Abstract
The results of a study on pMOS dosimeters manufactured by Tyndall National Institute, Cork, Ireland and their sensitivity on radiation doses used in radiotherapy are presented. Firstly, we deal with analysis of defect precursors created by ionizing radiation, responsible for increase in fixed and switching traps, which are further responsible for threshold voltage shift as a dosimetric parameter. Secondly, influence of some parameters, such as gate bias during irradiation, gate oxide thickness and photons energies, on threshold voltage shift is presented. Fading of irradiated pMOS dosimeters and possible application of commercial MOSFETs in ionizing radiation dosimetry are also presented.
Keywords
- fading
- MOSFET
- pMOS dosimeter
- radiation dose
- threshold voltage shift
1. Introduction
External radiotherapy is a well‐accepted and established therapeutic modality for cancer treatment [1]. In this technique, radiation beams, generated by either radiation source or linear accelerator, are specifically optimized to cause the death of the tumor cells without having a greater impact on the healthy tissues. It is estimated that dose precision in radiotherapy is approximately ±5%. However, in order to ensure proper dose delivery to the designated area and appropriate intensity, a sophisticated radiation oncology Quality Assurance (QA) program is required [1, 2]. Also, the verification of the final dose delivered to the patient, which can only be carried out by in vivo dosimeters, is very important and should basically be used for all patients undergoing radiation treatment [3].
In vivo dosimetry can be measured by thermoluminescent dosimeters (TLDs) [4, 5], diode dosimeters [6, 7] and MOSFET (Metal‐Oxide‐Semiconductor Field Effect Transistor) dosimeters [8, 9]. TLDs characteristics include the following: cable‐free, accurate, small volume and tissue‐equivalence. However, an important drawback of TLDs is the reading procedure because information is lost during the reading. Currently, TLDs are most popular dosimeters for QA radiotherapy despite the relatively high cost of the readout equipment and the requirement of a highly trained operator.
Diode dosimeters provide instantaneous readout; however, diodes must be connected to cable for applied voltage during radiation. Even though diode dosimeters are sensitive to the temperature and dependent on the radiation beam, the correction and calibration factors are generally well known.
The concept of radiation sensitive MOSFETs as dosimeter is based on converting the threshold voltage shift as a dosimetric parameter into radiation dose. Ionizing radiation creates positive charge in the MOSFETs oxide and interface trap at silicon dioxide‐silicon interface leading to a transistors threshold voltage shift. In p‐channel MOSFETs, both the positive charge in the oxide and interface traps contributes to threshold voltage shift in the same direction. This is reason why p‐channel MOSFETs instead n‐channel MOSFETs are usually used as dosimeters. p‐channel MOSFETs can be application in low‐field mode (without gate bias during irradiation) and in high‐field mode (with gate bias during irradiation). High‐field mode leads to the sensitivity increase in MOSFET dosimeters.
The p‐channel MOSFET as integrating dosimeters has been proposed in 1970 [10] and results being verified in 1974 [11]. This further leads to the production of radiation sensitive p‐channel MOSFETs, also known as RADiation‐sensitive Field Effect Transistor (RADFET) or pMOS dosimeter [12]. Besides, radiotherapy pMOS dosimeters could be used for radiation space monitoring [13, 14], irradiation of food plants [15] and in personal dosimetry [16].
A major advantage of the MOSFET as a radiation sensor is that the radiation‐sensitive region, the oxide film, is very small [11]. The sensing volume is much smaller than competing integral dose measuring devices, such as the ionization chamber or TLD. The MOSFETs sensitive volume is typically
In radiotherapy, the radiation oncologist determines the radiation dose depending on many factors such as the type and size of tumor, location in the body, how close the tumor is to other radiation sensitive tissues, how deep into the body the radiation need to penetrate, the patient general health and medical history, whether the patient will have other type of cancer treatments (e.g., chemotherapy) and other factors such as patient age and medical conditions. Cumulative dose range used in radiotherapy ranges from 20 to 70 Gy [24], while typical radiation dose for one fraction is from 1 up to 5 Gy.
This chapter presents some of the results obtained in our laboratory, which considers the influence of some parameters to pMOS dosimeters sensitivity and fading. Dosimeters were manufactured in Tyndall National Institute, Cork, Ireland. Sensitivity results are also presented for commercial MOSFETs in order to investigate their possible application in radiotherapy.
2. Mechanisms responsible for threshold voltage shift during irradiation
The dosimetry of ionizing radiation using radiation‐sensitive MOSFETs is based on the threshold voltage shift, conversion into absorbed radiation dose
The
Positive charge is formed in oxide by holes trapping, while electrons trapping lead to creation of negative charge. The concentration of positive charge in oxide is much higher since the hole trapping centers are more numerous compared to electron trapping centers. Moreover, trapped electrons and holes near Si/SiO2 interface have the strongest impact on channel carriers, hence on MOSFET characteristics.
Amphoteric defects
Positive trapped charge in the oxide is called fixed traps (FT), and positive trapped charge near Si/SiO2 interface is called switching traps (ST) [27], where FT represents traps in the oxide that without the ability to exchange the charge with the channel within the MOSFET transfer/subthreshold characteristic measurement time frame. On the other hand, ST represents traps created near and at Si/SiO2 interface, and they do capture (communicate with) the carrier from the channel within the transfer/subthreshold characteristic measurement time frame. Furthermore, one can differentiate between slow switching traps (SST) created in the oxide near Si/SiO2 interface and fast switching traps (FST) created at Si/SiO2 interface also known as true interface traps (
Threshold voltage shift
where
3. Response of pMOS dosimeters to gamma and X‐ray radiation
3.1. Important pMOS dosimetric parameters
The most important parameters that characterize the pMOS dosimetric radiation response are sensitivity, dose linearity and room temperature long‐term stability [35, 36]. Sensitivity represents threshold voltage shift
In practical applications, it is most convenient for pMOS dosimeters to have a linear response of threshold voltage shift
where
Positive gate bias during irradiation reduces the recombination of produced electron‐hole pairs in SiO2 and as a consequence the pMOS dosimeters response becomes more linear and sensitive [33, 41].
Room‐temperature long‐tem stability of irradiated pMOS dosimeters can be observed by calculating fading
where
3.2. Influence of gate bias on threshold voltage shift during irradiation
Figures 1 and 2 show the threshold voltage shift
Figure 3 shows the threshold voltage shift
Figure 5 shows the
Figure 6 shows the sensitivity
The increase in
3.3. Influence of gate oxide thickness on threshold voltage shift during irradiation
Figure 7 shows the threshold voltage shift
The
3.4. Influence of photon energy on pMOS dosimetry sensitivity
Figure 9 shows the threshold voltage shift
The
From Figures 9 and 10, it can be seen that increasing in
4. Fading of irradiated pMOS dosimeters
As a dosimeter radiation sensitive MOSFET must satisfy a crucial demand, which implies compromising between sensitivity to irradiation and stability with time after irradiation. Stability represent insignificant change in
Fading results for pMOS dosimeters with gate oxide thickness of 400 nm and 1 μm, at room temperature previously irradiated with X‐ray (energy 140 keV) up to 1 Gy for
The decrease in the positive trapped charge causes fading of pMOS dosimeters. This decrease originates from electron tunneling from Si into SiO2; once captured at positive oxide trapped charge, which lead to their neutralization/compensation and change in threshold voltage shift [45].
5. pMOS dosimeter reuse
For a while, it was widely thought that pMOS dosimeters could not be used for subsequent determination of radiation dose. They were, namely, just used to determine the maximum radiation dose, after which they would be replaced. However, studies on the pMOS dosimeter reuse are given in [46] for radiation dose 400 Gy. Recent work has shown that irradiated pMOS dosimeters manufactured in Tyndall National Institute, Cork, Ireland, could be annealed at room and elevated temperature and reused for ionizing radiation measurements. Figures 13 and 14 show the threshold voltage shift
6. Low‐cost commercial p‐channel MOSFETs as pMOS dosimeters
In recent years, many investigations were driven toward application of low‐cost commercial p‐channel MOSFETs as a dosimeter in radiotherapy [49]. Asensio et al. [50] show results of some most important dosimetric parameters (sensitivity, linearity, reproducibility and angular dependence) for power p‐channel MOSFETs 3N163. These transistors were irradiated by gamma‐rays originating from 60Co up to 55 Gy. These devices were irradiated without gate bias (
The possibility of vertical diffusion MOS also called double‐diffusion MOS transistor or simple DMOS as a sensor of electron beam was also investigated [51] These devices were DMOS BS250F, ZVP3306 and ZVP4525, manufactured by Diodes Incorporated (Plano, USA). The irradiation was performed by an electron beam of 6 MeV energy without gate bias. The same authors investigated the behavior of p‐channel MOS transistors from integrated circuit CD4007 (Texas Instruments, Dallas, USA and NXP Semiconductor Eindhowen, Netherlands) under 6 MeV energy electron beam. In Figure 16, the
7. Conclusion
The sensitivity of pMOS dosimeters manufactured in Tyndall National Institute, Cork, Ireland, with 100 nm, 400 nm and 1
Acknowledgments
The Ministry of Education, Science and Technological Development of the Republic of Serbia supported this work under contract no. 171007.
References
- 1.
Siebel OF, Pereira JG, Souza RS, Ramirez‐Fernandez FJ, Schneider MC and Galup‐Montoro CA. A very‐low‐cost dosimeter based on the off‐the‐shelf CD4007 MOSFET array for in vivo radiotherapy applications. Radiat. Meas. 2015; 75 :53–63. - 2.
Purdy JA, Klein E, Vijayakumar S, Perez CA and Levitt SH. Quality assurance in radiation oncology. In: Perez CA, Levitt SH, Purdy JA, Vijayakumar S (Eds). Technical Basis of Radiation Therapy: Practical Clinical Applications, Springer‐Verlag, Berlin 2005:395–422. - 3.
Rosenfeld AB. Electronic dosimetry in radiation therapy. Radiat. Meas. 2006; 41 :5134–5153. - 4.
Kron T, Butson M, Hunt F and Denham J. TLD extrapolation for skin dose determination in vivo. Radiother. Oncol. 1996; 41 :119–123. - 5.
Das R, Toye W, Kron T, Williams S and Duchesne G. Thermoluminescence dosimetry for in‐vivo verification of high dose rate brachytherapy for prostate cancer. Australas. Phys. Eng. Sci. Med. 2007; 30 :178–184. - 6.
Lancaster CM, Cosbe JC and Devis SR. In vivo dosimetry from total body irradiation patients (2000–2006); results and analysis. Australas. Phys. Eng. Sci. Med. 2008; 31 :191–195. - 7.
Saini AS and Zhu TC. Energy dependence of commercially available diode detectors for in vivo dosimetry. Med. Phys. 2007; 34 :1704–1711. - 8.
Scalchi P and Francescon P. Calculation of MOSFET detection system for 6‐MV in vivo dosimetry. Int. J. Radiat. Oncol. Biol. Phys. 1988; 40 :987–993. - 9.
Jornet N, Carrisio P, Jurada D, Ruis A, Eudaldo T and Ribas M. Comparison study of MOSFET detectros and diodes for entrace in vivo dosimetry in 18 MV x‐ray beam. Med. Phys. 2004; 31 :2534–2542. - 10.
Poch W and Holmes‐Siedle AG. The mosmeter‐a new instrument for measuring radiation dose. RCA Eng. 1970; 16 :56–59. - 11.
Holmes‐Siedle AG. The space charge dosimeter‐general principles of a new method of radiation dosimetry. Nucl. Instr. Methods. 1974; 121 :169–171. - 12.
Adams L and Holmes‐Siedle A. The development of MOS dosimetry unit for use in space. IEEE Trans. Nucl. Sci. 1978; 18 :1607–1612. - 13.
Kay K, Mullen E, Stopar W, Circle R and McDonald P. GRREs dosimetry results and comparison using the space radiation dosimeter and p‐channel MOS dosimeter. IEEE Trans. Nucl. Sci. 1992; 39 :1846–1850. - 14.
Scheik LZ, McNulty FJ and Roth DR. Dosimeter based on the ensure of floating gate in natural radiation environments. IEEE Trans. Nucl. Sci. 1998; 45 :2681–2688. - 15.
Faigon A, Lipovetzky J, Redin E and Kruscenski G. Expression of measurement range of MOS dosimeters using radiation induce charge neutralization. IEEE Trans. Nucl. Sci. 2008; 55 :2141–2147. - 16.
Sarrabayrouse G, Buchdahl D, Poliscuk V and Siscos S. Stacked‐MOS ionizing radiation dosimeters: potentials and limitations. Rad. Phys. Chem. 2004; 71 :737–739. - 17.
Rosenfeld AB, Kaplan GI, Kron T, Allen BJ, Dilmanian A, Orion I, Ren B, Lerch MLF and Holmes‐Siedle A. MOSFET dosimetry of an X‐ray microbeam. IEEE Trans. Nucl. Sci. 1999; 46 :1774–1780. - 18.
Gladstone DJ, Lu XQ, Humm JL, Bowman HF and Chin LM. A miniature MOSFET radiation probe. Med. Phys. 1994; 21 :1721–1728. - 19.
Kaplan GI, Rosemfeld AB, Allen BJ, Booth JT, Carolan MG and Holmes‐Siedle. A special resolution by MOFET dosimetry of an x‐ray microbeam. Med. Phys. 2000; 27 :239–244. - 20.
Jaksic A, Ristic G, Pejovic M, Mohammadzadeh A, Sudre C and Lane W. Gamma ray irradiation and post‐irradiation response of high dose range RADFETs. IEEE Trans. Nucl. Sci. 2002; 49 :1356–1363. - 21.
Price RA. Towards an optimum design of a P‐MOS radiation detector for use in high‐energy medical phantom beam and neutron facilities: analysis of activation materials. Radiat. Prot. Dosimetry. 2005; 115 :386–390. - 22.
Pejovic MM, Pejovic MM and Jaksic AB. Response of pMOS dosimeters on gamma ray irradiation during its re‐use. Radiat. Prot. Dosimetry. 2013; 155 :394–403. - 23.
Alchaikh S, Carolan M, Petasecca M, Lerch M and Metealfe AB. Direct and pulsed current annealing of p‐MOSFET based dosimeter, the MOSkin. Australas Phys. Eng. Sci. Med. 2014; 37 :311–319. - 24.
RCR. Available on www.rcr.uk. The royal college of radiologists, radiotherapy dose fractions. 2006. - 25.
Lipovetzky J, Redin EG and Fajgon A. Electrically erasable metal‐oxide‐semiconductor dosimeters. IEEE Trans. Nucl. Sci. 2007; 54 :1244–1250. - 26.
Moreno E, Picos R, Isern E, Roca M, Bota S and Suenoga K. Radiation sensor compatible with standard CMOS technology. IEEE Trans. Nucl. Sci. 2009; 56 :2910–2915. - 27.
Pejovic M, Osmokrovic, P, Pejovic M.M and Stankovic K. Influence of ionizing radiation ant hot carrier injection on metal-oxide-semiconductor transistors, In Nenoi M. (Ed), Current Topics in Radiation Research. INTECH, Institute for New Technologies, Maastricht (NT), 2012, Chapter 33, OCLC: 846871029 (accessed 0.6.06.15) - 28.
Griscom DL. Optical properties and structure of defects in silica glass. J. Ceram. Soc. Japan. 1991; 99 :923–941. - 29.
McLean FB. A framework for understanding radiation‐induced interface states in SiO2 MOS structure. IEEE Trans. Nucl. Sci. 1980; 54 :1651–1657. - 30.
Saks NS and Brown DB. Interface trap formation via the two‐stage H+ process. IEEE Trans. Nucl. Sci. 1989; 36 :1848–1857. - 31.
Griscom DL. Diffusion of radiolytic molecular hydrogen as a mechanism for the post‐irradiation buildup of interface states in SiO2‐on Si structures. J. Appl. Phys. 1985; 58 :2524–2533. - 32.
Poindexter EH. Chemical reactions of hydrogen species in the Si‐SiO2 system. J. Non‐Cryst. Solids. 1995; 187 :257–263. - 33.
Pejovic MM. Processes in radiation sensitive MOSFETs during and post irradiation annealing responsible for threshold voltage shift. Radiat. Phys. Chem. 2017; 130 :221–228. - 34.
Ma TP and Drensserdorfer PV, Ionizing Radiation Effects in MOS Devices and Circuits. J. Wiley; New‐York USA: 1989. - 35.
Pejovic SM, Pejovic MM, Stojanov D and Ciraj‐Bjelac O. Sensitivity and fading of pMOS dosimeters irradiated with X‐ray radiation doses from 1 to 100 cGy. Radiat. Prot. Dosimetry. 2016; 168 :33–39. - 36.
Pejovic MM. Dose response, radiation sensitivity and signal fading of p‐channel MOSFETs (RADFETs) irradiated up to 50 Gy with 60Co. Appl. Radiat. Isot. 2015; 104 :100–105. - 37.
Pejovic MM, Pejovic MM and Jaksic B. Radiation‐sensitive field effect transistor response to gamma‐ray irradiation. Nucl. Technol. Radiat. Protect. 2011; 26 :25–31. - 38.
Pejovic MM, Pejovic SM, Stojanov D and Ciraj‐Bjelac O. Sensitivity of RADFET for gamma and X‐ray doses used in medicine. Nucl. Technol. and Radiat. Protect. 2014; 29 :179–185. - 39.
Pejovic S, Bosnjakovic P, Ciraj‐Bjelac O and Pejovic MM. Characteristics of PMOSFET suitable for use in radiotherapy. Appl. Radiat. Isot. 2013; 77 :44–49. - 40.
Sarrabayrouse G and Gessinn FG. Thick oxide MOS transistors for ionizing radiation dose measurement. Radioprotection. 1994; 29 :557–572. - 41.
Rosenfeld AB. MOSFET dosimetry an modern radiation oncology modalites. Radiat. Prot. Dosimetry. 2002; 101 :393–398. - 42.
Pejovic MM, Pejovic MM and Jaksic AB. Contribution of fixed oxide traps to sensitivity of pMOS dosimeters during gamma ray irradiation and annealing at room and elevate temperature. Sens. Actuators A. 2012; 174 :341–345. - 43.
Pejovic MM. The gamma‐ray irradiation sensitivity and dosimetric information instability of RADFET dosimeter. Nucl. Technol. Radiat. Protect. 2013; 28 :415–421. - 44.
Ristic G, Jaksic A and Pejovic M. pMOS dosimeter transistors with two‐layer gate oxide. Sens. Actuators A. 1997; 63 :129–134. - 45.
McWhorter PJ, Miller SL and Miller WM. Modeling the anneal of radiation‐induced traps holes in a varying thermal environment. IEEE Trans. Nucl. Sci. 1990; 37 :1682–1689. - 46.
Kelleher A, McDonnell N, O'Neill B, Lane W and Adams L. Investigation into the re‐use of pMOS dosimeters. IEEE Trans. Nucl. Sci. 1994; 41 :445–449. - 47.
Pejovic MM, Pejovic MM, Jaksic AB, Stankovic KDj and Markovic A. Successive gamma ray irradiation and corresponding post‐irradiation annealing of pMOS dosimeters. Nucl. Technol. Radiat. Protect. 2012; 27 :341–345. - 48.
Pejovic MM, Pejovic MM and Jaksic AB. Response of pMOS dosimeters on gamma‐ray irradiation during its re‐use. Radiat. Prot. Dosimetry. 2013; 155 :394–403. - 49.
Aristru J, Calvo F, Martinez R,, Dubois M, Fisher S and Azinovic I. Lung cancer. In EBRT with or without IORT. From: Current Oncology: Intraoperative Irradiation: Techniques and Results, Ed. by Gunderson F. et al, Humana Press, Inc, Totowa, NJ., 1999 - 50.
Asensio LJ, Carvaial MA, Lopez‐Villaneva JA, Vilches M, Lallena AM and Palma AJ. Evaluation of a low‐cost commercial mosfet as radiation dosimeter. Sens. Actuators A. 2006; 125 :288–295. - 51.
Martines‐Garcia MS, Simancos F, Palma AJ, Lallena AM, Banqueri J and Carvajal MA. General purpose MOSFETs for the dosimetry of electron beams used in intra‐operative radiotherapy. Sens. Actuators A. 2014; 210 :175–181.