Open access peer-reviewed chapter

Radon Calibration System

Written By

Ayman M. Abdalla, Ahmed M. Ismail and Tayseer I. Al-Naggar

Submitted: 14 November 2016 Reviewed: 26 May 2017 Published: 29 November 2017

DOI: 10.5772/intechopen.69900

From the Edited Volume

Radon

Edited by Feriz Adrovic

Chapter metrics overview

1,842 Chapter Downloads

View Full Metrics

Abstract

In this chapter, an irradiation radon system was explained in detail. This system is based on soil gas as a natural radon source. The radon system was used to determine both the calibration factor of the radon detector and the equilibrium factor between radon and its short-lived daughters. Also, the calibration factor has been calculated theoretically and experimentally. The effect of humidity upon the calibration factor has been investigated. The diffusion of radon through polyethylene membrane has been determined using new nuclear method. This method depends upon the physical decay of radon.

Keywords

  • radon detection techniques
  • calibration system

1. Introduction

Radon R86222n is a naturally occurring noble element, radioactive gas, colorless, odorless, soluble in water, and chemically inert radioactive element. The importance of radon gas comes from its benefits in the field of uranium exploration, testing the permeability of gases in polymer membranes and its harmful effects on health. In this chapter, we will try to discuss briefly radon sources, methods of radon detection, and radon calibration system including the methods of determining the calibration factor for radon measurement experimentally and theoretically. The measurement of radon permeability through polyethylene membranes using scintillation detectors will also be discussed as a useful application in the field of radon measurement.

Advertisement

2. Radon sources

The three naturally produced isotopes, radon (222Rn from 238U series), thoron or thorium emanation (220Rn from 232Th series), and action or actinium emanation (219Rn from 235U series), are all disintegrated by producing alpha particles (Table 1) [1].

IsotopeCommon nameHalf lifeDecay series
222RnRadon3.8 days238U
220RnThoron54.5 s232Th
219RnActinon3.92 s235U

Table 1.

Radon isotopes.

Radon produced from uranium decay series. It is known that all types of rocks contain various amounts of uranium especially mountainous areas [2]. Radium is produced through radioactive decay of the three natural chains (238U, 235U, or 232Th), and radon isotopes are produced as a result of radium decay (Figure 1a–c).

Figure 1.

Three natural decay series [3].

Advertisement

3. Radon measurement techniques

Measurement, and detection of radon can done by two ways directly by itself (222Rn only) or indirectly by its progenies. Detection and measurement of radon can be accomplished with α, β, or γ radiation detection. One of the following techniques can be used for detection: nuclear emulsion, adsorption, solid scintillation, gamma spectroscopy, beta monitoring, solid-state nuclear track detector (SSNTD), electrometer or electroscope, ionization chambers, surface barrier detectors, thermoluminescent phosphors, electret, and collection. Radon-222 can be measured by passive method when radon entering the detection volume by just diffusion, or active, which contains the pushing of gas into or through a detecting instrument.

3.1. Passive devices

The detection sensor is commonly put inside a vessel that has an opening to allow radon enter it. The vessel is intended to keep the detector and to make area about the detector for a sensitive size huge enough so as to have as many α-particles created and detected in as small time as possible, and in the case of direct measurement of the radiation, an inverted plastic or paper cup in various versions [4] or more advanced container can be also used.

3.1.1. Solid-state nuclear track detector

Alpha particle is detected in a SSNTD by a particular path or “trail” of damage, which after chemical etching produces narrow channel, and it is seen under the microscope. SSNTDs such as CR-39, cellulose esters, and polycarbonates like bis-phenol-A polycarbonate are sensitive to α-particles within the energy range of the particles produced by radon. The SSNTDs are more sensitive to charged particles, on the other side, SSNTDs are more insensitive to β and γ rays, so beta and gamma rays do not serve an etchable discrete trails or tracks. However, preirradiation with nontrack forming radiation has also been found to increase sensitivity and etching speed [5]. There are various types of polymer detectors such as polyallyldiglycol carbonate (CR-39), cellulose nitrate (LR-115), and polycarbonate (Lexan, Makrofol) [6, 7]. SSNTDs do not need an energy source to work, so their detecting property is an essential quality of the substance they are made of. They are consequently very well suitable for long-term measurements in the field, so they have received special awareness. The methods in which SSNTDs are used for the detection of radon and/or radon progenies which, change generally in the geometry of the arrangement, commonly the use or not of diffusion membranes; and the use (or not) of filters to keep progenies out of equipment [8]. In other states, the detector is located vertically [9]. Some procedures have two measuring chambers, detached or not by a filter and/or a membrane. In some cases, SSNTDs are used in conjunction with another type of detector. For instant SSNTDs can be combined with electrets. In this way, a prompt response can be obtained together with an energy differentiation [10].

3.1.2. Alpha card

The alpha card technique is a passive radon detecting process which supplies a means for accurate measurements of radon in soil gas, with dimensions 4.5 cm × 5 cm, and diameter hole (205 cm) punctured in the interior of it. The hole is completely closed with a slight film. Silicon detectors are used to measure the activity of alpha card in terms of alpha particles. The evaluation of reading method consists of two silicon detectors facing one another, between which the card is placed. A slightly good counting efficiency is attained because the assembling membrane is very thin, which alpha particles issuing from both sides of the card impinge upon either one of the silicon detector. The alpha card reader is a small, movable with operated battery supply that can be used in the field, and the common reading time is about 5 min [11].

3.1.3. Electrets detector

The electret radon monitor is a new technique which is used in the field to measure the radon concentration [12]. Electret is a portion of dielectric material that displays a nearly lasting electrical charge if not otherwise worried. This charge yields a robust electrostatic field which is able to assemble ions of the contrary sign and the overall charge of the electret reductions. This property can be used to make gamma radiation detectors away from the electrets [13]. The concept is not enough till the modern progress of high dielectric fluorocarbon polymers, which have turned electrets to be trustworthy, electronic constituents until under high temperature and moisture provisions. The main components of a radon monitor electret are steel canister and the electret dosimeter which is stable on the internal upper. At the lower of such canister little inlet allows the radon gas to reach the assembly during a filter. The ions are produced from radon which produces ionizing particles inside the volume of the device. The ions are assembled, and the full charge of the electrets changes; after some time has passed, the surface potential of the electret is measured by the shutter method [1416] using a battery-operated, lightweight piece of equipment.

Detector sensibility and dynamic field are based on the size of the chamber of the equipment and of the thickness of the sensitive substance. As for solid-state nuclear track detector (SSNTD), varied geometrical preparations have been designed. For case, the measurement can be actuated time wise by having a dynamic segment translating, the electret in the lodging [17]. Detection of radon alone is achieved by using a 4π-geometry chamber, on the wall of which another electret is placed for daughter detection [18]. In some other instances, a positively charged electrode can be coated with a SSNTD or with a TL phosphor for completeness of detection and measurement.

3.1.4. The mechanism of chemical absorption

In chemical absorption mechanism, the radon dimension structure constructed on the chemical absorption of the gas on an adsorption medium such as charcoal, and then on the determination the activity of the restricted daughters 214Pb and 214Bi. When the adsorbing medium is actuated charcoal, the system is well with name ROAC (radon on activated charcoal) technique. The instrument for the chemical absorption of the radon is aplastic even when the can is placed inside the place contain radon gas, with its lid open, and left in place for a few days, typically from 4 to 12 days. The container is then returned, and the cover locked. It is then transported to the laboratory where gamma activity is specified using NaI scintillation counter. Radon activity can be achieved at the check point from the decay activity curves, after adequate calibration.

By this method, thoron assay is disappeared, and the count rate must be for radon only [19] found it is possible to determine the ratio between thoron and radon emanation from a soil. Another way of determining the amount of radon is to desorb it into scintillation liquid and to count the alpha, and beta activity by liquid scintillation [20].

3.1.5. Thermoluminescent detectors

When an ionizing particle moves through a crystal, some electrons liberate from its main sites into the conduction band. The electrons are then restricted into faults of the crystal and kept there in a very constant site, and the electron can leak if the crystal is heated. As a result of recombination with the holes in the valence band, or with recombination centers, by process of radioactive, light is liberated [21, 22]. Thermoluminescent dosimeters (TLDs) are built on this attitude, in the state of radon discovering, which depends upon the alpha activity of the radon progenies. Radon can inside a detection volume include the TLD.

At a small distance, the steel plate is placed facing TLD. Radon progenies summated on the plate and eventually decay, thus creating energy storage in the TLD. After prober exposure to the radon-rich atmosphere, the TLD is recovered and “read” in a TLD apparatus. The concentration of radon is appearing on a TLD device after suitable exposure to the radon abundant air. This device is basically a little furnace, and here the heat is advanced under a nitrogen atmosphere at a very fully controlled rate. The TLD produces quantity of light that falls on the photocathode of a photomultiplier and then is transported into an electrical signal. TLD spectra are noted. Every peak relates to a particular trap, and the amplitude and/or the area below a peak is proportional to the radiation dose. In many commercially available devices, various types of detectors containing TLD sensors have been used field measurements, and in the sphere of different applications. They are using them for detecting oil gas reservoirs [23].

The advanced detection method, which really keeps an electret detector for gamma-background gauge, has three TLDS [24]. One is a preirradiated TLD which provides information on the natural fading of the signal recorded in the two others. The signal related to the radioactivity created by the radon, which has spread to the detection bulk, and the next makes a cross measurement of the γ-field. In certain cases, the housing of the TLD sensor is made such that the full detector can resist the heat at which the light is emitted from the TLD when reading it [25].

3.1.6. Solid-state electronic detectors

Solid-state electronic detectors (SSEDs) are not widely used for radon detection because it needs considerable quantity of energy not simply obtainable in the field, and also it is slightly costly and brittle. One may mention, for instance, the Alpha meter, which the active part of the equipment is a Si(Li) detector associated with a counter unit, both operated on batteries. The counting is directly displayed on the instrument [1]. A special version of the radon devices allows water level, temperature, and conductivity measurements to be made in the meantime. The detector is especially produced for indoor measurements [26]. The measuring instrument with a hemispherical bell chamber is fitted with a solid-state alpha detector built by Ref. [27]. Several active cells for counting radon daughters use by Ref. [28]. The device based on light measurement with solid-state detectors is built by Ref. [29].

3.1.7. Ion chambers and electrometers

Apparatuses depend on the measurement of ion pairs produced through the way of α-particles in the air. The measurement can be achieved by using ion chambers and electrometers. At the first status of detection part, the amount of electrical charges can record, which have been formed without laying stable voltage. Through a second state, one assembles the electrical charges that have been produced between two electrodes in an electrical field [1].

3.2. Active devices

In active apparatuses, radon gas is either pumped from the surrounding or removed by means of a gas or liquid. In addition, the lively part for measuring radon or its progenies results is an electrical or electronic device in public. By applying a fan or a pump, many of the apparatuses up could be turned into an active apparatus. The detector where radon is propelled from the ground soil is built by Ref. [30]. The gas drift pushes through the electronic meter to the external of the instrument. Radon concentration is then determined. Direct scintillation can also be used for detecting the radon level by forcing air through the detector [31]. The other type of active detection is radon sampling or extract sampling. After sampling, the water or air or the solvent that holds radon is taken to the laboratory for detection. Sampling methods are various, but they all need one to be specially wary not to pollute the sample. Sampling can be accomplished by syringes, compressor, vacuum container, water sampling, degassing on the spot, suction action of pumps, and so on.

Advertisement

4. Radon irradiation chamber and its applications

Radon is a worldwide problem because inhalation of radon engages more than 50% from total natural radiation dose to the world population [3]. Radon and its short-lived decay products caused lung cancer in the USA according to the Environmental Protection Agency (EPA) [32], and the short-lived decay products of radon are responsible for most of the hazard by inhalation. The main environmental source is the soil gas emanation in dwelling areas or from building materials [33]. Solid-state nuclear track detector (CR-39) is used widely to determine the concentration of radon [34]. The calibration factor converts the track density to radon concentration and can be calculated from the relation below [35, 36]:

K=ρc0t0E1

where ρ: track density (track cm−2), C0:radon concentration (Bq m−3), t0: exposure time (day).

The annual effective dose rate (ED, mSv/y) of inhaled radon and its progeny can be estimated according to the following equation [37]:

ED=C0FTDE2

where F: equilibrium factor, D: dose conversion factor (9 nSv h−1/Bq m−3), T: average indoor occupancy time per person (7000 h y−1), C0: radon activity concentration in air (Bq m−3).

The equilibrium factor is defined as the ratio of potential alpha energy concentration (PAEC) of actual air radon mixture (also radon progeny) to the PAEC in secular equilibrium with radon [38].

F=(PACE)air radon mixture(PACE)equilibriumE3

There are two methods to estimate the equilibrium factor: one method uses silicon surface barrier detector by the following equation [39]:

F=γ1C1C0+γ2C2C0+γ3C3C0+γ4C4C0E4

where C0, C1, C2, C3, and C4 are the activity concentrations of 222Rn and its short-lived progenies 218Po, 214Po, 214Bi, and 214Po, respectively, and the coefficients γi are constant.

In another method, a solid-state nuclear track detector was used to evaluate the equilibrium factor by the following equations [40, 41]:

D=K(C0+C1+C4)E5
D0=K(C0)E6

Eqs. (5) and (6) are related the respective track densities D and D0 to the activity concentrations of radon and its progeny in air, and K is the detector sensitivity coefficient.

The equilibrium factor can be calculated by using the LR-115 detector, using the following equation [40]:

F=0.47(DD0)0.47E7

For the CR-39 track detector, the equilibrium factor can be determined by the following equation [42]:

F=ae[b(D0D)]E8

where the values of two constants a, b are 14.958 and 7.436, respectively.

4.1. Preparing a radon source

The amount of 226Ra and 222Rn produced in rocks and soils depends on its 238U contents, and many studies considered a variable alternative of using soil gas as a radon source in the experimental studies [43]. High-performance gamma-ray spectrometry (DSPECjr2.0) was used to measure the activity of 226Ra in the soil sample; 0.424 kg of soil sample was fitted into a Marinelli beaker of 250 cm3 volume. The gamma detector has resolution 2.2 KeV and relative efficiency 55% at 1.33 MeV of 60Co. The recommended operating bias negative 4500 V. 176Lu testing adapter and 137Cs were used for energy calibration.

The specific activity of radium was measured using the following equations [44, 45]:

As=N(E×Pr×M)E9

where N: counting rate for a specific gamma line, E: detector efficiency, Pr: absolute transition probability of γ-decay, M: mass of the sample (kg). In this work, the average specific activity of 226Ra in the soil sample was 182 ± 13 Bq kg−1. Metallic cylindrical drum with 24.7 l contains different amount of soil from the environment, used as radon source, and was covered with a polyethylene membrane to make the thoron (220Rn) gas outside the system. The concentration of radon was changed using different amounts of soil.

4.2. Radon irradiation system

A radon system consists of the following parts as shown in Figure 2:

  1. 1—plastic vessel, 2—monitor, 3—lucas cell, 3, 4—printer, 5—radon source, 6, 7—PID controlled furnaces, 8, 9, 13, 14—inlet and outlet tubes, 10—internal pump, 11, 12—valves, 15—desiccation column, 16, 17—pored filter of polyethylene, 18, 19—circular port, 20—inactive dosimeter, 21—flowmeter, 22—computer interface.

  2. The lucas cell 300A offers a proper tool for detection of radon. The cell was ready at a great-efficiency scintillator of silver-activated zinc sulfide. Lucas cell volume is nearly 272 ml, sensitivity of 0.037 cpm/Bq/m3, MDA of 27.4 Bq m−3, and Eα of 4.5–9 MeV [46]. The pump of AB-5’s can be used to withdrawal air into the Lucas cell by the porous polyethylene filter, and any radon gas in the spread air decays into its daughter products, some of which are liberating α-particles. Alpha particles were hit the scintillator, which lines the inner of the Lucas cell, and create light pulses which they are amplified and 30 counted in the AB-5. A microprocessor in the AB-5 is used to convert the signal into an average count value per minute (C), and the net count value (N) is obtained by correcting the background [47].

Radon gas concentration was determined by the general formula used by the AB-5 in the radon mode:

Figure 2.

Radon irradiation system.

C0=NCPMSE10

where, C0: radon gas concentration (Bq m−3), NCPM: the net count per minute, S: the counting sensitivity value (cpm/Bq/m3).

An electronic thermometer with a liquid crystal display was used to measure the temperature inside the chamber.

4.3. Calibration factor assessment

The calibration factor of the CR-39 detector can be assessed by three cups that were used with CR-39 sheets, which had an area of 1.0 cm × 1.0 cm and a thickness of 250 μm. The plastic cups 8 cm in height and 7 cm in diameter were used in this work, and then, the detectors were etched in 6.25 N NaOH solutions at 65°C for 6 h. The etched CR-39 samples were washed in running water and dried in air. The tracks were analyzed using an optical microscope at 500×, and also the background tracks were registered.

4.4. Comparative studies with the track detector

Films from LR-115 have 12 μm thickness, which were mounted on a 100-μm transparent polyester foil, were used for two different setups was used in this work. The filtered setup is the first one, where a piece of LR-115 was stable at the end of a plastic cup. Two cups were placed in the radon irradiation system for 1 week. The exposed LR-115 sheets were etched in 2.5 N NaOH solutions at 60°C for 100 min [48]. The track density in each sheet was calculated to obtain D and D0 for the open and closed plastic cups, respectively. The identical procedure was repeated using the CR-39 track detector that was etched at the optimum etching condition [49]. Figure 3a and b show a typical optical image of alpha particle tracks in the LR-115 and CR-39 detectors. The equilibrium factor was measured using Eqs. (7) and (8) for the LR-115 and CR-39 detectors, respectively.

Figure 3.

Track optical image of alpha particles for (a) LR-115 detector and (b) CR-39 detector.

Figure 4 shows the rate of track density as a function of the radon activity concentration for the diffused CR-39 track detector. From the slope of the line, it is found that the calibration factor is 0.165 track cm2/Bq m3 d with uncertainty approximately 20%. The obtained experimental value is considerably consistent with the reported experimental values in Farid (0.199 track cm2/Bq m3 d) [50] and Mansy et al. (0.177 track cm2/Bq m3 d) [51].

Figure 4.

Rate of track density as a function of radon activity concentration.

Table 2 demonstrates the equilibrium factor between radon and its progeny, which was determined using three different methods: solid-state nuclear track technique, surface barrier method, and our new radon chamber. From Table 2, the results obtained using our new radon chambers are considerably consistent with the results of the surface barrier detector method [7]. We can obtain a stable equilibrium factor between radon and its daughters in a closed environment of a radon system which emulates a closed environment of a room.

MethodOur new radon systemSSNTDSurface barrier detector method
Equilibrium factor0.210.17 (LR-115)
0.20 (CR-39)
0.19 [5]

Table 2.

Equilibriums factor obtained using three different methods.

As shown in Figure 5a, the equilibrium factor tends to decrease as the temperature increases (temperature range of 20–50°C for 24 h) and this is due to the temperature impact on the equilibrium factor by two chief causes: (1) the temperature reduces the equilibrium factor by effects on the radon emanation rate and (2) the temperature rises the equilibrium factor by affecting the number of radon daughters in the air [52, 53]. Also Figure 5b shows that the equilibrium factor at 20°C is equivalent to 2.2 times the coefficient of equilibrium at 50°C and 1.5 times the coefficient of equilibrium at 32°C. Also, the influence of the flow rate through irradiation on the equilibrium factor between radon and its daughters was obvious. The system was utilized in the flow rate of 0.2–0.6 l/min for 24 h at stable heat. Figure 5b presents the equilibrium factor for radon, and its daughters are influenced by the variation in flow rate of radon-bearing air. The equilibrium factor rises with rising flow rate. The value of equilibrium gives 0.25 at the optimum flow rate of 0.5 l/min. These consequences indicate that the measurement of an apparent equilibrium factor as a function of the temperature and flow rate is important for determining the annual effective dose rate.

Figure 5.

Equilibrium factor as a function of (a) temperature and (b) flow rate of air in the lucas cell.

Advertisement

5. Radon permeability measurement through polyethylene membrane using scintillation detector

Activated charcoal technique was used to study the permeability of 222Rn through polyethylene membranes. The permeability constant of 222Rn through low-density polyethylene, linear low-density polyethylene, and high-density polyethylene samples has been measured, and the diffusion coefficient of radon in charcoal as well as solubility of radon in polyethylene membrane has been taken into consideration. The two gases 222Rn (238U-series) and 220Rn (232Th-series) have differences in their radioactive half-lives 3.82 days and 56 s, respectively, which provide suitable properties for their separation using proper membranes. The difference in the decay rates leads to difference in the mean diffusion distances [54, 55].

The permeability (K) of a polymer to a gas can be defined by

K=SDE11

where S: the solubility, D: the diffusion coefficient of the gas in the polymer.

The significant factor which determines the appropriateness of a membrane for separation radon isotopes is its permeability constant. Ramachandran et al. [56] suggested a method to determine the permeability constant that depends on SSNTD technique, as shown in Figure 6, and it is expressed by the following relation:

Figure 6.

Schematic diagram shows solid-state nuclear track technique.

R=n2n1=KA(V1+V2)KA(V1+V2)+V1V2λδE12

where R: ratio between the concentrations of gaseous in volume V2 with a membrane and the concentration without any membrane in V1, K: the permeability constant (cm2 s−1), d: the thickness of the membrane (in cm), A: the surface area of the membrane (cm2), and λ: the decay constant (s−1).

The process used is based on exposing activated charcoal canister to radon concentration inside radon chamber [57], as illustrated in Figure 7.

Figure 7.

A schematic for radon calibration chamber.

The radon activity adsorbed by a diffusion barrier charcoal canister (DBCC) is described by the following relations given by Refs. [5860]:

Qi=βmCi=VCiE13
dCidt=(KAδV+λ)Ci+KAδVC0E14
τ=(KAδV+λ) if P=KAδ and ψ=1τ,  Then ψ=PV+λE15

where Qi: the activity of radon that passes through the membrane and adsorbed by DBCC (Bq), V: equivalent to a volume of air available to the activated charcoal (cm3), b: the adsorption coefficient of radon by activated charcoal (Bq g−1 per Bq cm3), m: the mass of activated charcoal in the canister (g), Ci: the radon concentration within the canister (Bq m−3), C0: is the radon concentration in the environment air (Bq m−3), Λ: the decay constant of radon (s−1), K: the radon permeability (cm2 s−1),, A: cross section of the diffusion barrier (cm2), D: thickness of the diffusion barrier (cm), and S: is the integration time constant of radon in the DBCC (s).

For Ci = Ci0, C0 = 0 at t = 0, the solution of the Eq. (14) is

Ci=Ci0exp(tτ)E16

The radon permeability for radon in the tested membrane can be easily evaluated by using the following equation:

K=VLA(ψλ)E17

Permeability of radon was measured at room temperature and a comparative moisture of 20% at exposure was finished. The canisters were closed and left for 3 h. Gamma spectroscopy 3″ × 3″ NaI (TI) was used to determine the gamma spectrum from each canister, as illustrated in Figure 8. The range of energy for gamma lines between 295 and 609 keV was achieved, and the activity of radon in the canister was determined.

Figure 8.

Schematic of the scintillation spectrometer.

Figure 9 presents the difference of the relative net count per minute as a function of time for different samples.

Figure 9.

Rate of counting as a function of time for different polyethylene membranes.

For SSND, pieces of CR-39 track detectors of 250 mm (TASTRAK supplied by Bristol, UK) were used for two diverse setups. In the first setup (named: filtered setup), a piece of CR-39 was stabled in the end of conical flask (250 ml), and then, the conical flask was closed with a polyethylene membrane which acts as a filter. Secondly, named bare (opened) setup, in which a piece of CR-39 was stabled at the end of opened conical flask. Filtered and open conical flasks were put in radon chamber for a term of 1 month. A diagram setup of SSNTD technique is represented in Figure 7. After that the CR-39 sheets have been etched in 6.25 normal NaOH solutions at 70°C .ffor 6 h. The etched samples were washed in running water and dried in air. The track density in each sheet is calculated to be n1 and n2 in the case of open and closed conical flasks, respectively. The radon permeability can be measured using Eq. (11).

Table 3 shows the radon permeability (cm2/s) of polyethylene membrane using three different methods for high-density polyethylene samples, and the permeability of radon is stabled fined by our offered method as the same value measured by Arafa system [61] for great density polyethylene samples. For low-density polyethylene sample, there is substantial covenant between the value of radon permeability constant determined by our offered method and that of SSNT technique.

SampleThickness (cm)Track techniqueArafa method (2002)Proposed method
LLDb0.002501.27E−070.49E−071.5E−07
LDa0.002451.2E−070.57E−070.9E−07
HDc0.002951.1E−070.46E−070.46E−07

Table 3.

Radon permeability (cm2/s) of polyethylene membrane using three different methods.

LD: Low-density polyethylene


LLD: Linear low-density polyethylene


HD: High density polyethylene


Advertisement

6. Theoretical calculation of the calibration factor

The calibration factor of radon dosimeter (CR-39 track detector) for radon and its daughters has been determined theoretically. Calibration factor (K) is the quantity, which is used for converting the observed track density rates to activity concentrations of the species of interest.

If ρ is the track densities observed on a SSNTD due to exposure in a given mode to a concentration C for a time t, then:

ρ=KCtE18

The signal measured by etched track dosimeters is the integrated track density (tracks cm−2). If the track density is uniform over the detector surface, the signal is defined by: [62]

ρ=Kτr0teC(t)dtE19

where τr is the radon decay time (5.5 days), C(t) is the radon concentration in the air around the detector (atom m−3) at time t, and te is the exposure time, K denotes the dosimeter response (calibration coefficient) and is defined by:

K=τrC(t)dρdt=1A0(t)dρdtE20

where A0 (t) is the activity concentration of radon in air (Bq m−3) at time t and /dt is the track density production rate (tracks cm−2 d−1).

For a diffusion chamber-type dosimeter with a long exposure time (te >> .3 h), the response can be calculated by: [36]

K=ρA¯0teE21

where A¯0 denotes the average radon activity concentration and te is the exposure time. In general, the total track density ρ(r) at the point of observation in the detector, defined by r, is the result of an integration of all α-emissions which can produce observable tracks. It can be written as:

ρ(r)=iρiv (r)+iρip (r)E22

where iρiv (r): track density due to radon, Po: isotopes existing in the air, and ρip (r): track density due to radon decay progenies.

Ilic reported that for every α-produced nuclide existing the air is calculated by [63]:

ρiv(r)=otedtV1Ai(r, t)cosθ4π|rr`|2dVE23

where Ai(r, t): activity concentration of ith nuclide for time t and at a point r, θ: angle between the particle’s trajectory and the normal to the detector surface, |rr`|: distance between the detector surface and the source volume dV, and Vi: the sensitive to volume for ith nuclide.

The stable rate of etching is calculated by the critical angle of etching. θc,i in this event can be found as:

ρiv=0.25AiteRicos2θc,iE24

Eq. (23) gives the estimation of track density as a function of the radon activity and intrinsic detector efficiency [64]. K0 can be calculated by the flowing relation:

K0=0.25R0cos2θ=0.88 cm=3.2 tracks cm2KBqm3 hE25
Advertisement

7. Future of radiation detection

In the last few years, radiation detection has emerged as an area of interest for quantum dots (QDs) application [65]. However, there have been very few published studies on the radiation detection based on colloidal QDs. The emission of QDs is size dependent, so the output wavelength can be tuned to the sensitivity of the PMTs or avalanche photodiodes (APD). However, the stopping power of most QDs is low and their scintillation luminescence is very weak [66]. The combination of high stopping power of inorganic scintillator with QDs could potentially lead to a new class of scintillator [67]. The aim of our future work is to develop the sensitivity and counting efficiency of radiation detector using nanoparticle.

Advertisement

Acknowledgments

This project was funded by King Abdulaziz City for Science and Technology, the kingdom of Saudi Arabia, grant number: 14-ENV1162-60. Financial support by King Abdulaziz City for Science and Technology is gratefully acknowledged.

References

  1. 1. Durrani SA, Ilic R, editors. Handbook of Radon Measurements by Etched Track Detectors: Applications in Radiation Protection, Earth Science and the Environment. World Scientific Publishing; 1997. p. 387. ISBN: 9810226667
  2. 2. Hararah SMAI. Investigation of radon pollution in groundwater in the southern part of Gaza Strip—Palestine [MSc thesis]. Gaza, Palestine: The Islamic University; 2007
  3. 3. Human Health Fact Sheet. EVS, Argonne National Laboratory; August 2005. https://www.remm.nlm.gov/ANL_ContaminantFactSheets_All_070418.pdf
  4. 4. Fleisher RL, Mogeo-Campero A. Mapping of integrated radon emanation for detection of long-distance migration of gases within the earth: Techniques and principles. Journal of Geophysical Research. 1978;83:3539–3549
  5. 5. US Atomic Energy Commission. Personal dosimeter for radon and daughter products. U.S. Patent No. 749. 725. 680. 802;1968
  6. 6. Cassou RM, Benton EV. Properties and applications of CR-39 polymeric nuclear track detector. Nuclear Track Detection.1978;2:173–179
  7. 7. Tommasino L. Radon monitoring by alpha track detection. In: Tommasino L, et al., editors. Proceedings of the International Workshop on Radon Monitoring in Radioprotection, Environmental Radioactivity and Earth Science;1981; Trieste. Singapore: World Scientific; 1990. pp. 123–132
  8. 8. Yoder RC. Radon gas detector with sensing film. U.S. Patent No. 211, 516, 880, 624;1988
  9. 9. Wheeler RV. Detector for radon gas and its decay products. U.S. Patent No. 012; 266; 1987
  10. 10. Pretzsch G, Borner E, Lehmann R, Sarenio O. Radon diffusion chamber with passive integrating detector. German (former Democratic Republic) Patent No. 276, 690, 850, 528; 1985
  11. 11. Durrani SA, Ilic R, editors. Handbook of Radon Measurements by Etched Track Detec- tors: Applications in Radiation Protection, Earth Science and the Environment. World Scientific Publishing; 1997. p. 387. ISBN: 9810226667
  12. 12. Kotrappa P, Dempsey JC, Hickey JR, Stieff LR. An electret passive environmental 222Rn monitor based on ionization measurement. Health Physics. 1988;54(1):47–56
  13. 13. Marvin HB. How to measure radiation with electrets. Nucleonics. 1955;13:82–87
  14. 14. Kotrappa P, Dua SK, Pimpale NS, Gupta PG, Nambi KSV, Bhagwat AM, Soman SD. Passive measurement of radon and thoron using TLD or SSNTD on electrets. Health Physics. 1982;43(3):399–404
  15. 15. Kotrappa P, Dua SK, Gupta PC, Pimpale NS, Khan AH. Measurements of potential alpha energy concentration of radon and thoron daughter using an electret dosimeter. Radiation Protection Dosimetry. 1984;5:49–51
  16. 16. Gupta PC, Kotrappa P, Dua SK. Electret personal dosimeter. Radiation Protection Dosimetry. 1985;11:107
  17. 17. Ramsey PM, Stieff LR. Programmable control exposure radon measurement system. U.S. Patent No. 920: 421; 1992
  18. 18. Gast H. Measuring chamber for determining alpha activity of radon gas in carrier gas especially air. German Patent No. 200: 308, 920, 109; 1992
  19. 19. Megumi K, Mjamuro T. Method of measuring radon and thoron exhalation from the ground. Journal of Geophysical Research.1977;77(17):3052–3056
  20. 20. Perlman D. Method and apparatus for radon detection. European Patent No. 359: 770, 900, 328; 1990
  21. 21. Horowitz YS. TL Dosimetry. Boca Raton, FL, USA: CRC Press; 1984
  22. 22. McKeever S. Thermoluminescence of Solids. Cambridge: Cambridge University Press; 1985
  23. 23. Hochman MB, Ypma PJ. Location of oil and gas reservoirs. Australian Patent No.01: 988, 788, 000; 1988
  24. 24. Harley NH, Maiello ML. Passive measurement of environmental radioactivity. U.S. Patent No. 021, 450, 870, 304; 1987
  25. 25. Seidel JL. Thermoluminescent dosimeter for monitoring subsurface radiation. U.S. Patent No. 941, 995, 780, 913; 1982
  26. 26. Kotlyarov AA, Krivashcheev SV. Measuring chamber of volume activity of radon surrounding air samples. Russian Patent No. 2, 008, 694, 940, 228; 1994
  27. 27. Matthies D. Radon concentration measuring instrument. German Patent No.940, 908, from 93DE-307, 015, 930, 305; 1994
  28. 28. Thomson I. Radon detector. U.S. Patent No.4, 983, 843, 910, 108; 1991
  29. 29. Dwcuir J. Radon monitoring device. European Patent No. 357, 697, 900, 314; 1990
  30. 30. Iimasti V. Device for measuring radon gas concentration. Finnish Patent No.002, 717, 870, 618; 1987
  31. 31. US Secretary of Interior Monitoring radon gas in air. U.S. Patent No. 653, 314, 760, 129; 1976
  32. 32. Tirmarche M, Harrison JD, Laurier D, Paquet F, Blanchardon E, Marsh JW. Lung Cancer Risk from Radon and Progeny and Statement on Radon CRP Publication 115 Ann. ICRP 40(1):2010
  33. 33. Environment Protection Agency of USA. A Citizen’s Guide to Radon. 4th ed. 2002
  34. 34. Imtiaz N, Faheem M. Determination of 238 U contents in ore samples using CR-39-based radon dosimeter—disequilibrium case Radiation Measurements 41(4):471-476
  35. 35. Matiullah J. Determination of the calibration factor for CR-39 based indoor radon detector. Journal of Radioanalytical and Nuclear Chemistry. 2013;298(1):369–373
  36. 36. Sutej T, Ilic R, Najzer M. Response of track-etch dosimeters to environmental radon. Nuclear Tracks and Radiation Measurements. 1988;15(45):547–550
  37. 37. UNSCEAR. United Nations Scientific Committee on the Effects of Atomic Radiation Sources, Effects and Risks of Ionizing Radiation. New York: United Nations; 2000
  38. 38. Dorsche lB, Piesch E. Radiation Protection Dosimetry. 1993;48(2):145
  39. 39. Jamil K, Rehman F-U, Ali S, Khan HA. Nuclear Instruments and Methods in Physics Research Section A.1997;388:267–272
  40. 40. Planinic J, Rdolic V, Faj Z, Suvelajak B. Nuclear Instruments and Methods in Physics Research Section A. 1997;396:414–417
  41. 41. Yu KN, Leung SYY, Nikezic D, Leung JKC. Radiation Measurements. 2008;43:S357–S363
  42. 42. Faj Z, Planinc J. Radiation Protection Dosimetry. 1991;35:265–268
  43. 43. Al-Azmi D. Radiation Measurements. 2009;44:306–310
  44. 44. Noorddin I. Natural activities of 238U, 232Th and 40K in building materials. Journal of Environmental Radioactivity. 1999;43:255–258
  45. 45. Wafaa A. Journal of Environmental Radioactivity. 2004;75:315–327
  46. 46. Pylon Electronics Inc. Models110A&300A. Instruction Manual Number A900071;2009. p. 12
  47. 47. Pylon Electronics Inc. Model AB-5. Instruction Manual Number A900024 Rev. 6; 2002
  48. 48. DeCicco F, Pugliese M, Roca V, Sabbarese C. Applied Radiation and Isotopes. 2013;78:18–112
  49. 49. Ashry AH, Abou-Leila M, Abdalla AM. Radiation Measurements. 2011;46:149–152
  50. 50. Farid SM. Journal of Environmental Radioactivity. 1997;34:29–36
  51. 51. Mansy M, Sharaf MA, Eissa HM, El-Kamees SU, Abo-Elmag M. Radiation Measurements. 2006;41:222
  52. 52. Chu T-C, Liu H-L. Applied Radiation and Isotopes. 1996;47:543–550
  53. 53. Singh K, Singh M, Singh S, Sahota HS. Radiation Measurements. 2005;39:213
  54. 54. Tanner AB. Radon migration in the ground: A Review. In: Adams JAS, Lowder WL, editors. The Natural Radiation Environment. Chicago: University of Chicago Press; 1964;161–190
  55. 55. Tanner AB. Radon migration in the ground: A supplementary review. In: Gesell TF, Lowder WM, editors. The Natural Radiation Environment III, Symposium Proceedings, Houston, Texas, April 1978, Vol. 1. U.S. Dep. Environ., Natl. Tech. Info. Serv., Washington, DC, Rep. CONF-780422, 1980
  56. 56. Ramachandran TV, Lalit BY, Mishra UC. Measurement of radon permeability through some membranes. International Journal of Radiation Applications and Instrumentation. Part D. Nuclear Tracks and Radiation Measurements.1987;13(1):81–84
  57. 57. El-Sammane H, Arafa W, Abdalla A. Temperature and humidity consideration for calculating air born 222Rn using activated charcoal canisters. Health Physics. 2002:97–105
  58. 58. Cohen BL, Cohen ES. Theory and practice of radon monitoring with charcoal adsorption. Health Physics. 1983;45:501–508
  59. 59. Cohen BL, Nason R. A diffusion barrier charcoal adsorption collector for measuring radon concentrations in indoor air. Health Physics. 1986;50:457–463
  60. 60. Montero Cabrera ME, Sujo LC, Villalba L, Peinado JS, Jiménez AC, Miranda AL, Herrera Peraza EF. Calibration of diffusion barrier charcoal detectors using a semi-empirical expression. Applied Radiation and Isotopes. 2003;59:281–287
  61. 61. Arafa W. Permeability of radon-222 through some materials. Radiation Measurements. 2002;35:207–211
  62. 62. Fleischer RL, Giard WR, Mogro-Compero A. Health Physics.1980;39:957
  63. 63. Ilic R. Environmental Radioactivity and Earth Sciences. 1990;133
  64. 64. Fleischer RL, Mogro-Compero A. Journal of Geophysical Research. 1978;83:3539
  65. 65. Hossu M, Liu Z, et al. Applied Physics Letters. 2012;100(1):013109
  66. 66. Yao M, Zhang X, et al. Journal of Applied Physics. 2010;108(10):103104
  67. 67. Sahi S, Chen W. Radiation Measurements. 2013;59:139–143

Written By

Ayman M. Abdalla, Ahmed M. Ismail and Tayseer I. Al-Naggar

Submitted: 14 November 2016 Reviewed: 26 May 2017 Published: 29 November 2017