Open access peer-reviewed chapter

Design and Control of an EMG Driven IPMC Based Artificial Muscle Finger

By R.K. Jain, S. Datta and S. Majumder

Submitted: February 3rd 2012Reviewed: May 10th 2012Published: October 17th 2012

DOI: 10.5772/48814

Downloaded: 2786

1. Introduction

The medical, rehabilitation and bio-mimetic technologydemands human actuated devices which can support in the daily life activities such as functional assistance or functional substitution of human organs. These devices can be used in the form of prosthetic, skeletal and artificial muscles devices (Andreasen et al., 2005; Bitzer & Smagt, 2006; DoNascimento et al., 2008). However, we still have some difficulties in the practical use of these devices. The major challenges to overcome are the acquisition of the user’s intention from his or her bionic signals and to provide with an appropriate control signal for the device. Also, we need to consider the mechanical design issues such as lightweight and small size with flexible behavior etc (Arieta et al., 2006;Shenoy et al., 2008). For the bionic signals, the electromyography (EMG) signal can be used to control these devices, which reflect the muscles motion, and can be acquired from the body surface. We are familiar with the fact that ionic polymer metal composite (IPMC) has tremendous potential as an artificial muscle. This can be stimulated by supplying a small voltage of 3V and shows evidence of a large bending behavior (Shahinpoor & Kim, 2001; 2002; 2004; Bar-Cohen, 2002).In place of the supply voltage from external source for actuating an IPMC, EMG signal can be used where EMG electrodes show a reliable approach to extract voltage signal from body (Jain et al. 2010a;2010b; 2011). Using this voltage signal via EMG sensor, IPMC can illustrate the bio-mimetic behavior through the movement of human muscles. Therefore, an IPMC is used as an artificial muscle finger for the bio-mimetic/micro robot.

The main objective of this chapter is to discuss the design and control of an IPMC based artificial muscle finger where this finger is actuated by EMG signal via movement of human finger. The movement is sensed by EMG sensor which provides signal for actuating the IPMC. When designing IPMC artificial muscle finger based micro gripper for handling the light weight components in an assembly, IPMC bending behaviour is utilized to hold the object. During holding the object, stable EMG signal is required. For this purpose, stable EMG signal is sent through proportional–integral–derivative (PID) controller to the system. Experimentally, it is found that IPMC based artificial muscle finger achieves similar movement like human index finger. This IPMC based artificial muscle finger attains deflection upto 12 mm. By developing a prototype of IPMC artificial muscle finger based micro gripper, it is demonstrated that EMG driven system like IPMC artificial muscle finger based micro gripper can be applicable in handling of light weight components. The major advantages of such system are that IPMC based artificial muscle finger tip shows the compliant behavior and consumes less energy for actuation. Therefore, EMG driven system shows enough potential to substitute for conventional mechanism in micro manipulation and rehabilitation technology.

This chapter is organized as follows: Section 2 describes the prior research related to EMG applications in robotics and bio-mimetics. Section 3.1 explains the basic design of IPMC artificial muscle finger based micro gripper which is driven by EMG signal. The basic tendon of index finger is studied in section 3.2 where muscles are identified for actuation of IPMC based artificial muscle finger. In section 3.3, a model for controlling the EMG signal is highlighted. Different types of control system are implemented for achieving stable data from EMG signal via index finger which is sent to IPMCbased artificial muscle finger. Section 4 discusses experimental testing setup for activation of IPMC based artificial muscle finger by human finger through EMG. In section 5, the results are discussed and the conclusions are drawn in section 6.The future work is recommended in section 7.

2. Prior research related to EMG applications

In the past, some researchers have reported work related to shape memory alloy (SMA) and other similar actuators to develop the bio-mimetic fingers but IPMC artificial muscle based finger related work is limited. Pfeiffer et al. (1999) have designed artificial limbs and robot prostheses that are lightweight, compact and dexterous. This mimics the human anatomy and maintains a high lifting capability. EMG control is used for SMA actuated fingers in robot prostheses. De Laurentis & Mavroidis (2002) have discussed the design of a 12 degree-of-freedom (DOF) SMA actuated artificial hand where the SMA wires are embedded intrinsically within the hand structure. Cocaud & Jnifene (2003) have investigated the use of artificial muscles as SMA actuators for robot manipulators. A solution is established in order to determine the optimal position of a muscle in various musculoskeletal configurations. Herrera et al. (2004) have also designed and constructed a prosthesis where linear actuators are used for designing the mechanical system and EMG sensors are introduced for designing the electrical control system. Bundhoo et al. (2005, 2008) have reported the design of artificially actuated finger by SMA towards development of bio-mimetic prosthetic hands. Different finger joints are actuated through SMA wires via EMG and the relationship between elongation/contraction of the SMA wires & the finger joints have been obtained. O’Toole & McGrath (2007) have also focused on mechanical design of a 12 DOF SMA actuated artificial hand. The SMA material is used for combination of high strength polymers such as polytetrafluoroethylene (PTFE), polyether ether ketone (PEEK) and low density metals such as titanium.Lau (2009) have carried out research work on a design and development of an intelligent prosthetic hand based on hybrid actuation through DC motor & SMA wires. These are controlled by myoelectric signal. Two novel features are introduced in the new prosthetic hand. Firstly, its hybrid actuation mechanism has the advantage of increasing the active degrees of freedom and secondly, using only two myoelectric sensors, has the potential for controlling more than three patterns of fingers movements. Pittaccio & Viscuso (2011) have developed a SMA wire device for the rehabilitation of the ankle joint where active orthosis powered by two rotary actuators like, NiTi wire are used to obtain ankle dorsiflexion and EMG signal is used to control the orthosis and trigger activation from muscle. Stirling et al. (2011) have shown the potential of SMA wire for an active, soft orthotic in the knee where NiTi based SMA wires is also used. A prototype is tested on a suspended, robotic leg to simulate the swing phase of a typical gait. Thayer & Priya (2011) have designed a biomimetic dexterous humanoid hand where the dexterity of the DART hand have been measured by quantifying functionality and typing speed on a standard keyboard. The hand consists of 16 servo motors dedicated to finger motion and three motors for wrist motion where some of joints are activated through SMA wires.

Some of the researchers have focused on the design of a biomechatronic robotic hand using EMG. Cheron et al. (1996) have found the relationship between EMGand the arm kinematics through dynamic recurrent neural networks (DRNN) method whereasHudgins et al. (1997) have focused on a new control scheme, based on the recognition of naturally myoelectric signal patterns, transfers the burden of multifunction myoelectric control from the amputee to the control system. Rosen et al. (2001) have developed a myosignal-based exoskeleton system. This is implemented in an elbow joint, naturally controlled by the human. The human–machine interface is set at the neuromuscular level, by using the neuromuscular signal (EMG) as the primary command signal for the exoskeleton system. The EMG signals along with the joint kinematics are fed into a myoprocessor (Hill-based muscle model) which in turn predicts the muscle moments on the elbow joint. Banks (2001) has given remarkable effort towards design and control of an anthropomorphic robotic finger with multi-point tactile sensation whereas Light et al. (2002) have emphasized on intelligent multifunction myoelectric control of hand prostheses. Peleg et al. (2002) have extracted multiple features via EMG signal from hand amputees which is selected by help of a genetic algorithm. Fukuda et al. (2003) have developed a prosthetic hand where human-assisting manipulator system based on the EMG signals is utilized. Wheeler (2003) has presented a neuro-electric interface method for virtual device control. The sampled EMG data is taken from forearm and then is fed into pattern recognition software that has been trained to distinguish gestures from a given gesture set. Krysztoforski et al. (2004) have given remarkable effort towards recognition of palm finger movements on the basis of EMG signals with the application of wavelets.

Crawford et al. (2005) have used EMG signals for classifying in real-time with an extremely high degree of accuracy in a robotic arm-and-gripper. A linear support vector machines (SVM) based classifier and a sparse feature representation of the EMG signal are used. Hidalgoet al. (2005) have proposed a design ofrobotic arm employing fuzzyalgorithms to interpretEMG signals from the flexor carpi radialis, extensor carpi radialis and biceps brachiimuscles. The control and acquisition systems are composedof a microprocessor, analog filtering, digital filtering & frequency analysis, and finally a fuzzy control system.Mobasser & Hashtrudi-Zaad (2005) have estimated rowing stroke force with EMG signal using artificial neural network method from upper armmuscles which is involved in elbow joint movement,sensed elbowangular position and velocity.Gao et al. (2006) have focused on acquiring the data from the upper limb of the body for robotic arm motion using EMG whereas Frigo et al. (2007) have detected EMG signal from voluntarily activated muscles which is controlled for functional neuromuscular by electrical stimulation. A comb filter (with and without a blanking window) is applied to remove the signal components synchronously correlated to the stimulus. Roy et al. (2007) have compared the performance of different sEMG signal at various conditions. These performances depend on the electro-mechanical stability between the sensor and its contact with skin. Zollo et al. (2007) have put a remarkable effort on the control system of biomechatronic robotic hand and on the optimization of the hand design in order to obtain human like kinematics and dynamics. By evaluating the simulated hand performance, the mechanical design is iteratively refined. The mechanical structure and the ratio between numbers of actuators to the number of DOF have been optimized. Yagiz et al. (2007) have developed a dynamic model of the prosthetic finger where a non chattering robust sliding mode control is applied to make the model follow a certain trajectory. Wege & Zimmermann (2007) have shown the potential of EMG control for a hand exoskeleton device. The device has been developed with focus on support of the rehabilitation process after hand injuries or strokes. Itoh et al. (2007) have studied the hand finger operation using sEMG during crookedness state of the finger. Two electrodes (Ag/AgCl electrodes) are sticked randomly on the forearm muscles and the intensity of EMG signals at different muscles is measured for each crooked finger.

Hao et al. (2008) have studied the design of pneumatic muscle actuator based robotic hand where its compliance and dexterity handling are attempted.A single finger is controlled by fuzzy & PID controller and comparative studies are discussed. Murphy et al. (2008) have explored the micro electro-mechanical systems based sensor for mechanomyography system whereas Saponas et al. (2008) have also explored the feasibility on muscle-computer interaction methodology that directly senses and decodes human muscular activity rather than relying on physical device actuation or user actions. Andrews (2008) has determined an effective approach to finger movement classification in typing tasks using myoelectric data which are collected from the forearm. Cesqui et al. (2008) have explored the use of EMG signals for post-stroke and robot-mediated therapy. In this work, a pilot study has been reported under young and healthy subjects where experiments are conducted to determine whether it is possible to build a static map to cluster EMG activation patterns for horizontal reaching movements. Chen et al. (2008) have implemented an EMG feedback control method with functional electrical stimulation cycling system (FESCS) for stroke patients. The stroke patients often suffer from low limbs paralysis. By designing the feedback control protocol of FESCS, the physiological signal is recorded with help of FPGA biomedical module, DAC and electrical stimulation circuit. Leeet al. (2009) have described a development procedure of bio-mimetic robot hand and its control scheme where each robot hand has four under-actuated fingers, which are driven by two linear actuators coupled together. Dalley et al. (2009) have given emphasis of an anthropomorphic hand prosthesis that is intended for use with a multiple-channel myoelectric interface. The hand contains 16 joints, which are differentially driven by a set of five independent actuators. Hu et al. (2009) have presented a comparison between electromyography-driven robot and passive motion device on wrist rehabilitation for chronic stroke patients. By comparative study, it was found that the EMG-driven interactive training had a better long-term effect than the continuous passive movement (CPM) treatment.

Blouin et al. (2010) have focused on control of arm movement during body motion as revealed by EMG whereas Luo &Chang (2010) have explored a feasibility study on EMGsignal integrated with multi-finger robot hand control for massage therapy applications. The forearm EMG of a person massaged by the human hands is recorded and analyzedstatistically. Khokhar et al. (2010) have showed the potential of EMG applications whereSVMclassificationtechnique is suitable for real-time classification of sEMG signals. This technique is effectivelyimplemented for controlling an exoskeleton device.Huang et al. (2010) have designed a robust EMG sensing interfacefor pattern classification. The aim of this study was to design sensor fault detection (SFD) module through the sensor interface to provide reliable EMG pattern classification.This module monitors the recorded signals from individual EMG electrodes and performs a self-recovery strategy to recover the classification performance when one or more sensors are disturbed.Naik et al. (2010) has studied the pattern classification of myo-electric signal during different maximum voluntary contractions using BSS techniques for a blind person whereas Artemiadis & Kyriakopoulos (2010 & 2011) have presented a switching regime model for the EMG-based control of a robot arm where decode the EMG activity of 11 muscles has a continuous representationof arm motion in the 3-Dspace.The switching regime model isused to overcome the main difficulties of the EMG-based controlsystems, i.e. the nonlinearity of the relationship between the EMGrecordings and the arm motion, as well as the non-stationary ofEMG signals with respect to time.Vogel et al. (2011) have demonstrated the robotic arm/hand system that is controlled in real time in 6 dimension Cartesian space through measured human muscular activity via EMG. DLR Light-weight Robot III is used during demonstration of impedance control. Li et al. (2011) have presented a robot control system using four different gestures from an arm. These are achieved by EMG signal using phase synchrony features.The phase synchrony analysis using the recent multivariateextensions of empirical mode decomposition(MEMD) is carried out. Joshi et al. (2011) have focused on brain-muscle-computer interface using a single sEMG signal. Initial results show that the human neuromuscular system can simultaneously manipulatepartial power in two separate frequency bands of a sEMG powerspectrum at a single muscle site.Matsubara et al. (2011) have proposed an interface to intuitively control robotic devices using myoelectric signals. Through learning procedure, a set of myoelectric signals is captured from multiple subjects in the system and it can be used as an adaptation procedure to a new user after only a few interactions.

Recently, Ahmad et al. (2012) have presented a review report on different techniques of EMG data recording where condition of an ideal pre-amplifier, signal conditioning and its amplification are discussed. Sun et al. (2012) have conducted an isokinetic exercise to realize the characteristics of femoral muscles in human knee movement through EMG where a mechanical model of muscle for human knee movement is established.Qi et al. (2012) have developed algorithms for muscle-fatigue detection and muscle-recruitment patterns in routine wheelchair propulsion scenarios, e.g., daily practice where for analysis purpose two speeds of muscular behavior are chosen. Gandole (2012) has developed an artificial intelligent model using focused time lagged recurrent neural network (FTLRNN) method with a single hidden layer. FTLRNN method reduces noise intelligently from the EMG signal. Chan et al. (2012) have developed an assessment platform for upper limb myoelectric prosthetic devices using EMG. The assessment platform consists of an acquisition module, a signal capture module, a programmable signal generation moduleand an activation & measurement module. The platform is designed to create a sequence of activation signals from EMG data captured from a patient.

An EAP actuator based design for IPMC fingers have been discussed by Biddiss&Chau (2006). This shows the potential of electroactive polymeric sensors within an operating range of voltage (3V) whereas Kottke et al. (2007) have reported on how to stimulate and activate a non-biological muscle such as an IPMC. Lee et al. (2006, 2007) have also demonstrated the potential of an IPMC actuating system with a bio-mimetic function using EMG signals. A mean absolute method is used for achieving the filtered EMG signal. Aravinthan et al. (2010) have designed a multiple axis prosthetic hand using IPMC. EMG signal through programmable interface controller (PIC) is sent to the IPMC prosthetic material to perform the required actions. By doing experiments, the potential of prosthetic hand using IPMC is shown.After that, we have also demonstrated actuation of IPMC through EMG via forearm muscles where potential of IPMC based micro robotic arm has been shown for lifting the object (Jain et al., 2010a; 2010b; 2011; 2012). For further application of EMG driven system, we are discussing detailed analysis of EMG signal control point view of IPMC based artificial muscle finger for micro gripper in this chapter.

3. Design and control of IPMC based artificial finger driven by EMG signal

3.1. Basic design of IPMC artificial muscle finger based micro gripper using EMG

For designing an IPMC artificial muscle finger based micro gripper using EMG, an IPMC strip (Size 40 mm 10 mm 0.2 mm) that imitates human finger movement, is assumed to be artificial muscle finger. This is fixed with holder and another plastic based finger of similar size is made for supporting the micro object as shown in Fig. 1. When human index finger moves up and down, it creates potential difference by its movements. This potential difference is transferred through EMG electrodes into the artificial muscle finger so that this finger is able to move accordingly and holdthe object. The main function of EMG electrode is to detect the voltage from human muscles since human muscles generate few milli-voltages when they are contracting or expanding during movement. For transferring this voltage signal to actuate the artificial muscle finger, it needs the amplification setup which is discussed in section 4. Therefore, IPMC based artificial muscle finger is activated using EMG and it allows holding an object for micro assembly operation.

Figure 1.

Schematic diagram of IPMC artificial muscle finger based micro gripper driven by EMG

Figure 2.

Schematic diagram of the actuation mechanism of IPMC (Chen et al., 2011)

During development of an EMG driven IPMCbased artificial muscle finger, a typical IPMC strip (Procured custom made from Environmental Robots Inc., USA) is used which has a thin (approximately 200 μm) perfluorinated ion exchange base polymer membrane (Nafion-117) with metal electrodes of platinum (5–10 μm) fused on either side. As a part of the manufacturing process, this base polymer is further chemically coated with metal ions that comprise the metallic composites. It responds in wet/dry condition. An IPMC is usually kept in a hydrated state to ensure proper dynamic operation. When the material is hydrated, the cations will diffuse toward an electrode on the material surface under an applied electric field. Inside the polymer structure, anions are interconnected as clusters providing channels for the cations to flow towards the electrode (Chen et al., 2011). This motion of ions causes the structure to bend toward the anode as shown in Fig. 2. An applied electric field affects the cation distribution within the membrane, forcing the cations to migrate towards the cathode. This change in the cation distribution produces two thin layers, one near the anode and another near the cathode boundaries. The potential is generated by changing the potential electric field on cluster of ionic strips that provides the actuation of the strip.

3.2. Basic tendon of index finger for identification of EMG signal

To examine the bio-mimetic behavior of IPMC based artificial muscle finger, it is important to study the physiological structure of human finger. An internal structure of human index finger is shown in Fig. 3. The index finger is actuated by three intrinsic muscles and four extrinsic muscles. The intrinsic muscles consist of two interosseous (IO 1 and IO 2) muscles & one lumbrical (LU) muscle and four extrinsic muscles connected through long tendons i.e. extensor digitorum communis (EDC), extensor indicis proprius (EIP), flexor digitorum superficialis (FDS) and flexor digitorum profundus (FDP) (Bundhoo & Park, 2005). For heavy lifting & holding purpose, EDC and EIP are responsible in tendon network. Consequently, EMG electrodes are placed at these two positions on the human finger so that we can achieve direct actuation of IPMC based artificial muscle fingerthrough said muscles.

Figure 3.

Basic tendons of the index finger (Bundhoo& Park, 2005)

3.3. Model for controlling the EMG signal for IPMC based artificial muscle finger

For acquiring the data from said muscles, EMG electrodes are placed for measuring the electric potential produced by voluntary contraction of muscle fiber on the human finger. The frequency range of the EMG signal is within 4 to 900 Hz. The dominant energy is concentrated in the range of 95 Hz and amplitude of voltage range is ±1.2 mV according to muscle contraction and the voltage function Vin(t) in term of signal sample time (t) is given below,

Vin(t)=Vinosin(2πft)E1

Where, Vinois amplitude of EMG voltage (0.0012V); f is frequency of EMG signal (95 Hz).

In Laplace domain, EMG input signal is written as

Vin(s)=2.96s2+3.51e5E2

Using these parameters, the circuit for filtered EMG signal is designed using MATLAB SIMULINK software as shown in Fig. 4.

Figure 4.

Block diagram of EMG signal behaviour from human index finger

In block diagram, the active EMG signal is taken from index finger muscle and uniform noise is considered. The electric potential is first amplified with gain 32 dB and then band pass filter (BPF) is used within specified frequency range (4 to 900 Hz). Using two band stop filters (BSF and BSF1), noise signal (60 Hz) that arises due to AC coupled power is eliminated. The signal is then passed through an amplifier with gain 60 dB. Subsequently, three integrators (Integrator, Integrator1 and Integrator2) are used for achieving better damped signal. The output of EMG signal with sampling time of 10-4 seconds after filtering is shown in Fig. 5.

Figure 5.

Acquired data from finger muscles via EMG signal

Thus, the total output duration of EMG signal for sampled data is 0.1 second. The general solution of acquired EMG voltage (VEMG) through curve fitting method is obtained as given below,

VEMG(t)=V0sin(2πf0t+δ0),
0t0.1E3

where, V0 is the amplitude of EMG voltage of sine function (0.0012 V); f0 is average frequency of sine function (4.7 Hz); 0 is initial phase difference of sine function (1.66 rad); t is signal sample time in second.

After adjustment of root mean square (RMS) value of this sine function, the EMG voltage VEMGrms(s) after filtering is written as

VEMGrms(s)=1.16e-5(s2+1.17e-4)(s20.014s+1.07e-4)(s20.014s+1.07e-4)E4

For controlling purpose, single input single output (SISO) control system tool in MATLAB is used and the output VEMG(s) data is obtained. The initial overall transfer function of EMG voltage VEMGinitial (s) is obtained through output signal from (4) to input signal from (2) and is given as

VEMGinitial(s)=3.92e-6(s2+1.17e-4)(s2+3.5e5)(s20.0146s+1.07e-4)(s2+0.0146s+1.07e-4)E5

After that, Nyquist criterion is applied to check the stability of EMG signal. The root-locus and bode scheme are plotted as shown in Fig. 6.

Figure 6.

Root-locus and bode plot behaviour of EMG signal via finger in initial condition (Jain et al., 2011)

It is found that the zero-poles have real value on both sides of the real axis which does not meet theNyquist stability criteria. Also from Fig. 6,gain cross-over frequency (GCF=2.53 rad/s) is greater than phase cross-over frequency (PCF=1.18 rad/s), indicating that this voltageobtained from EMG signal is unstable.

For achieving the stable EMG signal, different configurations of PID are analysed through SISO control tool of MATLAB software. By applying PD control with proportional control gain factor (Kp=1) and derivative control gain factor (Kd=1), EMG voltage VEMGfinal 1(s) is obtained as given below,

VEMGfinal1(s)=(s2+1.17e-4)(s2+3.516e5)(s+2.546e5)(s2- 9.91e-9s+1.17e-4) (s2+1.38s + 1.38)E6

The root locus and bode plot are shown in Fig. 7. This indicates that the obtained data from EMG signal is unstable but through this control system the data is converging towards stability from initial condition.

Figure 7.

Root-locus and bode plot behaviour of human finger through PD controller

In case of PI controller, the unit proportional control gain factor (Kp=1) and integrator control gain factor (Ki=1) parameters are used. After applying this control system, EMG voltage VEMGfinal 2(s) is obtained as given below,

VEMGfinal2(s)=3.92e-6 s (s2+1.17e-4) (s2+3.51e5)(s+0.724)(s2+ 9.91e-9s + 1.17e-4)(s2-0.724s+1.90) E7

The bode plot and root locus for this system are plotted as shown in Fig. 8.From this figure, it shows that the data from EMG signalis again unstable but the response has a better prospect of converging towards stability than previous configuration.

Figure 8.

Root-locus and bode plot behaviour of EMG signal via finger through PI controller

After that, PID controller is used where proportional control gain factor (Kp= 0.5), integrator control gain factor (Ki= 1) and derivative control gain factor (Kd=1) are given in compensator toattain the stability of EMG voltage. For applying the PID control system, the SIMULINK block diagram is modified as shown in Fig. 9.

Figure 9.

Block diagram of EMG signal via human finger after applying PID control system

Thereafter, the model is again simulated in MATLAB software, upon which, the data from finger muscles shows the all zero-poles in left hand side of real axis which satisfies the Nyquist criteria shown in Fig. 10. The final EMG voltage VEMGfinal (s) is acquired as given below,

VEMGfinal(s)=(s2+1.17e-4)(s2+3.51e5)(s+2.54e5)(s+1.52)(s2+1.83e-8+1.17e-4)(s20.14s+0.90)E8

Also, GCF (1rad/s) is less than PCF (1.34 rad/s). Hence, a stable EMG voltage data is achieved. This filtered EMG signal is stable enough to provide necessary voltage signal across the IPMC for proper functioning during operation. The major advantage of this control system is that it is stable with least amount of noise. This signal is sent to artificial muscle finger for holding the object.

Figure 10.

Root-locus and bode plot behaviour of EMG signal via finger through PID controller (Jain et al., 2011)

4. Experimental testing setup

In order to examine the bio-mimetic behaviour of IPMC based artificial muscle finger through EMG signal, EMG electrodes (Ag/Agcl based) are positioned at EDC and EIP muscles on the index finger. The input EMG signal from the muscle movement varies in the range of 1.2mV (which is observed through oscilloscope). But activation of IPMC based artificial muscle finger needs the voltage of 3V and current rating of 50-200 mA which is only possible by amplification of voltage and current. For desired output voltage to the artificial muscle finger, EMG signal is transferred through analog-digital convertor (ADC) card and PXI system (PXI-1031 along with NI-6289) in real time environment. EMG signal through electrodes are sent to an input channel at specified ports of the ADC card. Then this signal is amplified through amplification factor of 2550 using express VI of Labview 8.5 and sent to DAC output. But DAC output signal cannot provide enough current (50-200 mA) to drive an IPMC based artificial muscle finger. For achieving this current rating, the current amplification is done using customized IPMC control circuit by combining operational amplifier (Model: LM-324), transistor (Model: TIP 122) and resistances (1k and 10). Noise interference is eliminated by enabling low-pass filtering with PID control system to achieve the stability during operation of the artificial muscle finger as shown in Fig. 11. The IPMC size 40mm 10 mm 0.2 mm is used for testing purpose.

Now, in order to prevent the abrupt physio-chemical change of the IPMC nature and subsequent shortening of the actuation operating time of the IPMC material due to irreversible electrolysis (caused when the voltage applied across the two faces of an IPMC exceeds a maximum limit), two warning flags are used. One warning flag is placed at input EMG signal and another warning flag is placed at output voltage of DAC card where IPMC control circuit is connected. This limited voltage imposed on the IPMC based artificial muscle finger aborts the execution of the program when the warning flag has a high output. The flow chart for actuation of IPMC based artificial muscle finger is shown in Fig. 12.During operation, artificial muscle finger bends in a similar manner as that of the index finger. For generating force, this finger is held in cantilever configuration on the fabricated work bench. Aload cell is used to collect the data at different angles of the index finger. The current and voltage analysis of the human muscles are also done through oscilloscope. Thus an IPMC artificial muscle finger based micro gripper driven by EMG is developed and the holding behaviour is demonstrated.

Figure 11.

Basic testing layout for actuation of IPMC based artificial muscle finger

5. Results and discussion

During experimentation, the input parameter from muscle ranging from ±1.2 mV is taken through referenced single-ended (RSE) signal along with continuous sampled pulses. Thesepulses are amplified with the help of a PXI system (amplification factor 2550). The desiredoutput voltage range is generated through a DAC output port. The output signal is connected to IPMC based artificial muscle finger. Due to amplified output voltage from the DAC, an IPMC strip bends in one direction for holding the object. By changing the movement of human finger in an opposite direction, reverse behaviour of IPMC is obtained. The characteristics of IPMC based artificial muscle finger are traced on a graph paper and plotted as shown in Fig. 13. It shows that IPMC based artificial muscle finger gives similar bending behaviour as a human finger (Fig. 14). It is also observed that the deflection of IPMC based artificial muscle finger changes with voltage upto 12 mm in one direction. When this finger moves reverse direction, the characteristic of artificial finger does not attempt the same behaviour. It shows the error between two paths is 0.5 mm. The deflection characteristic of IPMC based artificial muscle finger () in term of voltage (V) with cubic behaviour for holding is given below,

δ(V)=0.67×V3-1.9×V2+3.7×V -0.17(in mm)E9

Figure 12.

Flow chart of for actuationIPMC based artificial muscle finger using EMG signal

Figure 13.

Deflection behaviour of IPMC based artificial muscle finger with different voltages

Figure 14.

Control of IPMC based article muscle finger through EMG signal (Jain et al., 2011)

Figure 15.

Load test setup for IPMC based artificial muscle finger using load cell

For generating force by the artificial muscle finger, a testing setup with load cell is used. The IPMC based artificial muscle finger is placed in cantilever mode, and a load cell is placed under the tip of the IPMC which produces the reactive force. When human finger moves downward, the generated force by IPMC artificial muscle finger increases accordingly in the load cell as shown in Fig. 15.

By controlling the movement of human finger, the generated force varies accordingly. The generated force characteristic with voltage shows a cubic polynomial behaviour as shown in Fig. 16 when IPMC artificial muscle finger touches the load cell. This happens due to compliant behaviour of IPMC. The generated force (F)by IPMC based artificial muscle finger in term of voltage (V) is given below,

F(V)=0.11×V3+0.25×V2+1.6×V-0.11(in mN)E10

Figure 16.

Generated force by IPMC based artificial muscle finger at tip with voltage

The maximum generated force of 10 mN is achieved by IPMC based artificial muscle finger at 450 angle of index finger with cubic polynomial behaviour as shown in Fig. 17. This happens due to human finger behaviour where EIP and EDC muscles are connected with DIP and PIP joints. The generated force (F)by IPMC based artificial muscle finger at tip in term of human finger angle () is also obtained as under

F(θ) = - 0.000083×θ3+ 0.01×θ2- 0.063×θ + 0.045(in mN)E11

For observing the real time IPMC based artificial muscle finger behaviour with moving human finger angle, experiments are conducted and data are plotted as shown in Fig. 18 and it shows almost proportional behaviour with quadratic relationship. This occurs due to human finger behaviour where EIP and EDC are connected with IO and LU muscles (Fig. 3). The relationship between IPMC based artificial muscle finger displacement () and human finger angle () is given below,

δ(θ) = - 0.0016×θ2+ 0.34×θ0.15(inmm)E12

Figure 17.

Generated force by IPMC based artificial muscle finger with human finger angle

Figure 18.

Relationship between IPMC based artificial muscle finger and human finger angle

For actuation of IPMC based artificial muscle finger, the analysis of activated muscles is carried out as given in Table 1. We have analysed different conditions during contraction of different muscles of index finger like intrinsic and extrinsic which are responsible for actuation of IPMC based artificial muscle finger so that they can be used to hold an object.

CasesIntrinsic musclesExtrinsic musclesStatePolarity
IO 1IO 2LUEDCEIPFDSFDPSide ASide B
1OFFOFFOFFOFFOFFOFFOFFNoneNoneNone
2ONOFFONOFFOFFONONAdduction+ive-ive
3OFFONONONONOFFOFFAbduction-ive+ive
4ONONONONONONONNoneNoneNone

Table 1.

Analysis of different condition of muscles

The two surfaces of IPMC are denoted as side A and side B. In case of intrinsic muscles, the adduction is possible. When IO 1 or IO 2 are in either “on” or “off” condition along with LU muscle in “on” condition then it shows the abduction state. In case of extrinsic muscles, EDC and EIP muscles both are in “off” condition to achieve the adduction state when FDS and FDP muscles both are in “on” condition and for attaining the abduction state EDC and EIP both are in “on” condition when FDS and FDP both are in “off” condition. In rest cases, no power is achieved. Therefore, IPMC based artificial muscle finger is activated in above mentioned conditions from muscles.

The voltage characteristic behavior is taken from EMG and fed to IPMC based artificial muscle finger for actuation in real time environment as shown in Fig. 19. It is found that the trend of IPMC actuation voltage is similar to EMG voltage with amplification factor.

Figure 19.

Different voltage responses in real time environment

The EMG voltage VEMG (t) and IPMC actuation voltage VIPMC actuation (t) equations are respectively given below,

VEMG(t)=V0Esin(2πf0Et+δ0E),
0t0.1E13
and VIPMCactuation(t)=V0Isin(2πf0It+δ0I),
0t0.1E14

Where, V0E is average value of EMG voltage (V); f0E is EMG frequency range (Hz) ; t is signal sample time (s) ; 0E is phase difference when signal is taken through EMG (rad); V0I is average value of IPMC actuation voltage (V); f0I is IPMC actuation frequency range (Hz); 0I is phase difference when signal is given to IPMC (rad). For finding the frequency range of each signal, the experimental data are taken and solved through MATLAB curve fitting tool. The numerical values are V0E= 0.0014510.0002707 V, f0E= 4.70.006201 Hz, 0E= -1.7360.036 rad, V0I= 2.4930.208, f0I= 48.50.65 and 0I= -10.57320.6556 rad. From these data, it is found that EMG frequency range (f01) is similar to simulated data and IPMC actuation frequency range is 48.50.65 Hz which is in between human muscle frequency range (48-52Hz).

Figure 20.

IPMC based artificial muscle finger based micro gripper driven by EMG

After these analyses, an IPMC artificial muscle finger based micro gripper is developed which is driven by EMG as shown in Fig. 20 where one IPMC based artificial muscle finger and other plastic based finger are fixed with double sided tape within one holder. The IPMC based artificial muscle finger is connected through copper tape and wire with EMG sensor so that an IPMC based artificial muscle finger is activated by EMG signal via human finger. The packing tape is also placed on the tip of IPMC based artificial muscle finger so that this finger perfectly holds the object like micro pin for assembly. After these observations, it is understood that IPMC finger behaves as an artificial muscle and this characteristic is implemented in the development of IPMC artificial muscle finger based micro gripper for holding the object through EMG. The major advantages of EMG-driven IPMC based artificial muscle finger are low voltage in man-machine interface, large bending amplitude and simple control that are applied in development of micro/bio robot.

6. Conclusion

In order to develop the micro/bio-mimetic robot for micro assembly, the potential of an EMG driven artificial finger is discussed in this chapter. An artificial finger for micro assembly is designed using IPMC where IPMC is used as an active artificial finger for holding the object. An IPMC has several advantages such as actuating through a small voltage (3 V), light in weight, flexible in nature and does not involve sophisticated controllers for operation. For activating the IPMC based artificial finger, voltage is taken from human index finger through EMG sensor instead of battery source as this is used as a man-machine interface device. Principally, EMG sensor acquires the signal from body during expansion or contraction of muscles. These movements are transferred into an IPMC based artificial muscle finger. For achieving the stable data from EMG, different configurations of control methods are analysed. A PID control system is implemented for attaining the noiseless and stable signal from the user’s myoelectric signal. While acquiring the data, a differential amplification technique is applied where data is filtered through a band pass filter and noise is eliminated through three band stop filters. For sending this signal to the IPMC, an algorithm has been developed in Labview software which gives emphasis on following points:

  • Acquire voltage data (1.2mV) from human index finger using EMG sensor through data acquisition system

  • Amplify the continuous EMG signal through DAQ assistant enabling the filter and domain frequency range options

  • Activate IPMC finger through amplified data (3V) using interface device for functioning as artificial muscle finger

Experimentally, it is demonstrated that IPMC based artificial muscle finger is capable of adopting this voltage from EMG signal and mimics as a human finger. From application point of view, an IPMC artificial finger based micro gripper is developed and its capability is also verified. Through this demonstration, it is proved that IPMC can be activated through EMG signal and is applicable as flexible and compliant finger for holding the object in the fields of micro manipulation. IPMC based artificial muscle could also be a replacement of an electro-mechanical system like electric motors in the application field of rehabilitation technology.

7. Future direction

In future, we will focus on developing well equipped EMG driven micro robotic system where IPMC based micro robotic arm along with multiple IPMC artificial fingers will be used. IPMC based micro robotic arm will be operated through human fore arm movement for lifting and manipulation. Multiple IPMC fingers will be used for robust application like grasping, holding and mimicking of a human hand. Therefore, this new generation of robotic system can be really operated in real world through humans using EMG signals.

Acknowledgement

The authors are grateful to the Director, Central Mechanical Engineering Research Institute (CMERI), Durgapur, West Bengal, India for providing the permission to publish this book chapter.This work is financially supported by the Council of Scientific and Industrial Research (CSIR), New Delhi, India under eleventh five year plan on “Modular Re-configurable Micro Manufacturing System (NWP-30)”.

How to cite and reference

Link to this chapter Copy to clipboard

Cite this chapter Copy to clipboard

R.K. Jain, S. Datta and S. Majumder (October 17th 2012). Design and Control of an EMG Driven IPMC Based Artificial Muscle Finger, Computational Intelligence in Electromyography Analysis, Ganesh R. Naik, IntechOpen, DOI: 10.5772/48814. Available from:

Embed this chapter on your site Copy to clipboard

<iframe src="http://www.intechopen.com/embed/computational-intelligence-in-electromyography-analysis-a-perspective-on-current-applications-and-future-challenges/design-and-control-of-an-emg-driven-ipmc-based-artificial-muscle-finger" />

Embed this code snippet in the HTML of your website to show this chapter

chapter statistics

2786total chapter downloads

3Crossref citations

More statistics for editors and authors

Login to your personal dashboard for more detailed statistics on your publications.

Access personal reporting

Related Content

This Book

Next chapter

Application of Surface Electromyography in the Dynamics of Human Movement

By César Ferreira Amorim and Runer Augusto Marson

Related Book

First chapter

Micro Macro Neural Network to Recognize Slow Movement: EMG based Accurate and Quick Rollover Recognition

By Takeshi Ando, Jun Okamoto and Masakatsu G. Fujie

We are IntechOpen, the world's leading publisher of Open Access books. Built by scientists, for scientists. Our readership spans scientists, professors, researchers, librarians, and students, as well as business professionals. We share our knowledge and peer-reveiwed research papers with libraries, scientific and engineering societies, and also work with corporate R&D departments and government entities.