Type and principle of functioning of developed electrochemical biosensors.
1. Introduction
There is a growing need of highly efficient compact devices for a wide range of applications in several fields. Among the candidate materials, porous silicon (PSi) has attracted an increasing research interest, apart from its obvious potentially straightforward integration with standard Si technologies, thanks to its unique properties, and its present applications span from biomedicine (Anglin et al. 2008) to biosensing, from photonics (Huy et al. 2009) to photovoltaic devices (Xiong et al. 2010).
After its discovery (Uhlir 1956), porous silicon hasn’t attracted much attention until the discovery of its room temperature luminescence properties (Canham 1990). However, the porous silicon-based photonics with all-silicon light-emitting devices never showed, to date, high enough efficiency for real applications. Nevertheless, many other applications have since been the object of much research, thanks to the many advantages of porous silicon: the flexibility of its formation process (Föll et al. 2002), the extensive tailoring of its structural properties (Lehmann et al. 2000), the very large specific surface (Halimaoui 1993) and also its biocompatibility, mandatory for both drug delivery devices and several biosensing applications (Low et al. 2009; Park et al. 2009). The deep knowledge of silicon chemistry is easily applicable to porous silicon (Buriak 2002; Salonen and Lehto 2008) and allows functionalization of the internal pore surface for the chemical bonding of the biological molecules of interest or for a better stabilization of the structure. The relevance for the field of PSi biosensors of a well controlled surface chemistry has been detailed by Lees and coworkers and by Kilian and coworkers (Lees et al. 2003; Kilian et al. 2009) and will be treated more deeply later in this article.
The very large internal surface of porous silicon proved to be a great advantage since, like other porous materials, it allows the bonding of active molecules over a large surface in a small volume (e.g. a 20 µm thick PSi sample with a specific surface area of 500 m2/cm3 may offer 1 m2 of developed surface with only 1 cm2 of external surface), with a sensible increase of the efficiency of the devices. Moreover, the large internal surface is important when there is the need of dispersing the active molecules, as happened in the case of laser dye dispersed in a PSi matrix (Setzu et al. 1999). Differently from other materials, however, PSi may be easily prepared either in powder or wafer, depending on the specific application. This allows for the fabrication of devices that can be dispersed in a given medium or that can be reusable. Devices integrating PSi layers with specific enzymes or with molecules with specific target allow the realization of label free biosensors. Examples of this kind of devices have been demonstrated, for instance, for DNA sensing (Rong et al. 2008) and for triglycerides quantitative measurements (Setzu et al. 2007).
Porous silicon was successfully used in the development of a quite large variety of new biosensors, mainly using optical detection (Chan et al. 2001; Jane et al. 2009). It is surprising that porous silicon electrochemical sensors didn’t get as much attention as the optical sensors. This is likely due to the fact that much research efforts have been devoted to the development of optical PSi devices in the field of optoelectronics. Then, a natural transfer of this knowledge to the field of biosensing has occurred when PSi-based biosensors become an interesting research field. However, the general field of electrochemical sensors (Bakker and Telting-Diaz 2002; Privett et al. 2008) is the most developed sensor branch and also PSi electrochemical sensors have been developed showing interesting characteristics and sensitivity properties.
Several reviews describe the state of the art of PSi biosensors (Anglin et al. 2008; Jane et al. 2009; Kilian et al. 2009) mainly based on optical signal transduction. The remarkable development of optical PSi biosensor has been triggered by the ability to modulate the porous silicon refractive index in the etch direction, and therefore to tailor the optical properties of the devices to one’s needs, that has stimulated the research in the field of signal transduction by optical means. Optical transduction through either Fabry-Perot fringes, microcavity resonators or rugate filters has been widely investigated (Lin et al. 1997; Chan et al. 2001; Sailor 2007). There are several examples of quantitative determination of a given DNA strain by using optical detection on a single or multiple-layer PSi sample (Chan et al. 2000; De Stefano et al. 2007). Other examples show in-body detection of drug release (Anglin et al. 2008). Optical detection may be extremely useful when there is the need of a very high sensitivity for small amount of molecules to be detected, since the refractive index of the porous layer is highly affected by a change in the refractive index of the liquid in the pores (Anderson et al. 2003). The measured sensitivity of the optical biosensors strongly depends on the chosen optical structure and on the analyte (Haes and Van Duyne 2002; DeLouise et al. 2005).
It is, however, rather surprising that very little research has been made in the field of electrochemical porous silicon-based sensors even if electrochemical sensors have several important advantages: low cost and high sensitivity, together with a low power requirement and relatively simple detection instruments. Moreover they can be miniaturised more easily than optical biosensors. All these considerations make particularly worthwhile to review the advantages of PSi electrochemical biosensors. This is the main aim of the present chapter.
2. Porous silicon formation, oxidation, functionalisation, and biomolecules immobilisation
2.1. Formation process and main properties of porous silicon
Porous silicon samples are mainly produced by an electrochemical etch in the dark of a bulk single crystalline silicon substrate (Lehmann 1996). Fig. 1 shows a typical electrochemical cell used for PSi formation (Fig. 1a). A PSi sample is also shown (Fig. 1b). There are also various alternative preparation methods (Kolasinski 2005), i.e. chemical vapour etching (Ben Jaballah et al. 2005), metal-assisted etching (Harada et al. 2001; Chattopadhyay et al. 2002), and stain etching (Steckl et al. 1993; Ünal et al. 2001).
These methods give, at present, less reproducible results with respect to electrochemical etch, even though stain etch PSi is already commercially available. The etching solutions are prepared using HF, ethanol and pure water in different concentrations. The concentration of HF in the solution is one of the parameters controlling the structural properties of the samples and gives different porosities and pores’ densities for a given current density used in the fabrication process. The HF concentration is also a fundamental parameter for the porosity range available. Halimaoui (Halimaoui 1993) studied the PSi layer porosities as a function of the applied current density for different HF concentrations, and observed that the porosity range between the lowest and highest current densities available for the porous layer formation varied for different HF concentrations. It has also been demonstrated that the PSi characteristics depend on the HF concentration of the etching solution (Dian et al. 2004; Kumar et al. 2009) from pores shape to density. In particular, Dian and coworkers showed that, for the same formation current density, varying the HF concentration leads to layers with different characteristics and porosities whose variations may reach about 30% in their experimental conditions. Kumar and coworkers studied the variations of the physical and electronic properties of PSi layers prepared using etching solutions with various HF contents by means of a combination of volumetric sorption isotherms, visual colour observation, photoluminescence, scanning electron microscopy, and Raman spectroscopy.
2.2. PSi layers morphology and design
Critical parameters for the defining of the PSi layers pores morphology are the doping type and the doping level of the crystalline silicon substrates used for the preparation of the samples (Föll et al. 2002). These parameters affect the kind of porosity, starting from the pores’ diameter that can span from nanopores (a few nm) to mesopores (a few tens of nm to a few hundreds of nm) up to macropores (a few µm).
Only the nanoporous
the formation condition (Gruning and Lehmann 1996; Ouyang and Fauchet 2005). The pores’ diameter can be further modified after formation by means of one or more of the several available techniques able to enlarge the pores’ diameter to better adapt to the size of the molecule of interest. Tinsley-Bown and co-workers (Tinsley-Bown et al. 2000) described a method based on immersing the porous layers in an ethanol-rich alkali (KOH) solution, whose effect on the pores’ diameter may be controlled by the immersion time. Hamm and coworkers (Hamm et al. 2003) used a similar method by immersing the PSi samples in a NaOH etching (0.1 and 1 M) solution. To avoid problems due the hydrophobic nature of the PSi walls, an ethanol drop is put on the samples’ surface before the immersion in the NaOH solution. These methods add a significant flexibility in the tailoring of the PSi layers structural properties in view of the realization of biosensors. More details about the dependence of the kind, shape and size of the PSi pores on the substrate doping may be found in the scientific literature (Smith and Collins 1992; Lehmann et al. 2000; Föll et al. 2002).
A relevant characteristic of porous silicon is that once the porous layer is formed it will play no longer any role in the ongoing formation process: the electrochemical etch is essentially an interface process taking place only at the porous/crystalline interface (Chazalviel et al. 2000). Since the structural and optical features of the formed layers depend on the formation parameters, and in particular on the formation current density, it is possible to design structures well tailored on the needs of the application. In particular, knowing that higher formation current densities lead to layers with higher porosities and lower refractive index, PSi multilayers with an in-the-depth modulation have been realised. A periodic variation of the formation current density will lead to a periodic variation of the porosity in the formation direction and then to a corresponding periodic refractive index variation (Setzu et al. 2000), leading to the realization of 1D photonic bandgap in the formation direction.
2.3. Oxidation of PSi surface
Surface chemistry of PSi is a feature of fundamental interest (Kilian et al. 2009). Chemical modifications at the air-solid interfaces strongly affect the biosensing performance (Dancil et al. 1999). Indeed, freshly prepared PSi is unstable due to the presence of highly reactive silicon hydride (Si-H
Anodic oxidation is a technique commonly used to stabilize the very reactive surface of fresh porous silicon (Bsiesy et al. 1991; Petrova-Koch et al. 1992; Cantin et al. 1996). Compared to other oxidation techniques, it allows the obtainment of highly reproducible oxidised samples. Recently, we reported a study where PSi layers were oxidised through this electrochemical procedure (Salis et al. 2010). Anodic oxidation is realised through a controlled procedure where a given current intensity is made to flow into an electrochemical cell until a fixed oxidising potential or a given oxidation time is reached. A typical anodic oxidation curve is shown in Fig. 3.
2.4. PSi surface functionalisation and biomolecules immobilisation
As reported above, the surface of fresh PSi is almost completely covered by highly reactive hydride species (Petrova-Koch et al. 1992). Indeed, when PSi is put in contact with alkaline solutions, and even with buffer solutions at physiological pH, it dissolves giving orthosilicic acid (Si(OH)4) (Anderson et al. 2003). Neglecting this phenomenon can lead to confuse a signal drift, due to the oxidation phenomenon, with the biosensing signal (Lin et al. 1997; Janshoff et al. 1998; Dancil et al. 1999). Besides oxidation, a strategy to improve PSi stability is the modification of the surface by mean of functionalising agents. This is usually done as a preliminary step which ends with the biomolecule immobilisation.
Xia and coworkers published a study where porous silicon surfaces were bio-functionalised by a simple three-step method (Xia et al. 2006). A scheme of the functionalising procedure is shown in Fig. 4. First the hydrogen-terminated porous silicon was oxidised and amino-silanised in a one-pot reaction by 3-aminopropyl(triethoxyl)silane (APTES) with the aid of an organic base, diisopropylethylamine. Then, the primary amine reacted with a two homobifunctional cross-linker, bis(N-succinimidyl)carbonate. By modulating the reaction conditions, a high surface coverage of linking groups, succinimidyl ester, could be obtained. A similar functionalisation could be done by using (N,N’-bis(p-maleimidophenyl)methylene instead of bis(N-succinimidyl)carbonate. Succinimidyl ester is an amino-reactive group, therefore mouse monoclonal antibody bearing amino groups was grafted (Xia et al. 2006). An enzyme (Horseradish peroxidise) linked immunosorbent assay was used to evaluate the surface density of antibody. The other linker, (N,N’-bis(p-maleimidophenyl)methylene, after reflux in acetonitrile with surface amines, resulted in maleimide-terminated surfaces. Then a reduced urokinase bearing accessible thiol groups was grafted and its enzymatic activity determined.
Another procedure was followed by Fernandez and coworkers (Fernandez et al. 2008). PSi surface was firstly oxidised then modified with APTES and with the bifunctional reagent gluharaldehyde. This procedure allowed to immobilise covalently
Although these procedures confer a high degree of chemical stability to PSi surface, they may have some drawbacks. As will be seen below, the presence of silica layers at solid/liquid interface confers sensitivity to pH variations. The functionalisation steps, which substitutes silica with other chemical groups may lead to the loss of pH sensitivity and hence to the impossibility to use it for the realisation of potentiometric biosensor (Hamann and Lewis 2006). This is not, in any case, a problem for other types of signal transduction.
As noted earlier, all porous materials have a large specific surface that can reach very high area values, such as 2000 m2/g e.g. in case of activated carbon (Kaneko et al. 1992). This high value, considering an activated carbon density of about 0.5 g/cm3 (it depends on the characteristics of the activated carbon), is analogous to that obtainable with porous silicon, whose specific surface may reach more than 900 m2/cm3 in the case of
DeLouise and coworkers (DeLouise and Miller 2004) studied the enzyme immobilisation capacity of porous silicon samples using glutathione-
The insertion of active molecules in nano- or mesoporous structures also showed an improvement of enzyme stability in several materials thanks to the spatial confinement and to the interaction with the pores’ walls (Sotiropoulou et al. 2005; Kima et al. 2006). This enhanced stability is relevant for increasing shelf life, allowing the reusability of the devices and improving the biosensor-related measurement reliability.
3. Electrochemical biosensors
There are two main types of electrochemical transduction in biosensors: potentiometry and amperometry/voltammetry. They have been used for the analysis of different kinds of substances as shown in table 1.
3.1. PSi-based potentiometric biosensors
Potentiometric biosensors use porous silicon as an electrode of an electrochemical cell (Thust et al. 1996). The PSi wafer has the function both to immobilise the biological element and to be sensitive to the chemical variations produced by the immobilised biomolecule in the electrolyte solution, transducting them in a detectable electric signal. The measured parameter is the potential difference between the cathode (i.e. a platinum electrode) and the anode (PSi electrode) of the electrochemical cell. This voltage is affected both by the chemical composition of the electrolyte solution (i.e. pH) and on the semiconducting properties of PSi, which depend on the flat-band potential (
where
Type of transduction | Type of PSi | Biological element | Analyte | Reference |
Potentiometry (C-V) | p-type | Penicillase | Penicillin | (Thust et al. 1996) |
Potentiometry (C-V) | p-type | Triglycerides | (Reddy et al. 2001; Reddy et al. 2003) | |
Potentiometry (C-V) | Lipase Urease | Triglycerides Urea | (Basu et al. 2005) | |
Potentiometry (OCP) | n+-type | Triglycerides | (Setzu et al. 2007) | |
Amperometry | p-type | Cholesterol oxidase, Biliriubin oxidase, Glutamate oxidase | Cholesterol/ bilirubin | (Song et al. 2007) |
Cyclic Voltammetry | p-type | Glutamate oxidase | ALT/AST | (Song et al. 2009) |
Cyclic Voltammetry | p-type | Laccase/Azurin | - | (Ressine et al. 2010) |
Cyclic Voltammetry | p-type | DNA | tDNA | (Jin et al. 2010) |
Conductivity | p-type | Tyrosinase | Catechol | (Tembe et al. 2008) |
The site binding model is commonly applied to metal oxide sites which may either bind with a hydrogen ion at the surface,
with equilibrium constant K
with equilibrium constant K
Indeed, silicon oxide surface in contact with aqueous solutions is positively charged, negatively charged or neutral depending on the pH of the solution and on the point of zero charge (pHpzc=1/2(p
Nernst law predicts an ideal shift of surface potential of 59 mV per pH unit at 298 K. The surface potential drop across the interface is described by the equation (Yates et al. 1974): Where At Calculating the derivative of surface potential respect to pH Since pHpzc depends on the intrinsic acid-base properties of the oxide surface (
Thust and coworkers immobilised penicillinase, an enzyme sensitive to penicillin, on oxidised
Following this pioneering work, Chadha and coworkers published a series of papers aimed to the development of a potentiometric PSi-based biosensor for the analysis of triglycerides. The active biomolecule was a
Finally, the system was implemented since
Most of the works described used
In this configuration the PSi sample was used as an electrode, while a platinum electrode was used as the second electrode. Moreover, a reference electrode (saturated calomel electrode) was used to take into account properly the variations of the cell’s open circuit potential (
aqueous emulsion of tributyrin was used to ascertain the activity of the immobilised lipase. The butyric acid formed, as a result of the biocatalyst action, produces a pH change that is measured by the variation of the open circuit potential at the electrodes of the electrochemical cell (Okorn-Schmidt 1999). The open-circuit potential, also referred to as equilibrium potential or rest potential, is the overall voltage generated in the electrochemical cell measured with respect to the potential of a reference electrode. This voltage is measured in the open circuit configuration using a high-impedance voltmeter, so that only a negligible current can flow through the cell. The
The absolute value of
3.1.1. Comparison of PSi triglycerides biosensors with others literature detectors
It is interesting to compare the performances of PSi biosensors with other kinds of sensors described in the scientific literature. We will do here a brief comparison with a few literature examples of devices using completely different methods, with no pretention of exhaustivity and limited to triglycerides biosensors, to better understand how PSi bisensors place themselves with respect to other devices for the same applications. The literature values are to be compared to those reported for PSi. The PSi potentiometric detectors show, in the paper of Setzu and coworkers (Setzu et al. 2007), a linear regime, dependent on the amount of lipase immobilised within the PSi pores, up to about 10 mM. In the paper of Reddy and co-workers (Reddy et al. 2003) the authors show a calibration curve up to a few tens of mM.
These results can also be compared with the human blood normal triglycerides level which is below 1.69 mM (150 mg/dL) in the normal regime and above 5.56 mM (500 mg/dL) for a very high risk.
Method | Sensitivity | Detection range | Reference |
Potentiometric - Immobilised lipase in PSi | - | <15mM | (Setzu et al. 2007) |
Potentiometric, PSi electrode, lipase in solution | 30mV/pH unit | 5-40mM | (Reddy et al. 2003) |
Impedimetric - Polyanilyne nanotubes | 2.59 × 10−3 KΩ−1 mg−1 dL | 25–300 mg dL−1 | (Dhand et al. 2009) |
pH measurements – lipase on silica beads - tributyrin | 0.478 pH/mM | <4 mM | (Pijanowsk et al. 2001) |
pH measurements – lipase on silica beads - triacetin | 0.022 pH/mM | <30 mM | (Pijanowsk et al. 2001) |
Potentiometric - EISCAP | 500 µM | 8mM | (Fernandez et al. 2009) |
Micromechanical | 10 µM | 500µM | (Fernandez et al. 2009) |
ENFET – lipase on magnetic nanoparticles | - | 5-30mM | (Vijayalakshmi et al. 2008) |
Dhand and coworkers (Dhand et al. 2009) using a polianilyne nanotube film obtained a linear response range of 25–300 mg dL−1, a sensitivity of 2.59 × 10−3 KΩ−1 mg−1 dL, with a response time of 20 s and regression coefficient of 0.99. Pijanowsk and coworkers studied a pH detector based on silica gel beads immobilised with a lipase in a microreactor (Pijanowsk et al. 2001). The authors tested three different methods for immobilising the lipase and compared the detector sensitivity on three different triglycerides. The higher sensitivity was obtained for tributyrin with 0.478 pH/mM for concentration < 4 mM, while the widest linear range, up to 30 mM, was obtained using triacetin but with a significantly lower sensitivity (0.022pH/mM). Fernandez and coworkers compared two kind of Si-based biosensors (Fernandez et al. 2009).
The first one was an EISCAP (electrolyte–insulator–semiconductor capacitor) sensor built depositing silicon nitride on
3.2. Other types of PSi-based electrochemical biosensors
PSi can be used for the realisation of amperometric and voltammetric biosensors, besides potentiometric ones. In this case the immobilised enzyme catalyses a redox reaction involving analyte oxidation/reduction which produces a flux of electrons measured, in terms of current intensity, by the electrodes of the electrochemical cell. In these systems, PSi is usually one of the involved electrodes. In fact, due to its low conductivity properties, the PSi surface has to be modified using a conducting material as gold (Ressine et al. 2010), platinum (Song et al. 2007), or a conductive polymer (Jin et al. 2010; Zhao and Jiang 2010). In that case the main function of PSi is to act as a high surface area substrate to enhance the sensitivity of biosensing. Indeed, it has been reported that porous electrodes can increase biosensor sensitivity in comparison with flat surface electrodes. Song and coworkers (Song et al. 2006) conducted a series of experiments using a planar silicon electrode and a porous silicon electrode. Cholesterol oxidase was covalently immobilised on each electrode by silanisation. The calculated effective surface area and sensitivity of the porous electrode were about 3.1-fold larger than those of the planar electrode. Successive works of the same group were devoted to develop an electrochemical biosensor array system for the diagnosis and monitoring of liver diseases (Song et al. 2007; Song et al. 2009). This biosensor array was able to simultaneously detect four different biomarkers of liver disease, namely: cholesterol, bilirubin, alanineaminotransferase (ALT) and aspartateaminotransferase (AST) levels in aqueous fluid samples. The aim of the work was to design a miniaturised measurement system using electrodes processed by porous silicon and array technology. Enzymes (cholesterol oxidase, bilirubin oxidase and glutamate oxidase) were covalently immobilised to Pt thin-film electrodes based on the PSi substrate. The immobilisation procedure involved few steps, namely: Pt oxidation; silanisation of hydroxilated Pt film electrode with APTES (aminopropyltriethoxysilane); silanised surface activation with glutaraldehyde; covalent binding of enzymes through reaction of amino groups of lysines with aldehydic group of activated electrode surface (Fig. 7). All the immobilised enzymes catalyse the oxidation of their respective analytes yielding hydrogen peroxide. This molecule is oxidised at the Pt thin film electrode, according to:
The control of the level of these analytes in human serum is important in the diagnosis and monitoring of liver diseases and myocardial infarction.
Ressine and coworkers fabricated a novel three-dimensional chip with high surface area from double side porous silicon wafers by a sequential two-step anodic dissolution process (Ressine et al. 2010). Double-sided three-dimensional porous silicon chips, 6 mm × 6 mm, covered with a 40 nm gold (nano)layer, were fabricated from a porous silicon wafer. These devices, characterised by scanning electron microscopy along with electrochemical characterisation, showed ample conductivity, mechanical stability, and high surface area (10 times higher electrochemically active surface area compared with the geometric area). The three-dimensional gold coated silicon chips were further modified with thiol layers, followed by immobilisation of a simple copper-containing redox protein, azurin, or a complex multicopper redox enzyme, laccase. The bioelectrochemical studies showed very high surface concentrations of azurin and laccase,
Jin and co-workers fabricated and electrochemically characterised a label-free DNA sensor based on a porous silicon substrate (Jin et al. 2010). A
Tembe and coworkers developed a conductivity-based catechol biosensor using porous silicon as the immobilisation matrix for enzyme tyrosinase (Tembe et al. 2008). The enzyme was immobilised in an electrochemically etched surface of
4. Conclusions
In this chapter an overview of the recent advances on the use of porous silicon as a matrix for the realisation of biosensors is given. The main advantages of PSi biosensors, both with electrochemical or optical transduction, are the large developed surface, the relative simplicity of the fabrication process, the easy integration with standard technology, the wide range of possible applications. We focused here our attention on electrochemical (particularly potentiometric) biosensors since, despite their easy realisation, for both potentiometric and amperometric/voltammetric biosensors, few research efforts have been devoted to these devices compared with optical biosensors. Nevertheless, there is a number of promising works concerning PSi-based electrochemical biosensors now available in the literature.
The key points discussed here concern the techniques used for PSi formation, the importance of surface stabilisation by oxidation techniques, the different strategies for enzyme/biomolecule immobilisation, and the realisation of the electrochemical biosensors. Concerning the last point, we showed that the realisation of potentiometric biosensors is based on the oxidation of the PSi pores’ walls needed for both stabilising the PSi layers and confer them pH sensitivity. The transduction is due the formation of and acidic/basic substance as the result of the reaction catalysed by a suitable enzyme immobilised on the PSi surface.
Amperometric/voltammetric PSi-based biosensors, instead, measure the current intensity produced by the action of a redox reaction on the analyte of interest by an enzyme immobilised on the electrode surface. The importance of PSi is due the sensitivity enhancement due to the higher surface area of PSi compared to flat electrodes. The main difficulty in these biosensors is due to the low conductivity of PSi compared to metal electrodes. This can be overcome by coupling PSi with a Pt, Au, or a conductive polymer thin film to enhance its conductivity.
In conclusion, porous silicon constitutes an interesting matrix for the realisation of biosensing devices. Due to the recent growing interest it is very likely that previously neglected electrochemical transduction will reach a similar level of development as that obtained with optical transduction.
Acknowledgments
MIUR, PRIN 2008 grant number 2006030935 and Fondazione Banco di Sardegna are thanked for financial support.
References
- 1.
Anderson M. A. Tinsley-Bown A. Allcock P. Perkins E. Snow P. Hollings M. Smith R. G. Reeves C. Squirrell D. J. Nicklin S. Cox T. I. 2003 Sensitivity of the optical properties of porous silicon layers to the refractive index of liquid in the pores ,197 528 533 - 2.
Anderson S. H. C. Elliot H. Wallis D. J. Canham L. T. Powell J. J. 2003 Dissolution of different forms of partially porous silicon wafers under simulated phsiological conditions. ,197 331 335 - 3.
Anglin E. J. Cheng L. Freeman W. R. Sailor M. J. 2008 Porous silicon in drug delivery devices and materials . ,60 1266 1277 - 4.
Bakker E. Telting-Diaz M. 2002 Electrochemical Sensors. ,74 2781 2800 - 5.
Basu I. Subramanian R. V. Mathew A. Kayastha A. M. Chadha A. Bhattacharya E. 2005 Solid state potentiometric sensor for the estimation of tributyrin and urea . : Chemical,107 418 423 - 6.
Ben Jaballah. A. Hassen M. Hajji M. Saadoun M. Bessaïs B. Ezzaouia H. 2005 Chemical vapour etching of silicon and porous silicon: silicon solar cells and micromachining applications. ,202 1606 - 7.
Bsiesy A. Vial J. C. Gaspard F. Herino R. Ligeon M. Muller F. Romestain R. Wasiela A. Halimaoui A. Bomchil G. 1991 Photoluminescence of high porosity and of electrochemically oxidized porous silicon layers . ,254 195 200 - 8.
Buriak J. M. 2002 Organometallic Chemistry on Silicon and Germanium Surfaces. ,102 1271 1308 - 9.
Canham L. T. 1990 Silicon quantum wire array fabrication by electrochemical and chemical dissolution of wafers . ,57 1046 - 10.
Cantin J. L. Schoisswohl M. Grosman A. Lebib S. Ortega C. von Berdeleben H. J. Vázsonyi É. Jalsovszky G. Erostyák J. 1996 Anodic oxidation of p- and p+-type porous silicon: surface structural transformations and oxide formation. Thin Solid Films,276 76 79 - 11.
Chan S. Fauchet P. M. Li Y. Rothberg L. J. Miller B. L. 2000 Porous Silicon Microcavities for Biosensing Applications. (a),182 541 546 - 12.
Chan S. Li Y. Rothberg L. J. Miller B. L. Fauchet P. M. 2001 Nanoscale silicon microcavities for biosensing . ,15 277 282 - 13.
Chattopadhyay S. Li X. Bohn P. 2002 In-plane control of morphology and tunable photoluminescence in porous silicon produced by metal-assisted electroless chemical etching . ,91 6134 - 14.
Chazalviel J. N. Wehrspohn R. B. Ozanam F. 2000 Electrochemical preparation of porous semiconductors: from phenomenology to understanding . ,69-70 ,1 10 - 15.
Cullis A. Canham L. Calcott D. 1997 The structural and luminescence properties of porous silicon . ,82 909 965 - 16.
Dancil K. P. S. Greiner D. P. Sailor M. J. 1999 A Porous Silicon Optical Biosensor: Detection of Reversible Binding of IgG to a Protein A-Modified Surface . ,121 7925 7930 - 17.
De Stefano L. Arcari P. Lamberti A. Sanges C. Rotiroti L. Rea I. Rendina I. 2007 DNA optical detection based on porous silicon technology: from biosensors to biochips . ,7 214 221 - 18.
De Louise L. A. Kou P. M. Miller B. L. 2005 Cross-correlation of optical microcavity biosensor response with immobilized enzyme activity. Insights into biosensor sensitivity. ,77 3222 3230 - 19.
De Louise L. A. Miller B. L. 2004 Quantatitive assessment of enzyme immobilization capacity in porous silicon. ,76 6915 6920 - 20.
Dhand C. Solanki P. R. Sood K. N. Datta M. Malhotra B. D. 2009 Polyaniline nanotubes for impedimetric triglyceride detection . ,11 1482 1486 - 21.
Dian J. Macek A. Niznansky D. Nemec I. Vrkoslavc V. Chvojka T. Jelinek I. 2004 SEM and HRTEM study of porous silicon-relationship between fabrication, morphology and optical properties . ,238 169 174 - 22.
Fernandez R. E. Bhattacharya E. Chadha A. 2008 Covalent immobilization of Pseudomonas cepacia lipase on semiconducting materials . ,254 4512 4519 - 23.
Fernandez R. E. Hareesh V. Bhattacharya E. Chadha A. 2009 Comparison of a potentiometric and a micromechanical triglyceride biosensor . ,24 1276 1280 - 24.
Föll H. Christophersen M. Carstensen J. 2002 Formation and application of porous silicon . ,39 93 141 - 25.
Gruning U. Lehmann V. 1996 Two-dimensional infrared photonic crystal based on macroporous silicon . ,276 151 - 26.
Haes A. J. Van Duyne R. P. 2002 ,124 10596 10604 - 27.
Halimaoui H. 1993 Electrochemical and chemical behavior of porous silicon layers: the role of the material wettability and its high specific surface area . In: , D. C. Bensahel (Ed.), The Netherlands, Kluwer Academic Publisher:11 EOF - 28.
Hamann T. W. Lewis N. S. 2006 Control of the Stability, Electron-Transfer Kinetics, and pH-Dependent Energetics of Si/H/O Interfaces through Methyl Termination of Si(111) Surfaces. B,110 22291 22294 - 29.
Hamm D. Sakka T. Ogata Y. H. 2003 Etching of porous silicon in basic solution . ,197 175 179 - 30.
Harada Y. Li X. Bohn P. Nuzzo R. 2001 Catalytic amplification of the soft lithographic patterning of Si. Nonelectrochemical orthogonal fabrication of photoluminescent porous Si pixel arrays. ,123 8709 - 31.
Huy B. Hoi P. V. Khoi P. H. Van Binh D. K. P. T. Cham T. T. 2009 Porous silicon as a low dimensional and optical material . ,187 pp. Article012033 - 32.
Jane A. Dronov R. Hodges A. Voelcker N. H. 2009 Porous silicon biosensors on the advance . ,27 230 239 - 33.
Janshoff A. K. Dancil P. S. Steinem C. Greiner D. P. V. S. Lin Y. Gurtner C. Motesharei K. Sailor M. J. Ghadiri M. R. 1998 Macroporous p-Type Silicon Fabry-Perot Layers. Fabrication, Characterization, and Applications in Biosensing . ,120 12108 12116 - 34.
Jin J. H. Alocilja E. Grooms D. 2010 Fabrication and electroanalytical characterization of label-free DNA sensor based on direct electropolymerization of pyrrole on p-type porous silicon substrates . ,17 169 176 - 35.
Kaneko K. Ishii C. Ruike M. kuwabara H. 1992 Origin of superhigh surface area and microcrystalline graphitic structures of activated carbons . ,30 1075 1088 - 36.
Karlsson L. M. Tengvall P. Lundstrom I. Arwin H. J. 2003 Penetration and loading of human serum albumin in porous silicon layers with different pore sizes and thicknesses. ,266 40 47 - 37.
Kilian K. A. Boecking T. Gooding J. J. 2009 The importance of surface chemistry in mesoporous materials: lessons from porous silicon biosensors . ,630 640 - 38.
Kima J. Grate J. W. Wang P. 2006 Nanostructures for enzyme stabilization.61 1017 1026 - 39.
Kolasinski K. 2005 Silicon nanostructures from electroless electrochemical etching . ,9 73 83 - 40.
Kumar P. Lemmens P. Ghosh M. Ludwig F. Schilling M. 2009 Effect of HF Concentration on Physical and Electronic Properties of Electrochemically Formed Nanoporous Silicon . ,7 pp. - 41.
Lees I. N. Lin H. Canaria C. A. Gurtner C. Sailor M. J. Miskelly G. M. 2003 Chemical Stability of Porous Silicon Surfaces Electrochemically Modified with Functional Alkyl Species . ,19 9812 9817 - 42.
Lehmann V. 1996 Porous silicon formation and other photoelectrochemical effects at silicon electrodes anodized in hydrofluoric acid . ,106 402 405 - 43.
Lehmann V. Stengl R. Luigart A. 2000 On the morphology and the electrochemical formation mechanism of mesoporous silicon . ,69-70 ,11 22 - 44.
Lin V. S. Y. Motesharei K. K. Dancil P. S. Sailor M. J. Ghadiri M. R. 1997 A Porous Silicon-Based Optical Interferometric Biosensor . ,278 840 843 - 45.
Low S. P. Voelcker N. H. Canham L. T. Williams K. A. 2009 The biocompatibility of porous silicon in tissues of the eye . ,30 2873 2880 - 46.
Madou M. J. Loo B. H. Frese K. W. Morrison S. R. 1981 Bulk and surface characterization of the silicon electrode . ,108 135 152 - 47.
Nakato Y. Ueda T. Egi Y. Tsubomura H. 1987 Decomposition Potentials of Crystalline Silicon as Related to the Photocurrent Stability of p-n Junction Silicon Semiconductor Electrodes . ,134 353 358 - 48.
Okorn-Schmidt H. F. 1999 Characterization of silicon surface preparation processes for advanced gate dielectrics . ,43 351 365 - 49.
Ouyang H. Fauchet P. M. 2005 Biosensing using porous silicon photonic bandgap structures. ,600508 600515 - 50.
Park J. H. Gu L. Maltzahn G. v. Ruoslahti E. Bhatia S. N. Sailor M. J. 2009 Biodegradable luminescent porous silicon nanoparticles for in vivo applications . ,8 331 336 - 51.
Petrova-Koch T. Muschik A. Kux B. K. Meyer F. Koch Lehman V. 1992 ,61 943 - 52.
Pijanowsk D. G. Baraniecka A. Wiatera R. Ginalska G. Lobarzewski J. Torbicz W. 2001 The pH-detection of triglycerides.78 263 266 - 53.
Privett B. J. Shin J. H. Schoenfisch M. H. 2008 Electrochemical sensors. ,80 4499 4517 - 54.
Reddy R. R. K. Basu I. Bhattacharya E. Chadha A. 2003 Estimation of triglycerides by a porous silicon based potentiometric biosensor . ,3 155 161 - 55.
Reddy R. R. K. Chadha A. Bhattacharya E. 2001 Porous silicon based potentiometric triglyceride biosensor . ,16 313 317 - 56.
Ressine A. Vaz-Domínguez C. Fernandez V. M. De Lacey A. L. Laurell T. Ruzgas T. Shleev S. 2010 Bioelectrochemical studies of azurin and laccase confined in three-dimensional chips based on gold-modified nano-/microstructured silicon . ,25 1001 1007 - 57.
Rong G. Ryckman J. D. Mernaugh R. L. Weiss S. M. 2008 Label-free porous silicon membrane waveguide for DNA sensing . ,93 161109 - 58.
Sailor M. J. 2007 Color me sensitive: amplification and discrimination in photonic silicon nanostructures. ,1 248 252 - 59.
Salis A. Cugia F. Setzu S. Mula G. Monduzzi M. 2010 Effect of oxidation level of n+-type mesoporous silicon surface on the adsorption and the catalytic activity of candida rugosa lipase. ,345 448 453 - 60.
Salonen J.-V. Lehto P. 2008 Fabrication and chemical surface modification of mesoporous silicon for biomedical applications . ,137 162 172 - 61.
Setzu S. Ferrand P. Romestain R. 2000 Optical properties of multilayered porous silicon . ,69-70 ,34 42 - 62.
Setzu S. Salis S. Demontis V. Salis A. Monduzzi M. Mula G. 2007 Porous silicon-based potentiometric biosensor for triglycerides . ,204 1434 1438 - 63.
Setzu S. Solsona P. Létant S. Romestain R. Vial J. C. 1999 Microcavity effect on dye impregnated porous silicon samples . ,7 59 63 - 64.
Smith R. Collins S. 1992 Porous silicon formation mechanisms. Journal of App lied Physics,71 1 22 - 65.
Song M.-J. Yun D.-H. Jin J.-H. Min N.-K. Hong S.-I. 2006 Comparison of Effective Working Electrode Areas on Planar and Porous Silicon Substrates for Cholesterol Biosensor. Japanese Journal of Applied Physics,45 7197 7202 - 66.
Song M. J. Yun D. H. Hong S. I. 2009 An Electrochemical Biosensor Array for Rapid Detection of Alanine Aminotransferase and Aspartate Aminotransferase . Biotechnology and Biochemistry,73 474 478 - 67.
Song M. J. Yun D. H. Min N. K. Hong S. I. 2007 Electrochemical Biosensor Array for Liver Diagnosis Using Silanization Technique on Nanoporous Silicon Electrode . ,103 32 37 - 68.
Sotiropoulou S. Vamvakaki V. Chaniotakis N. A. 2005 Stabilization of enzymes in nanoporous materials for biosensor applications . ,20 1674 1679 - 69.
Steckl A. J. Xu J. Mogul H. C. Mogren S. 1993 ,62 1982 - 70.
Tembe S. Chaudhari P. S. Bhoraskar S. V. D’Souza S. F. Karve M. S. 2008 Conductivity-Based Catechol Sensor Using Tyrosinase Immobilized in Porous Silicon . ,8 1593 1597 - 71.
Thust M. Schöning M. J. Frohnhoff S. Arens-Fischer R. Kordos P. Lüth H. 1996 Porous silicon as a substrate material for potentiometric biosensors . ,7 26 29 - 72.
Tinsley-Bown A. M. Canham L. T. Hollings M. Anderson M. H. Reeves C. L. Cox T. I. Nicklin S. Squirrel D. J. Perkins E. Hutchinson A. Sailor M. J. Wun A. 2000 Tuning the pore size and surface chemistry of porous silicon for immunoassays . ,182 547 553 - 73.
Uhlir A. 1956 Electrolytic Shaping of Germanium and Silicon. ,35 333 347 - 74.
Ünal B. Parbukov A. N. Bayliss S. C. 2001 Photovoltaic properties of a novel stain etched porous silicon and its application in photosensitive devices . ,17 79 82 - 75.
van de Lagemaat J. Vanmaekelbergh D. Kelly J. J. 1998 Photoelectrochemical characterization of 6H-SiC . ,83 6089 6095 - 76.
Vijayalakshmi A. Tarunashree Y. Baruwati B. Manorama S. V. Narayana B. L. Johnson R. E. C. Rao N. M. 2008 Enzyme field effect transistor (ENFET) for estimation of triglycerides using magnetic nanoparticles ,23 1708 1714 - 77.
Xia B. Xiao S.-J. Guo D.-J. Wang J. Chao J. Liu H. B. Pei J. Chen Y.-Q. Tang Y. C. Liu J.-N. 2006 Biofunctionalisation of porous silicon (PS) surfaces by using homobifunctional cross-linkers. Journal of Materials Chemistry,16 570 578 - 78.
Xiong Z. Zhao F. Yang J. Hu X. 2010 Comparison of optical absorption in Si nanowire and nanoporous Si structures for photovoltaic applications . ,96 181903 - 79.
Yates D. E. Levine S. Healy T. W. 1974 Site-binding model of the electrical doble layer at the oxide/water interface. ,70 1807 1818 - 80.
Zhao Z. Jiang H. 2010 Enzyme-based Electrochemical Biosensors . In: , P. A. Serra, InTech.
Notes
- The surface potential drop across the interface is described by the equation (Yates et al. 1974):