2CmIl1ImIl2E38
Zin2=1C2+1R2+L2+RLE39
Z2=1ω2Cm2Zin2E40
Zin1=1C1+1Z2+L1+R1E41

To analyze the input impedance, the extended π-type circuit with a separate capacitance is shown in Figure 10. The input impedance of each stage can be expressed as Eqs. (38), (39), and (40).

Figure 10.

π-type circuit with a separate capacitance for electric field coupling with resonance.

When operating frequency is equal to the resonant frequency on the secondary side, the imaginary part of the Zin2 becomes zero. Then, the resonant frequency can be acquired as Eq. (42).

ω2=1L2C2R2+RLL22E42

In the same way, when the operating frequency is equal to the resonant frequency on the primary side, the imaginary part of the Zin1 becomes zero. The L1 can be obtained as Eq. (43).

L1=C1R2+RL2ω12C2R2+RL2+ωL22L2L2C1C1C2Cm2R2+RL2+ωL22ω14L2C1C1C2Cm2R2+RL2+ωL222+ω12C12R2+RL2E43

To understand the power efficiency, the ratio of energy loss for the primary-side internal resistance, secondary-side internal resistance, and load can be denoted by PR1, PR2, and PRL as Eq. (44).

PR1:PR2:PRL=I12R1:I22R2:I22RLE44

The square ratio of I2 and I1 can be calculated as Eq. (45), according to Eq. (38) [16].

I1I22=ω2C1C2R2+RL2+ωL22+C121ω2L2C2Cm2E45

The power ratio can be rewritten as Eq. (46), according to Eqs. (44) and (45).

PR1:PR2:PL=ω2C1C2R2+RL2+ωL22+C121ω2L2C2R1:Cm2}R2:Cm2RLE46

Finally, the power efficiency can be expressed as Eq. (47).

ηω=PLPin=PLPR1+PR2+PL=Cm2RLω2C1C2R2+RL2+ωL22+C121ω2L2C2R1+Cm2R2+Cm2RLE47
Advertisement

5. WPT system overview

The basic topology of magnetic field resonant coupling is shown in Figure 11 [16, 20, 21]. The transmitter and receiver sides are coupled through the magnetic field (Figure 11a) and electric field (Figure 11b). The DC power supply generates the source power. The inverter is responsible for DC/AC conversion to produce the high-frequency AC power. The AC power is transferred to the receiver side and converted from AC to DC power by a rectifier. Then, the power is transferred to the load. In the WPT operating frequency for implantable devices, the common frequency is a at low frequency because of the reduction of tissue heating effect [22]. The typical frequency is 13.56 MHz within the license-free industrial, scientific, and medical (ISM) band [23].

Figure 11.

System overview for (a) magnetic field coupling and (b) electric field coupling methods.

The class E power amplifier is commonly applied in MHz-range WPT applications because of its excellent power efficiency from the properties of zero voltage switching and zero voltage derivative switching [24]. The class-E power amplifier can be depicted in Figure 12. In theory, it has a maximum efficiency of 100% [25], according to Eq. (48). The overall circuit consists of the power supply Vdc, RF-chuck, and passive load network. The RF-chuck with high reactance ensures DC voltage pass through the system. The passive load network is responsible for minimizing the overlap between drain voltage and drain current and further reducing the output power loss [26].

Figure 12.

Circuit diagram of a class E power amplifier.

Pdissipation=1T0T0IDSVDSdt=0E48

Some group has proposed a class E power amplifier-based WPT system for implantable biomedical in practical applications. Mutashar et al. [27] designed a wireless magnetic resonant coupling system using a class-E power amplifier and inductive power links via two spiral transmitters and receiver coils (Figure 13a). The 13.56 MHz carrier frequency within the ISM band range was selected to avoid tissue damage. The proposed class-E power amplifier can reach up to 87.2% power efficiency. The overall inductive coupling system achieves a power transmission efficiency of 73%. The compact structure could be implemented in bio-implants in the future. In practical applications for the electrical field coupling method, Narayanamoorthi et al. [24] developed a class-E-based capacitive coupled wireless power transfer system for biomedical implants (Figure 13b). They optimize the class E-based capacitive coupling WPT systems at an operating frequency of 13.56 MHz. Moreover, the matching circuit consisted of inductive-capacitive-inductive (LCL) impedance was used to improve the power efficiency of the power amplifier up to 96.34%.

Figure 13.

WPT system by (a) magnetic field coupling [27] and (b) electric field coupling for implantable applications [24].

Advertisement

6. WPT in emerging biomedical topics

WPT systems are rapidly evolving for biomedical implants, including emerging topics of implantable pacemaker devices, implantable optogenetic devices, and implantable impedance plethysmography (IPG).

6.1 Wireless charge-based implantable pacemaker

Heart failure is one of the most killers worldwide. It occupies one-third of all global death. Nowadays, the implantable cardioverter pacemaker is a common approach to directly suppress heart failure [28, 29]. However, the battery life could cause health risks and patient inconvenience. Parinaz Abiri et al. [30] developed an inductively powered wireless pacing device to realize remote cardiac stimulation. They designed a Class E-based magnetic field resonant coupling system with a resonant frequency of 13.56 MHz to pace rhythm on Yucatan miniature pig.

6.2 WPT-based implantable optogenetic device

Optogenetics is an emerging field that provides an alternative for direct electrical stimulation to modulate membrane voltage. By genetic modification, the ion channels in the cell can be sensitive to certain wavelengths of light [31]. Nowadays, several advanced biomedical topics have involved the optogenetic topic in different applications, including neuroscience to control animal behavior [32, 33] and cardiology to find out a solution for cardiovascular disease [34, 35]. Yu et al. [36] implement an implantable optogenetic device implanted into the dog’s chest. WPT-based optogenetic modulation of cardiac sympathetic nerve activity can prevent ventricular arrhythmias.

6.3 WPT-based implantable bio-impedance sensor

Bio-impedance sensing is a noninvasive sensing technology based on Ohm’s law. It was commonly applied to cell structural depiction, human composition, and physiological measurement [37, 38, 39, 40]. The main advantage of bio-impedance sensing is providing wearable and low-cost applications in biomedical engineering. The bio-impedance technique commonly uses the two-pair electrodes to detect arterial pulsation by Ohm’s law. One pair of electrodes is responsible for current excitation in our body. Furthermore, the other pair electrodes are responsible for sensing voltage signals. Theodor et al. [41] developed an implantable bio-impedance sensor to measure the artery activity from the domestic pig for additional cardiovascular parameters extraction such as pulse rate and blood pressure [42].

Advertisement

7. Conclusion

This chapter has presented the WPT methodology for implantable medical devices. The common configuration of WPT and its principle were demonstrated, including magnetic field coupling and electric field coupling. Most of all, the recent emerging biomedical topics such as wireless-charging pacemakers, implantable optogenetic devices, and implantable bio-impedance sensors were demonstrated. Overall, this chapter thoroughly presented the electric principle of WPT and extended it to emerging biomedical applications.

References

  1. 1. Sun L, Ma D, Tang H. A review of recent trends in wireless power transfer technology and its applications in electric vehicle wireless charging. Renewable and Sustainable Energy Reviews. 2018;91:490-503. DOI: 10.1016/j.rser.2018.04.016
  2. 2. Bi Z, Kan T, Mi CC, Zhang Y, Zhao Z, Keoleian GA. A review of wireless power transfer for electric vehicles: Prospects to enhance sustainable mobility. Applied Energy. 2016;179:413-425. DOI: 10.1016/j.apenergy.2016.07.003
  3. 3. Hoang H, Lee S, Kim Y, Choi Y, Bien F. An adaptive technique to improve wireless power transfer for consumer electronics. IEEE Transactions on Consumer Electronics. 2012;58(2):327-332. DOI: 10.1109/TCE.2012.6227430
  4. 4. Kiani M, Ghovanloo M. An RFID-based closed-loop wireless power transmission system for biomedical applications. IEEE Transactions on Circuits and Systems II: Express Briefs. 2010;57(4):260-264. DOI: 10.1109/TCSII.2010.2043470
  5. 5. Iqteit N, Yahya K, Khan SA. Wireless power charging in electrical vehicles. In: Zellagui M, editor. Wireless Power Transfer. Recent Development, Applications and New Perspectives. London, UK: IntechOpen; 2021
  6. 6. Meng E, Sheybani R. Insight: Implantable medical devices. Lab on a Chip. 2014;14(17):3233-3240. DOI: 10.1039/C4LC00127C
  7. 7. Basar MR, Ahmad MY, Cho J, Ibrahim F. Application of wireless power transmission systems in wireless capsule endoscopy: An overview, Sensors (Basel, Switzerland). 2014;14(6):10929-10951. DOI: 10.3390/s140610929
  8. 8. DeForge WF. Cardiac pacemakers: A basic review of the history and current technology. Journal of Veterinary Cardiology: the Official Journal of the European Society of Veterinary Cardiology. 2019;22:40-50. DOI: 10.1016/j.jvc.2019.01.001
  9. 9. Wang T-W, Lin T-T. Electromagnetic compatibility issues in medical devices. In: Kishk A, editor. Electromagnetic Compatibility. London, UK: IntechOpen; 2021
  10. 10. Kanaan AI, Sabaawi AM. Implantable wireless systems: A review of potentials and challenges. In: Al-Rizzo H, Abushamleh S, editors. Antenna Systems. London, UK: IntechOpen; 2021
  11. 11. Shadid R, Noghanian S. A literature survey on wireless power transfer for biomedical devices. International Journal of Antennas and Propagation. 2018;2018:4382841. DOI: 10.1155/2018/4382841
  12. 12. Khan SR, Pavuluri SK, Cummins G, Desmulliez MPY. Wireless power transfer techniques for implantable medical devices: A review. Sensors. 2020;20(12):3487. DOI: 10.3390/s20123487
  13. 13. Etemadrezaei M. Wireless power transfer. In: Rashid MH, editor. Power Electronics Handbook. 4th ed. Oxford, UK: Butterworth-Heinemann; 2018. pp. 711-722
  14. 14. Garnica J, Chinga RA, Lin J. Wireless power transmission: From far field to near field. Proceedings of the IEEE. 2013;101(6):1321-1331. DOI: 10.1109/JPROC.2013.2251411
  15. 15. Wang TW, Sung YL, Lin SF. Cardiac influence of repetitive transcranial magnetic stimulation in small animals. IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology. 2020;4(4):279-285. DOI: 10.1109/JERM.2019.2958686
  16. 16. Imura T. Wireless Power Transfer: Using Magnetic and Electric Resonance Coupling Techniques. Singapore: Springer Nature; 2020
  17. 17. Wang TW, Zhang H, Lin SF. Influence of capacitive coupling on high-fidelity non-contact ECG measurement. IEEE Sensors Journal. 2020;20(16):9265-9273. DOI: 10.1109/JSEN.2020.2986723
  18. 18. Wang TW, Lin SF. Negative impedance capacitive electrode for ECG sensing through fabric layer. IEEE Transactions on Instrumentation and Measurement. 2021;70:1-8. DOI: 10.1109/TIM.2020.3045187
  19. 19. Wang T-W, Lin S-F. Non-contact capacitive sensing for ECG recording in small animals. Measurement Science and Technology. 2020;31(12):125703. DOI: 10.1088/1361-6501/ab8cfc
  20. 20. Kindl V, Zavrel M, Drabek P, Kavalir T. High efficiency and power tracking method for wireless charging system based on phase-shift control. Energies. 2018;11(8):2065. Available from: https://www.mdpi.com/1996-1073/11/8/2065
  21. 21. Yi K. Capacitive coupling wireless power transfer with quasi-LLC resonant converter using electric vehicles’ windows. Electronics. 2020;9(4):676. DOI: 10.3390/electronics9040676
  22. 22. Freeman DK, Byrnes SJ. Optimal frequency for wireless power transmission into the body: Efficiency versus received power. IEEE Transactions on Antennas and Propagation. 2019;67(6):4073-4083. DOI: 10.1109/TAP.2019.2905672
  23. 23. Sanchez-Montero R, Lopez-Espi P-L, Alen-Cordero C, Martinez-Rojas J-A. Bend and moisture effects on the performance of a U-shaped slotted wearable antenna for off-body communications in an Industrial Scientific Medical (ISM) 2.4 GHz band. Sensors. 2019;19(8):1804. DOI: 10.3390/s19081804
  24. 24. Narayanamoorthi R, Vimala Juliet A, Chokkalingam B, Padmanaban S, Leonowicz ZM. Class E power amplifier design and optimization for the capacitive coupled wireless power transfer system in biomedical implants. Energies. 2017;10(9):1409. DOI: 10.3390/en10091409
  25. 25. Lim A, Tan A, Kong Z-H, Ma K. A design methodology and analysis for transformer-based Class-E power amplifier. Electronics. 2019;8(5):494. DOI: 10.3390/electronics8050494
  26. 26. Sokal NO, Sokal AD. Class E-A new class of high-efficiency tuned single-ended switching power amplifiers. IEEE Journal of Solid-State Circuits. 1975;10(3):168-176. DOI: 10.1109/JSSC.1975.1050582
  27. 27. Mutashar S. Efficiency improvement of wireless power transmission for bio-implanted devices. International Journal of Biomedical Engineering. 2013;7:721-724
  28. 28. Kirk JA et al, Pacemaker-induced transient asynchrony suppresses heart failure progression, Science Translational Medicine. 2015;7(319):319ra207. DOI: 10.1126/scitranslmed.aad2899
  29. 29. Khazanie P, Hellkamp AS, Fonarow GC, Curtis LH, Al-Khatib SM, Hernandez AF. Permanent pacemaker use among patients with heart failure and preserved ejection fraction: Findings from the Acute Decompensated Heart Failure National Registry (ADHERE) National Registry. American Heart Journal. 2018;198:123-128. DOI: 10.1016/j.ahj.2017.12.020
  30. 30. Abiri P et al. Inductively powered wireless pacing via a miniature pacemaker and remote stimulation control system. Scientific Reports. 2017;7(1):6180. DOI: 10.1038/s41598-017-06493-5
  31. 31. Deisseroth K. Optogenetics. Nature Methods. 2011;8(1):26-29. DOI: 10.1038/nmeth.f.324
  32. 32. Mahmoudi P, Veladi H, Pakdel FG. Optogenetics, tools and applications in neurobiology. Journal of Medical Signals and Sensors. 2017;7(2):71-79
  33. 33. Boyden ES. Optogenetics and the future of neuroscience. Nature Neuroscience. 2015;18(9):1200-1201. DOI: 10.1038/nn.4094
  34. 34. Williams JC, Entcheva E. Optogenetic versus electrical stimulation of human cardiomyocytes: Modeling insights. Biophysical Journal. 2015;108(8):1934-1945. DOI: 10.1016/j.bpj.2015.03.032
  35. 35. Entcheva E, Kay MW. Cardiac optogenetics: A decade of enlightenment. Nature Reviews Cardiology. 2021;18(5):349-367. DOI: 10.1038/s41569-020-00478-0
  36. 36. Yu L et al. Optogenetic modulation of cardiac sympathetic nerve activity to prevent ventricular arrhythmias. Journal of the American College of Cardiology. 2017;70(22):2778-2790. DOI: 10.1016/j.jacc.2017.09.1107
  37. 37. Wang TW, Chen WX, Chu HW, Lin SF. Single-channel bioimpedance measurement for wearable continuous blood pressure monitoring. IEEE Transactions on Instrumentation and Measurement. 2021;70:1-9. DOI: 10.1109/TIM.2020.3035578
  38. 38. Wang TW et al. Single-channel impedance plethysmography neck patch device for unobtrusive wearable cardiovascular monitoring. IEEE Access. 2020;8:184909-184919. DOI: 10.1109/ACCESS.2020.3029604
  39. 39. Wang T-W et al. Bio-impedance measurement optimization for high-resolution carotid pulse sensing. Sensors. 2021;21(5):1600. DOI: 10.3390/s21051600
  40. 40. Wang TW, Sung YL, Chu HW, Lin SF. IPG-based field potential measurement of cultured cardiomyocytes for optogenetic applications. Biosensors & Bioelectronics. 2021;179:113060. DOI: 10.1016/j.bios.2021.113060
  41. 41. Theodor M et al. Implantable impedance plethysmography. Sensors. 2014;14(8):14858-14872. DOI: 10.3390/s140814858