Open access peer-reviewed chapter

On an Overview of Nonlinear and Chaotic Behavior and Their Controls of an Atomic Force Microscopy (AFM) Vibrating Problem

By José Manoel Balthazar, Angelo Marcelo Tusset, Atila Madureira Bueno and Bento Rodrigues de Pontes Junior

Submitted: February 1st 2012Reviewed: July 24th 2012Published: October 24th 2012

DOI: 10.5772/51834

Downloaded: 1239

1. Introduction

It is known that in Brazil and in the whole world, the development of mathematical modeling techniques for the study of problems involving new technologies and the design of nonlinear structures is an emerging area in Engineering Science. To background to the general problems of non-linear dynamics, see the comprehensive monographs: (Awrejcewicz,1991) and (Awrejcewicz and Lamarque, 2003).

This chapter addresses this issue, involving the dynamical behavior of Electrical Mechanical Systems (EMS), that are systems responsible for electromechanical energy conversion. It is common knowledge that an electromechanical system can be classified in three different groups, according to scales dimensions: i) Macro-Systems >1mm, ii) Micro-Systems <1mm and >0.1μm, and iii) Nano-Systems<0.1 μm and >0.1nm.

The Macro-Systems are characterized by being visible without using microscope and they are present in our Day to Day activities. The Micro-Systems are generally constructed on silicon chips, and are employed in manufacturing integrated electronic circuits. The Nano-Systems can only be seen and manipulated with the use of the Atomic Force Microscope.

In this research field - MEMS and NEMS -, the researches are based on the fact that, in opposition to traditional macro-scale problems, the atomic forces effects and surface phenomena are preponderant over the forces of inertia and gravity. It is also known that the manufacture of these new products have been based on the binomial consuming time and an enormous amount of financial resources. Thus, it becomes necessary some investment in an interface, which addresses the design and manufacturing that, gives designers of "MEMS" and "NEMS" the proper tools of mathematical modeling and simulations as well as qualitative dynamic analysis. It should be considering a macro-model, which agrees with the simulation results on a physical level, and with the experimental results, obtained from test structures in a lab. Many surveys have been developed in major research centers in Brazil and throughout the world in this direction. It is announced that several research groups in Brazil are involved with this theme, but most of the research carried out or in development focuses on experimentation or in the use of the numerical method - called finite element method. It is known that the nonlinear phenomena are prevalent in MEMS and NEMS, which is a strong motivation for the work of our research group.

The main idea of this chapter lies in the study of "State of the Art" of non-linear dynamic models and the design of their controls, since this new area of research is of paramount importance. This chapter deals with two boundary problems belonging to this line of research, in development at the university of Rio Claro and Bauru UNESP, for us, through research groups and development projects for Graduate (Masters) and Post–Doctor. Mainly for the Atomic Force Microscope.

2. General aspects of the Atomic Force Microscope (AFM)

In Binnig et al. (1986) the atomic force microscope ("AFM") was presented. The AFM has been described as one of the most efficient tools for obtaining high-resolution images of the samples, by exploiting its surface, both in air and in liquid media. At the core of the "AFM", there is a tip, mounted on the end of a micro cantilever, while the tip vibrates to scan the sample, its vibration is detected by a laser system, which emit signals to a photo detector that generate images of the object under examination at high resolution. This movement may vary depending on the need and on the type of material being analyzed. These variations include techniques such as static contact mode and dynamic techniques, such as non-contact mode and intermittent or tapping mode. The principle of operation of the "AFM" is to the measure the deflection of the microcantilever on whose free end the probe is mounted. The deflections (analyzed during the scan) are caused by forces acting between the tip and sample. These forces act on medium to large distances - typically ≥ 100 Å – such as attractive Van der Waals forces, magnetic forces and Coulomb forces (Eisenschitz, 1930). It can be said that when the tip approaches to the sample, it is first attracted by the surface, due to a wide range of attractive forces in the region. This attraction increases when the probe is very close to the sample. However, when the atoms of the tip and sample become very close, the orbital electrons begin to repel. The forces are canceled when the distance between the atoms is of the order of several angstroms (the distance of the order characteristic of a chemical bond). When the forces become positive, it can be said that the atoms of the tip and the sample are in contact and ultimately the repulsive forces dominate.

In Figures: 1, 2 and 3, the main characteristics and the operation of an AFM can be observed. Figure 3; display the basic configuration of an AFM. The micro cantilever is V-shaped or rectangular with a sharp tip. This tip is usually of pyramidal or conical shape (Figure 2).

Figure 1.

AFM photo (source: Bowen and Hilal 2009)

Figure 2.

source: Bowen and Hilal 2009)

Figure 3.

Control of the positioning of the scanner. Adapted from : (Cidade et al. 2003)

Typically, probes are made predominantly of silicon nitride (Si3N4); its upper surface is coated with a thin reflective surface, generally gold (Au) or aluminum (Al). The probe is brought to inside and out of contact with the sample surface, by using a piezo-crystal (Hilal and Bowen, 2009). In the illustration of a typical micro-manipulator, shown in Figure 3, it is possible to observe that it consists of a movable stage mounted under an optical microscope. The movement of the base can be controlled by an electronic control console. Large deflections are required to achieve high sensitivity. Therefore, the spring should be very "soft" (slightly stiff).

2.1. Operation modes in AFM

The probe mounted on the AFM performs the scan on the sample in a raster fashion. The movement of the microcantilever over the sample is carried out by the piezoelectric scanner, which comprises piezoelectric material that expands and contracts according to the applied voltage. There are several modes of operation for scanning and mapping surface. These modes include non-contact, contact and intermittent contact modes. These three modes of operation differ from each other, basically, by the tip and sample distance.

The Lennard-Jones potential describes the relationship of the tip and sample interaction forces as depending on the tip and sample surface distance, considering the potential energy of a pair of particles, and is given by:

U(r)=4ε[(σr)12(σr)6]E1

whereεand σare constants depending on the sample properties, σis approximately equal to the diameter of the particles involved. Deriving potential function (U) in relation to the distance (r) gives an expression for the force (F) versus distance (r) (Equation (2)). This force is represented in Figure 4.

F(r)=Ur=24ε[2σ12r13σ6r7]E2

The region above the r-axis corresponds to the region where the repulsive forces dominate (contact region). The region below the r-axis corresponds to the region where attractive forces dominate (non-contact region). Also in red, it can be seen the distance region that the tapping-mode technique is applied.

3. Mathematical modeling of the Atomic Force Microscope (AFM)

The mathematical models governing the dynamics of the AFM micro-beams, usually result from the discretization of the classical beam equations, based on its mode of vibration, leading to one or more degrees of freedom. Several models of this kind are described in the literature, e.g., Wang et al. (2009), Garcia and San Paulo, (2000); (Laxminarayana and Jalili, 2004), (Hu and Raman, 2007) (Raman et al. 2008), (Ashab et al. 1999), (Farrokh et al. 2009), (Lozano and Garcia, 2008) among others. Most of the mathematical models are linear mass-spring-damper systems incorporating nonlinear interaction between the tip and sample (Paulo & Garcia, 2002). Different AFM techniques provide a number of possibilities for topographical images of the samples, generating a wide range of information. In this chapter models for intermittent contact mode are discussed.

Figure 4.

Force (F) versus Distance (r), (Source: Cidade et al., 2003).

3.1. Mathematical modeling of AFM: with inclusion of the cubic (spring) term

The physical model of the AFM tip-sample interaction can be considered as shown in Figure 5 (Wang, Father and Yau, 2009). The microcantilever-tip-sample system is regarded as a sphere of radius Rsand massms, suspended by a spring of stiffnessk=kls+knls, where klsandknlsare the linear and nonlinear stiffness. The van der Waals potential for the sample-sphere system is given by:

P=AcRc6(Zb+X)+12klsX2+14knlsX4E3

The energy of the system scaled by the mass of the cantilever is given byE(X,X',Z):

E=12X˙2+12ω12X2+14ω22X4Dω12(ZbX)E4

Replacing X1=XandX2=X˙, then, from equation (3) results:

X'1=EX2X'2=EX1E5

Figure 5.

Model of an AFM (Source: Wang, Father and Yau (2009))

The dynamic AFM system in Figure 5 is obtained replacing (2) and (3) into (4):

{X˙1=X2X˙2=ω12X1ω22X13Dω12(Zb+X1)2E6

whereZbthe distance from the equilibrium position. The molecular diameter isD=AHR6k, where Ah is the Hamaker constant and R is the sphere radius. Considering only attractive Van der Waals force, and that the cantilever is being excited bymfcos(wt), where wis the natural frequency, the system equations can be written as:

X˙1=X2X˙2=ω12X1ω22X13Dω12(ZbX1)2fcoswtϕX2E7

WhereϕX2=(a4cosba5X2)'is the damping force. Considering the relations:x1=X1Zs, x2=X2ωsZs, z=ZbZs, Zs=32(2D)13andτ=wt, the system (6) may be rewritten in the following dimensionless form:

x˙1=x2x˙2=a1x1a2x13b(z+x1)2+csinτE8

3.2. AFM Mathematical modeling: intermittent mode and hydrodynamic damping

The microcantilever schematic diagram of the AFM operating in intermittent mode can be seen in Figure 6. The base of the microcantilever is excited by a piezoelectric actuator generating a displacementfcos(wt). According to (Zhang et al., 2009), considering only the first vibration mode, the (AFM) can be modeled as a spring-mass-damper, as shown in Figure 7. The "tip" is considered as being a of radius R and Z0 is the distance from the equilibrium position of the cantilever to the sample. The position of the cantilever is given by x, measured from the equilibrium position. According to (Rutzel et al., 2003) the tip-sample interaction can be modelled as a sphere- flat surface interaction, given by:

U(x,z0)=A1R1260(z0+x)7A2R6(z0+x)E9

whereU(x,z0)is the Lennard-Jones potential(LJ), A1=π2ρ1ρ2c1and A2=π2ρ1ρ2c2are the Hamaker constant for the attractive and repulsive potential, respectively, with ρ1and ρ2the densities of the interacting components, and c1andc2are constants from interaction. Itshould be noted that, when the "cantilever" is close to the sample, attractive van der Waals

Figure 6.

Microcantilver-tip-sample system

Figure 7.

Physical model (Source: Zhang et al., 2009)

force must be considered. These forces can be represented as the sum of the attractive and repulsive forces (Rutzel et al., 2003), expressed by:

F=U(x+z0)=A1R180(z0+x)8A2R6(z0+x)2E10

In the intermittent mode (TM-AFM) the probe touches the surface of the sample at the point of maximum amplitude of oscillation. During the scanning the microcantilever is driven to oscillate according to the forcefcoswt, resulting that the tip-sample contact generates the forceF. The contact between the tip and sample is delicate, and this mode of operation is suitable for fragile samples. Then, considering the Lagrangian L=TVand the Euler-Lagrange equationddt(Lq˙i)Lqi=Qi, where

T=12mx˙2,V=12klx2+14knlx4andQi=F+cx˙+csx˙+fcoswtE11

are the kinetic energy, the gravitational potential energy, and nonconservative forces, respectively, the equation of motion for the microcantilever tip displacement x is given by:

mx¨+cx˙+klx+knlx3=A1R180(z0+x)8A2R6(z0+x)2+μeffB3L(x+z0)3x˙+fcoswtE12

with:

csx˙=μeffB3L(x+z0)3x˙E13

whereμeffis an effective coefficient of viscosity, B the width of the cantilever and L is the length of the cantilever (Zhang et al. (2009)).

Defining:

Zs=(23)(2D)13E14

whereD=A2R6kand considering the following relations in (11)

τ=w1t, y=xzs, y˙=x˙w1zs,a=z0zs, b=klkat, c=knlkatzs2,d=427, e=2405(azs)6, g=f0katzs, p=μeffB3Lmw1zs3, Ω=ww1,r=1Q. Equation (11) can be rewritten in the dimensionless form:
y¨+ry˙+by+cy3=d(a+y)2+e(a+y)8+gcosΩτp(a+y)3y˙E15

Writing equation (14) into state space form results:

x˙1=x2x˙2=rx2bx1cx13d(a+x1)2+e(a+x1)8+gcosΩx3p(a+x1)3x2E16

where:x1=y, x2=y˙andx3=τ.

Considering the values of parameters: Ω=1;r=0.1;b=1;c=0.35;d=4/27;e=0.0001;g=0.2; p=0.005e a=1.6(obtained by Zhang et al. (2009)).The displacement can be seen in Figure 8a and the phase portrait can be seen in Figure 8b.

Figure 8.

a): Tip displacement (b): Phase Portrait

4. Chaos in mathematical model with cubic spring

According to Ashhab (1999) the chaotic behavior of AFM depends on the damping of the excitation and on the distance between the tip and the sample, suggesting that a feedback control of the states can be used to eliminate the possibility of chaotic behavior. According to that, considering the system (7) in nondimensional form:

x˙1=x2x˙2=αa1x1αa2x13b(z+x1)2+δsinτE17

with the parameters:α=0.14668;b=0.17602, δ=2.6364ez=2.5, a1=1, a2=14.5, numerically simulation results can be seen Figure 9. Additionally, The FFT and the Lyapunov exponents (λ1=0.336,λ2=0.336) are shown in Figure 10.

Figure 9.

Phase diagram

Figure 10.

a): FFT, (b): Lyapunov exponent

5. Chaos with hydrodynamic damping in TM-AFM

The elastic constant of the cantilever kcmust be less than the effective elastic constant of the interatomic coupling katof the sample. Thus the elastic constant of the spring must beK<Kat, withKat=wat2mat. Typical atomic vibration frequencies are ωat=1013Hzand atomic masses are of order 1025kg and K<10[N / m]. Considering the case of K<Katand rewriting the equation into state space results:

x˙1=x2x˙2=rx2bx1cx13d(a+x1)2+e(a+x1)8+gcosΩx3p(a+x1)3x2E18

wherex1=y,x2=y˙. The phase diagram can be observed in Figure 11a. For the parameters values: Ω=1;r=0.1;b=0.05;c=0.35;d=4/27;e=0.0001;g=0.2; p=0.005anda=1.6. The Lyapunov exponents (λ1=0.23, λ2=0;λ3=0.1), can be seen in Figure 11b, indicating that the system has a chaotic attractor.

Figure 11.

a):Phase diagram (b): Lyapunov exponents

6. Scanner position control

A laser beam focus on the top of the microcantilever and the reflection is detected by a photodiode. The light is converted into an electrical signal, and stored in the computer as a reference. An oscillation of the microcantilever deflects the laser beam on the photodiode, allowing the system to compute the microcantilever motion. The error signals are then forwarded, and the piezoelectric scanner moves vertically to scan the sample, as shown in Figure 3. The control techniques are diverse: PID or PD, sliding mode, LQR, or other.

6.1. Feedback control for the model with hydrodynamic damping

Considering the model (17) with the inclusion of control:Fu.

x˙1=x2x˙2=rx2bx1cx13d(a+x1)2+e(a+x1)8+gcosΩx3p(a+x1)3x2+FuE19

and defining a periodic orbit as a function ofx˜(t). The desired regime is given by:

y˜¨=ry˜˙by˜cy˜3d(a+y˜)2+e(a+y˜)8+gcosΩτp(a+y˜)3y˜˙+u˜E20

Since u˜control the system in the desired trajectory, and x˜(t)is a solution of (19), without the term controlFu, thenu˜=0, resulting:

y˜=a0+a1cos(Ωτ)+b1sen(Ωτ)+a2cos(2Ωτ)+b2sen(2Ωτ)+...E21

The feedforward control u˜is given by:

u˜=y˜¨+ry˜˙+by˜+cy˜3+d(a+y˜)2e(a+y˜)8gcosΩτ+p(a+y˜)3y˜˙E22

Replacing (19) in (18) and defining the deviation from the desired trajectory as:

e=[x1x˜1x2x˜2]E23

The system (18) can be represented as follows:

e˙1=e2e˙2=re2be1c(e1+x˜1)3+cx˜13d(a+e1+x˜1)2+d(a+x˜1)2+e(a+e1+x˜1)8e(a+x˜1)8p(e2+x˜2)(a+e1+x˜1)3+px˜2(a+x˜1)3+uE24
withu=Fuu˜, and the feedback control:u=Ke. The system (23) can be represented in deviations as:
e˙=Ae+g(x)g(x˜)+BuE25

Representing the system (24) in the form:

[e˙1e˙2]=[01br][e1e2]+g(x)g(x˜)+[01]uE26

where

g(x)g(x˜)=[0c[(e1+x˜1)3x˜13]d(a+e1+x˜1)2+d(a+x˜1)2]+[0e(a+e1+x˜1)8e(a+x˜1)8p(e2+x˜2)(a+e1+x˜1)3+px˜2(a+x˜1)3]E27

Defining the desired trajectory as the periodic orbit, with amplitude less than (a) and frequency equal to (Ω), then:

x˜=1.3sin(t)E28

Considering the parameters values: Ω=1;r=0.1;b=0.05;c=0.35;d=4/27;e=0.0001;g=0.2; p=0.005e a=1.6the matrices A and B, are given by:

A=[010.050.1],B=[01], and definingQ=[100010], R=[1], and using the matlab(R) the control u is obtained. In Figure 12 the tip displacement without and with control are shown.
u=k11e1k12e=3.1127e13.9293e2E29

Figure 12.

Displacement of the tip to the system without control and with control

6.2. Feedback control for a model with cubic spring

Considering the following parameters:α=0.14668, b=0.17602, δ=2.6364, z=2.5, a1=1, a2=14.5, and the control U in (16) results:

x˙1=x2x˙2=αa1x1βa2x13b(z+x1)2+δsinb+UE30

where:U=u˜+u, uis the feedback control, u˜is the feedforward control, given by :

u˜=x˜˙2+αa1x˜1+αa2x˜13+b(z+x˜1)2δsinbE31

Replacing (29) in (28) and defining the deviation from the desired trajectory as:

y=(xx˜)E32

wherex˜is the desired orbit, and rewriting the system in deviations, results:

y˙1=y2y˙2=αa1y1αa2(y1+x˜1)3+αa2x˜13b(z+y1+x˜1)2+b(z+x˜1)2+uE33

Considering the system (31) in the following way:

y˙=Ay+g(x)g(x˜)+BuE34

where:y=[y1y2];A=[01αa10];B=[01]and

g(x)g(x˜)=[0αa2(y1+x˜1)3+αa2x˜13b(z+y1+x˜1)2+b(z+x˜1)2]. The control u is obtained by solving the following equation:
u=R1BTPyE35

whereP is a symmetric matrix, solution of the reduced Riccati equation:

PA+ATPPBR1BTP+Q=0E36

Defining the desired trajectory as:

x˜=2cos(t)E37

The matricesA=[010.146680], B=[01], and defining Q and R asQ=[2500020], R=[0.1], results after using the matlab(R) to obtain u:

u=49.8535y117.3120y2E38

In Figure 13 it can observed the tip displacement with and without control.

7. AFM mathematical modeling with Phase-Locked Loops (PLLS)

As mentioned above, the Atomic Force Microscopy started in 1986 when the Atomic Force Microscope (AFM) was invented by Binnig in 1986. Since then many results have been obtained by simple contact measurements. However, the AFM cannot generate truly atomic resolution images, by simple contact measurement, in a stable operation. Besides, since 1995, using noncontact techniques, it was possible to obtain atomic resolution images, with stable operation, under attractive regime at room temperature (Giessible, 1995; Morita et. al., 2009). Noncontact AFM operates in static and dynamic modes. In the static mode the tip-sample interaction forces are translated into measured microcantilever deflections, and the image is a map z(x,y,Fts) with Fts constant.

Figure 13.

a): Tip displacement without control (b): Tip Displacement with control

In the dynamic mode the microcantilever is deliberately vibrated. The Amplitude Modulated AFM and the Frequency Modulated AFM are the most important techniques. In both AM-AFM and FM-AFM the amplitude and frequency of the microcantilever are kept constant by two control loops. The AGC (Automatic Gain Control) and the ADC (Automatic Distance Control). The AGC controls the amplitude of oscillation and the ADC controls the frequency by adjusting the distance between tip and sample. The oscillatory behavior of the microcantilever is illustrated in Figure 14. The FM-AFM block is shown in Figure 15.

Figure 14.

Microcantilever oscillatory behavior. Source: (Bueno et al., 2011).

In the FM-AFM the control signal of the AGC loop is used to generate the dissipation images and the ADC control signal is used to generate the topographic images. The FM-AFM improved image resolution and for surface studies in vacuum is the preferred AFM technique (Morita et. al., 2009; Bhushan, 2004). From Figure 15 it can be seen that the PLL generates the feedback signal for both control loops, therefore the PLL performance is vital to the FM-AFM. The PLL is a closed loop control system that synchronizes a local oscillator to a sinusoidal input. The PLLs are composed of a phase detector (usually a multiplier circuit), of a low-pass filter and of a VCO (Bueno et. al., 2010; Bueno et al., 2011), as it can be seen in Figure 16, and additionally, shows the PM and AM outputs used in the AFM system. The AGC loop also depends on the amplitude detector output, shown in Figure 17. The amplitude detector is composed of diode followed by a first-order low-pass filter. The circuit holds the output A(t) for a while, allowing the AGC to determine the control signal.

Figure 15.

Block Diagram of the FM-AFM control system. Source: (Bueno et al., 2011).

Figure 16.

PLL block diagram

7.1. Mathematical model of the FM-AFM

The mathematical model of the FM-AFM considers the microcantiler dynamics, the tip-sample interaction, the amplitude detector circuit and the PLL. The microcantilever is assumed to be a second order system with natural frequency ωcand damping coefficientγ. Concerning to tip-sample interaction, there are short, medium and long range forces. Since the FM-AFM operates in long-range distance, the predominant force is the Van Der WaalsAH6(d(t)+z(t))2, where AH is the Hamaker constant and d(t) is the tip-sample distance, Figure 7.1. Besides, the microcatilever is excited by an external forcing signal with a previously determined amplitude. The signal is a sinusoid with phase ωct+φ(t)and amplituder(t)vo, where r(t)is the AGC signal, and vois constant. The tip-sample interaction forces cause modulations both in the amplitude and in the frequency of oscillation of the AFM microcantilever. The modulations are detected by the PLL and used by the AGC and by the ADC, in order to control the microcantilever, drivin it to oscillate according toz(t)=A(t)sen(ωct+φc(t)). The microcantilevermathetamical model is given by equation 37.

Figure 17.

Amplitude detector. Source (Bueno et al., 2011)

The mathematical model of the amplitude detector is given by equation 38, where τd=1RCandzd={z(t),z(t)>00,z(t)0. The PLL model can be seen in many works in the literature. The mathematical model in equation 39 follows Bueno et al., 2010 and 2011. Considering the filter transfer function f(t)=α0s2+β1s+β0and the gainG=12kmkovoAc, where kmis the phase detector gain, kois the VCO gain, vois the VCO output amplitude and Acis the nominal microcantilever amplitude of oscillation, Figure 14. Equations 40 and 41 represent the AGC and ADC, respectively. Equations 37 to 41 are the model of the FM-AFM.

z¨+γz˙(t)+ωc2z(t)=r(t)vosen(ωct+φo(t))+AH6(d(t)+z(t))2E39
A˙(t)+τdA(t)=τdzd(t)E40
φo(t)+β1φ¨o(t)+β0φ˙o(t)+α0Gsen(φo(t)φc(t))=0E41
r(t)=ΦAGC(AcA(t))E42
d(t)=ΦADC(Δωcφ˙o(t))E43

7.2. Local stability and PLL design for FM-AFM

From equation 39, and considering the phase errorϑ=φoφc, results that:

ϑ+β1ϑ¨+β0ϑ˙+α0Gsen(ϑ)=φc+β1φ¨c+β0φ˙cE44

that represents the phase erros between the microcantilever oscillation and the PLL. The PLL behavior analysis is conveniently perfomed considering the cylindric state space, consideringϑ(π,π]. In that case, the synchronous state, corresponding to an asymptocally stable equilibrium point of equation 42 (See Bueno et al., 2010 and Bueno et al., 2011), corresponds to a Constant phase error ϑand to null frequency and acceleration errors, i.e.,ϑ=ϑ¨=ϑ˙=0. For small phase erros it can be considered that sen(ϑ)ϑin (42). In addition, consideringφc=Ωt, (42) can be rewritten as:

ϑ+β1ϑ¨+β0ϑ˙+α0Gϑ=β0ΩE45

that represents the PLL linear response to a frequency shift (step) of amplitudeΩ. The local stability of equation 42 can be determined by the position of the poles of equation 43, or by the Routh-Hurwitz criterion (See Bueno et al., 2010 and Bueno et al., 2011; Ogata, 1993). Therefore, considering that the coefficients of the filter are all positive and real, the poles of equation 43 have negative real parts if:

G<β0β1α0E46
.

Considering the filter coefficients α0=β0=ωn2and β1=2ξωn, where ξis the damping factor and ωnthe natural frequency, then, from equation 44, results:

G<2ξωnE47
.

Equation 45 establishes a design criterion that assure the local stability of the PLL, i.e., for small phase and frequency steps the PLL synchronizes to the microcantilever oscillation. Additionally, from the design parameters ξand ωnthe loop gain Ghas a superior bound, and can be determined in order to satisfy the requirements of performance and stability.

Despite the good transient response and high frequency noise rejection - such as the double frequency jiter -, provided by the all pole filter the steady state response may need improvement. The PLL must demodulate an FSK (Frequency Shift Keying) signal, that actually is a frequency step. In order to track a frequency shift the loop filter must have at least a pure integration, i.e., the PLL must be at least a type 2 system (Bueno et al., 2010 and Bueno et al., 2011; Bueno, 2009, Ogata, 1993). According to that, and considering the loop filter F(s)=as+1s(bs+1)the stability of the PLL is assured only ifa>b.

Figure 18 illustrates the PLL response to an FSK signal, showing the PLL FM output (figure 16). After the transient the mean value of the FM output is the same value of the FSK signal. The oscillation is due to the Double frequency jitter (Bueno et al., 2010 and Bueno, 2009). This shows that the PLL design must provide strong damping to noise and to the double frequency jitter. Besides, if the PLL is not at least of type 2 the PLL presents steady state error in the FSK demodulation, impairing the AFM imaging process. The PLL perfomance was analysed and presented under the FM-AFM perspective. Besides, a PLL design method was shown and illustrated by simulation, making clear the PLL performance importantance in the AFM control system.

Figure 18.

PLL response to a FSK signal.

8. Conclusions

This chapter deals with emergent problems in the Engineering Science research, presenting study and research related to NEMS systems, specially microcantilevers with many modes of vibration, for which the tip-sample interaction forces are highly nonlinear, impairing the stability of the latent image, while the others modes of vibration can be explored in order to improve the AFM performance.

In the context of this work, the following specific problem have been approached: The understanding of the relations of the properties and the structure of the nanoscopic and molecular materials, through the atomic force microscopy, using microcantilevers, giving subsidiary information to next generation of microscopy instrumentation.

© 2012 The Author(s). Licensee IntechOpen. This chapter is distributed under the terms of the Creative Commons Attribution 3.0 License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

How to cite and reference

Link to this chapter Copy to clipboard

Cite this chapter Copy to clipboard

José Manoel Balthazar, Angelo Marcelo Tusset, Atila Madureira Bueno and Bento Rodrigues de Pontes Junior (October 24th 2012). On an Overview of Nonlinear and Chaotic Behavior and Their Controls of an Atomic Force Microscopy (AFM) Vibrating Problem, Nonlinearity, Bifurcation and Chaos - Theory and Applications, Jan Awrejcewicz and Peter Hagedorn, IntechOpen, DOI: 10.5772/51834. Available from:

chapter statistics

1239total chapter downloads

More statistics for editors and authors

Login to your personal dashboard for more detailed statistics on your publications.

Access personal reporting

Related Content

This Book

Next chapter

Dynamics of a Pendulum of Variable Length and Similar Problems

By A. O. Belyakov and A. P. Seyranian

Related Book

First chapter

Numerical Solution of Many-Body Wave Scattering Problem for Small Particles and Creating Materials with Desired Refraction Coefficient

By M. I. Andriychuk and A. G. Ramm

We are IntechOpen, the world's leading publisher of Open Access books. Built by scientists, for scientists. Our readership spans scientists, professors, researchers, librarians, and students, as well as business professionals. We share our knowledge and peer-reveiwed research papers with libraries, scientific and engineering societies, and also work with corporate R&D departments and government entities.