Abstract
A proposed model consisting of two coupled van der Pol models is considered as a description of the heart action potential. A system of ordinary differential equations is used to recreate pathological behaviour in the conducting system of the heart such as Wolff-Parkinson-White (WPW) syndrome and the most common tachycardia: atrioventricular nodal reentrant tachycardia (AVNRT). Part of the population has abnormal accessory pathways: fast and slow. These pathways in the atrioventricular node (AV node) are anatomical and functional excipients of supraventricular tachycardia. However, the appearance of two pathways in the AV node may be an excipient of arrhythmia—the WPW syndrome. The difference in the conduction time between these pathways is the most important factor. This is the reason to introduce three types of couplings and delay to our system in order to reproduce different types of the AVNRT. In our research, the result of introducing the feedback loops and couplings entails the creation of waves which can correspond to the re-entry waves which occur in the AVNRT. Our main aim is to study solutions of the equations of the system and to take into consideration the influence of feedback and delays which occur in the pathological modes. The proposed models made it possible to reproduce the most important physiological properties of the discussed pathologies. Since the model is phenomenological, the results are accurate as far as a simple model can describe the potential found in one of the more complex oscillators found in biology.
Keywords
- van der Pol model
- ordinary differential equations
- delay
- feedback
- couplings
- action potential
- WPW syndrome
- AVNRT
- numerical analysis
1. Introduction
In this chapter, the research on the electrical conduction system of the human heart is concerned. The parts of this system are pacemaker centres made of special cells similar to embryonic cells which form the following concentrations: sino-atrial node (SA), atrioventricular node (AV) and His-Purkinje system [1]. The key elements of the conduction system considered include SA node and AV node. Modelling the formation and conduction of electrical impulses in the heart is one of the most developed areas of mathematical biology. For years, the most popular models of action potentials which occur in the heart have included the Hodgkin-Huxley model and the Purkinje cells model. These models, even though very effective, are not very suitable for studying the dynamics of the system, which is described in view of high dimensionality of the phase space and very broad spatial parameters. In this chapter, we propose ordinary differential system which is based on the van der Pol models and which makes it possible to reconstruct pathological behaviours in the system of the heart, such as the WPW syndrome and different types of AVNRT. The motivation for writing this chapter was that there is a problem with making the appropriate diagnosis and therefore with treating the disease effectively. This kind of problem is observed mainly in different types of AVNRT. The reasons of those problems are not fully known and the mechanisms of these pathologies are not fully understood. Also, the symptoms are often mistakenly taken for other heart diseases. It should be emphasized that the clinical pictures of these disease are also non-specific. In the past few years, we understood that the structure of the AV node has a multi-level architecture in which there may be many pathways (slow and fast) at different locations in the AV node [2, 3]. This helped to recognize many types of AVNRT, which previously were understood as one, although the mechanisms of action were different. In literature, there are no mathematical models that would specifically model the various types of AVNRT. There were attempts to model AVNRT only as a single pathology having the slow and fast pathways [4, 5]. Part of the population has abnormal accessory pathways: fast and slow, cf. [1, 6]. The pathways in the AV node are anatomical and functional contributions of the most popular supraventricular tachycardia, which is a re-entry tachycardia from the AV node. The atrioventricular nodal re-entrant tachycardia is caused by re-entries. A condition for AVNRT to occur is that two electric pathways occur in and around the AV node (for example, slow and fast pathways). This gives way to the occurrence of re-entry. We can distinguish five different forms of the AVNRT (typical: slow/fast, atypical: fast/slow and other forms: slow/slow, more than two re-entries waves, one fast pathway with depolarization of slow pathway) [7, 8, 9]. This depends on the multi-level architecture of the AV node. The AVNRT circuit involves larger areas including atrioventricular junction, adjacent atrial structures and in particular so-called atrial inputs including at least antero-superior and postero-inferior entries, and sometimes also the left atrial entry. Based on the van der Pol equation, we study the influence of feedback which occurs in the normal heart action mode as well as in pathological modes. Particularly, it is important to introduce the time delay into this feedback. Delay values used in our research correspond to those which occur in the electrical conduction system, for example, in the case of an accessory conducting pathway, so-called Wolff-Parkinson-White syndrome. In the conduction system of the heart, the only correct electrical conduction pathway between atria and ventricles is the bundle of His. Part of the population has abnormal accessory pathways through which electrical pulses are directly conducted from atria to ventricles. The extra electrical pathway is presented at birth and is fairly rare. The WPW syndrome is detected in about 4 out of every 100000 people. People of all ages, including infants, can experience the symptoms related to the WPW syndrome. Additional path conducts electrical impulses to the ventricles in the case of WPW. This conduction takes place without proper synchronization. Usually, this extra conduction pathway does not cause serious disturbances. Sometimes, however, there is a reflection of the electrical impulse, and it returns to the atria after each heartbeat. This causes the coupling in which each atrial contraction is followed by chambers contraction and then atrial contraction again. The heart rate can reach about 200 beats per minute when the normal rhythm at rest is about 72 beats per minute.
As already mentioned, synchronization is very important for a correctly functioning heart in the sense of cardiovascular conduction. The impact of different types of couplings related to the pathologies which are presented in this research and how do they affect the synchronization of the two oscillators will be discussed in this chapter. The analysis of synchronization of various modifications of the van der Pol model has been considered in many works. For example, paper [10] presents synchronization areas near the main parametric resonance and transition conditions from regular to chaotic motion. In paper [11], authors analyse the phenomenon of complete synchronization in a network of four coupled oscillators. Reference [12] considered mechanisms of various bifurcation phenomena occurring in Bonhoffer van der Pol neurons coupled with a time delay through the characteristics of synaptic transmissions. The phenomenon of synchronization in van der Pol oscillators coupled by a time-varying resistor is researched in ref. [13]. However, in these papers there are no examples of application of this model for recreating a pathological behaviour of the electrical-conduction system of the human heart, and therefore the considered ranges of parameters are wider than those which can be used in medical applications. The van der Pol oscillator provides rich dynamical behaviour [14] and also synchronization phenomena. In order to better understand the mechanisms that cause these arrhythmias in paper are proposed models of AVNRT and WPW syndrome.
2. Modelling techniques
The van der Pol model was used in this research because it is a two-dimensional model with small number of parameters and it is a relaxation oscillator. The van der Pol equation is often used to describe an action potential occurring in the heart. The model with a delay and often with a term including a coupling coefficient has been a topic of many articles, e.g. [15–17]. However, in these articles there are no exact examples of application of this model for recreating a pathological behaviour of the electrical conduction system of the human heart such as the AVNRT and the WPW, and therefore the range of parameters considered by them is wider than that which can be used in medical applications. Below, the author will present one of the most important existing mathematical models of the action potential. Basing on the knowledge of van der Pol model given below will be carried out construction of the proposed models.
2.1. Van der Pol model
Since every node is a self-exciting pacemaker, it can be described using the van der Pol oscillator which is the relaxation oscillator. It was first introduced by van der Pol and van der Mark as a model in the electronic circuit theory in 1927, [19]:
where
where
In this way, he introduced modifications which allowed maintaining the required structure of the phase space. Such model can be applied to model either SA or AV node. The main property of a modified relaxation oscillator is the mutual interaction of a limit cycle which is presented around an unstable focus with a saddle and a stable node. This allows reproducing correctly the refraction period and non-linear phase sensitivity of an action potential of node cells.
A solution of Eq. (3) in time presents the action potential whereas a solution in velocity makes it possible to obtain very important phase portrait. We can easily see that the main qualitative difference between Eqs. (1) and (3) is the appearance of two additional steady states,
The system that we use as a reference one is given in following form:
where
3. Construction and analysis of proposed models
In this section we turn to constructing new models for specific pathologies we would like to describe. All the models we propose below consist of two modified van der Pol models. The system with delayed feedback describes various pathologies observed in the heart action, e.g. WPW syndrome and different types of AVNRT.
A numerical model was created using Dynamics Solver and a program which was written in C++. The explicit Runge-Kutta formula with a variable integration step was used. Results were confirmed and plots were made with Matlab standard dde23 solver.
For the single node model (Eq. (4)) of an electrical conduction system with no feedback, a periodic potential and the corresponding limit cycle (with a transient state) are obtained like in Figures 1 and 2. Potential period equals 1.4.
The reference parameters have values which are given below:
The addition of coupling (especially feedback with delay) to the
Parameters have the following values:
Now we consider the AVNRT pathology. We would like to propose a mechanism which reproduces the dynamics of action potential occurring at the phenomenon of AVNRT. In this case, we have a coexistent conduction by two or more pathways: slow and fast. In this model, we assume that the fast pathway is described by the first van der Pol model, but the slow pathway is treated as an action potential in other characteristic, so we use the second model of van der Pol to describe the pathway.
Next, we focus on such a type of AVNRT in which only one fast pathway with depolarization of slow pathway occurs. The conduction is only out of the fast pathway, because descending depolarization of slow pathway prevents conduction. To describe that situation, we add only one unidirectional coupling to our model. This coupling does not change the main rhythm
where
Now, we consider a situation where the difference in the refractive state of both pathways leads to an excitation of one of them causing the re-entry wave [20]. In Figure 5, the result for a typical AVNRT (fast/slow) is presented. This type of AVNRT is described by the following system:
The mechanism of the typical AVNRT is similar to this which is observed in the WPW syndrome. Also, we have a feedback with delay in the first part of the given model.
The atypical AVNRT (slow/fast type of AVNRT) is described in the following way:
In both of those types of AVNRT (fast/slow and slow/fast) we obtain regular fast rhythm, which is a typical behaviour for this kind of pathology (during this type of tachycardia, the rhythm of the heart is about 35% more frequent than normal rhythm, which is in accordance with our results).
The last case is for the situation where there are two re-entry waves but both go through slow pathways (slow/slow type of AVNRT), which we reflect by the following system of equations:
In this variant, we also observe shortening (like in Figure 5) of the period of oscillations. In this pathology, the presence of more than two conduction paths is possible which is associated with more re-entry waves. Increasing the number of feedbacks modelling re-entry waves of slow pathways causes a progressive shortening of the period of oscillation, while the rhythm remains regular. The conclusion is that a regular excitation may evoke a regular answer.
After examining the influence of excitation with modified van der Pol oscillator wave, we have analysed behaviour of the modified van der Pol oscillator with feedback excited with a single rectangular pulse. In this way, two excitations co-exist in the system—the wave brought by the oscillator’s feedback and the one from the external pulse. The pulse is defined by a step function H for the amplitude
Influence of the pulse is treated as an appearance of the accessory current in the system. For this reason, this pulse is introduced to the equation describing the potential change (the
where
The influence of the pulse, which results in the period length change, depends on in which phase of oscillations it is applied. As mentioned before, the excitation influences the system having the regular behaviour.
Numerical solution of the system with feedback (
The system is phase-sensitive, and that, depending on the phase, excitation may change the potential period length. Models with feedback are much less phase-sensitive, Figure 7. The feedback strongly modifies the properties of the modified van der Pol model.
The human heart rhythm with feedback about AV node is low-variable. The results of this study were consulted with the medical doctor from the Institute of Cardiology in Anin near Warsaw, who confirmed that the very stable heart rhythm is typical for patients with AV nodal re-entrant tachycardia.
One of the most important features of oscillators is synchronization. The behaviour of cardiac pacemaker cells resembles those relaxation oscillators. A characteristic property of relaxation oscillators is that they may be synchronized by an external signal, if the latter has a periodicity similar to spontaneous frequency of the oscillator [14]. We investigate a phenomenological model for the heartbeat which consists of two coupled van der Pol oscillators. The coupling between these oscillator (action potential) can be either unidirectional or bidirectional, with or without feedback. Below, we present our model with possible couplings and feedbacks and try to analyse all of them for behaviour of given system and its synchronization.
If we consider physiological coupling between nodes, then the
But for bigger value (e.g. 10 or 100), parts of amplitudes are synchronized in-phase with
This causes the attempt of fitting the AV frequency to SA which slows down the heart rate.
We study here the synchronization properties of such an association with respect to the nature and intensity of coupling.
In Figures 11–13, we can observe an influence of unidirectional coupling for strength of synchronization of given oscillators (Eq. (11)). We must remember that complete synchronization of oscillators, characterized by
Finally, the influence of the bidirectional coupling for behaviour of the model was studied. It would seem that the bidirectional coupling should operate so that when we enter it in relation 2:1 between oscillators, the rhythm of the fast pathway should be slowed down, while the rhythm of the fast pathway should accelerate.
In the model without feedback, periodic potential
However, with bigger values of
We investigated the synchronization between oscillators themselves despite that the generation of cardiac dysrhythmias are associated with a lack of synchronization between autonomous pacemakers [22]. But we showed that we can obtain certain types of synchronization: phase synchronization, partial synchronization or lag synchronization. This depends on the strength of coupling coefficients. Corresponding large coefficient may cause synchronization, but often setpoints are not the values of the physiological range.
4. Conclusions
Main aim of the paper was to propose the system of differential equations describing the dynamics of action potential that accompanies the WPW syndrome and five different forms of AVNRT. In this work, by using the proposed models we were able to reproduce the most important physiological properties of the discussed pathologies. In literature, there are no mathematical models of AVNRT. By using one of the proposed models, we can try to add an equation which is responsible for modelling the pharmacological treatment of the pathologies of our interest. These type of models then can help to determine an optimal treatment. On the other hand, we should keep in mind that the proposed models are phenomenological, so the results are accurate as far as a simple model can describe the potential found in one of the more complex oscillators appearing in real biological phenomena. We can continue the validation of these models with the use of medical data. However, to collect the necessary data, only the invasive methods can be used, which constitutes an important difficulty.
References
- 1.
S. Konturek, editor.The human physiology. The cardiovascular system.II. Cracow: Jagiellonian University Press; 2001. - 2.
D. G. Katritsis, M. E. Josephson. Classification of electrophysiological types of atrioventricular nodal re-entrant tachycardia: a reappraisal. Europace. 2013; 15 :1231-1240. - 3.
K. Malaczynska, K. Blaszczyk. Atrioventricular nodal reentrant tachycardia. Polish review of the Cardiology. 2012; 14 :196-203. - 4.
B. Zduniak, M. Bodnar, U. Foryś. A modified van der Pol equation with delay in a description of the heart action. Int. J. Appl. Math. Comput. Sci. 2014; 24 :853-863. - 5.
P. Podziemski, J. J. Żebrowski,. A simple model of the right atrium of the human heart with the sinoatrial and atrioventricular nodes included. J Clin Monit Comput. 2013; 27 :481-498. - 6.
A. Fujiki. Junctional rhythm associated with ventriculoatrial block during slow pathway ablation in atypical atrioventricular nodal re-entrant tachycardia. Europace. 2008; 10 :928-987. - 7.
B.Dabrowska, P. Gajewski. Postepowanie u chorych z nadkomorowymi zaburzeniami rytmu Wytyczne American College of Cardiology, American Heart Association European Society of Cardiology. Medycyna Praktyczna. 2004; 6 :1-62. - 8.
R.A. Freedman, J.W. Mason. Sustained ventricular tachycardia, clinical aspects. Cardiac Pacing and Electrophysiology. Third Edition. P. Samet and N. El. Sherif, eds., Philadelphia: W. B. Sanders Co. 1991; 247-264. - 9.
W. G. Stevenson. Exploring postinforction reentrant ventrivular tachycardia with entertainment mapping. J. Am. Coll. Cardiol. 1997; 29 : 1180-1189. - 10.
J. Warminski. Synchronization effects and chaos in the Van der Pol-Mathieu oscillator. J. Theoretical. Appl. Mechanics. 2001; 4 (39): 861-884. - 11.
P. Perlikowski, A. Stefanski, T. Kapitaniak. Discontinuous synchrony in an array of Van der Pol oscillators. Int. J. Non-Linear. Mechanics. 2010; 45 :895-901. - 12.
K. Tsumoto, T. Yoshinaga, H. Kawakami. Bifurcations of synchronized responses in synaptically coupled Bonhoffer Van der Pol neurons. Phys. Rev. E. 2002; 65 : 036330-1-036230-9. - 13.
Y. Uwate, Y. Nishio. Synchronization phenomena in van der Pol oscillators coupled by a time-varying resistor. Int. J. Bifurc. Chaos. 2007; 17 (10):3565-3569. - 14.
L. Henk van der Tweel, F. L. Meijler, F. J. L. van Capelle. Synchronization of the heart. J. Appl. Physiol. 1973; 34 (2): 283-287. - 15.
A. Maccari. The response of a parametrically excited van der Pol oscillator to a time delay state feedback. Nonlinear Dynamics. 2001; 26 :105-119. - 16.
J. Xu, K.W. Chung. Effects of time delayed position feedback on a van der Pol – Duffing oscillator. Physica D. 2003; 180 (1):17-39. - 17.
D.V. Ramana Reddy, A. Sen, G.L. Johnston. Experimental evidence of time-delay-induced death in coupled limit-cycle oscillators. Phys. Rev. Lett.. 2000; 85 :3381–3384. - 18.
A. Hodgkin, A.F. Huxley. A quantitative description of membrane current and its application to conduction and excitation in nerve. J. Physiol. 1952;117: 500-544. - 19.
Ch. Letellier, editor. Chaos in Nature. New Jersey: World Scientific Publishing Co. Ltd.; 2013. pp. 1-377. - 20.
D. Postnov, S. K. Han, H. Kook. Synchronization of diffusively coupled oscillators near the homoclinic bifurcation. Phys. Rev. E. 1999; 60 (3): 2799-2807. - 21.
K. Grudziński. Modeling the Electrical Activity of the Heart Conduction [thesis]. Warsaw, Poland; 2007. - 22.
S.Bocaletti, J. Kurthus, G. Osipov, D. Valladares, C.S. Zhou. The synchronization of chaotic systems. Phys. Rep. 2002; 366 (1):1-101.