Abstract

In this chapter, we will review the applications of augmented reality (AR) and virtual reality (VR) in diagnostic radiology. We begin the chapter by discussing state of the art medical imaging techniques to include scanner hardware, spatial resolution, and presentation methods to the radiologist or other medical professionals. We continue by discussing the methodology of a technique called Depth-3-Dimensional (D3D) imaging, which transforms cross-sectional medical imaging datasets into a left and right eye images that that can be viewed with state of the art AR or VR headsets. We include results of the D3D processed AR/VR imaging and conclude with a discussion of the path forward.

Keywords

  • augmented reality
  • virtual reality
  • diagnostic radiology
  • D3D
  • depth perception

1. Introduction

The purpose of this chapter is to review the applications of augmented reality (AR) and virtual reality (VR) in diagnostic radiology. This introduction section will provide a brief general summary of AR technologies, VR technologies, AR/VR applications in medicine and AR/VR applications diagnostic radiology. The remainder of the paper will discuss state-of-the-art medical imaging systems, current methods of viewing medical images, the imaging processing techniques for generating an AR/VR image, AR/VR imaging results and path forward.

1.1. Introduction to AR/VR

AR technologies can be classified into AR systems or mixed reality (MR) systems. In both the AR and MR systems, the user wears a head mounted display (HMD), which provides the simultaneous display of a virtual image and the scene of the real-world surroundings [1]. The differences are that in AR, the virtual image is transparent like a hologram and in MR the virtual image appears solid. Examples of AR systems include the Meta and DAQRI systems. An example of an MR system is the Microsoft HoloLens. In AR systems, the user can view the virtual image and interact with the real-world scene.

VR technologies may be classified as fully immersive, semi-immersive or non-immersive [2]. In fully immersive VR systems such as the Oculus Rift and the HTC Vive, the HMD displays a virtual image and the real-world surroundings are completely occluded from the user’s field of view [3]. In semi-immersive VR systems such as the Samsung Gear VR, the HMD displays the virtual image, but the real-world scene is only partially occluded from the user’s field of view [3]. In VR systems, the user can maneuver through the virtual world by head movements via HMD tracking or walking via external camera tracking systems. Additional ways that a user can interact with the virtual environment include voice gestures or through handheld devices with haptic feedback. The HMDs for the AR and VR systems display a unique image to each eye; therefore, stereoscopic imaging and depth perceptions is achieved. Both AR and VR are rapidly growing fields. Worldwide revenues for AR/VR were $5 billion in 2016, but are expected to increase to $162 billion by 2020 [3].

1.2. Applications of AR/VR in medicine

In 2015, the United States spent 17% of its gross domestic product on healthcare. 32% of the healthcare spending was during hospital stays [4]. Nearly half of hospital costs are for surgical care [5]. Therefore, approximately 3% of the gross domestic product includes costs related to surgery. There is a drive to improve efficiency in the operation room to both improve patient care and drive down costs. AR/VR holds promise in accomplishing these goals through improving pre-operative planning and enhancing intra-operative surgical procedures [6].

1.3. Applications of AR/VR in diagnostic radiology

In the United States, the total cost of diagnostic imaging has been estimated to be $100 billion in 2006 alone [7]. Utilization rates of diagnostic imaging are on the rise [8, 9, 10]. Currently, there is no United States Food and Drug Administration (FDA) approved AR/VR system used in diagnostic radiology [11]. AR/VR provides enhanced viewing including depth perception and improved human machine interface (HMI) [12, 13]. AR/VR HMDs provide unique images to each eye yielding depth perception. AR/VR systems leverage advanced gaming controllers and joysticks to improve HMI. Because of these features of AR/VR and others discussed later in this chapter, we believe there will be increasing applications of AR/VR in diagnostic radiology in the future.

Advertisement

2. Current state-of-the-art diagnostic medical imaging

Diagnostic radiology plays a major role in medicine as it provides precise anatomic and physiologic information to physicians enabling diagnosis of complex disease and monitoring response to treatment. This section will be organized into three subsections including medical imaging equipment, conventional imaging techniques and the advanced 3D rendering methods.

2.1. Medical imaging equipment

The field of diagnostic radiology includes a wide variety of imaging equipment including systems that generate inherently 2D images (e.g., chest radiograph) as well as systems that generate volumetric medical imaging datasets (e.g., computed tomography (CT), magnetic resonance imaging (MRI), and positron emission tomography (PET)). In this section, we will focus on the latter.

2.1.1. Computed tomography (CT)

In order to perform a CT scan (also known as a CAT scan), the patient is placed in the horizontal position on the CT scanner table. See Figure 1. The table is then translated through a donut shaped device containing both an X-ray tube and X-ray detector. Multiple projection images are acquired as the X-ray tube and detector assembly rotate around the patient.

Figure 1.

Photograph of a computed tomography (CT) scanner. The patient is being transferred from onto the mobile CT scanner table that will translate through the donut-shaped device during the CT scan and take multiple projection images.

Image reconstruction algorithms, such as filtered back projection, are performed in order to generate cross-sectional images in the axial plane (x-y plane). Since sequential, contiguous axial images can be obtained, coronal plane (x-z) and sagittal plane (y-z) plane images can be reconstructed. Data is stored in digital imaging and communication in medicine (DICOM) files with a typical matrix of a CT scan is 512 × 512 pixels. A pixel in the axial plane is a 2D object with a discrete length in the x-direction and discrete length in the y-direction. A voxel is a 3D object created with a pixel by adding a third dimension to create a volume.

Each pixel has an associated gray-scale value called a Hounsfield Unit (HU), which is a function of the density and composition of the tissue. As a reference, water has a HU of 0. Soft tissues (e.g., brain, kidney, muscle, etc.) are slightly denser than water and have a HU of approximately 30–40. Compact bone can have a density of 400. Fat is slightly less dense than water and has a HU of approximately −100. Air is significantly less dense than water and has a HU of −1000.

A process called “windowing and leveling” is performed by the radiologist to set the window “level” and window “width.” The window “level” refers to the HU number where mid-gray can represent. The window “width” is the range of shades of gray such that any value more extreme than the range is white (if more dense) or black (if less dense). See Figure 2.

Figure 2.

Axial CT scan of the head in “bone window” (A), “brain window” (B) and “scalp window” (C) at the same level within the brain. In bone window (A), the window level is set to 400 (to optimize visualization of bone which has a density of approximately 400 HU) and the width is set to 3000. In the brain window (B), the window level is set to 30 (to optimize visualization of brain which as a density of approximately 30 HU) and the width is set to 80. Any pixels that are below −10 (30 (window level) minus 40 (1/2 window width)) will be black. Air with a HU of −1000 is black because it is less than −10. Fat (yellow star) with a HU of −100 is black because it is less than −10. Any pixels that are higher than 70 (30 (window level) plus 40 (1/2 window width)) will be white. In this image, the bone is white. In the scalp window (C), the window level is set to 30 and the window width to 800. Note that fat (blue star) with a HU of −100 is now gray since it falls within the range of the window level of 30 and width of 800, which would be 30±400 or −370 to 430.

Modern CT scans can be perform a total body head-to-toe scan in less than 15 seconds with spatial resolution of less than 1 mm. Thus, CT scans can be invaluable in the setting of trauma since it enables the radiologist to diagnose critical injuries anywhere in the body, such as traumatic brain injury [14]. This provides critical information to the neurosurgeon who can then perform life-saving interventions.

2.1.2. Magnetic resonance imaging (MRI)

In order to perform an MRI scan, the patient is placed in the horizontal position on the MRI scanner table within a large cylindrical shaped device. A large magnetic field is directed through the long axis of the cylinder. Transmit coils direct a radiofrequency (RF) pulse into the patient’s body and receive coils process the returning electromagnetic signal from the body to create an image. Similar to CT, contiguous planar images can be stacked and axial, sagittal and coronal reformats can be reconstructed. The imaging data for a MRI scan is similar to that of a CT scan in matrix size and the fact that each pixel has a comparable gray scale.

Unlike CT scanners, MRI scanners do not employ ionizing radiation and are therefore ideal for young children or pregnant women who are more vulnerable to radiation. Furthermore, MRI scans have the ability to perform exceptional contrast resolution between tissues of similar density and can diagnose certain types of traumatic brain injury that cannot be seen on CT scans [15]. Modern MRI scans require significantly more time to perform than CT scans, nearly one-hour of image acquisition time for a MRI scan of the brain.

2.1.3. Positron emission tomography (PET) scanner

In order to perform a PET scan, a radiopharmaceutical (e.g., fludeoxyglucose F-18) is administered to the patient. Then, the patient is placed in the horizontal position on the PET scanner table within a donut shaped device. As the radiopharmaceutical decays, photons are emitted from within the patient and are received by the PET detector crystals. As with CT and MRI, axial images can be stacked and sagittal and coronal images can be reconstructed.

A typical matrix of a PET scan is 128 × 128, which is smaller than that of CT or MRI. However, it is similar in that each pixel has an associated numerical value associated with it indicating gray scale. In the case of PET, the gray scale corresponds to the amount of radioactivity emitted from that location. Since the radiopharmaceuticals can target certain structures in the body, it is possible for improved diagnosis of conditions like Alzheimer’s disease compared with MRI [16].

2.2. Current methods of viewing medical imaging

In the previous section, we discussed three types of medical imaging scanners that generate volumetric data. In this section, we will review the current methods of viewing the volumetric data.

2.2.1. Conventional viewing of the volumetric data

The conventional viewing method for reviewing volumetric datasets is a slice-by-slice viewing method for axial, sagittal and coronal imaging planes or on occasion oblique reformats. It is estimated that most radiologists spend more than 95% of their total time on cross-sectional imaging datasets using this conventional slice-by-slice approach [17]. Certain anatomical structures are easier to visualize on particular imaging planes. As an example, some radiologists have a preference for viewing the midline structures in the brain on the sagittal images, as shown in Figure 3.

Figure 3.

Image of the brain from an MRI scanner. This is of an axial T1-weighted sequence (A) with coronal (B) and sagittal (C) images reconstructed.

Occasionally, the radiologist needs to view an abnormality in another plane, other than the axial, sagittal, coronal imaging planes. In these instances, oblique plane reformats can be used with the images still viewed in a conventional slice-by-slice approach. See Figure 4. Curved planar reformats can also be performed and have been shown to be beneficial [18].

Figure 4.

CT scan of the chest in lung windows showing oblique reformats (A, B) and coronal reformat (C). Note that the image (A) is the oblique plane image corresponding to the yellow line in (C). Note that the image (B) is the oblique plane image corresponding to the green line in (C). Note that image (C) is the coronal plane image corresponding to the red line in (A) and (B).

A standard viewing method includes a flat screen, high-resolution diagnostic imaging monitor with keyboard and mouse, as shown in Figure 5. Typically, the radiologist will use the wheel on the back of the mouse to scroll through the stacks of images. Note that the radiologist also uses a microphone to dictate the radiology reports.

Figure 5.

Image of the typical diagnostic radiology workstation displaying images of a breast MRI. Note that four flat-screen monitors are present. Methods for human machine interaction (HMI) to manipulate the images include a mouse and keyboard.

2.2.2. Challenges in conventional viewing of the volumetric data

First, is the challenge of information overload. The dramatic improvements in spatial resolution (commonly smaller than 1 mm) of CT and MRI coupled with large portions of the body imaged generate immense datasets and the radiologists face the challenge of information overload. As an example, a CT scan of the chest with an axial matrix of 512 pixels (x-direction) by 512 pixels (y-direction) would have 262,144 pixels on a single slice. Thin-cut imaging of the chest provides 500 axial slices, each containing the 262,144 pixels, or roughly 131 million pixels in the data set.

Second, is the challenge of detecting small lesions. One example of this is the challenge is the identification a small pulmonary nodule, which is a topic of great concern for radiologists and a top cause of litigation [19, 20]. Identifying a tumor at an small size and corresponding early stage is important in order to improve patient survival and reduce cost of treatment. A very deliberate slice-by-slice method takes considerable time.

Third, is the challenge of mentally building a 3D image from reviewing slices [21]. Depending on the clinical scenario and body part imaged, the radiologist can be tasked with following certain twisting and turning structures through the body such as following a blood vessel or loops of intestines.

2.2.3. Advanced (non-AR/VR) viewing methods of the volumetric data

Most radiologists spend a small fraction (<5%) of their total time in interpreting their imaging scans with advanced viewing methods [17]. Such non-AR/VR techniques include surface rendering and volume rendering.

2.2.3.1. Surface rendering

The first 3D rendering technique to display the human body’s anatomy was surface rendering (also known as shaded surface display). Through segmentation techniques such as thresholding to display only a prescribed set of pixels, apparent surfaces are displayed within the body. A virtual light source is used to provide surface shading. In surface rendering, only a single surface is used. An advantage of only displaying a single surface is the fact that surface rendering techniques are typically not limited by overlapping tissues within the human body. However, there are a few limitations.

One limitation of only displaying a single surface is the fact that thresholding is used for one tissue type at a time and it can be difficult to understand the anatomic relationship of multiple different organ systems when only a single organ system is displayed. Another limitation is the fact that many organs are of similar density to their surroundings and it can be difficult to segment these structures out. Finally, since surface rendering images have been displayed on flat screen images true depth perception is not achieved. See Figure 6.

Figure 6.

Shaded surface display (SSD) of the bones of the head and neck in three different projections (A, B, C). Note that while the density of the skull is intrinsically the same HU, the gray scale displayed changed due to the apparent light source. Note that true depth perception is not achieved because this figure is displayed on a flat computer screen or paper.

2.2.3.2. Volume rendering

The technique of volume rendering has been researched for many years by the computer graphics industry and has recently been applied to diagnostic radiology [22]. In volume rendering, a transfer function is applied to assign a color and opacity to each intensity value. As an example, voxels that correspond to the density of blood vessel are colored red and voxels that correspond to the density of bone are colored white. This has significantly helped radiologists visualize complex 3D structures [23]. See Figure 7.

Figure 7.

Volume rendering images of the bones and blood vessels of the head and neck in three different projections (A, B, C). Note that this differs from the shaded surface display since these volume rendering images showed both the bones and the blood vessels and the SSD only displayed the bones. Note that the carotid artery can be seen in (A). Note that in image (C), the shading from the apparent light source has been removed.

One of the key limitations of volume rendering is the overlapping structures [24, 25]. This limitation is significantly worse in settings such as viewing of the vasculature of the brain. See Figure 8.

Figure 8.

Volume rendering image of an magnetic resonance angiogram (MRA), which is a type of magnetic resonance imaging (MRI) of the brain. Note the extensive areas of overlap, which limits evaluation.

Advertisement

3. Review of AR/VR in diagnostic radiology

In the first sections of this chapter, we have reviewed the medical imaging equipment, conventional slice-by-slice techniques and the advanced 3D rendering methods including surface rendering and volume rendering. We will now review AR/VR in diagnostic radiology.

The previously discussed limitation of overlapping structures can be minimized through a process called depth-3-dimensional (D3D) imaging by providing unique images to each eye and display on AR/VR HDUs. D3D transforms cross-sectional imaging datasets and displays them on AR/VR head display units (HDUs). In doing so, an overall immersive viewing experience is created. In AR/VR radiology, the radiologist will wear a head display unit, which can be either a VR, AR or MR system. The basic concept is outlined in Figure 9.

Figure 9.

Overview of the system. Cross-sectional imaging data is sent via the digital recording device to the general purpose processor and then sending a unique image to the left eye and a unique image to the right eye a head display unit (HDU) [26].

This section of the paper will be organized into three sections. First, we will discuss the process for generating an AR/VR image from a cross-sectional medical imaging dataset. Second, we will include results. Third, we will discuss the path forward and future opportunities of AR/VR in diagnostic radiology.

3.1. Image processing for AR/VR

To optimize visualization, the D3D software suite must provide the capability for visualizing medical imagery rendered in a true 3 dimensional representation [26, 27]. In order to accomplish this, input imagery is converted from 2 dimensional images into a 3 dimensional voxel space, segmented into distinct tissue types; and then filtered. Finally, rendering is then performed wherein the rendering engine computes a left and right view. This allows the operator to visualize the data using the same stereopsis that our eyes and brains have spent a lifetime interpreting and processing.

3.1.1. 3D segmentation

During the 3D segmentation process, each image pixel in the input imagery is treated as a voxel, a three dimensional entity with length, width, and height. Each voxel is read into a large 3D array. The 3D array is analyzed for similarity to its neighbors, and for context clues that imply similarity to historical training data.

Mean and variance statistics largely drive the similarity analysis. Every voxel is compared to all the neighboring voxels in the 3x3 (or 5x5 or larger) nearest neighbor region. The most similar voxels are assigned the same class designation. The 3x3 variance is used to determine the width of the local intensity distribution. The width of the local intensity distribution in turn determines the threshold applied for declaring similarity between voxels. Simple local features on the neighborhood are computed around each voxel to describe the local average intensity or texture surrounding a voxel, which provides a classifier algorithm with measurements that are used to distinguish one class from another, where classes correspond to the different tissue types.

Historical data from numerous similar imaging exams will be collected and “ground truthed” so that a neural network or deep learning classifier can be used to improve classification accuracy. Techniques will be used to first classify the input imagery into categories such as skeletal, fat, normal breast tissue, breast cancer, etc. [26, 27]. The user will need to input the tissue-type selection until an intelligent system will automatically determine the anatomy of the input imagery. User input of the tissue-type selection will be used as historical data to drive the classifier.

3.1.2. 3D filtering algorithm

The goal of the 3D filtering algorithm is to make the most understandable representation of the imagery to the radiologist [26, 27]. In order to accomplish this goal, it utilizes the image tissue-type from the 3D segmentation algorithm to determine which tissue types should be given highest priorities.

Current graphic hardware have limitations on how many voxels that can be displayed. Therefore, it is necessary to allocate percentages to the various tissue types based on priority. Those tissue voxels that must be seen from a clinically importance perspective (e.g., tumor) are given the highest percentage. Other tissues that are not as clinically important (e.g., subcutaneous fat) are sparsely sampled. There are two benefits of this process. First of all, this allows the operator to understand the high priority tissue (e.g., tumor) in proper context (e.g., tumor is touching the spine). It can be difficult or impossible to understand exact position of the tumor when all the surrounding tissue has been removed. Second of all, this lower priority tissue needs to be somewhat more transparent so the higher priority tissue can be seen through the lower priority tissue. However, transparency allocated may not necessarily correspond to tissue density. Additionally, thinning the tissue between the viewing perspectives and the area of interest makes it much easier to see the area of interest accurately.

User selected filtering is employed to enable the operator to slice away tissue sections to facilitate viewing of areas of interest. Additional implementation of various geometric volumes and surfaces to temporarily remove sections of the voxel cloud may be beneficial. The operator can also enable or disable display of specific tissue types.

3.1.3. 3D rendering engine

The goal of the 3D rendering is to replicate, insofar as possible, what a radiologist would see if looking with his/her own two eyes through the skin and into the body were possible. A slightly different image would be seen from each eye from its particular vantage point. This difference allows one to see depth. The basic concept is illustrated in Figure 10. The geometry behind the approach is illustrated in Figures 11 and 12.

Figure 10.

The 3D volume shows a gray voxel and a black voxel. In volume rendering (A), the gray voxel is hidden behind the black voxel and only the black voxel is displayed on the 2D volume rendering image. In depth-3-dimensional (D3D), both the gray and the black voxels are projected onto the imaging plane; furthermore, the user can distinguish that the gray voxel is farther than the black voxel because the D3D process displayed with AR/VR HDUs provides depth perception.

Figure 11.

This is an overview of the depth-3-dimensional (D3D) processing system for the left eye viewing perspective (LEVP). Reprinted from the Douglas et al. [12] with permission from Dove Medical Press Ltd.

Figure 12.

This is an overview illustrating the fact that the angle from the left eye viewing perspective (LEVP) to voxel (0, 64, 64) is different from the angle from the right eye viewing perspective (REVP) to voxel (0, 64, 64). This voxel (0, 64, 64) will be displayed on a different spot on the left eye display as compared with the right eye display within the HDU.

The Left Eye Viewing Perspective (LEVP) would be at an observer/ radiologist selected point to inspect the volume. The Right Eye Viewing Perspective (REVP) would be the inter-ocular distance to the right of the LEVP and similar math would be applied to generate the slightly different image presented to the right eye display. The rendering engine generates the left and right eye views to provide a true 3D data visualization to be displayed in the AR/VR HDUs [26, 27]. This component relies heavily on the Graphics Processing Units (GPU) built into the graphics card. The rendering engine supports a convergence depth adjustment for fine-tuning the operator’s focal point as shown in Figure 13.

Figure 13.

Figure illustrates left eye viewing perspective (LEVP) and right eye viewing perspective (REVP). The angles from the LEVP through the volume of interest can converge to a convergence point “c” to enhance visualization of a small volume of interest [26, 27].

The interocular distance can of course be altered to change stereopsis. The rendering engine allows the operator to comfortably move the viewing position, zoom, and rotate about the pitch, roll or yaw axes. The angular field of view can be changed. The volume of interest (VOI) can be rotated. The viewing perspectives can also be rotated. See Figure 14. And to improve visibility, the operator can select any number of color palettes for different tissue classes. The rendering engine also incorporates a graphical overlay for image markup.

Figure 14.

Five examples of augmented reality/virtual reality viewing options with the D3D technology. Notes: (A) initial viewing angle into the volume of interest. (B) Illustrates increasing the interocular distance, which provides for increased binocular disparity. (C) Illustrates changing of the angular FOV, so rather than an α FOV of 40°, it changes to an α FOV of 10° and rather than a β FOV of 40°, it changes to a β FOV of 10°. This serves to focus on a particular region within the volume. (D) Illustrates rotation of the VOI, so that the radiologist can have a different viewing perspective. (E) Illustrates rotation of the viewing perspective, which is similar to the radiologist turning the head to see new features of the image, allowing for improved HMI. Note that in (A–E), the center pixels (i.e., α = 0° and β = 0°) for both LEVP and REVP converge at the center of the VOI (i.e., voxel [32, 32, 32]), such that the VOI is optimally presented to the user. Abbreviations: FOV, field of view; HMI, human machine interface; LEVP, left eye viewing perspective; REVP, right eye viewing perspective; VOI, volume of interest. Reprinted from the Douglas et al. [12] with permission from Dove Medical Press Ltd.

3.2. AR/VR control systems used in diagnostic radiology

The field of AR/VR is evolving with new and changing HDUs and gaming control systems coming on the market. In order to illustrate what we have used in our prior research, we illustrate example control systems shown and how various control buttons correspond with basic maneuvers and advanced maneuvers. See Figures 1517.

Figure 15.

Control system used by depth-3-dimensional (D3D) technologies. See example buttons and functions.

Figure 16.

This figure illustrates a graphical user interface (GUI) used by D3D technologies programmed with basic movements, which are defined as a single controls input at one time.

Figure 17.

This figure illustrates a graphical user interface (GUI) used by D3D technologies programmed with complex movements, which are defined as a single controls input at one time.

3.3. Initial results of AR/VR in radiology

As previously discussed, there have been numerous applications of AR/VR in surgery to include pre-operative planning, education, and intra-operative assistance [6]. However, AR/VR is not yet FDA approved in diagnostic radiology, but is being actively researched by DXC Technologies/D3D Technologies with an initial focus in breast cancer imaging [11, 12, 13].

Breast Cancer is one of the leading causes of death in women [28, 29]. Breast calcifications are extremely common and are present in up to 86% of mammograms [30]. The calcifications are classified by distribution with a linear and branching pattern suspicious for ductal carcinoma in situ (DCIS) [31, 32]. Standard mammographic views may not reveal the true linear and branching pattern due to suboptimal view point and lack of depth perception. Therefore, the D3D AR imaging system was tested on a simulated set of microcalcifications. The radiologist who rated the AR system found that when the microcalcifications appeared as a cluster when viewed from a single perspective, but with rotation and the AR HDU appeared as a linear and branching pattern. See Figure 18.

Figure 18.

Simulated breast microcalcifications as white dots on a black background viewed using the D3D system. Notes: The microcalcifications are shown in (A), (B), and (C), but in different projections. (A) Initial single projection viewed by the radiologist (EW) was thought to represent a cluster as indicated by the circle. Note that some of the dots are bigger, which is due to the fact that they are closer to the viewing perspective and the calcifications subtend a larger angular resolution from the viewing perspective. (B) The cluster is now rotated and viewed with both an LEVP and an REVP. The top portion (arrows) was thought to represent a linear pattern by the radiologist (EW) with the linear portion indicated by the arrows. (C) Same cluster as (A) and (B), but now rotated once again. The bottom portion (arrows) was thought to represent a branching pattern by the radiologist (EW), where two branches originate at the bottom of the image as indicated by the arrows. In retrospect, the viewing perspective (A) was closest to the top of the microcalcifications as if one were looking down on linear, branching pattern seen from the side view on (B) and (C). Abbreviations: D3D, depth 3-dimensional; LEVP, left eye viewing perspective; REVP, right eye viewing perspective. Reprinted from the, Douglas et al. [12] with permission.

In addition to microcalcifications, breast cancer can also present as a mass on imaging. Characterization of the shape and margins is important in determining whether the breast mass is malignant or benign. Dedicated breast CT provides high spatial resolution of breast masses [33]. Recent data has shown the importance of characterizing tumor morphology [34]. Therefore, the D3D AR system was also tested in viewing of a known breast cancer [13]. Malignant features of spiculations were noted to be more conspicuous on the D3D system than the native CT. See Figure 19.

Figure 19.

(A) Contrast-enhanced breast CT demonstrates the mass with small spiculations extending from the margins. (B and C) same mass from breast CT exam as seen in (A), but viewed with D3D where (B) represents the left eye viewing perspective (LEVP) and (C) represents the right eye viewing perspective (REVP). The red box illustrates the 3D cursor used. (D&E) represent the same mass from the breast CT, but zoomed in and viewed from a different perspective with (D) representing the LEVP and (E) representing the REVP. Red arrows show spiculations extending from the margins of the mass. The red circle represents a spiculation sticking out toward the user, which was well seen when rotating with the D3D system. Reprinted from the Douglas et al., [13] with permission.

3.4. Future work of AR/VR in radiology

Both the fields of AR/VR and diagnostic radiology are large and rapidly growing. One of the most common reasons for performing diagnostic imaging is for cancer. Early determination of whether a particular therapy regimen is working would be extremely helpful to improve survival and would save costs. We outline a flow chart below as a recommended process for introducing AR/VR evaluating a cancer at multiple time points. See Figure 20.

Figure 20.

Overview of how D3D AR/VR system could be used to evaluate how a tumor at multiple time points.

It is foreseeable that AR/VR will one day play a major role in diagnostic radiology. Computer aided diagnosis (CAD) will help to identify the abnormalities. The role of the radiologist will include assessing an abnormality in great detail to appreciate subtle changes to diagnose accurately and assess treatment response. Features of AR/VR including depth perception, head tracking, improved GUIs create an overall immersive environment allowing for new opportunities in diagnostic radiology. Radiologists will interact with medical images in ways never before including voice commands, gestures or through handheld devices with haptic feedback.

Advertisement

4. Conclusions

Both diagnostic radiology and AR/VR are large and rapidly growing fields. In this chapter, we reviewed diagnostic medical imaging equipment, data storage, conventional slice-by-slice analysis and advanced 3D rendering techniques including surface rendering and volume rendering. We then introduced D3D processing of images, so volumetric medical imaging can be displayed in AR/VR HDUs. We showed a variety of GUIs including controllers and joysticks with a variety of functions achieved by various functions. Imaging cases were illustrated with a specific focus on breast cancer. We concluded with a discussion on future techniques in comparing how a tumor changes appearance over multiple time points.

Advertisement

Conflict of interest

All authors either have direct financial interest or are employees of D3D Technologies or DXC Technologies.

References

  1. 1. Lovo EE, Quintana JC, Puebla MC, Torrealba G, Santos JL, Lira IH, et al. A novel, inexpensive method of image coregistration for applications in image-guided surgery using augmented reality. Neurosurgery. 2007;60(4 Suppl 2):366-371; discussion 71-2
  2. 2. Baus O, Bouchard S. Moving from virtual reality exposure-based therapy to augmented reality exposure-based therapy: A review. Frontiers in Human Neuroscience. 2014;8:2
  3. 3. Merel T. The 7 Drivers of the $150 billion AR/VR Industry. Aol Tech; 2015. https://techcrunch.com/2015/07/08/the-7-drivers-of-the-150-billion-arvr-industry/
  4. 4. Statistics NCfH. Health, United States, 2016: With chartbook on long-term trends in health. 2017
  5. 5. Insider B. Available from: http://www.businessinsider.com/20-most-expensive-surgeries-in-the-us-2014-2
  6. 6. Douglas DB, Wilke CA, Gibson D, Petricoin EF, Liotta L. Virtual reality and augmented reality: Advances in surgery. O Biologico. 2017;2(5):1-8
  7. 7. Mitchell JM, LaGalia RR. Controlling the escalating use of advanced imaging: The role of radiology benefit management programs. Medical Care Research and Review. 2009;66(3):339-351
  8. 8. Mettler FA Jr, Wiest PW, Locken JA, Kelsey CA. CT scanning: Patterns of use and dose. Journal of Radiological Protection. 2000;20(4):353-359
  9. 9. Mitchell DG, Parker L, Sunshine JH, Levin DC. Body MR imaging and CT volume: Variations and trends based on an analysis of medicare and fee-for-service health insurance databases. American Journal of Roentgenology. 2002;179(1):27-31
  10. 10. Boone JM, Brunberg JA. Computed tomography use in a tertiary care university hospital. Journal of the American College of Radiology. 2008;5(2):132-138
  11. 11. Douglas DB, Wilke CA, Gibson JD, Boone JM, Wintermark M. Augmented reality: Advances in diagnostic imaging. Multimodal Technologies and Interaction. 2017;1(4):29
  12. 12. Douglas DB, Petricoin EF, Liotta L, Wilson E. D3D augmented reality imaging system: Proof of concept in mammography. Medical Devices (Auckland). 2016;9:277-283
  13. 13. Douglas DB, Boone JM, Petricoin E, Liotta L, Wilson E. Augmented reality imaging system: 3D viewing of a breast cancer. Journal of Natural Sciences. 2016;2(9):215
  14. 14. Douglas DB, Chaudhari R, Zhao JM, Gullo J, Kirkland J, Douglas PK, et al. Perfusion imaging in acute traumatic brain injury. Neuroimaging Clinics of North America; 2017;28:55-65
  15. 15. Douglas DB, Iv M, Douglas PK, Anderson A, Vos SB, Bammer R, et al. Diffusion tensor imaging of TBI: Potentials and challenges. Topics in Magnetic Resonance Imaging: TMRI. 2015;24(5):241-251
  16. 16. Douglas D, Goubran M, Wilson E, Xu G, Tripathi P, Holley D, et al. Correlation between arterial spin labeling MRI and dynamic FDG on PET-MR in Alzheimer’s disease and non-Alzhiemer’s disease patients. EJNMMI Physics. 2015;2(Suppl 1):83
  17. 17. David Douglas MDDbcnnmppiC, USA. Dual Board Certified Neuroradiologist Nuclear Medicine Physician Practicing in California USA; 2017
  18. 18. Ochi T, Shimizu K, Yasuhara Y, Shigesawa T, Mochizuki T, Ikezoe J. Curved planar reformatted CT angiography: Usefulness for the evaluation of aneurysms at the carotid siphon. American Journal of Neuroradiology. 1999;20(6):1025-1030
  19. 19. White CS, Salis AI, Meyer CA. Missed lung cancer on chest radiography and computed tomography: Imaging and medicolegal issues. Journal of Thoracic Imaging. 1999;14(1):63-68
  20. 20. Gurney JW. Missed lung cancer at CT: Imaging findings in nine patients. Radiology. 1996;199(1):117-122
  21. 21. Ferroli P, Tringali G, Acerbi F, Schiariti M, Broggi M, Aquino D, et al. Advanced 3-dimensional planning in neurosurgery. Neurosurgery. 2013;72(suppl_1):A54-A62
  22. 22. Hohne K-H, Bomans M, Tiede U, Riemer M, editors. Display of Multiple 3D–Objects Using the Generalized Voxel-Model. Medical Imaging II. Newport Beach:  SPIE; 1988
  23. 23. Calhoun PS, Kuszyk BS, Heath DG, Carley JC, Fishman EK. Three-dimensional volume rendering of spiral CT data: Theory and method. Radiographics: A Review Publication of the Radiological Society of North America, Inc. 1999;19(3):745-764
  24. 24. Fishman EK, Ney DR, Heath DG, Corl FM, Horton KM, Johnson PT. Volume rendering versus maximum intensity projection in CT angiography: What works best, when, and why. Radiographics: A Review Publication of the Radiological Society of North America, Inc. 2006;26(3):905-922
  25. 25. Johnson PT, Heath DG, Kuszyk BS, Fishman EK. CT angiography with volume rendering: Advantages and applications in splanchnic vascular imaging. Radiology. 1996;200(2):564-568
  26. 26. Douglas D, inventor; US Patent Office, assignee. US 8,384,771 Method and Apparatus for Three Dimensional Viewing of Images. USA; 2013
  27. 27. Douglas D, inventor; US Patent Office, assignee. US 9,349,183 Method and Apparatus for Three Dimensional Viewing of Images. USA; 2016
  28. 28. DeSantis C, Ma J, Bryan L, Jemal A. Breast cancer statistics, 2013. CA: a Cancer Journal for Clinicians. 2014;64(1):52-62
  29. 29. Ferlay J, Shin HR, Bray F, Forman D, Mathers C, Parkin DM. Estimates of worldwide burden of cancer in 2008: GLOBOCAN 2008. International Journal of Cancer Journal International du Cancer. 2010;127(12):2893-2917
  30. 30. Kopans D. Analyzing the Mammogram-Calcifications. Philadelphia, PA: Lippincott-Raven; 1998
  31. 31. Lee KS, Han BH, Chun YK, Kim HS, Kim EE. Correlation between mammographic manifestations and averaged histopathologic nuclear grade using prognosis-predict scoring system for the prognosis of ductal carcinoma in situ. Clinical Imaging. 1999;23(6):339-346
  32. 32. Burnside ES, Ochsner JE, Fowler KJ, Fine JP, Salkowski LR, Rubin DL, et al. Use of microcalcification descriptors in BI-RADS 4th edition to stratify risk of malignancy. Radiology. 2007;242(2):388-395
  33. 33. Boone JM, Lindfors KK. Breast CT: Potential for breast cancer screening and diagnosis. Future Oncology. 2006;2(3):351-356
  34. 34. Mukhtar RA, Yau C, Rosen M, Tandon VJ, I-Spy T, Investigators A, et al. Clinically meaningful tumor reduction rates vary by prechemotherapy MRI phenotype and tumor subtype in the I-SPY 1 TRIAL (CALGB 150007/150012; ACRIN 6657). Annals of Surgical Oncology. 2013;20(12):3823-3830

Written By

David B. Douglas, Demetri Venets, Cliff Wilke, David Gibson, Lance Liotta, Emanuel Petricoin, Buddy Beck and Robert Douglas

Submitted: 09 October 2017 Reviewed: 23 January 2018 Published: 08 March 2018