Open access peer-reviewed chapter

Investigation of Bioluminescence at an Optical Fiber End for a High-Sensitive ATP Detection System

By Masataka Iinuma, Ryuta Tanaka, Eriko Takahama, Takeshi Ikeda, Yutaka Kadoya and Akio Kuroda

Submitted: May 12th 2012Reviewed: August 27th 2012Published: June 13th 2013

DOI: 10.5772/52747

Downloaded: 1221

1. Introduction

In biological research, the luminescience from fluorescent proteins or luminescent enzymes is widely applied for monitoring a change of environment at a cell. Biomolecules used as the probe, such as Green Fluorescence Protein (GFP) or luciferase molecules found in fireflies can respond to the existence of specific molecules or ions and subsequently emit a photon. The detection of a specific molecule can then be confirmed by detecting the emitted photons efficiently with a photon detector. A highly efficient detection of the luminescence is normally essential to realization of a high sensitivity to the specific molecules or ions and an improvement of the sensitivity can upgrade the capability of detection in a low concentration of sample solution. Therefore, there are many efforts to improve the efficiency of the collection of emitted photons and of the optical coupling to the photon detector.

A straightforward method is to directly detect the luminescence from the sample solution in a test tube with a single photon detector via simple coupling optics as shown in Fig. 1. This detection system is very simple and easy-operational, so that it has been widely used for various applications so far. To realize high efficiency detection, however, this method needs a single photon detector with the wide photon-sensitive area, which is ideally larger than a photon-emission area in the test tube. Here, we are introducing an alternative method, where the luminescent biomolecules are immobilized at an optical fiber end and the luminescence is detected by a photon detector which is optically coupled to the other optical fiber end. Fig. 2 illustrates the optical fiber-based system. This method has been investigated for application to a fiberoptic biosensor, which is constructed by immobilizing either an enzyme or an antibody. A review of this method is given in reference [1] and [2].

Figure 1.

Direct detection system of luminescence

Figure 2.

Optical fiber-based system of luminescence

This method has three merits. The first one is to permit a local detection within the sample solution, because the optical fiber end functions as a needle-like probe. Meanwhile, the method as shown in Fig. 1 is suitable for detecting the luminescence from a large area in the sample solution. The second one is that the detection scheme does not require that the photon detection is very close to the sample solution. This feature makes it easier to mount the sensing parts in integrated bioengineering, such as μ-TAS. The third merit is that single photon detectors with a small sensitive area are also available, because the photon-emission area, which is almost identical to the cross section of the core part in the optical fiber, is small. In general, single photon detectors have lower dark counts for smaller sensitive area. Low noise is very important, because it essentially gives the upper limit of the sensitivity of photon detection. Recently, single photon detectors using avalanche photo diodes (APDs) have become widely available with good performance, but their sensitive area is small and typically 0.1 mm. The luminescence detection with the optical-fiber based system allows us to fully use the merits of compactness, high quantum efficiency, and low noise of these APD detectors. On the other hand, single photon detectors using compact and cooled photomultiplier tubes (PMTs) are also available, since they have the characteristics of low noise and much larger sensitive area than the APD’s. Moreover, their quantum efficiency is not so much lower than the APD’s. The fiber-based system with these PMT detectors can make a fully use of the merits of large sensitive area and low noise of the PMTs, although the quantum efficiency is lower than the APD’s.

We have built detection systems of bioluminescence at an optical fiber end and investigated the sensitivity of Adenosine triphosphate (ATP) detections by using an APD-type as described in [3] and [4]. In this chapter, results with a PMT-type detector are presented in comparison with the results by using the APD-type photon detector. We also discuss the reason of limiting the present sensitivity in the system with the PMT-type detector. ATP is a good indicator of biochemical reaction or life activity, since ATP is considered as the universal currency of biological energy for all living things. Therefore, there are many efforts to develop ATP-sensing techniques for compact and efficient ATP detection in reference [5-7]. In particular, high-sensitivity detection of ATP can indicate the existence of microorganisms even in low numbers. Thus, a compact, simple, and easy-operational system with extremely high sensitivity has been desirable.

One well-known and powerful method for highly sensitive ATP detection is to use the chemical reaction involved in thebioluminescence, the luciferin-luciferase reaction in reference [8]. In this reaction, after one ATP molecule and one luciferin molecule are bound to one luciferase molecule, and the luciferin molecule is oxidized using the energy of ATP. As consequence, one photon is emitted during the transition from the excited state to the ground state of the oxidized luciferin molecule bound to the luciferase molecule. The emission of one photon indicates the use of the energy of one ATP molecule. In the method using the luciferin-luciferase reaction, the efficient detection of the bioluminescence is essential for high-sensitivity detection of ATP.

The oxidation of luciferin is catalysed by the enzyme luciferase, so that the immobilization of luciferase molecules on solid probes of various sizes allows highly sensitive and local detection of ATP. Three types of immobilization have been used: firstly attachment to the cell surface in [9], secondary attachment to small particles, such as nanoparticles in [10], glass beads and rods in [11], thirdly attachment to extended objects with a size in the centimeter range, such as strips in [12] and [13], and films in [14]. For the ATP-detection on the intermediate scale below 1 millimeter, a fiberoptic probe employing immobilized luciferase in [2] as well as microchips in [15] and [16] is utilizable. Therefore, the efficient detection system of bioluminescence at an optical fiber end can achieve the local detection of ATP. The realization of highly sensitive detection of ATP potentically provides the local detection of extremely low number of microorganisms. Thus, it is desirable to construct a highly efficient detection system of the bioluminescence at an optical fiber end and to evaluate the detection limit with the system. In order to explore possibilities for improving the detection limit, moreover, it is also necessary to investigate the bioluminescent reaction at an optical fiber end.

The rest of this chapter is organized as follows. In sec. 2, we describe a concept for the construction of the optical fiber-based system and show how to construct the detection systems by using the PMT detector. In sec. 3, we describe the sensitivity test with the constructed system and show the results are consisitent with the APD, but also show that the sensitivity can not reach the expected detection limit. In sec. 4, we present the results of kinetic properties obtained from experimental data on the bioluminescence and clarify a dominant reason of restricting the detection limit.Sec. 5 summarizes prensent results and future problems.

2. Construction of the optical fiber-based system

2.1. General concept

For the construction of an efficient detection of a bioluminescence, it is necessay to consider a collection efficiency of the luminescence at the optical fiber end and a coupling efficiency between the other optical fiber end and a photon detector as described in [4]. Using the optical fiber with a core diameter ϕ0and a numerical aperture NA0, the collection efficiency of the luminescence ηat the optical fiber end depends only on NA0as shown in Fig.3.

Figure 3.

Luminescence at the optical fiber end. θm is a maximum opening angle for light propagation in the optical fiber.

From the simple calculation of the solid angle with a maximum open angle θm, η(NA0)can be expressed as,

η(NA0)=12(1cosθm)=12[11(NA0nw)2],E1

where nwis the refraction index of the substance surrounding the optical fiber end. In immersing this optical fiber end into water, its value should be identical to the one of water, which is about 1.33.The caluculated values of the collection efficiency ηNA0using Eq. (1) at nw=1.33are shown in Fig.4 as a function of NA0. It is easy to see that ηNA0monotonically increases with NA0.

Figure 4.

Calculated values of the collection efficiency η as a function of NA0 at nw=1.33.

In the following, let us consider the situation where the other optical fiber end is optically coupled to a photon detector with a detection window of diameter ϕ4and with a circular sensitive area having a diameter ϕ3and a numerical aperture NA3. The coupling efficiency εbetween the optical fiber end and the photon detector depends on ϕ0, NA0of the optical fiber and ϕ3, NA3, ϕ4of the photon detector used.The number of emitted photons is proportional to the square of ϕ0and ηNA0monotonically increases with NA0, so that the number of the transmitted photon to the other optical fiber end is proportional to ϕ02and ηNA0. On the other hand, the coupling efficiency εgenerally decreases as ϕ0or ηNA0increases for the fixed ϕ3, NA3, and ϕ4. Thus, we can define the following formula for a figure of merit (FOM) and optimize ϕ0, NA0and parameters of the coupling optics xifor the fixed values of ϕ3, NA3, and ϕ4to maximize the FOM:

FOM=ϕ2ηNA0εϕ0,NA0,xi,ϕ3,NA3, ϕ4E2

It should be noted that the coupling efficiency εcan be ideally 100 % under the condition of ϕ0ϕ3and NA0NA3.In many cases, the condition NA0NA3is satisfied when using the typical photon detector. Under the condition NA0NA3, thus, we can categorize into two cases: case (1) is ϕ0ϕ3and case (2) is ϕ0>ϕ3. In the case (1), εis constant and can be ideally 100 %, so that the detection efficiency excluding the quantum efficiency of the photon detector is only limited to ηNA0. Therefore, the optimization of ϕ0and NA0is not necessary. The conditions of ϕ0=ϕ3and NA0=NA3both maximize the FOM and the sensitivity becomes highest. In the case (2), however, the optimization of ϕ0and NA0and a specific design of the coupling optics are necessary, because the coupling efficiency εdecreases as ϕ0or NA0increases.

2.2. Construction with a cooled PMT detector

2.2.1. Photon detectors

To construct the optical fiber-based system, a choice of a single photon detector is very important. Photon detectors generally have two significant factors contributing to the sensitivity of detection for weak light: the efficiency and the dark counts of the detector. Recently, two types of single photon detectors, which are a cooled APD and a small size of cooled PMT, are available. The cooled APD which can detect for single photons is mostly used because of the high quantum efficiency and the low dark counts. The sensitive area must be very small ( ϕ0~0.2mm ), but the quantum efficiency is several times larger than that of a typical PMT. Furthermore, it has the useful characteristics of compactness, easy operation, and durability compared to a typical PMT detector. On the other hands, the compact size of cooled PMT is also useful for the optical fiber-based system. The quantum efficiency is typically lower than the APD, but the sensitive area is roughly 10 times larger than the APD’s in spite of the same dark counts as the cooled APD. Therefore, it is very easy to construct a coupling optics to the sensitive area of the detector with the high sensitivity.

We selected the PMT counting head (H7421) manufactured by Hamamatsu Photonics K.K. for this system. Its characteristics are summarized in Tab. 1. For comparison, we also present the characteristics of the APD-type photon counting module (SPCM-AQR-14) provided by Perkin Elmer. Ltd. It has already been verified that this APD is applicable to the fiber-based system by our previous investigation as discussed in [3], [4].

Quantum efficiency(ηqe )Dark noisesensitive area(ϕ3)NumericalAperture(NA3)Detection window(ϕ4)
PMT (H7421)40% at 550nm100 s-15.0 mm0.1237.2 mm
APD (SPCM-AQR-14)55% at 550nm100 s-10.175 mm0.786.16 mm

Table 1.

Characteristics of photon detectors in reference [17] and [18]

The value of NA3can be calculated from the geometrical structure between the sensitive area and the photon detection window.

2.2.2. Coupling efficiency

As showing in section 2.1, under the condition of ϕ0>ϕ3, it is necessary to optimize ϕ0, NA0and to design the coupling optics for maximal sensitivity. In the use of the PMT detector, it is not absolutely necessary to optimize ϕ0, NA0, because the characteristics of ϕ3=5.0mmsufficiently satisfies the condition of ϕ0<ϕ3for the typical optical fiber. For easy construction, here, the optical fiber end was directly connected to the attachment of the PMT counting head without the additional coupling optics as shown in Fig. 5.

Figure 5.

Geometrical structure of the PMT counting head

The coupling efficiency εϕ0,NA0in such geometrical structure can be obtained by the statistical method with matrix formalism in paraxial optics, which can describe the propagation of light. The light at the initial state ri,r'i, where riis a distance from an optical axis and r'iis a slope of the light direction, can be transferred by some matrices. In the case of Fig. 5, the matrix expressing free-space propagation can transfer the initial state ri,r'iat the optical fiber end to the final state rf,r'fat the PMT sensitive area. We obtained the value of εϕ0,NA0by counting the number of the final states inside the sensitive area for many initial states selected with random numbers, which were generated by using the software package based on algorithm of Mersenne Twister in reference [19]. The calculated results are shown as a function of NA0in Fig. 6. Since the value of ϕ3is approximately 5 times as large as one of ϕ0, the coupling efficiency εϕ0,NA0is independent of ϕ0, but monotonically decreasing with NA0. Therefore, we calculated the FOM for confirming the existence of optimal conditions for ϕ0,NA0. Fig. 7 shows the plot of the calculated FOM as a function of NA0and obviously indicates that the FOM is almost constant to NA0. Thus, we determined NA0=0.37and ϕ0=1.0 mmbecause of availability and flexibility of the optical fiber. These values easily give the coupling efficiency εϕ0,NA0of 21.7% and FOM×100of 0.420.

Figure 6.

Calculated values of the coupling efficiency ϕ0 as a function of NA0. The solid circles represent the values at ϕ0=0.8  mm, the solid triangles are the values at ϕ0=1.0  mm, and the solid inverse triangles are the values at ϕ0=1.5  mm.

On the other hands, in the use of the APD, the optimization of NA0=0.37, ϕ0=0.6 mmand the design of the optimal coupling optics are absolutely necessary. A simple optically coupling way is shown in Fig. 8. With the above APD, the condition of FOM×100and εϕ0,NA0maximizes the FOM and the value of ηNA0is 0.234. The determined design parameters of the coupling optics give the maximum εof 33.3% and NA0of 1.97%. The detailed description about the determination of the design parameters is given in [3] and [4].

Figure 7.

Calculated values of FOM as a function of NA0. The solid circles represent the values at ϕ0=0.8  mm, the solid triangles are the values at ϕ0=1.0  mm, and the solid inverse triangles are the values at ϕ0=1.5  mm.

Figure 8.

One example of additional coupling optics

The relative sensitivity for the bioluminescence detection can be compared by using a product of FOM and the quantum efficiency ηqeof the detector as an indicator. For the PMT, the value of (FOMηqe)×100is 0.168 and 0.129 for the optimal values of the above APD.In such direct connection, the sensitivity with the PMT is almost same as with the APD. For the system with the PMT, furthermore, it should be noted that the coupling efficiency εat NA0=0.37and ϕ0=1.5 mmcan be 100% using the additional coupling optics as shown in Fig. 8. These values give (FOMηqe)×100= 1.77, which is about 10 times larger than the present value.

3. Sensitivity test for ATP detection

3.1. Luciferin-luciferase reaction

Bioluminescence in living organisms, such as fireflies and some marine bacteria, typically occurs due to the optical transition from the excited state to the ground state of oxidized luciferin molecules produced by the luciferin-luciferase reaction under the catalytic activity of luciferase molecules. This reaction can be expressed by the following squential of reaction steps:

E+ S + LH2C1+PPi

C1+O2C2+AMP+CO2+ γ

C2 E+P ,

where E indicates luciferase, SATP, LH2luciferin, PPipyrophoric acid, C1is an enzyme-substrate compound ELH2-AMP, AMPadenosine monophosphate, Poxyluciferin ( oxidized luciferin ), C2a luciferase-oxyluciferin compound, and γa photon in reference [20]. The emission of one photon at the position of luciferase molecule corresponds to the use of the energy of one ATP molecule. Against the second reaction which is a rate-limiting reaction in the above reaction chain, it is known that the following reaction is competitive,

C1+O2C3+H2O2

C3 E+L-AMP ,

where L-AMPrepresents dehydroluciferyl-adenylate, C3is an enyzme-substrate compond EL-AMPas described in [21]. This reaction does not induce the photon emission. Dehydroluciferyl-adenylate L-AMPas well as oxyluciferin is known as a competitive inhibitor to the equilibrium reaction in [22].

In the presence of enough luciferin molecules, the immobilization of luciferase molecules at the optical fiber end allows us to sense the present of ATP around the fiber end using single photon counts. For this purpose, we used a compound protein containing a silica-binding protein ( SBP ) molecule and a luciferase molecule ( SBP-luciferase ), which were recently synthesized by Taniguchi and co-workers in [23]. This protein makes it possible to immobilize a lucuferase molecule on the optical fiber end through a SBP molecule retaining its activity. The spectrum of the emitted photons shows a central wavelength of 550 nm and a width of about 100 nm in reference [24] and [25]. Both photon detectors of the APD and the PMT have the large quantum efficiency at 550 nm, the photon counting detectors are suitable for ATP sensing.

3.2. Michaelis-Menten formula

In the solution containing nonlocalized homogeneously dispersed luciferase and ATP, the Michaelis-Menten formula is applicable to the enzyme reaction as descibed in [26]. In the presence of sufficient luciferin molecules in the solution, a rate of emitted photons at steady state vγcan be expressed as the Michaelis-Menten formula,

vγ=Vmaxss+KM,E3

where Vmaxis a maximum reaction rate which is equivalent to a product of a concentration of luciferase molecules and a reaction constant h1from C1to C2, KMis the Michaelis constant, and sis the ATP concentration. In the fiber-based system for sensing dispersed ATP molecules, on the other hands, an ATP-flow generated by a gradient of ATP concentration around the luciferase-terminated fiber end carries ATP molecules to the vicinity of immobilized luciferase molecules. One of ATP molecules is bound to one immobilized luciferase molecule near them and subsequently used for the luciferin-luciferase reaction at this fiber end. By solving reaction-diffusion equations, as described in [4], we have confirmed that an ATP diffusion rate is not a rate-limiting process when the number of immobilized luciferase molecules is 1011order and the ATP diffusion constant is D=0.5×10-3mm2/sgiven in [27]. Therefore, the Michaelis-Menten formula Eq. (3) can be also applied for the fiber-based system without change.

3.3. Mesurement of the sensitivity

3.3.1. Experimental setup

Fig.9 shows the experimental setup for the sensitvity test and the investigation of bioluminescence at the optical fiber end. One optical fiber end was optically connected to the PMT as describing in the previous section. On the other fiber end, the luciferase molecules were immobilized via SBP molecules and the bioluminescent reaction occurs by immersing the luciferase-terminated fiber end into a sample solution. The emitted photons were transmitted to the PMT through the optical fiber and TTL pulses outputted from the PMT were counted by a PC card installed in a personal computer(PC). To reducing the background light, the whole system was put into the dark box.

For the observation of the bioluminescence rising, a test tube containing the sample solution was fixed on the Z-stage, which is a motorized stage and externally controllable. By raising the test tube for immersing the luciferase-terminated fiber end after starting the data acquisition system, the photon counts rising from the background level and subsequently reaching a maximum were observed with time. The numbers of detected photons during 0.0298 s or during 0.1 s were recorded every 0.0321s or every 0.1s by the PC, respectively. These values were obtained from a calibration test of data acquisition system. The details on the experimental setup and the measurements are described in reference [28].

Figure 9.

Experimental setup for investigation of bioluminescence at the optical fiber end

3.3.2. Immobilization of luciferase

Before immobilizing luciferase molecules, we cut the optical fiber and cleaned the cut surface with ethanol and Tris buffer (0.25mM Tris-HCl with 0.15 M NaCl). Different from the previous experiments with the APD, we cleaved the optical fiber for making a flat surface on the fiber end, which can reproduce the number of immobilized luciferase. The flat surface also allows us to indivisually evaluate the number of immobilized luciferase molecules by using element analysis, although the sensitivity with the flat surface is about 10 times lower than the one with the appropriately irregular surface cut without the cleaving technique as described in [29]. The cut surface was direcly observed by using the fiber scope and checked its flattness by eyes. After cleaning, the surface was immersed in a solultion of SBP-luciferase and was left at a temperature of 3to 6for a period of about two hours.

For evaluating the number of immobilized luciferase molecules, element analysis to the fiber end was carried out by using Xray Photoemission Spectroscopy (XPS). We measured a spectrum including peaks from nitrogen in the SBP-luciferase and from silicon on the surface of the fiber end which made with the silica and obtained the ratio of the area of the nitrogen-peak to the one of the silicon-peak. Utilizing the absolute number of silicon on the surface of the fiber end, the surface density of immobilized luciferase molecules was determined to be 2.0×1010mm-2. By repeating the same measurement and analysis, the error of the surface density was estimated to be 15 %.

3.3.3. Sample solutions

The samples were a 1:4:4:31 mixture of 20 mM D-luciferin solution, Tris buffer solution( 250 mM Tris-HCl mixed with 50 mM MgCl2 ), ATP solution, and distilled water. Several solutions of ATP with different ATP concentrations were made by diluting the ATP standard in ATP Bioluminescence Assay Kit CLS II manufactured by Roche Co. Ltd. A series of sample solutions with different ATP concentrations were prepared in advance. To obtain a background before the ATP measurements, an additional sample without ATP was also produced by mixing distilled water instead of the ATP solution.

3.3.4. Results

The time dependence of photon counts per 0.1-s interval were measured in immersing the luciferase-terminated fiber end into the sample solutions with various ATP concentration and converted to the values of photon counting rate. A typical result for 100 μlat the ATP concentration of 1.65×10-6M is shown in Fig. 10.

Figure 10.

Time dependence of measured counting rate for 100 μl at the ATP concentration of 1.65×10-6 M.

The photon counts rise up, reach a maximum at about 100 s, and decrease toward the background level with time scale of 1000 s after the immersion. The background level was about 130 s-1, which was essentially determined by the dark counts. For observing the detection limit of the ATP concentration, we obtained the integrated counts of detected photons over the time range from 0 to 100 s for various ATP concentrations. The result is shown as a function of the ATP concentration in Fig. 11.

Figure 11.

Number of photons integrated over the time range from 0 to 100 s as a function of ATP concentration.

Statistical errors were estimated as one standard deviation assuming Poisson distribution. From Fig. 11, the sensitivity in this system is limited to 1.65×10-9M, whichcorresponds to a number of ATP molecules of about 10-14mol in the 10 μlsolution. This value is about 10 times higher than the one in the previous experiment with the APD as described in [3] and [4], although the FOM of this system is almost same as of the APD system. In this experiment, the surface flatness of the luciferase-terminated fiber end makes us identify the effective area as the cross section of the cut surface, as while the effective area of the cut surface in the previous experiment with the APD system was enlarged due to a surface asperity.On the effect of such different cutting ways, we have already confirmed that the sensitivity in the flat surface is about 10 times lower than the one in the appropriately irregular surface cut without the cleaving technique. Therefore, the above results are consistent with our previous results by using the APD.

To check the ATP concentration dependence of the photon counting rate at maximum, the average of counts in sixteen 1-s intervals around the time at which the counting rate become maximal was calculated for each ATP concentration. The results are indicated by the solid circles in Fig. 12. By the analysis of fitting data points in Fig. 12 to Eq. (3), we obtained the Michaelis constant of KM=6.47×10-5M and the maximum reaction rate of Vmax=7.38×104s-1.

Figure 12.

Measured photon counting rate as a function of ATP concentrations. Solid line is a curve obtained by fitting data with the Michaelis-Menten formula.

3.3.5. Discussion

The detection limitis essentially determined by both of the parameter Vmaxand the dark noise of the photon detector. Since the Vmaxcan be expressed as Vmax=εtotalh1e0, where h1is a reaction rate of one luciferase molecule, e0is a total number of immobilized luciferase molecules on the optical fiber end and εtotalis a total detection efficiency, we can calculate the value of Vmaxwith εtotal=ηεηqewhich is 0.00171 at 550 nm, ϕ0=1mm, the surface density of immobilized luciferase given by 2.0×1010mm-2, and h1=0.125 s-1in reference [30]. The prediction of Vmaxis 3.36×106s-1, which is two orders of magnitude larger than the obtained value of 7.38×104s-1. In our previous experiment with the APD, the predicted value of Vmaxwas also one order of magnitude rather than the obtained one. Possible reasons are a reduction of h1, or an existance of inactive luciferase molecules, or both of them as discussed in reference [4] and [31]. To clarify the reason, it is necessary to individually evaluate the number of active immobilized luciferase molecules eaand the reaction rate h1, respectively, from experimental data.

In the PMT system, it is noted that the improvement of two orders of magnitude for Vmaxis promising by using the optimal optics coupled to the optical fiber with ϕ0=1.5 mmand NA0=0.37and the optical fiber end with the appropriately irregular surface cut without the cleaving technique for immobilizing the luciferase.

4. Investigation of bioluminescence at the optical fiber end

4.1. Measurement of the bioluminescence with high time resolution

To obtain the reaction rate h1, the number of acitve luciferase molecules ea, and other kinetic parameters from experimental data, the counts of photons with high time resolution and the advaced analysis to such data are necessary. The data acquisition system has a capability of recoding the numbers of detected photons every 0.0321s. By making a full use of this specification, the time dependence of detected photons with high time-resolution at the ATP concentration of 1.65×10-4M was obtained as shown in Fig. 13.

Figure 13.

Time dependence of detected photons with the resolution of 0.0312 s at the ATP concentration of 1.65×10-4 M. The upper figure (a) shows the result in the solution of 100 μl. Solid line represents an extrapolation of the fitting curve with the parameters obtained by fitting the data from 0 s to 30 s. The lower figure (b) shows a magnified plot around the peak.

Fig. 13 (a) shows the result of the detected photons with the immobilized SBP-luciferase molecules at the optical fiber end into the solution of 100 μlat the ATP concentration of 1.65×10-4M. For comparison, we also measured the time depencence of the photons with homogeniously dispersed SBP-luciferase molecules in the solution of 500 μl. The direct measurement of bioluminescence was carried out with an other type of cooled PMT detector having a huge sensitve area of 1 cm×1 cmto detect the bioluminescence from a large area of the solution. The details on the direct measurement and the data analysis for dispersed luciferase molecules are given in reference [28].

4.2. Analysis

4.2.1. Reaction model including inhibitors

For obtaining kinetic parameters of bioluminescient reaction, we consider the rate equations including the effects of inhibitors. In the luciferin-luciferase reaction, two kinds of products, oxyluciferin and L-AMPare strong cometitive inhibitors to substrates. Each equilibrium constant for exyluciferin Ki and for L-AMPKj has been measured and the values of Ki=0.5 μMand Kj=3.8 nMare given in reference [22]. For simplicity of the model, we assume that two inhibitors contribute the competitive inhibition to the equilibrium reactions between luciferase and ATP in the presence of enough luciferin molecules. Fig. 14 shows the reaction steps of luciferin-luciferase reaction including the effects of competitive inhibitors.

Figure 14.

Enzyme reaction including the inhibitors

Here, s, e, p, nγ, c1, c2, c3in Fig. 14 represents a concentration of ATP, luciferase, oxyluciferin, photon, ELH2-AMP, EP, and EL-AMP, respectively, k+, k-, ki+, and ki-are kinetic coefficients for equilibrium and h1and h2are reaction rates. Since the inhibition by L-AMPis much stronger than the ones by oxyluciferin, it can be assumed that the enzymes in the state of the enzyme-susbstrate compound do not release L-AMPmolecules and concequently lose the activation in the time scale of our experiment.

In the use of the immobilized luciferase molecules for sensing dispersed ATP molecules, it is natural to consider that the reaction occures in a volume ΔVwhich is the vicinity of the luciferase-terminated optical fiber end. Therefore, the serise of the rate equations describing the enzyme reaction shown in Fig. 14 in the volume ΔVcan be expressed as

dedt=-k+es+k-c1+ki-c2+ki+epE4
dc1dt=k+es-k-+h1+h2c1E5
dc2dt=h1c1-ki-c2+ki+epE6
dc3dt=h2c1E7
dnγdt=h1c1E8
NAV0dsdt=-k+se+k-c1E9
NAΔVdpdt=-ki+ep+ki-c2,E10

where the variable e, nγ, c1, c2, c3is a number of each kind of molecules or photon in the volume ΔV. The variable sand pis the concentration of ATP and oxyluciferin, respectively, and their unit is M. The NAis Avogadro number and the V0is a volume of the solution. The unit of k-, ki-, h1, h2is s-1and of k+, ki+is M-1s-1. The volume V0is explicitly utilized into Eq. (9), since the concentration of ATP in the volume of V0is almost same as in the volume of  ΔVbecause of a rapid diffusion rate. The volume ΔVcan be approximately considered as a cylinder with a diameter of ϕ0=1 mm and a height of 10 nm because the size of SBP-luciferase molecules is about several nanometer, and its value is ΔV=7.85×10-12 l.

In addition to the above rate equations, the following conditions described as,

ea=e+c1+c2+c3E11
nγ=c2+NAΔVpE12

should be satisfied. The condition of Eq. (11) shows that the total number of active luciferase molecules eais constant and Eq. (12) means that the total number of photons is equivalent to that of oxyluciferin molecules. The Michaelis constant Kmand the equilibrium constant of oxyluciferin Kican be expressed as

Km=k-+h1+h2/k+E13
Ki=ki-/ki+E14

Using Eq. (11), Eq. (12), Eq. (13), and Eq. (14) as boundary conditions and inputting the constant values of ea, h1, h2, k+, ki-, Km, Kiand the initial values for variables, we can numerically solve the rate equations and obtain the time evolution of nγas the numerical solution at each time step. To simplify the numerical calculation, we assumed that oxyluciferin molecules instantaneously move out from the volume of ΔVbecause of the fast diffusion rate of the oxyluciferin molecules. Therefore, the kinetic constant ki+can be practically zero. This treatment means that the influence of the competitive inhibition by the oxyluciferin can be neglected.

In the solution containing non-localized homogenous dispersed SBP-luciferase, the volume of ΔVin the equations is replaced with the volume of the solution V0and the kinetic constant ki+is treated as non-zero.

4.2.2. Results of analysis

The five parameters aea,k+,ki-,h1,h2were treated as fitting parameters and the data was fitted to numerical solution of nγfor obtaining the values of a. As the first step, by inputting the initial values of ato the rate equations, the emitted photon Nthiawas calculated every time step of 0.0321 s with Ki=0.5 μMgiven in [22] and Km=6.47×10-5M, which was deduced from the data analysis shown in Fig. 12. The data-set of Nthia, the n0number of measurend counts per 0.0321 s Nexpi, and the background counts Nexp0enable us to calculate a chi-square χ2with the formula given by

χ2a=i=1n0Nexpi-Nexp0-Nthia2Nexpi+Nexp0E15

The chi-square χ2ais a good indicator for fitting data and its minimum gives the optimal combination of probable values in a. Solving the series of the rate equations was executed by using the software package RKSUITE based on the Runge-Kutta method [32] and the minimization of the chi-square was performed with routines of MINUIT package provided by CERN software [33].

The result of fitting the data from 0 s to 30 s is represented as a solid line in Fig. 13 (a), which is extrapolated to 60 s using the obtaind parameters. This result is not reproduced completely in the time range from 0 s to 60 s, because the effect of the competitive inhibition of oxyluciferin is not considered and the fitting fuction includes only the contribution of the deactivation process. The inhibition of the oxyluciferin weakens with time due to its diffusion process, but this diffusion effect is not considered in this analysis. In contrast, the contribution of the deactivation process, which was evaluated by fitting the data around the peak, is concequently overestimated compared to the actual contribution. Therefore, the effect of the relatively strong evaluation for the deactivation process appears in the time range after 30 s.

The parameters obtained by fitting the data are summarized in table 2 together with the results of the dispersed luciferase for comparison. The parameter rrepresents the activation ratio of the SBP-luciferase, which can be defined as a ratio of the number of active luciferase molecules eato the total number of immobilized luciferase molecules e0, which can be calculatedfrom the surface density of immobilized SBP-luciferase or from the concentration of the dispersed SBP-luciferase. In addtion, the parameter k-and ki+was derived from Eq. (13) and Eq. (14), respectively. The statistical error was 3 % at the maximum, but the systematical error was estimated to be 20 % taking account of the errors of the numerical calculation and the parameters used.

dispersed luciferaseimmobilized luciferase
Volume of solution500 μl100 μl
Region used for fitting0 – 60 s0 – 30 s
r ( ea/e0)0.440.010
k+1.7×104M-1s-12.1×104M-1s-1
h10.21s-10.61s-1
h20.090s-10.073s-1
ki-0.090s-10.25s-1
k-0.83s-10.68s-1
ki+1.8×105

Table 2.

Summary of obtained parameters

4.3. Discussion

In table 2, it is easily seen that the kinetic parameters in the immobilized luciferase are almost same as in the non-localized dispersed luciferase except the reaction rate of h1. Since the rising time is approximately given by 1/k+s, it is useful for checking a consistency of k+with the reference. For dispersed luciferase, the rising time of 0.29 s is obtained with k+=1.7×104M-1s-1and for immobilized luciferase, the rising time of 0.36 s is given by the value of k+=2.1×104M-1s-1. Both of them are close to the value of 0.3 s in reference [34], so that their values are consistent with the reference.

On the reaction rates, the obtained value of h1=0.61 s-1for immobilized lucifrase is about 5 times larger than the reference value 0.125s-1given in [30], but the order of both values is the same. A more precise comparison is not meaningful, because the surrounding environment of luciferase is not exactly same as in the reference.On the other hands, a branching ratio to the deactivation process, which is given by h2/h1+h2, can be estimated to be 30 % for immobilized luciferase and 10 % for dispersed luciferase, respectively. Since the reference value of 20 % is given in [22], both of them are close to the refrence value. Thus, we can consider the obtained values of the parameters are almost consistent with the values which had been measured so far.

From table 2, the activation ratio ris 44 % for the dispersed SBP-luciferase and 1 % for the immobilized SBP-luciferase, respectively. In contrast, the reaction rate h1is the same order as the value we expected. Therefore, the detection limit for ATP detection results in two orders of magnitude larger than the expected one. As a concequence, the results of the sensitivity test described in Section 3 can be explained from the reduction of the activation ratio for the immobilized luciferase molecules.

5. Summary

We introduced a method of high-sensitivity detection of bioluminescence at an optical fiber end for an ATP detection as an efficient alternative to direct detection of bioluminescence for a sample solution. For investigation of the bioluminescence, we constructed an optical fiber-based system, where the luciferase molecules are immobilized on the optical fiber end and the other end is optically coupled to a compact size of cooled PMT-type photon counting head which has a large sensitive area. Although the sensitivity for the bioluminescence is not optimal, it is almost same as the system which had been constructed with an APD-type photon counting detector. We have evaluated the sensitivity for ATP detection and verified the detection limit of 10-9M which is consistent with the previous results with the APD-type detector. This detector limit allows us to detect the absolute ATP number of 10-14mol in a 10 μlsolution, but it is two orders of magnitude larger than the expected one. For clarifying the reason, we have performed measurements with high time resolution and analyses of data by using an enzyme reaction model including inhibitors to individulally obtain an activation ratio and a reaction rate of the immobilied luciferase. As the results, the reaction rate of 0.61 s-1and the activation ratio of 1 % have been obtained and these results have explained the reason of two orders of magnitude higher than the expected one. For reducing the detection limit more, it is necessary to improve the activtion ratio of the immobilized luciferase on the optical fiber end as well as the enlargement of the effective area of the cut surface based on a surface asperity and the increase of the FOM with the optimal values of a core diameter ϕ0, a numerical aperture NA0and parameters of the coupling optics.

Acknowledgements

We are grateful to Prof. Hiroyuki Sakaue for supporting the element analysis with XPS and Prof. Kenichi Noda for useful supports to the experiments. This work has been partially supported by the International Project Center for Integration Research on Quantum, Information, and Life Science of Hiroshima University and the Grant-in-Aid for Scientific Research (C)(19560046) of Japanese Society for the Promotion of Science, JSPS.

How to cite and reference

Link to this chapter Copy to clipboard

Cite this chapter Copy to clipboard

Masataka Iinuma, Ryuta Tanaka, Eriko Takahama, Takeshi Ikeda, Yutaka Kadoya and Akio Kuroda (June 13th 2013). Investigation of Bioluminescence at an Optical Fiber End for a High-Sensitive ATP Detection System, Current Developments in Optical Fiber Technology, Sulaiman Wadi Harun and Hamzah Arof, IntechOpen, DOI: 10.5772/52747. Available from:

chapter statistics

1221total chapter downloads

More statistics for editors and authors

Login to your personal dashboard for more detailed statistics on your publications.

Access personal reporting

Related Content

This Book

Next chapter

Smart Technical Textiles Based on Fiber Optic Sensors

By Katerina Krebber

Related Book

Frontiers in Guided Wave Optics and Optoelectronics

Edited by Bishnu Pal

First chapter

Frontiers in Guided Wave Optics and Optoelectronics

By Bishnu Pal

We are IntechOpen, the world's leading publisher of Open Access books. Built by scientists, for scientists. Our readership spans scientists, professors, researchers, librarians, and students, as well as business professionals. We share our knowledge and peer-reveiwed research papers with libraries, scientific and engineering societies, and also work with corporate R&D departments and government entities.

More about us