1. Introduction
This chapter is focused on emerging robotic techniques for improving conventional colonoscopy.
Video-colonoscopy is considered the gold-standard for the diagnosis of colonic diseases, and it is included as first line choice in colon-rectum cancer screening program in high-risk populations. However, this diagnostic technique shows some technical limitations, such as invasiveness and patient discomfort, which limit the adherence to the procedure.
To facilitate the conventional colonoscopy procedure, robotic colonoscopy solutions have been proposed. State of the art of robotic colonoscopy has been thus summarized. In details, Endotics System and Invendoscope are presented.
The
The
Limitations and advantages of the two devices are reported compared to conventional colonoscopy. In the last part of the chapter are presented data of pilot studies both in healthy volunteers and patients in terms of technical aspects (cecal intubation, pain score, sedation) and clinical results (lesions detection).
2. Why and how robotic systems in colonoscopy?
The first colonoscopy procedures go back to the 1960’s, when in Japan a device for the examination of the left colon was developed (Niwa et Al, 1969). In the 1970’s further progresses were made, and colonoscopy devices able to explore the whole colon were available (Classen et Al., 2010).
Since then, research efforts were focused towards improvements of the vision system, of the degrees of flexibility and of the localization systems. Nevertheless, the main characteristics of the devices, based on a CCD camera or a fiber optic camera on a flexible tube passed through the anus, remained unchanged till the 1990’s. In those years, robotic technologies grew up enough to allow an increasing number of robots to be realized and used in various fields of medicine. The main reasons for including colonoscopy were to try to overcome the existing limitations of the standard devices, quite rigid, requiring high experience of the doctor to perform difficult maneuvers to proceed along the tortuous colon walls, and constructed of materials that could be damaged by heat, pressure, and moisture used during the decontamination processes. (Sturges & Laowattana, 1993) The stated above limitations made, and still make, standard colonoscopy a quite invasive technique, with risks related to perforation, sedation and cross-infections, far for being accepted by massive percentages of patients as needed in colorectal cancer screening programs. Screening as a matter of fact can find non-cancerous colorectal polyps and remove them before they become cancerous. If colorectal cancer does occur, early detection and treatment dramatically increase chances of survival. The relative 5-year survival rate for colorectal cancer when diagnosed at an early stage before it has spread is about 90%, but since screening rates are low, less than 40% of colorectal cancers are found early. Once the cancer has spread to nearby organs or lymph nodes, the 5-year relative survival rate goes down, and if cancer has spread to distant organs (like the liver or lung) the rate is about 11%, and as many as 60% of deaths from colorectal cancer could be prevented if everyone age 50 and older were screened regularly.
Moreover, painless colonoscopy, besides being a remarkable achievement for the patient, and avoiding any risk related to sedation, has major fallout in terms of prevention. As matter of fact, colonoscopy could be largely used for screening purposes of healthy and asymptomatic patients, less willing to feel pain because of an invasive procedure. Nowadays, colonoscopy is used as screening test just in first-level demonstrative studies and pilot projects. One of the main limitations to use this survey as primary screening, besides the feasibility related to allocation of facilities and complications rates, is the acceptance of the procedure. Participation in the first-level FOB (faecal occult blood) screening test is always above 50% (Faivre et Al., 2004), while the few available data in literature about compliance to colonoscopy as primary screening is in a range from 15% to 90% (Swaroop et Al., 2002). Compliance for second level screening programs, which in principle should be very high since this second examination takes place after positive results of the first one, is in a range from 30 to 60%, as reported in a study of from AIGO - Oncology Group Study.
For the above listed reasons, it appears clear how big can be the impact, in terms of survival rate, of new devices able to perform painless and safe diagnostic colonoscopic procedures. Thus robotics, as a science that tries to find and develop methodologies that enable machines to perform specific tasks, could make the difference in developing endoscopes that pulled themselves, with no risk for stretching the colonic wall outward and causing painful cramps. The main challenge to building such devices involved clutching onto the slippery walls of the colon in a way that did not damage them. The new endoscopes had to be disposable, highly flexible, with a particularly suited internal locomotion, and with a direct vision of the colon tissue, to solve acceptance problems and maintain quality of gold standard.
2.1. Robotic colonoscopy: state of the art
As for robotic colonoscopy, first studies go back to 1995, with the locomotion system "inch-worm". Subsequently, a lot of research work was carried out aiming at devising several robotic colonoscopes based on different types of locomotion such as “snake”, “earthworm”, “continuum” and “caterpillar”, or other different concepts.
The Inchworm robots were inspired to the caterpillar Geometridae, whose mode of locomotion is to firmly attach the rear portion of its body to a surface via its foot pads, extending the remainder of its body forward, attaching it to the surface and bridging the rear part of its body to meet the forward part. On this principle is based the Endotics system. (Slatkin et Al., 1995)
The
The Snake robots took inspiration from the sinuous movement of the snake, based on the temporal shifting of positions and angles of subsequent parts of its body. Movements starts from the head that, bending and moving forward, is able to avoid obstacles. In robotics, this is translated in devices with a finite number of independent segments where, during the progress, the position and the angle of the distal part is encoded by an algorithm and then associated with the next segment. All segments are associated with the same geometrical parameters of the previous segment. NES, the NeoGuide System is based on this principle of functioning.
The
Peristalsis motion like an earthworm has attracted attention because the movement is useful to progress in small spaces (Saito et Al., 2009). The Earthworm robots were based on the earthworm’s locomotion, thus moving not only changing length, but also changing thickness (Zuo et Al., 2005). The Continuum robots worked according to the moving principle of the elephant trunk or the tongue, i.e. structures without rigid constraints able to perform complex movements (Hu et Al., 2009). Finally, the Caterpillar robots were characterized by wheels and caterpillar tracks and their locomotion was similar to a tank. Other technique of locomotion not based on bio-mimicking, but having relations with robotics in terms of sensors or automated movements have been developed (Swain, 2009).
The
A rectal introducer, consisting of a hollow tube with a stationary balloon attached to its outer surface, is inserted through the anus and, when the stationary balloon is inflated, seals the orifice. An electro-optical capsule is embedded in the front of a lightweight balloon vehicle, while low pressure colon insufflation with CO2, between stationary and vehicle balloon, propels the vehicle balloon, causing it to glide along the ‘slippery' colon walls. Computer controlled pressure management, coupled with sensors in the workstation, adjust balloon size and shape to changing bowel anatomy, thus allowing the pressure-propelled balloon to find its path. The Aer-O-Scope visual system provides simultaneous 360 degree viewing of the colon mucosal surface. (Pfeffer et Al., 2006)
The
The
The above stated overview of the state of the art of robotic devices developed for colonoscopy procedures shows that the main priority was to realize a system with an internal locomotion action, able to advance in a hostile environment. The movement of the device is always under computer control by means of mechanical, electrical, or computer-algorithm based sensors. Moreover, the robotic colonoscopes generally include the following sub-systems:
A probe, with at least a disposable part (the one in contact with the colonic mucosa);
A vision system located at the tip of the device;
A PC-based workstation or a hand-held unit which controls the propulsion of the probe in the colon.
Some of the characteristics of the above listed sub-systems, as well as other aspects related to personnel and sedation during the procedures have of course an impact on the cost saving issue. Moreover, also the related timings have to be considered, including e.g. the preparation which non-disposable devices have to undergo to prevent from cross-infections, the duration of the procedures themselves, the gaining of time in the turn-over of patients and the recovery time for the patients to go back to work.
Even if several robotic colonoscopes have been tested in vitro and seemed to be ready to be used in pilot studies on human beings, very few completed the engineering phase and went through the certification steps, and only one became available off-the-shelf as a real product. In particular, in the following further details will be presented on the Endotics System, whose core component is a disposable probe with inchworm locomotion, currently used in clinical practice in a few hospitals, and on the Invendoscope, based on inverted sleeve technology, not yet commercially available, but with most recent news than others.
|
|
|
|
Neoguide | 2006 | Transformed in laparoscopic device | No |
Aer-o-Scope | 2007 | DDW 2007 | No |
Colonosight | 2007 | Closed business 2008 | No |
Endotics | 2011 | Fismad 2011 | Yes |
Invendoscope | 2011 | - | No |
|
DEVELOPER/ COMPANY NAME |
|
|
|
Neoguide | NeoGuide Systems Inc | USA | Abandoned * | n.a |
Aer-o-Scope | GI View Ltd | Israel | Almost abandoned * | n.a. |
Colonosight | Sightline Technologies Ltd | Israel | Almost abandoned * | n.a. |
Endotics | Era Endoscopy Srl | Italy | Off the shelf product | Europe |
Invendoscope | invendo medical GmbH | Germany | Filed 510(K) notice submission with the FDA (08/03/2010) | USA |
2.2. The endotics system
The excessive stretching of bowel and mesenteries and the air insufflation are the main reasons of the pain felt by the patient during this uncomfortable procedure. In order to avoid perforation risks, addressed both to pushing action exerted by the endoscopist during the intubation phase and to the rigidity in the pushing direction of the traditional colonoscope, it is essential to realize a system with an internal locomotion action, able to advance in a hostile environment. The ideal screening investigation should be as non invasive as possible and safe while maintaining a high diagnostic accuracy. Thus, a device extremely flexible and soft, which gently deforms just locally the colon tissue, represents the optimal solution.
The innovative systems require a lower amount of insufflations and do not stress on mesenteries resulting in a real painless colonoscopy. Moreover, infective risks, due to sterilization procedure’s limits, are definitively eliminated by disposable endoscopes.
From a technical point of view, the simplest inchworm device consists of two clampers at the ends, used to adhere securely onto the “terrain”, and one extensor as its midsection that brings about a positive displacement. The device of the robot was focused towards a disposable device, totally pneumatically driven, and very soft and flexible, able to adapt its shape to the configuration of the bowel. The probe is composed of two main parts: an active one, including the head, the steering and the flexible extensible body, and the passive components of the devise including the tail and the tank containing eventual body fluids, and the connector used to fix the disposable probe to the workstation. The overall dimensions of the active part are: a diameter of 17 mm and a variable length from 24 to 40 cm, considering the inchworm movements. The head hosts both a vision system, including a CMOS camera and LED light sources, and channels for water jet and air in order to provide rinsing and suction/insufflations, respectively. As the Endotics system requires air insufflations only in the immediate proximity of the head lens, an accurate automatic insufflations-suction balance prevents painful bowel stretching.
The passive component, a very thin and extremely flexible plastic tail, has a diameter of 7,5 mm and a length of 180 cm. The workstation allows the endoscopist to easily and fully control the disposable probe by means of a hand-held console and to visualize on a screen real time images. Thanks to the electro-pneumatic steerable tip, the operator can steer the head of the robotic colonoscope of 180º in every direction, elongate the soft body of the probe in order to move it forward and backward following the shape of the intestine, and control rising, insufflations and suction.
The locomotion phase begins with the automatic adherence of the proximal clamper of the probe to the colon walls. The next phases can be described as follows:
the midsection is elongated under control of the operator;
the distal clamper adheres to the colon walls (automatic);
the proximal clamper is released (automatic);
the midsection is contracted (automatic);
the proximal clamper adheres to the colon walls (automatic);
the distal clamp is released (automatic).
The purposely developed patented clamping system allows to hold the colonic tissue by means of a combined vacuum-mechanical action. The clamping mechanism does not create neither lesions in the bowel wall, nor mucosal lacerations.
Diagnostic accuracy and patient acceptance of robotic colonoscopy have been evaluated in a first pilot multicentre study, in 40 consecutive volunteers (27 men and 13 women) who underwent standard colonoscopy also. This pilot study showed that the Endotics System has a diagnostic accuracy equivalent to the one achievable through the standard colonoscope. Moreover, the Endotics System was able to visualize two small polyps (sized below 2 mm), in two different cases, not seen using standard colonoscopy. This probably due to the fact that during standard colonoscopy a bigger amount of air was insufflated causing a flattening of the small polyps. Considering the patient acceptance issue, the Endotics colonoscopy was unanimously rated strongly better than conventional colonoscopy: in a scale from 0 to 10 for pain and discomfort the procedure performed by means of the Endotics System scored on average 0.9 and 1.1 (mode 0 for both), compared to 6.9 and 6.8 (mode 9 and 8) of the standard colonoscopy, respectively. (Cosentino et Al., 2009)
In a second pilot study which involved 71 patients (40 men and 31 women), diagnostic accuracy and enhanced patient acceptance of Endotics system compared with standard colonoscopy, was confirmed showing a sensitivity equal to 93,3% (95% C.I.:68.0-99.0), a specificity equal to 100% (95% C.I.:88.0-99.9), a predictive positive and negative values, PPV and NPV respectively, equal to 100% and 97.7%. No patients requested sedation during the Endotics procedure, while 14 subjects (19.7%) requested the administration of midazolam and meperidine during standard colonoscopy. In this study has been used a slightly different Endotics probe version from the one used in the previous pilot study (25 cm length in the contracted form and 43 cm in the elongated form, with respect to 23 and 37 cm, respectively, of the previous version) (Tumino et Al., 2010)
In a third study 12 patients with inflammatory bowel diseases were enrolled in order to compare the diagnostic performance and tolerability of the Endotics System with standard colonoscopy for the staging of ulcerative colitis. Mean pain/discomfort on a 0-10 scale was 2.08 (SD 1.67) for Endotics system and 4.17 (SD 1.74) for standard colonoscopy, with a statistically significant difference (p = 0.066) favoring Endotics system.
In conclusion, the Endotics System is a diagnostic instrument comparable to the gold standard and highly suitable for screening purposes due to the extremely high level of patients acceptance. (Pallotta et Al., 2011)
2.3. The invendoscope
This computer-assisted colonoscope is based on inverted sleeve technology, where the outer side of the inverted sleeve stays in position, and the inner side is pulled forward below the distal tip, moving the colonoscope into the colon by 10 cm each time. Wheels are rolled on the inner side of an inverted sleeve, so that the sleeve is rolled inside out, drawing the colonoscope deeper into the colon. With this mechanism there is no relative movement to the colon wall, and in combination with the small bending diameter minimizes the forces on the colon walls and prevents looping, minimizing pain and discomfort for the patient. The device is equipped with a centralized 3.2 mm working channel with the support of the deflectable electrohydraulic tip; therefore it can be also used for routine therapeutic procedures such as polypectomy. Both the vision capabilities and the working channels are similar to those of conventional colonoscopes. All endoscopic activities are controlled by a hand-held unit.
First pilot study (Roesch et Al, 2007) was focused on capability of the device of reaching cecum measuring time needed and pain/discomfort rate. Were enrolled 24 patients reaching cecum in 79% of cases, with a mean time 26 minutes. Participants rated the examination on an overall score (1.77 points; range, 1-3), using a self assessed pain scale (pain scale range was from 1 = no discomfort to 6 = severe pain).
A second single-arm, pilot study (Invendoscope 1st) on 39 paid healthy volunteers was carried out. Again, cecum reaching rate and time were the focus of study, with some attention towards the patient acceptance. The cecal intubation rate was of 82% (95% C.I., 66-92). Two incomplete colonoscopies had to be stopped at the sigmoid colon because of pain, and in other four volunteers the procedure was terminated at the hepatic flexure. Bloating was reported in four volunteers after that an endoscopy intravenous administration of 20 to 40 mg of hyoscine butylbromide was allowed to facilitate endoscope passage. It should be noted that only limited time was spent on inspection of the mucosa while withdrawing the instrument. The volunteer rating showed a mean score of 1.96 (range 1-6; 1 = no discomfort). Study was divided into two phases. On the basis of experience during phase 1, the instrument was made longer (from 170 to 180 or 200 cm) and a few other modifications (e.g., stiffening below the endoscope tip, improved coating) were also made to achieve better performance in the right colon, however, a comparison between the two instruments was not the main aim of the pilot study. To date, no data concerning diagnostic accuracy and comparison with conventional colonoscopy are available.
Moreover, a third prospective single-arm study (Invendoscope 2nd) on 61 paid healthy volunteers was conducted. There were 34 men and 27 women with a mean age of 57.5 years (range 50 – 70) and a mean body mass index of 26.3 kg/ m 2 (19.5 – 36.8). Main outcome parameters were safety, as measured by the frequency and severity of device related adverse events, and device effectiveness, as shown by cecal intubation rate. Secondary outcome parameters were utility of the device in the documentation and biopsy of pathological findings, and pathological findings. Pain/discomfort rating and introduction/withdrawal timing were also recorded. Comparison with standard colonoscopy was not included in the parameters of the study. Cecal intubation was reached in 60 volunteers (98,4%): introduction mean time was 16.4 min as also withdrawal mean time. Abdominal compression and/or position change were used in approximately two-thirds (66%) of the patients, to help in further advancing the scope. Sedation was used in three participants (4.9%); the Propofol doses used were 120, 130, and 180 mg. The mean ratings from the screenees, immediately after colonoscopy, for overall assessment and pain/discomfort were 1.6 (range 1– 3) and 2.3 (range 1– 6). A rating of 6 was automatically given immediately after the procedure in cases where sedation was used. CO2 was used for insufflation in all cases. Water immersion, administered via a foot pump, was used during insertion at the discretion of the endoscopist. Follow-up at 24 h and 7 days was complete for all the study participants. The mean overall ratings at 24 h and at 7 days were 1.4 and 1.3 (range 1–5). The mean pain/discomfort ratings at 24 h and at 7 days were 1.5 and 1.3 (range 1–6). Only three screenees had previous colonoscopy. (Groth et Al., 2011)
2.4 Endotics Vs Invendoscope: a data comparison coming from published results
Before to compare the two technologies it is mandatory to uniform data recovered from studies:
In the calculation of cecum reaching rate Endotics included procedures where device had technical problems, while Invendoscope not. For the calculation, procedures with technical failures are excluded also for Endotics.
Both systems presented two models, where the second one was intended as a ameliorative system. For the calculation of caecum reaching rate and time for both devices is considered the second device
According to observational studies (Rex et Al., 2002 - a) and as reported in guidelines (National Guidelines Clearinghouse-NGC 4969, 2006), cases in which procedures are aborted because of poor preparation or severe colitis need not be counted in determining cecal intubation rates. Thus, such procedure are not counted in the presented data. Moreover, because of the protocol of third study afferent Endotics system is focused on the assessment of ulcerative colitis endoscopic activity with Endotics system, related data are not included in the comparison.
Pain score is calculated on the basis of different ranges, from 0 to 10 for Endotics and from 1 to 6 for Invendoscope. Pain score is indicated in the following table as percentage of maximum value of the respective range. Moreover, should be noted that Endotics system used air for insufflation, while Invendoscope, in the last paper, described the use of advanced reduction discomfort techniques, such as CO2 insufflation instead of air and water immersion during insertion.
For all devices, comprising standard colonoscope, “time to cecum” does not include the time spent to carefully analyze the colonic mucosa with diagnostic purposes, except for Endotics, that, due to its working principle, makes observations useful for diagnosis while proceeding towards cecum. Thus, to make a consistent comparison, data related to colonoscopy completion timing should be taken into account. As regards the Endotics system, time to cecum and time to complete diagnosis is slightly different, while for other colonoscopes that make diagnosis during withdrawal is substantially different. According to ASGE guideline, physicians performing a colonoscopy should have an average withdrawal time of six minutes or more for a thorough exam (ASGE-Media Backgrounder, 2010). Moreover, colonoscopist with a low miss rate of lesions have a mean withdrawal time of about nine minutes (Rex et Al., 2000) (Simmons et Al, 2006) (Barclay et Al., 2006) (Overholt et Al, 2010)
Endotics | Invendoscope | |||
Disposable | Yes | Yes | ||
Tool channel | No | Yes | ||
# Studies | 3 | 3 | ||
# Patients enrolled | 123 | 124 | ||
|
||||
Endotics | Invendoscope |
Invendoscope 2nd | ||
Comparison with Standard Colonoscopy | Yes | No | No | |
Asymptomatic volunteers | No | Yes | Yes | |
Paid volunteers | No | Yes | Yes | |
Mean Age | 51.9 | 49.7 | 57.5 | |
One to One procedure* | Yes | No | No | |
Sedation (Propofol) | 0% | 0% | 4.9% | |
Antispasmodic given | 0% | 79% | Not mentioned | |
Insufflation of CO2 | No | No | Yes | |
Water immersion technique | No | No | Yes | |
Pain range | 9% fs | 32.6% fs | 26.6% f.s. | |
Discomfort range | 11% f.s. | Not mentioned | 38.3% f.s. | |
Abdominal compression | 0% | Occasionally | 66% | |
Sensitivity | 93.3% | n.a.** | n.a.** | |
Specificity | 100% | n.a.** | n.a.** | |
NPV | 97.7% | n.a.** | n.a.** | |
PPV | 100% | n.a.** | n.a.** | |
Cecal intubation rate | 93,6% | 90% | 98.4 % | |
Mean time to cecum (min) | n.a. | 23 | 16.4 | |
Mean completion time procedure (min) | 45,3 | n.a. | 32.8 |
Moreover it has to be considered that studies carried out with Endotics have eligibility criteria that include mainly people with prior diagnosis of colorectal diseases (about 70% of patients in the second study) thus procedures’ timing should be compared with a similar study population where completion of the procedure is reached in a mean time of 33 min (range 10-80) ( Rex. et Al. 2002 – b).
The problem with studies reporting a very high completion rate is that they are screening endoscopies in asymptomatic individuals. These populations are different from normal daily practice. Patients undergo colonoscopy for all kinds of clinical indications (Loffeld et Al., 2009). For this reason it is very important, in order to fully understand carry out an exhaustive comparative analysis a comparison between different clinical trials, to study also eligibility and exclusion criteria adopted. The Endotics and Invendoscope systems are described and compared with parameters advocated to predict a difficulty colonoscopic procedure. Parameters are listed in table 3 (Anderson et Al., 2001). Sometimes difficulty’s parameters could be described in different ways, e.g. “
Exclusion Criteria | |||
Difficulty’s parameters | Endotics | Invendoscope 1st | Invendoscope 2nd |
Age "/ 50 | No | Not described | No |
History of abdominal surgery | No | Not described | Yes |
History of pelvic surgery | No | Not described | Yes |
History of diverticular disease | No | Not described | Yes |
Body mass index | No | Not described | Yes |
Inflammatory bowel disease | Yes | Not described | Yes |
3. Conclusion
In this chapter main reasons for including robotic colonoscopy in common practice of colonoscopy screening have been considered. Standard devices are quite rigid, require high experience of the doctor to perform difficult maneuvers to proceed along the tortuous colon walls, and are constructed of materials that could be damaged by heat, pressure, and moisture used during the decontamination processes. The stated above limitations, that could be overcome with robotic colonoscopies, made, and still make, standard colonoscopy a quite invasive technique, with risks related to perforation, sedation and cross-infections, far for being accepted by massive percentages of patients as needed in colorectal cancer screening programs. Nowadays, colonoscopy is used as a matter of fact as screening test just in first-level demonstrative studies and pilot projects. Participation in the first-level FOB (fecal occult blood) screening test is above 50%, and compliance for second level screening programs, is very low compared to expected values, since it is in a range from 30 to 60%. Other important issues that have to be taken into account are related to timing and personnel. As a matter of fact, timing affects all the procedure phases, starting from the preparation which not-disposable devices have to undergo to prevent from cross-infections, and including the duration of the procedures themselves, the gaining of time in the turn-over of patients and the recovery time for the patients to go back to work. As for the personnel, both the number of operators needed and their specific competences.
A state of the art related to working principles of robotic devices have been described as well as main robotic devices proposed for pilot studies. Among these devices two robotic colonoscopes are deeply described and compared: Endotics System and Invendoscope. The comparison included:
If the device is disposable or not
Age of the patients and if they are asymptomatic and/or paid
Presence of tool channel
Number of studies, with related number of patients enrolled
Comparison with standard colonoscopy, in terms of pain range, sensitivity, specificity, NPV and PPV
Sedation or antispasmodic administration
Procedure details, such as cecal intubation rate and timing, abdominal compression maneuvers
An additional table related to the difficulty’s parameters in colonoscopy and exclusion criteria adopted in clinical trials has been filled. In this table parameters related to the age of patients, their surgical and/or colonic disease history, and their BMI are considered.
Endotics system appears to be a promising diagnostic instrument comparable to the gold standard and highly suitable for screening purposes due to the extremely high level of patients’ acceptance even without the adoption of advanced discomfort reducing techniques like CO2 insufflation and water immersion during insertion.
The introduction of this diagnostic instrument in clinical practice could facilitate the adoption of colonoscopy as first-level screening, with a further reduction in the incidence of the colon cancer, estimated in the order of 76-90%. In conclusion a painless colonoscopy, besides being a remarkable achievement for the patient and avoiding any risk related to sedation, has major fallout in terms of prevention.
References
- 1.
Anderson J. C. Messina C. R. Cohn W. Gottfried E. Ingber S. Bernstein G. Coman E. Polito J. 2001 Factors predictive of difficult colonoscopy, Gastrointestinal Endoscopy54 N. 5, 2001 - 2.
Berclay R. L. Vicari J. J. Doughty A. S. Johanson J. F. Greenlaw R. L. 2006 Colonoscopic Withdrawal Times and Adenoma Detection during Screening Colonoscopy N Engl J Med 2006;355 2533 41 - 3.
Classen M. Tygat G. N. J. Lightdale C. J. 2010 Two Centuries of digestive tract endoscopy: a concise report. Gastroenterological Endoscopy, Thieme, 2010 - 4.
Cosentino F. Tumino E. Rubis Passoni. G. Morandi E. Capria A. 2009 Functional evaluation of the Endotics System, a new disposable self-propelled robotic colonoscope: in vitro tests and clinical trial. The International Journal of Artificial Organs /32 8 517 527 - 5.
Eickhoff A. Jakobs R. Kamal A. Mermash S. Riemann J. F. van Dam J. 2006 In vitro evaluation of forces exerted by a new computer−assisted colonoscope (the NeoGuide Endoscopy System), Endoscopy 2006; 38: 1224±1229 - 6.
Ell C. Riemann J. F. Pochon T. Sakai P. Yamamoto H. 2008 GI Endoscopy- Standards and Innovations. Proceedings of the Falk Symposium 166 held in Mainz, Germany, September18 19 Published by Springer, Dordrecht, The Netherlands, 2009, 154 pages - 7.
Faivre J. Dancourt V. Lejeune C. MA Tazi Lamour. J. Gerard D. Dassonville F. Bonithon-Kopp C. 2004 Reduction in colorectal cancer mortality by fecal occult blood screening in a French controlled study. Gastroenterology. 2004 Jun;126 7 1674 80 - 8.
Groth S. Rex D. K. Rösch T. Hoepffner N. (201 2011 High Cecal Intubation Rates With a New Computer-Assisted Colonoscope: A Feasibility Study. American Journal of Gastroenterology advance online publication, 8 March 2011;doi: ajg.2011.52 - 9.
Hu H. Li M. Wang P.i. Feng Y. Sun L. 2009 Development of a continuum robot for colonoscopy, High Technology Letters, 2009,2 Page115 119 - 10.
Levin B. Lieberman D. A. Mc Farland B. MD Smith R. A. Brooks D. Andrews K. S. Dash C. Giardiello F. M. Glick S. Levin T. R. Pickhardt P. Rex D. K. Thorson A. Winawer S. J. (200 2008 Screening and surveillance for the early detection of colorectal cancer and adenomatous polyps, 2008: a joint guideline from the American Cancer Society, the US Multi-Society Task Force on Colorectal Cancer, and the American College of Radiology. Gastroenterology 2008;134 1570 1595 - 11.
Loffeld R.J.L.F. & van der Putten A.B.M.M. 2009 The Completion Rate of Colonoscopy in Normal Daily Practice: Factors Associated with Failure. Digestion 2009;80 267 270 - 12.
Niwa H. Utsumi Y. Nakamura T. Yoshida A. Yoshitoshi Y. Fujino M. Kaneko E. Kasumi A. Matsumoto M. 1969 Clinical experience of colonic fiberscope. Gastroent. Endosc. 1969 Tokyo,11 163 173 - 13.
Overholt B. F. Brooks-Belli L. Grace M. Rankin K. Harrell R. Turyk M. Rosenberg F. B. Barish R. Gilinsky N. H. 2010 Withdrawal times and associated factors in colonoscopy: a quality assurance multicenter assessment. Journal of clinical Gastroenterology-April 2010-4-pp e80 e86 - 14.
Pallotta S. Tumino E. Manes G. Sacco R. Ardizzone S. Bresci G. De Franchis R. 2011 Assessment of ulcerative colitis endoscopic activity with novel robotic colonoscope: the Endotics system. XVII Congresso Nazionale delle malattie digestive Torino,5 9 Marzo 2011 - 15.
Perri F. Iacobellis A. Gentile M. Tumino E. Andriulli A. 2010 The intelligent, painless, “germ free” colonoscopy: A Columbus’ egg for increasing population adherence to colorectal cancer screening? Dig Liver Dis (2010),doi:10.1016/j.dld.2010.06.007 - 16.
Pfeffer J. Grinshpon R. Rex D. K. Levin B. Rösch T. Arber N. Halpern Z. 2006 Proof of New Colonoscope Concept in a Pig Model ´ Endoscopy 2006; 38: 144±148RexD.K. (2000) Colonoscopic withdrawal technique and adenoma miss rates. Gastrointestinal Endoscopy51 1 - 17.
Rex D. K. Bond J. H. Winawer S. Levin T. R. Burt R. W. Johnson D. A. Kirk L. M. Litlin S. Lieberman D. A. Waye J. D. Church J. Marshall J. B. Riddell R. H. 2002 Quality in the Technical Performance of Colonoscopy and the Continuous Quality Improvement Process for Colonoscopy: Recommendations of the U.S. Multi-Society Task Force on Colorectal Cancer The American Journal of Gastroenterology97 6 a - 18.
Rex D.K. & Goodwine B.W. 2002 Method of Colonoscopy in 42 Consecutive Patients Presenting After Prior Incomplete Colonoscopy. The American Journal of Gastroenterology97 5 b - 19.
Rex D.K. 2008 Achieving cecal intubation in the very difficult colon. Gastrointestinal Endoscopy67 6 - 20.
Roesch T. Adler A. Pohl H. Wettschureck E. Koch M. Wiedenmann B. Hoepffner . N. (200 2007 A prospective pilot study to assess technical performance of a new single use colonoscope with inverted sleeve technology. Gastrointest Endosc 2007, 65,AB340 EOF - 21.
Saito T. Kagiwada T. Harada H. Kawamura Y. 2009 Development of an Earthworm Robot with a Shape Memory Alloy and Braided Tube. Advanced Robotics,23 Numbers 12-13, 2009,1743 1760 Publisher: VSP, an imprint of Brill - 22.
Shike M. Fireman Z. Eliakim R. Segol O. Sloyer A. Cohen L. B. Goldfarb-Albak S. Repici A. 2005 New Technology ColonoSight OMED Colorectal Cancer Screening Meeting May 15, 2005 - 23.
Simmon D. T. Harewood G. C. Baron T. H. Petersen B. T. Wang K. K. Boyd-Enders F. Ott B. J. (200 2006 Impact of endoscopist withdrawal speed on polyp yield: implications for optimal colonoscopy withdrawal time. Alimentary pharmacology & therapeutics- 2006-Sep;6 965 971 - 24.
Slatkin B. A. Burdik J. (199 1997 The development of a robotic endoscope, Experimental Robotics IV, The 4th International Symposium, Stanford, California, June 30- July 2, 1995,223 161 169 - 25.
Sturges R. H. Laowattana S. 1993 A flexible, tendon-controlled device for endoscopy, 1993- The International Journal of Robotics Research12 2 121 131 - 26.
Swain P. 2009 Colonoscopy: New Designs for the Future Gastrointestinal Endoscopy Clinics of North America,15 4 839 863 - 27.
Swaroop V.S. & Larson M.V. 2002 Colonoscopy as a Screening Test for Colorectal Cancer in Average-Risk Individuals. Mayo Clin Proc. 2002;77 951 956 - 28.
Tumino E. Sacco R. Bertini M. Bertoni M. Parisi G. Capria A. (201 2010 Endotics system vs colonoscopy for the detection of polyps. World J Gastroenterol 2010;16 43 5452 5456 - 29.
Waye J. D. Rex D. K. Williams C. B. (200 2009 Colonoscopy: Principle and practice. Second edition. 2009 Published by Wiley-Blackwell - 30.
Zuo J. Yan G. Gao Z. 2005 A micro creeping robot for colonoscopy based on the earthworm, J Med Eng Technol. 2005 Jan-Feb;29 1 1 7