Open access peer-reviewed chapter

Anticorrosive, Antimicrobial, and Bioactive Titanium Dioxide Coating for Surface‐modified Purpose on Biomedical Material

Written By

Hsi‐Kai Tsou and Ping‐Yen Hsieh

Submitted: 24 November 2016 Reviewed: 28 March 2017 Published: 26 July 2017

DOI: 10.5772/intechopen.68854

From the Edited Volume

Application of Titanium Dioxide

Edited by Magdalena Janus

Chapter metrics overview

1,771 Chapter Downloads

View Full Metrics

Abstract

A multifunctional titanium dioxide (TiO2) coating was used to provide anticorrosive, antimicrobial, and bioactive properties for the surface modification of biomedical materials because TiO2 has a stable bonding structure, photocatalytic characteristics, and negatively charged surfaces in nature. For successful deposition, an arc ion plating technique was adopted to deposit the TiO2 coating. The antimicrobial activity values of anatase‐TiO2‐coated stainless steel specimens against Staphylococcus aureus and Escherichia coli were 3.0 and 2.5, respectively, which are far beyond the value designated in JIS Z2801:2000 industrial standard. TiO2 coatings on stainless steel also generate an increased (i.e., less negative) corrosion potential and decreased corrosion current in a sodium chloride solution, showing a reduced tendency and rate of substrate dissolution as well as a reduced coating of species into the electrolyte. In addition, TiO2 coatings, especially with rutile phase, satisfied the requirements for activating the biological property of a polymeric polyetheretherketone surface. Therefore, TiO2 is a promising surface modification for the biomedical materials used in surgical instruments and implants.

Keywords

  • anticorrosive
  • antimicrobial
  • bioactive
  • titanium dioxide
  • biomedical material

1. Introduction

Biomedical material is any substance that has been engineered to interact with biological systems for a medical purpose, which may be therapeutic (i.e., to treat, augment, repair, or replace malfunctioning tissue in the body) or diagnostic. Among the various types of biomedical materials, metallic materials are the most widely used because of their high load‐supporting capacity, desirable qualities of wear and friction, and acceptable biocompatibility. Stainless steel, titanium, and their alloys are considered especially promising materials for surgical instruments and implants of many types and sizes. Polymeric materials have also garnered considerable interest in research and development as soft‐ and hard‐tissue replacements, on the basis of the ease of manufacturing and modifying such materials, and their appropriate physical, chemical, and mechanical properties.

When biomedical materials come in contact with physiological tissue and body fluids, various interactions, such as corrosive reaction, inflammation, and host response, are triggered. For this reason, knowing and understanding the surface properties of biomedical materials are crucial. Unfortunately, metallic materials are easily influenced by corrosion damage due to electrochemical reactions; additionally, the bioinertness and hydrophobic surface properties render polymeric materials unfavorable for cell adhesion. Long‐term clinical experiments have also indicated that the primary causes of implant failure include not only unstable implant fixation to bone tissue, but also bacterial infection.

To overcome the aforementioned problems, a surface modification technique that uses a multifunctional titanium dioxide (TiO2) coating is introduced to provide anticorrosive, antimicrobial, and bioactive properties for the underlying biomaterial. These versatile natural features of TiO2 are attributed to its stable bonding structure, photocatalytic characteristics, and negatively charged surfaces. In this paper, a brief overview of TiO2 coating modification in the field of biomedical material is provided. The two main topics discussed in the next section are as follows:

  • Antimicrobial and anticorrosive titanium dioxide coating on stainless steel to reduce hospital‐acquired infection.

  • Bioactive titanium dioxide coating on polyetheretherketone for spinal implant application.

Advertisement

2. Antimicrobial and anticorrosive titanium dioxide coating on stainless steel to reduce hospital‐acquired infection

2.1. Background

The increasing incidence and host risk of device‐related infections that result in morbidity and even mortality have been noted for some time, particularly regarding the spread of antibiotic‐resistant bacteria, such as methicillin‐resistant Staphylococcus aureus and bursting Clostridium difficile. These hospital‐acquired infections are a worldwide problem [1]. The outbreaks of SARS and avian influenza have also drawn attention to novel preventative measures, including the development and application of antimicrobial materials, to enhance the conventional disinfection concept. This movement compelled us to develop an antimicrobial technique for medical implements in clinical use.

Antimicrobial or antibacterial refers to the inhibition of bacterial growth and reproduction [2]. Antimicrobial functions can be performed by essential materials themselves or through the use of coating materials. One example of an essential antimicrobial alloy material is stainless steel that has been doped with copper. This material forms when ε‐copper precipitates in a steel matrix; specifically, copper ions can be dissolved into a surface‐passivated chromium oxide film. Such creates an antimicrobial effect on the stainless steel surface, resulting in the inhibition of bacterial growth [3]. The similar antimicrobial metal alloys, such as copper‐containing ferritic stainless steel [4], martensitic stainless steel [5], and austenitic stainless steel [6], were also developed. On the other hand, for the antimicrobial purpose on coating materials, the idea of coatings containing with copper, silver, zinc, and other antimicrobial active metals was considered [7]. Unfortunately, such substance may induce the corrosion reaction because of the undesired Galvanic effect between two metals, which may be unsustainable during service. In this regard, TiO2 with anatase (A‐TiO2) phase may be the promising candidates for antimicrobial purposes.

The antimicrobial effects of TiO2 are attributed to its photocatalytic characteristics, as discovered by Fujishima and Honda [8]. The photocatalytic process of TiO2 involves the generation of electron‐hole pairs when the material is exposed to light that emits energy exceeding the band gap energy of TiO2. The aggressive superoxide ions (O2−) are generated by the electron attack, and the holes accelerate hydroxyl radical (OH) formation on the material surface [9, 10]. These active radicals subsequently inhibit the growth of germs and bacteria that are known to be antimicrobially active through the direct oxidation of intracellular coenzyme, reducing the respiratory activity and thereby causing cell death [11].

In the present study, arc ion plating (AIP) was used to deposit a TiO2 coating on common medical‐grade AISI 304 stainless steel. The antimicrobial efficacy of the TiO2‐coated stainless steel specimens was then evaluated according to the JIS standard. The corrosion resistance of the TiO2 coating was also examined to determine whether such films can be stable in a physiological environment. The results suggest that this modification may be effective as an antimicrobial surface coating for medical implements to reduce the risk of hospital‐acquired infections.

2.2. Preparation of antimicrobial and anticorrosive TiO2 films

TiO2 deposition was conducted using a typical AIP technique and involved three steps: argon ion bombardment, bottom titanium layer deposition, and TiO2 coating deposition. The ion bombardment was performed to clean and mildly preheat the substrate, followed by the bottom titanium layer deposition, which enhanced the adhesion between the substrate and TiO2 coating.

The wide acceptance indicates that an A‐TiO2 phase structure is the key factor for maximizing the antimicrobial efficiency of TiO2. This corresponds to a specific condition with 100% oxygen pressure at 0.5 Pa by using the AIP technique with a cathode target voltage of 20 V and a cathode target current of 90 A. Under this optimized deposition condition, the proportion of A‐TiO2 in the TiO2 coating has been reported to be 76.8% [1214].

2.3. Antimicrobial characteristics of TiO2‐coated stainless steel

The JIS Z2801:2000 [15] was employed as a standard to test the antimicrobial efficacy of TiO2‐coated stainless steel specimens. The bacterial strains used in this test were Gram‐positive Staphylococcus aureus (S. aureus, ATCC 6538P) and Gram‐positive Escherichia coli (E. coli, ATCC 8739) with an initial concentration of 4.0 × 105 bacteria/mL. In the antimicrobial test, the specimens were divided into three groups: group A and group B consisted of uncoated stainless steel specimens, and group C consisted of TiO2‐coated stainless steel specimens. The specimens in group A immediately underwent serial dilution and plate culture after inoculation, while the specimens of groups B and C were incubated with exposure to fluorescent lighting for 24 h. The fluorescent lamp used was a regular daily‐living light source that emitted mainly visible light and had a weak emission of 365 nm. Antimicrobial activity (R) of the specimens in all three groups was then calculated.

As revealed in Figure 1 [13], the petri dishes corresponding to groups A and B (the uncoated stainless steel specimens) presented significant numbers of S. aureus and E. coli bacterial colonies, respectively; by contrast, the TiO2‐coated stainless steel specimens in group C did not show a significant amount of bacterial colonies. This qualitatively describes the antimicrobial ability of the TiO2 coating. Although only one out of the three petri dishes corresponding to each group is pictured in Figure 1, those not shown revealed a similar situation; this confirms the statistical accuracy of the antimicrobial test.

Figure 1.

S. aureus and E. coli colonies formed on petri dishes after 24 h on the (a) group A stainless steel specimens, (b) group B stainless steel specimens, and (c) TiO2‐coated stainless steel specimens [13].

For both S. aureus and E. coli, the numbers of viable bacteria for groups A, B, and C are compared in Figure 2 [13]. The group A specimens showed 2.85 × 105 and 1.06 × 105 viable bacteria cells, respectively, for S. aureus and E. coli, whereas the group B specimens showed 1.04 × 104 and 1.36 × 104 viable bacteria cells, respectively, for S. aureus and E. coli. By contrast, the group C specimens showed no bacterial colonies (10 bacteria cells) for S. aureus and 4.30 × 101 viable bacteria cells for E. coli. Based on these results, the TiO2‐coated stainless steel specimens presented R values of 3.0 and 2.5, respectively, for S. aureus and E. coli. Such values are far beyond the index of 2 stipulated for the JIS test standard.

Figure 2.

Viable bacteria numbers of S. aureus and E. coli for (a) group A stainless steel specimens, (b) group B stainless steel specimens, and (c) TiO2‐coated stainless steel specimens [13].

To further investigate the antimicrobial mechanism of a TiO2 coating, the bacterial microstructure was observed using transmission electron microscopy (TEM; JEOL JEM‐1230). This closer examination revealed that most of the S. aureus cells were retained their integrity as the cells were inoculated on bare stainless steel with the exposure to fluorescent light for 24 h; moreover, the complete cell structure, including the cell wall, cytoplasmic membrane, cytoplasma, and nucleoid, was observed. The cells were undergoing mitosis, as presented in Figure 3(a) [14], was also found. These results indicate that the inoculated S. aureus cells on bare stainless steel were not deactivated by the fluorescent light. However, for the S. aureus cells on the TiO2‐coated stainless steel specimens, detachment of the cell wall from the cell membrane was frequently observed in the microscopic field (Figure 3(b) [14]). As has been noted elsewhere [1618], the cell walls in these specimens are attacked by superoxide ions and hydroxyl radicals, and lipid peroxidation caused polyunsaturated phospholipids in the cell membrane to be destroyed; similarly, the degeneration of the membranes in the present study caused the detachment of the cell walls from the cell membranes.

Figure 3.

Cell structures of S. aureus inoculated on (a) bare stainless steel and (b) TiO2‐coated stainless steel specimens, following continuous exposure to a fluorescent lamp for 24 h. (The arrow indicates detachment of cell wall from the cell membrane.) [14].

A high percentage of the E. coli cells inoculated on bare stainless steel and exposed to fluorescent light for 24 h also retained their integrity, as depicted in Figure 4(a) [14]. By contrast, a large amount of E. coli cell fragments were observed following inoculation on TiO2‐coated stainless steel specimens and exposure to fluorescent light for 24 h, as presented in Figure 4(b) [14]. This occurred because E. coli cell walls are too thin to protect against attack by superoxide ions and hydroxyl radicals, resulting in massive death. A closer examination of the E. coli cells reveals that the nucleoid structures in the cytoplasma tend to give way to features of condensation, as indicated by the arrow in Figure 4(b) The degeneration of E. coli in response to photocatalysis found in the present study is similar to the degeneration that was observed in response to the antimicrobial effects of silver ions [16].

Figure 4.

Cell structures of E. coli inoculated on (a) bare stainless steel and (b) TiO2‐coated stainless steel specimens, following continuous exposure to a fluorescent lamp for 24 h. (The arrows indicate the condensation features of the nucleoid) [14].

2.4. Anticorrosive characteristics of TiO2‐coated stainless steel

A potentiodynamic polarization test was carried out in a potentiostat (EG&G 263 A) according to the ASTM G44–99 standard [19] to evaluate the corrosion resistance of a TiO2 coating in a 3.5 wt.% sodium chloride electrolyte. A saturated silver/silver chloride electrode was used as the reference, with a platinum counter electrode; a TiO2‐coated stainless steel specimen was inserted as the working electrode.

Figure 5 illustrates the potentiodynamic polarization curves of bare stainless steel and TiO2‐coated stainless steel specimens [20]. The corrosive potential (Ecorr) and corrosive current (Icorr) were −0.54 V and 6.0 × 10−8 A/cm2, respectively, for the bare stainless steel specimens. Once the specimens had been coated with TiO2, the Ecorr and Icorr of the specimens were −0.42 V and 1.0 × 10−8 A/cm2, respectively. Notably, TiO2 is an inorganic compound, and its inertness in corrosive environments (e.g., a sodium chloride solution) helps reduce the tendency and rate of substrate dissolution and species coating in an electrolyte. This increases the corrosive potential and decreases the corrosive current, as noted herein.

Figure 5.

Polarization curves of bare stainless steel and TiO2‐coated stainless steel specimens in a 3.5 wt.% sodium chloride solution [20].

In summary, the research results show that A‐TiO2 adds effective antimicrobial characteristics to stainless steel. The key to providing efficient antimicrobial efficacy lies in the photocatalytic performance of the coating, which originates from the anatase phase. Furthermore, based on the TEM observation results, the antimicrobial mechanisms that inhibit S. aureus and E. coli bacteria under the photocatalytic action of A‐TiO2 are different; specifically, the antimicrobial efficacy of A‐TiO2 against E. coli is more thorough. The A‐TiO2 coating also reduces the overall rate of corrosion and increases the corrosion barrier, compared with the features of bare stainless steel.

Advertisement

3. Bioactive titanium dioxide coating on polyetheretherketone for spinal implant application

3.1. Background

Orthopedic implants have become one of the most highly developed fields in hard‐tissue replacement. Polyetheretherketone (PEEK) polymer, with its high chemical resistance, radiolucency to X‐ray scanning, and low elastic modulus similar to human cancellous bone, has become a highly preferred biomaterial, providing a promising alternative to metallic implants [21]. In particular, the elastic modulus can avoid the stress shielding effect, and prevent compression fractures and osteopenia syndrome; the X‐ray radiolucency characteristic does not present a medical image shielding problem. PEEK can also be sterilized and shaped by machining to fit the contour of bones [22]. Consequently, PEEK has been widely used for load‐bearing orthopedic applications, including dental implants, screws, and spinal interbody fusion cages [23, 24].

Despite these excellent properties, PEEK is still categorized as a bioinert material because of its hydrophobic feature and inertness with the surrounding tissue [21]. To overcome this problem, two primary strategies, bulk modification and surface modification, have been proposed to enhance the bone fusion ability of the PEEK. Bulk modification incorporates various bioactive materials, such as hydroxyapatite (HA) [25], strontium‐containing hydroxyapatite [26], β‐tricalcium phosphate [27], or TiO2 [28], into the PEEK matrix to form PEEK‐based biocomposites. However, their tensile strength and toughness decrease as more of the bioactive materials are added, resulting in a substantial increase in the elastic modulus of these biomedical composites; the biomechanical property of these PEEK‐based biocomposites is therefore no longer similar to that of human cancellous bone [21]. Conversely, surface modification only alters the surface properties of a material, without adversely affecting its bulk properties. In other words, surface modification is a more suitable approach for adapting PEEK to be used as implant. Consequently, various surface modification approaches have been developed to promote the hydrophilic and biological characteristics of PEEK, such as using plasma treatment to change the surface chemistry [29], using chemical treatment to graft functional groups [30], and using laser treatment to roughen the surface [31]. Moreover, adding a functional coating to PEEK to create a bioactive surface is a more effective method for enhancing osseointegration performance [3238]. Functional coating materials include HA [32], titanium [33, 34], TiO2 [3537], and diamond‐like carbon [38]. To date, by taking the advantage of good biocompatibility of titanium with human body, very thick titanium produced over PEEK surface via vacuum plasma spray for spinal implant has been clinically available.

It has been well established that under humid conditions, the surface of TiO2 generates hydroxyl groups (─OH), followed by the conjunction with calcium ions (Ca2+) and phosphate groups (PO43−) from physiological fluid. Therefore, bone‐like apatite compounds can be formed on the TiO2 surface to induce osteoblast cell adhesion and proliferation [39, 40]. Based on the results, TiO2 has been reported to exhibit excellent biocompatibility and further classified as a bioactive material [39, 40]. Furthermore, TiO2 demonstrated excellent osseointegration ability, according to the animal experiment study [41]. These biological characteristics render TiO2 film an even more promising material for the successful modification of PEEK surfaces, in comparison with regular titanium film.

In this research, the AIP technique was used to deposit TiO2 films with controllable A‐TiO2 and rutile (R‐TiO2) phases onto PEEK substrates. The investigation focused on determining the effects of introducing a TiO2 coating on the in vitro and in vivo characteristics of TiO2‐coated PEEK specimens, and evaluating the ability of the modified PEEK in a clinical application to shorten the osseointegration period for spinal implants and bone tissues.

3.2. Preparation of biocompatible TiO2 films

The detailed AIP‐TiO2 deposition work is described in Section 2.2. The deposition conditions used in this section are listed in Table 1; target current and substrate bias were systematically manipulated to achieve specific ratios of A‐TiO2 and R‐TiO2 in the deposited films, characterized by a fixed 100% oxygen pressure of 0.5 Pa and a cathode target voltage of 20 V.

Sample codeTarget current (A)Substrate bias (V)A‐TiO2 content (%)
60A0V60092.7
90A0V90076.8
90A20V90–2046.6
90A25V90–2521.5
90A30V90–309.1

Table 1.

Deposition conditions and the proportions of A‐TiO2 phases for TiO2 coatings.

Based on the microstructure characteristics results [12], the AIP process can successfully fabricate TiO2 films of varying A‐TiO2 and R‐TiO2 composition when appropriate coating parameters are used. Specifically, the A‐TiO2 phase in the deposited films ranged from 9.1% to 92.7% (Table 1). A low target current promotes the growth of A‐TiO2, whereas a high substrate bias induces the formation of R‐TiO2. The mechanism behind this outcome was previously investigated [12].

3.3. In vitro characteristics of TiO2‐coated PEEK

First, the MC3T3‐E1 osteoblast cell line was used in the osteoblast compatibility test to assess the cell adhesion test, cell proliferation test, cell differentiation test, and osteogenesis performance [namely quantification of osteopontin (OPN), osteocalcin (OCN), and calcium content]. Next, the cell morphology that had attached to the PEEK and TiO2‐coated PEEK specimens was observed using field emission scanning electron microscopy (FESEM; Hitachi S‐4800).

Figure 6 shows the osteoblast cell adhesion ability, cell proliferation ability, cell differentiation ability, and osteogenesis performance on the PEEK and TiO2‐coated PEEK specimens at various deposition conditions [36]. Notably, the osteoblast cell adhesion, proliferation, and differentiation abilities on TiO2‐coated PEEK specimens were superior to the bare PEEK specimens for all of the deposition conditions. This indicates that all of the obtained TiO2 coatings possessed cell induction capabilities, which led to accelerated cell adhesion and growth and increased cell proliferation and maturity. These three indicators confirmed the osteoblast compatibility of the TiO2‐coatings deposited on PEEK. Furthermore, the osteogenesis performance (revealed by OPN, OCN, and calcium content as shown in Figure 6(d)(f) [36], respectively) demonstrated that TiO2 coatings also significantly increased the osteogenesis performance. This suggests that TiO2 coatings enhance extracellular bone matrix growth. Figure 6 [36] also shows that the specimen 90A30V, which was the richest in R‐TiO2 phase, exhibited the most osteoblast compatibility.

Figure 6.

(a) Cell adhesion ability, (b) cell proliferation ability, (c) cell differentiation ability, (d) OPN, (e) OCN, and (f) calcium content of the osteoblast inoculated on bare PEEK and TiO2‐coated PEEK specimens with various deposition conditions [36].

Figure 7 shows the morphologies of the osteoblast cells after they were cultured for 0.5 and 48 h on PEEK and TiO2‐coated PEEK specimens at different deposition conditions [36]. Specifically, the morphology of osteoblast cells on the bare PEEK specimens remained spherical without the appearance of filopodium, suggesting the poor adhesion to the specimen. By comparison, osteoblasts on the TiO2‐coated PEEK specimens with the culturing time of 0.5 h showed a very comfortable adhesion features, that is, the filopodia extension and well‐developed lamellipodia on the cells; this was particularly notable on the films with high ratios of R‐TiO2 to A‐TiO2. Similar results were observed in the cells cultured for 48 h. Overall, these results further confirm that a deposited film with high R‐TiO2 content has superior osteoblast growth.

Figure 7.

Morphologies of the osteoblasts cultured for 0.5 and 48 h on (a) the bare PEEK specimens, and the TiO2‐coated PEEK specimens at different deposition conditions: (b) 60A0V, (c) 90A0V, (d) 90A20V, (e) 90A25V, and (f) 90A30V [36].

Furthermore, bare PEEK, and TiO2‐coated PEEK specimens were then immersed in a simulated body fluid (SBF) for 1, 3, 7, 14, and 28 days, to investigate the effect of TiO2 coating on the ability to induce HA formation. The TiO2 coatings that possessed A‐TiO2 and R‐TiO2 under the deposition conditions of 60A0V and 90A30V, respectively, were examined. This biomimetic immersion test is a valuable approach for evaluating bioactivity of a candidate bone implant material prior to an in vivo test [42].

Figure 8 illustrates the X‐ray diffraction (XRD) patterns of bare PEEK, A‐TiO2‐coated PEEK, and R‐TiO2‐coated PEEK specimens after immersion in the SBF for a varying number of days [43]. During the early immersion period, the diffraction peaks that are ascribed to PEEK showed no observable change, indicating that the growing layer was undetectable in all of the specimens. After 28 days of immersion, weak and broadened diffraction peaks that are ascribed to HA were found, as shown in Figure 8(a) [43]. This implies that a very poor crystalline or even amorphous calcium phosphate layer had formed on the PEEK specimens. By contrast, after only 7 days and 3 days of immersion in the SBF solution, diffraction peaks that are ascribed to HA could be observed in A‐TiO2‐ and R‐TiO2‐coated PEEK specimens, respectively. Over time, the intensity of these diffraction peaks increased significantly, as shown in Figure 8(b) and (c) [43], suggesting that additional crystalline HA was formed on them. Overall, these results suggest that HA growth in a SBF solution can be enhanced by adopting TiO2 coatings, and that the R‐TiO2 coating seems to exhibit a superior capability to induce HA formation. Therefore, the results of the biomimetic immersion tests agree well with the finding of in vitro characteristics from osteoblast compatibility tests.

Figure 8.

XRD patterns of the (a) bare PEEK, (b) A‐TiO2‐coated PEEK, and (c) R‐TiO2‐coated PEEK specimens immersed in a SBF for 1, 3, 7, 14, and 28 days [43].

3.4. In vivo characteristics of TiO2‐coated PEEK

Bullet‐shaped PEEK implants with a diameter of φ 4.0 mm × L 6.0 mm were used in an animal experiment. Bare PEEK, A‐TiO2‐coated PEEK, and R‐TiO2‐coated PEEK implants were inserted into the femurs of New Zealand white male rabbits to evaluate the in vivo osseointegration capacity through the push‐out test and histological observation.

The push‐out test can precisely quantify the degree of fixation between an implant and bone tissues [44]. Figure 9 shows the push‐out test results for the three implants after 4, 8, and 12 weeks [37]. Notably, the shear strength between the bone tissues and the implant increased as implantation time increased; at 12 weeks, the shear strength of the bare, A‐TiO2‐coated, and R‐TiO2‐coated PEEK implants was 2.54 MPa, 3.02 MPa, and 6.51 MPa, respectively. It was thus concluded that the bare PEEK implant had the poorest shear strength, but this could be enhanced by adding a TiO2 coating. Overall, the R‐TiO2 coating had the optimal fixation.

Figure 9.

Shear strength between bone tissues and mplant for the (a) bare PEEK implant, (b) A‐TiO2‐coated PEEK implant, and (c) R‐TiO2‐coated PEEK implant at 4, 8, and 12 weeks after implantation [37].

To identify the failure mode between the implant and bone tissues after the push‐out test, FESEM was adopted to observe the fracture morphology of the implant surface at 12 weeks, as shown in Figure 10 [37]. It was noted that new bone tissue had fully peeled off the surface of the bare PEEK implant (Figure 10(a) [37]), indicating that failure occurred at the bone/PEEK interface. Thus, the osseointegration capacity of a bare PEEK implant is poor. By contrast, when a TiO2 coating was applied to the implant, a large area of the residual bone tissue adhered to the surface of the implant (Figure 10(b) and (c) [37]). Additionally, a particularly large amount of residual bone tissue on the R‐TiO2‐coated PEEK implant surface was confirmed by elemental mapping, as revealed in Figure 10(d) [37]. These analytical results indicate that TiO2‐coated implants have a superior ability to induce bone growth and achieve bone ingrowth. The A‐TiO2‐coated PEEK implants experienced some coating detachment, resulting in a mixed adhesive failure between the A‐TiO2 coating and PEEK substrate, as well as cohesive failure of the bone itself. However, the R‐TiO2‐coated PEEK implant surfaces were almost completely covered with new bone tissue, almost no film detachment from the implants was observed, and thus, the failure can be regarded as cohesive failure by the bone tissue itself.

Figure 10.

Fracture morphology of the (a) bare PEEK implant, (b) A‐TiO2‐coated PEEK implant, and (c) R‐TiO2‐coated PEEK implant with (d) the composition analysis of its bone tissues and implant interface after the push‐out test conducted at 12 weeks [37].

Figure 11 depicts the histological sections of the three implants at 4, 8, and 12 weeks after implantation [37]. Notably, new bone tissue that was generated by bone remodeling had formed mature lamellar bone, and directly connected to the TiO2‐coated PEEK implants after 4 weeks, indicating excellent osseointegration performance. Thus, it was concluded that TiO2 coating exhibits strong osteoblast compatibility and rapidly activates bone remodeling. Subsequently, the coating induced adhesion and proliferation of osteoblasts on the implant surface, and differentiation into osteocytes for the production of new bone tissue and later bone bonding. Conversely, new lamellar bone on the surface of the bare PEEK implants was not completely mature and not fully bonded with the implant.

Figure 11.

Histological sections of the bare PEEK implant, A‐TiO2‐coated PEEK implant, and R‐TiO2‐coated PEEK implant at 4, 8, and 12 weeks after implantation [37].

The response of the TiO2‐coated PEEK implants in the marrow cavity (located far from the cortical bone) at 4 weeks indicated that regenerated bone tissues grew onto the implant surfaces; moreover, this new bone is the result of bone tissue repair, which proliferates from the endosteum of cortical bone. Due to the osteoconductive effect, the new bone tissues grew inward to the implant surfaces in the marrow [45]. These findings indicate that TiO2 coatings have excellent osteoconductivity and promote new bone growth on the TiO2‐coated PEEK implant surfaces, with connections to cortical bone. By contrast, the surfaces of the bare PEEK implant were covered with fibrous tissue, implying that bone bonding did not occur between the implant and the cortical bone. Fibrous tissue growth is likely caused by micro movement in the implant and poor stability during the early implantation period [46].

When the implant period was extended to 8 weeks, immature osteogenesis was observed in the cortical bone around the bare PEEK implant, and new bone tissue was maturing after 12 weeks. However, fibrous tissue was still identified at the interface between the implants and bone tissues, indicating that the osseointegration capacity of bare PEEK implants is very limited, even when the implantation period is extended. By contrast, 8 weeks after the implantation of the TiO2‐coated PEEK implants, histological sections in the marrow cavity revealed that the new bone tissue was maturing and osteocytes covered the their surface. In other words, the osteoconductive effect of TiO2 coating triggers quick bone remodeling. The new bone was fully mature and closely integrated with the TiO2 coating in the cavity after 12 weeks (Figure 11 [37]). However, a comparison of the TiO2 coatings with different phase structures indicated that the degree of bone bonding between new bone and the R‐TiO2‐coated PEEK implant was significantly better than that between new bone and the A‐TiO2‐coated PEEK implant. In addition, some gaps existed between the A‐TiO2 coating and the new bone in some areas; detachment of the A‐TiO2 coating was also noted.

In summary, the in vitro and in vivo characteristics can be improved by TiO2 coating because of its bioactivity; R‐TiO2 coatings perform particularly well, promoting biomimetic HA growth, osteoblast compatibility, and osseointegration. These phenomena are attributable to the abundance of negatively charged hydroxyl groups on the R‐TiO2 surface [3537].

Advertisement

4. Conclusions

In this chapter, TiO2 coatings prepared using the AIP technique to alter the surface properties of biomaterials were described. Specifically, it was found that introducing TiO2 coating to stainless steel and PEEK specimens adds various anticorrosive, antimicrobial, and bioactive surface properties to the materials, which were systematically reviewed herein. The following conclusions can be drawn:

  1. Owing to the efficient photocatalytic performance of its anatase phase structure, A‐TiO2‐coated stainless steel exhibits excellent antimicrobial efficacy against S. aureus and E. coli bacteria. The material could possibly serve as a new antimicrobial treatment for surgical instruments and medical implements to reduce the risk of hospital‐acquired infections.

  2. The high corrosion resistance of TiO2 coatings in a 3.5 wt% sodium chloride solution was postulated as a direct consequence of its ceramic nature, suggesting that TiO2 is electrochemically inert in the human body environment.

  3. Based on the in vitro and in vivo tests, the bioactivity and osseointegration of all TiO2 coatings were far superior to bioinert PEEK; moreover, R‐TiO2 coatings exhibited greater performance than A‐TiO2 coatings because of the abundance of negatively charged hydroxyl groups on its surface. Consequently, TiO2‐coated PEEK specimens are suggested for use in clinical applications.

  4. Overall, the aforementioned results prove that TiO2 coatings are highly suitable for surface modifications of biomedical materials.

References

  1. 1. Gottenbos B, Van der Mei HC, Klatter F, Grijpma DW, Feijen J, Nieuwenhuis P, Busscher HJ. Positively charged biomaterials exert antimicrobial effects on gram‐negative bacilli in rats. Biomaterials. 2003;24:2707-2710
  2. 2. Kourai H. Surface science and microbiology: Antimicrobial finishings. Journal of the Surface Science Society of Japan. 2001;22:663-670
  3. 3. Morihiro H, Katsuhisa M, Sadao H, Michiyuki K, Seiichis I. Improving method of antibacterial performance of copper‐containing stainless steel. Japan Patent. 1996;08:229107
  4. 4. Chungtsun TH, Tachiupao C, Kungnan KC. Antibacterial characteristic and material quality of ferrite antibacterial stainless steel “NSSAM‐1”. Nisshin Steel Technical Report. 1997;76:48-55
  5. 5. Chungtsun TH, Tachiupao CJ, Shanpen CJ. Corrosion quality of galvanized stainless steel. Nisshin Steel Technical Report. 1998;77:69-81
  6. 6. Shihyeh HH, Yutienchuan CK, Panching CN, Yuen KH. Antimicrobial pre‐painted steel sheet. Nisshin Steel Technical Report. 1998;78:90-96
  7. 7. Kim TN, Feng QL, Kim JO, Wu J, Wang H, Chen GC, Cui FZ. Antimicrobial effects of metal ions (Ag+, Cu2+, Zn2+) in hydroxyapatite. Journal of Materials Science: Materials in Medicine. 1998;9:129-134.
  8. 8. Fujishima A, Honda K. Electrochemical photolysis of water at a semiconductor electrode. Nature. 1972;238:37-38
  9. 9. Cheng Q, Li C, Pavlinek V, Saha P, Wang H. Surface‐modified antibacterial TiO2/Ag+ nanoparticles: Preparation and properties. Applied Surface Science. 2006;252:4154-4160
  10. 10. Sawunyama P, Yasumori A, Okada K. The nature of multilayered TiO2‐based photocatalytic films prepared by a sol‐gel process. Materials Research Bulletin. 1998;33:795-801
  11. 11. Matsunaga T, Tomoda R, Nakajima T, Wake H. Photoelectrochemical sterilization of microbial cells by semiconductor powders. FEMS Microbiology Letters. 1985;29:211-214
  12. 12. Chung CJ, Tsou HK, Chen HL, Hsieh PY, He JL. Low temperature preparation of phase‐tunable and antimicrobial titanium dioxide coating on biomedical polymer implants for reducing implant‐related infections. Surface and Coatings Technology. 2011;205:5035-5039
  13. 13. Chung CJ, Lin HI, Tsou HK, Shi ZY, He JL. An antimicrobial TiO2 coating for reducing hospital‐acquired infection. Journal of Biomedical Materials Research Part B: Applied Biomaterials. 2008;85B:220-224
  14. 14. Chung CJ, Lin HI, Chou CM, Hsieh PY, Hsiao CH, Shi ZY, He JL. Inactivation of Staphylococcus aureus and Escherichia coli under various light sources on photocatalytic titanium dioxide thin film. Surface and Coatings Technology. 2009;203:1081-1085
  15. 15. JIS Z2801:2000. Antimicrobial products—Test for antimicrobial activity and efficacy. Japanese Industrial Standard. 2001
  16. 16. Huang Z, Maness PC, Blake DM, Wolfrum EJ, Smolinski SL. Bactericidal mode of titanium dioxide photocatalysis. Journal of Photochemistry and Photobiology A: Chemistry. 2000;130:163-171
  17. 17. Maness PC, Smolinski S, Blake DM, Huang Z, Wolfrum EJ, Jacoby WA. Bactericidal activity of photocatalytic TiO2 reaction: Toward an understanding of its killing mechanism. Applied and Environmental Microbiology. 1999;65:4094-4098
  18. 18. Feng QL, Wu J, Chen GQ, Cui FZ, Kim TN, Kim JO. A mechanistic study of the antibacterial effect of silver ions on Escherichia coli and Staphylococcus aureus. Journal of Biomedical Materials Research. 2000;52:662-668
  19. 19. ASTM G44-99. Standard practice for exposure of metals and alloys by alternate immersion in neutral 3.5% sodium chloride solution. American Society for Testing and Materials. 2005
  20. 20. Chung CJ, Hsieh PY, Hsiao CH, Lin HI, Leyland A, Matthews A, He JL. Mutifunctional arc ion plated TiO2 photocatalytic coatings with improved wear and corrosion protection. Surface and Coatings Technology. 2009;203:1689-1693
  21. 21. Kurtz SM, Devine JN. PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials. 2007;28:4845-4869
  22. 22. Barton AJ, Sagers RD, Pitt WG. Bacterial adhesion to orthopedic implant polymers. Journal of Biomedical Materials Research. 1996;30:403-410
  23. 23. Williams DF, McNamara A, Turner RM. Potential of polyetheretherketone (PEEK) and carbon‐fibre‐reinforced PEEK in medical applications. Journal of Materials Science Letters. 1987;6:188-190.
  24. 24. Schwitalla A, Müller WD. PEEK dental implants: A review of the literature. Journal of Oral Implantology. 2013;39:743-749
  25. 25. Yu S, Hariram KP, Kumar R, Cheang P, Aik KK. In vitro apatite formation and its growth kinetics on hydroxyapatite/polyetheretherketon biocomposites. Biomaterials. 2005;26:2343-2352
  26. 26. Wong KL, Wong CT, Liu WC, Pan HB, Fong MK, Lam WM, Cheung WL, Tang WM, Chiu KY, Luk KDK, Lu WW. Mechanical properties and in vitro response of strontium‐containing hydroxyapatite/polyetheretherketone composites. Biomaterials. 2009;30:3810-3817
  27. 27. Petrovic L, Pohle D, Münstedt H, Rechtenwald T, Schlegel KA, Rupprecht S. Effect of βTCP filled polyetheretherketone on osteoblast cell proliferation in vitro. Journal of Biomedical Science. 2006;13:41-46
  28. 28. Wu X, Liu X, Wei J, Ma J, Deng F, Wei S. Nano‐TiO2/PEEK bioactive composite as a bone substitute material: In vitro and in vivo studies. International Journal of Nanomedicine. 2012;7:1215-1225
  29. 29. Ha SW, Kirch M, Birchler F, Eckert KL, Mayerv J, Wintermantel E, Sittig C, Pfund‐Klingenfuss I, Textor M, Spencer ND, Guecheva M, Vonmont H. Surface activation of polyetheretherketone (PEEK) and formation of calcium phosphate coatings by precipitation. Journal of Materials Science: Materials in Medicine. 1997;8:683-690
  30. 30. Pino M, Stingelin N, Tanner KE. Nucleation and growth of apatite on NaOH‐treated PEEK, HDPE and UHMWPE for artificial cornea materials. Acta Biomaterialia. 2008;4:1827-1836
  31. 31. Riveiroa A, Soto R, Comesaña R, Boutinguiza M, del Val J, Quintero F, Lusquiños F, Pou J. Laser surface modification of PEEK. Applied Surface Science. 2012;258:9437-9442
  32. 32. Wu GM, Hsiao WD, Kung SF. Investigation of hydroxyapatite coated polyether ether ketone composites by gas plasma sprays. Surface and Coatings Technology. 2009;203:2755-2758
  33. 33. Han CM, Lee EJ, Kim HE, Koh YH, Kim KN, Ha Y, Kuh SU. The electron beam deposition of titanium on polyetheretherketone (PEEK) and the resulting enhanced biological properties. Biomaterials. 2010;31:3465-3470
  34. 34. Walsh WR, Bertollo N, Christou C, Schaffner D, Mobbs RJ. Plasma‐sprayed titanium coating to polyetheretherketone improves the bone‐implant interface. The Spine Journal. 2015;15:1041-1049
  35. 35. Tsou HK, Hsieh PY, Chung CJ, Tang CH, Shyr TW, He JL. Low‐temperature deposition of anatase TiO2 on medical grade polyetheretherketone to assist osseous integration. Surface and Coatings Technology. 2009;204:1121-1125
  36. 36. Tsou HK, Hsieh PY, Chi MH, Chung CJ, He JL. Improved osteoblast compatibility of medical‐grade polyetheretherketone using arc ion plated rutile/anatase titanium dioxide films for spinal implants. Journal of Biomedical Materials Research Part A. 2012;100A:2787-2792
  37. 37. Tsou HK, Chi MH, Hung YW, Chung CJ, He JL. In vivo osseointegration performance of titanium dioxide coating modified polyetheretherketone using arc ion plating for spinal implant application. BioMed Research International. 2015;2015:328943
  38. 38. Wang H, Xu M, Zhang W, Kwok DTK, Jiang J, Wu Z, Chu PK. Mechanical and biological characteristics of diamond‐like carbon coated poly aryl‐etherether‐ketone. Biomaterials. 2010;31:8181-8187
  39. 39. Wu JM, Xiao F, Hayakawa S, Tsuru K, Takemoto S, Osaka A. Journal of Materials Science: Materials in Medicineacidic titanium tetrafluoride solution. Journal of Materials Science: Materials in Medicine. 2003;14:1027-1032
  40. 40. Lindberg F, Heinrichs J, Ericson F, Thomsen P, Engqvist H. Hydroxylapatite growth on single‐crystal rutile substrates. Biomaterials. 2008;29:3317-3323
  41. 41. Erli HJ, Rüger M, Ragoβ C, Jahnen‐Dechent W, Hollander DA, Paar OO, von Walter M. The effect of surface modification of a porous TiO2/perlite composite on the ingrowth of bone tissue in vivo. Biomaterials. 2006;27:1270-1276
  42. 42. Kokubo T, Kim HM, Kawashita M. Novel bioactive materials with different mechanical properties. Biomaterials. 2003;24:2161-2175
  43. 43. Chi MH, Tsou HK, Chung CJ, He JL. Biomimetic hydroxyapatite grown on biomedical polymer coated with titanium dioxide interlayer to assist osteocompatible performance. Thin Solid Film. 2013;549:98-102
  44. 44. Dhert WJA, Jansen JA. The validity of a single pushout test. In: An YH, Draughn RA, editors. Mechanical testing of bone and the bone–implant interface. New York: CRC Press; 2000. pp. 477-487
  45. 45. Dua C, Meijer GJ, van de Valk C, Haan RE, Bezemer JM, Hesseling SC, Cui FZ, de Groot K, Layrolle P. Bone growth in biomimetic apatite coated porous Polyactive® 1000PEGT70PBT30 implants. Biomaterials. 2002;23:4649-4656
  46. 46. Søballe K, Hansen ES, Rasmussen HB, Jørgensen PH, Bünger C. Tissue ingrowth into titanium and hydroxyapatite‐coated implants during stable and unstable mechanical conditions. Journal of Orthopaedic Research. 1992;10:285-299

Written By

Hsi‐Kai Tsou and Ping‐Yen Hsieh

Submitted: 24 November 2016 Reviewed: 28 March 2017 Published: 26 July 2017