1. Introduction
Advanced biomechanical models of biological tissues should be based on statistical morphometry of tissue architecture. A quantitative description of the microscopic properties of real tissue samples is an advantage when devising computer models that are statistically similar to biological tissues in physiological or pathological conditions. The recent development of X-ray microtomography (micro-CT) has introduced resolution similar to that of routine histology. The aim of this chapter is to review and discuss both automatic image processing and interactive, unbiased stereological tools available for micro-CT scans and histological micrographs. We will demonstrate the practical usability of micro-CT in two different types of three-dimensional (3-D)
2. Principles of micro-CT
This chapter covers the basic principles of micro-CT.
Certain devices operate with geometrical magnification only, in which the resolution increases with the distance between the sample and the detector. Unfortunately, increased geometrical magnification can result in blurriness, depending on the X-ray source spot size.
More sophisticated devices use microscope objectives for increased magnification and resolution. For optimal settings of the micro-CT scan, the spot size (SS) and the awaited pixel size (PS) are the parameters that set the source distance (SD) and detector distance (DD). When the SD and DD are smaller, higher photon counts can be achieved, thus reducing the time costs. The settings should fulfill equation 1; see also Fig. 2:
The source SS is strongly dependent on the power of the X-ray tube used. In today’s machines, the source SS ranges between 1-10 µm. For high resolution scanning of biological samples, it is advisable to operate with a low accelerating voltage (the typical range for the Xradia XCT 400 (Xradia, Pleasanton, CA, USA) is 20–60 kV used with a power of 4 W, at which the tube yields the lowest SS). The scanning time depends on the magnification and resolution required. For details smaller than 1 µm, 24 hours or more might be necessary, whereas an overview scan with a pixel size of approximately 10 µm can be achieved within an hour. Sample drifting might be an issue during a long scanning time. From the reconstructed images, the objects of interest are visualized, thresholded, traced and analyzed.
3. Current applications of micro-CT in biomechanics and medicine
Compared with standard human CT devices, which offer a resolution limit of approximately 0.4 mm, the micro-CT introduced a promising modality. However, the clinical use of this method is limited by its higher radiation exposure and longer scanning times. It is used either to visualize individual fine functional and anatomical structures of
Most organs have already been analyzed with micro-CT, including bones (Peyrin, 2011), heart and blood vessels (Schambach et al., 2010), lungs, kidney, liver, and cerebral structures (Schambach et al., 2010). Micro-CT devices can be used for the characterization of bone or vascular microarchitecture (Peyrin, 2011; Burghardt et al., 2011; Missbach-Guentner et al., 2011). This method also allows the precise detection of the margins of tumors and their vascularity (Ma et al., 2011; Missbach-Guentner et al., 2011). Tissue composition (e.g., bone mineralization) can be directly linked to 3-D tissue morphometry (Burghardt et al., 2011). Thus, 3-D micro-CT analysis becomes a method of choice for describing the spatial complexity of organ segmentation and the relationships between morphological and functional units (e.g., hepatic lobules and portal acini) (Schladitz, 2011). Micro-CT can also link the imaging of anatomical structures with functional and molecular imaging, e.g., tissue and organ perfusion, the flow rate of exocrine secretions within parenchymatous organs and glands (Marsen et al., 2006), or heart movements (Badea et al., 2005).
For the examination of soft tissues and vessels, contrast solutions are necessary. In
Other modifications of CT applicable in medical experiments are the mini-CT devices (voxel size 10-3 mm3, used for scans of whole organs or small animals) and the nano-CT devices (voxel size of 10-7 mm3) (Ritman, 2011; Müller, 2009).
The greatest progress in micro-CT exploration was acquired in the exploration of tumor microvascularization and the study of neoangiogenesis. The latter phenomenon is also important for the understanding of tumor growth and could be used in oncological treatment strategies, especially in patients treated by biological therapy with antibodies against vascular endothelial factor A (bevacizumab), which inhibits neoangiogenesis (Ma, 2011). Micro-CT imaging of pathological vascularity can provide new information, e.g., about changes in vessel walls in atherosclerosis or other pulmonary vascular diseases (Razavi, 2012).
A promising trend in experimental work is using hybrid methods that combine detailed anatomical information from micro-CT with information about cellular metabolism and structure from methods of nuclear medicine (micro-SPECT/CT and micro-PET/CT) (Ritman, 2011). Therefore, it is appropriate to combine or compare the results of micro-SPECT or micro-PET with, e.g., microscopic analysis of a specimen to estimate the correct anatomical orientation and acquire a satisfactory interpretation of the results. The new suggested tools would be able to use higher energy examination from more X ray sources and thus obtain results on the cellular or subcellular level. The development of new tissue-specific contrast solutions could also be promising for future research activities using micro-CT or its hybrid methods.
4. Micro-CT imaging of biological samples ex vivo
4.1. Biomechanics of bone scaffolds
The aim of this chapter is to illustrate the application of micro-CT in tissue engineering and in assessing the biomechanics of biocompatible collagen/hydroxyapatite bone scaffold samples (Prosecká et al., 2011). Tissue engineering is a promising interdisciplinary research field that aims to develop biological substitutes for the repair of damaged tissue. The typical strategy involves either the delivery of isolated and expanded cell populations within a tissue engineering construct or the recruitment of host cells local to the site of damage through the use of conductive scaffolds and inductive biological signals. The question of how to optimize the design of scaffolds for different tissues remains unsolved. To assess the suitability of polymer tissue scaffolds for use in regenerative medicine, methods to characterize scaffolds are needed (Renghini et al., 2009). The bone scaffolds should be stiff enough to withstand high forces in the bone after implantation but, conversely, should also be flexible enough to enable growth of the cells and changing of the original shape of the graft to meet the needs of the bone complex. Therefore, it is important to perform mechanical measurements and 3-D imaging of bone scaffolds before the seeding of mesenchymal stem cells.
Generally, bones and bone scaffolds can be mechanically tested using various types of techniques: tensile or torsion tests used for strip- or block-shaped tissue specimens; a pressure test used for block-shaped or cylindrical specimens; a ring test in which a ring of given thickness is cut from a tubular organ (typically a blood vessel), clamped into the jaws of special measurement devices and loaded by tension; and an intraluminal pressure inflation-deflation test of tubular organs. The choice depends on the physiological loading of the tissues. The aim is to be as close as possible to the real loading and thus to the real mechanical properties. In any case, regardless of the chosen technique, the result given by a measurement device is of the same nature: a stress-strain (tensile, pressure, ring test) or pressure-outer diameter (inflation-deflation test) curve. The stress-strain curve has a mostly nonlinear shape showing the tissue stiffening as loading increases. The stiffening is caused by various tissue components as they subsequently contribute to the tissue response. The soft component, mostly elastin, contributes to the mechanical response at low loading and is connected with the low stiffness of the tissue, whereas the curly and stiff collagen fibers are straightened as loading progresses and contribute to stiffening at high loading. The relevant portions of a stress-strain curve could be approximated (e.g., by a line), and thus, the mechanical parameters, such as Young's modulus of elasticity at small deformations (low loading) and at large deformations (high loading), the pressure-strain elastic modulus, the initial modulus of compression, the limit stress and the strain in the case of loading until tissue rupture, could be obtained.
To emulate the loading of bones, in which most of their parts are under pressure and only a small fraction of the tissue is under tension, pressure loading was applied to collagen/ hydroxyapatite composite bone scaffolds. Cylindrical specimens (approximately 12 mm in diameter and in height) of composite scaffolds containing various amounts of collagen and hydroxyapatite prepared according to Prosecká et al. (2011) underwent pressure mechanical loading with a loading velocity of 1 mm/min. The resultant stress-strain curves, and particularly the regions between 2% and 10% of the original specimen's height, were approximated by linear regression, and thus the initial moduli of compression were determined. (for details of the measurements and their evaluation, see Prosecká et al., 2011).
From a biomechanical point of view, the shape of the stress-strain curves was identical for all compositions of scaffolds. The beginning of each stress-strain curve was linear with low stiffness, and increasing loading led to scaffold stiffening. This stiffening was likely caused by the fact that, with increasing deformation of the porous specimens, the originally high amount of free space filled by air became smaller, and the stiff components became closer together and started to contribute more to the mechanical response. The decreasing porosity caused by the increasing pressure loading could thus be connected with the higher stiffness of the scaffold. The initial modulus of compression increased rapidly with an increase in the collagen concentration. From these two results, we can conclude that not only the composition but also the porosity may play crucial roles in the mechanical properties of collagen/hydroxyapatite composite bone scaffolds.
4.2. Imaging and biological applications of bone scaffolds
The 3-D porous structure and sufficient mechanical stiffness of the bone scaffolds are necessary conditions for the attachment, growth, and progress of mesenchymal osteoprogenitor stem cells (Prosecká et al., 2011); see Fig. 3.
Traditional methods for evaluating the osseointegration of tissue-engineered scaffold/cell constructs are based on 2-D histological and radiographic techniques. Sectioning followed by histology can image the scaffold interior but is destructive, lengthy and only semi-quantitative (Ho & Hutmacher, 2006). Fluorescence microscopy can be quantitative when high-throughput approaches are applied, e.g., producing 3-D images with confocal fluorescence microscopy (Tjia & Moghe, 1998). Colorimetric and fluorometric soluble assays are available for cell components, such as enzymes, protein or DNA (Ho & Hutmacher, 2006). However, these soluble assays are quantitative but do not provide information on cell distribution. In contrast to these methods, by micro-CT we can non-destructively obtain 3-D images that penetrate deep into the scaffold interior and produce inherently quantitative results (Tjia & Moghe, 1998). Micro-CT currently appears to be the most suitable approach for this task (Ho & Hutmacher, 2006; Mather et al., 2007; Mather et al., 2008; Cancedda et al., 2007). Clearly, in a structure as complex as bone, micro-CT provides a distinct advantage over conventional microscopy. Structures can be followed continuously, from the level of osteon to gross bone morphology.
Dorsey et al. (2009) explored the use of X-ray microcomputed tomography for observing cell adhesion and proliferation in polymer scaffolds. The ability of micro-CT to detect cells in scaffolds was compared with those of fluorescence microscopy and a soluble DNA assay. The researchers demonstrated that fluorescence microscopy had better resolution than micro-CT and that the soluble DNA assay was approximately 5 times more sensitive than micro-CT. However, by micro-CT, they were able to reveal the interiors of opaque scaffolds and to obtain quantitative 3-D imaging and analysis via a single, non-invasive modality. They observed that for quantitative micro-CT volume analysis, a cell density of more than 1 million cells/ml is required in the scaffolds. The results demonstrate the benefits and limitations of using micro-CT for the 3-D imaging and analysis of cell adhesion and proliferation in polymer scaffolds. Various bone engineering groups have noted the importance of micro-CT analyses in tissue engineering. Among them, Müller and Rüegsegger have investigated, and quantified, the architecture of cancellous bone using micro-CT (Müller, 1994; Müller et al., 1996, 1998; Rüegsegger, 2001).
In addition to these analyses, a number of morphological measures are being investigated, such as volume fractions of tissues, directivity of calcifying tissue, porosity, pore connectivity, putative vascularization, curvature and surface-to-volume measure (Bentley et al., 2002; Jorgensen et al., 1998).
Despite the advantages of using micro-CT, there are still several issues with both image segmentation and resolution that are exacerbated by the low image contrast due to the low X-ray attenuation of the materials being used (Mather et al., 2008). Morris et al. (2009) studied a method for the generation of computer-simulated scaffolds that resemble foamed polymeric tissue scaffolds. They showed that the quality of the images (and hence the accuracy of any parameters derived from them) may be improved by a combination of pixel binning and by taking multiple images at each angle of rotation. However, micro-CT is considered to be a standard technique in tissue-engineered bones (Cancedda et al., 2007).
5. Three-dimensional imaging of microvascular corrosion casts
5.1. Corrosion casting
Anatomical corrosion casts provide 3-D insights into the arrangement of hollow structures and organ cavities at both the macro- and microscopic levels. The most frequently used are vascular corrosion casts (Giuvărăşteanu, 2007), which, in combination with scanning electron microscopy, constitute the primary application of corrosion casting describing the morphology and anatomical distribution of blood vessels (Lametschwandtner et al., 2004; 2005). Many other examples are found in the literature - the airways (Hojo, 1993), bile ducts (Gadžijev & Ravnik, 1996), urinary tract (Marques-Sampaio et al., 2007) and lymphatic system (Fujisaka et al., 1996) have been visualized by means of corrosion casting. The steps for obtaining the corrosion casts are as follows: perfusion of the hollow target structures to remove the contents (in the case of vascular casts, prior heparinization is required for the maintenance of blood fluidity), injection of the casting media, its polymerization in the fully filled cavities and the subsequent removal of the surrounding tissues by a highly aggressive corrosive solution. The corrosion casting method has been known since the beginning of the 16th century, when Leonardo da Vinci injected dissolved wax into bovine cerebral chambers (Paluzzi et al., 2007). Over the following centuries, all of the steps in the method have been improved. However, the aim has remained the same – to obtain the most accurate replica of the biological structure.
5.2. Casting media
Because the highest authenticity is required in the proportions of 3-D corrosion casts, casting media must have sufficient viscosity (to pass through but not to penetrate); they must be capable of even, rapid polymerization with minimal shrinkage and of physicochemical resistance to the subsequent corrosion and dissection procedures. All of these properties are combined in methylmethacrylate (MMA) resin. The MMA polymerizing fluid can be produced by warming the MMA monomer with a catalyst, benzoyl peroxide. MMA easily penetrates the capillary network and flows out via the opposite vessels (Kachlík & Hoch, 2008). A polyurethane pigment paste of various colors can be added into the mass along with the X-ray contrast medium, e.g., barium sulfate (BaSO4) or mercury sulfide (HgS) (Gadžijev & Ravnik, 1996). There are also commercially produced, partially polymerized MMAs (e.g., Mercox or Dentacryl), but these products vary considerably in their rheological properties. Dentacryl was used mainly in the last century, originally in dental prosthetics. This compound is a commercially produced, colorless, partially polymerized MMA that is not, due to its high viscosity, suitable for casts of the microvasculature (Kachlík & Hoch, 2008). The greatest advantage of this casting medium is its low cost; thus, Dentacryl is used for more spacious casts, such as those of the bronchial tree (Havránková et al., 1989) or paranasal sinuses (Hajnis, 1988). Painting clays can be used to color Dentacryl. Mercox is a low-viscosity acrylic resin, a MMA monomer stabilized solution. Mercox II (Ladd Research, Williston, Vermont, USA) is commercially offered in two colors (blue and red); the kit contains an MMA resin and a catalyst (benzoyl peroxide). The Mercox II system combines excellent permeability through the entire vascular bed, excellent infiltration properties, short time of preparation and polymerization, minimal shrinkage and high chemical resistance. Although its price is rather high, it is a method of choice for the vascular corrosion casts (Bartel & Lametschwandtner, 2000; Minnich et al., 2002; Lametschwandtner et al., 2004; Kachlik & Baca, 2006, Kachlík & Hoch, 2008).
In our experience with obtaining corrosion casts of porcine liver vascular trees, it is helpful to administer 50,000 IU of heparin in 1 liter of saline and subsequently rinse the vascular bed with 5 liters of Ringer´s solution prior to sacrificing the animal. We found that even the lowest Mercox dilution recommended by the manufacturer (20 ml resin/0.4 g of 40% benzoylperoxide) polymerizes within a very short time (approximately 5 min), which makes it difficult to fill the complicated liver vasculature (Eberlová, 2012, unpublished results). After 5 minutes of stirring the resin with the catalyst, it was possible to fill a volume of approximately 20 ml before the Mercox started to harden. To prevent any vascular lesions and artifacts, the use of an infusion pump appears to be warranted (Minnich et al., 2002).
Unfortunately, Mercox does not offer sufficient X-ray-opacity for micro-CT. To fill and opacify microvessels
6. Automatic image processing, topology analysis and measurement of statistical features
This chapter describes an approach to automatic micro-CT image processing using computer vision techniques and the Mercox vascular corrosion casts of the intestinal mucosa described in the previous chapter. The input data were acquired with Xradia XCT 400 (Xradia, Pleasanton, CA, USA) by scanning Mercox vascular corrosion casts. The resulting data set consisted of a DICOM stack of approximately 1000 slices, each 1000 times 1000 pixels, such that the scanned volume contained 109 volume elements (voxels), each represented by a 16-bit signed number. The depicted volume in reality was a cube with an edge of approximately 35 mm, whereas the voxel edge was 36.53 µm. For image processing, we used the Insight Toolkit library for C++ (www.itk.org) and MATLAB with its Image Processing Toolbox (The MathWorks, Inc., Natick, MA, USA). The first step of the image processing was an analysis of the density distribution of the data, using a density histogram; see Fig. 4.
Segmentation is a process in computer vision that divides the voxels of the source image into two subsets – foreground and background – depending on their affinity to the objects of the real world (Sonka et al., 1998). In our case, the foreground was formed by the vascular corrosion cast, whereas all of the other data were considered to be background.
Thresholding is a common and widely used image segmentation method in machine vision. This technique outputs a binary image
In this equation,
After segmentation was performed, it was necessary to further pre-process the binary image to eliminate artifacts, e.g., holes and bays caused by irregular spreading of the casting material, and measurement errors. These errors were resolved by applying the morphological operation of hole filling, which converts cavities inside of objects into object voxels. The next step of the data analysis used the labeling algorithm, which assigned a unique label to each contiguous region of a binary image (Sonka et al., 1998). This procedure gave us a large amount of useful information about object counts and sizes in the volume. Part of the label image histogram is shown in Fig. 5.
The label histogram (Fig. 5) demonstrates that there are many objects of insignificant size that are produced simply by measurement noise. We can convert these objects into background and eliminate them from further processing. The size of the objects at which the segmentation of the blood vessel regions becomes unreliable, reveals also the resolution limits of the current micro-CT scan. Acquiring a reliable representation of all of the microvessels would therefore require another scan with a different objective and micro-CT settings. We can also deduct the count of dominant objects in the volume. For the purposes of visualization, in the subsequent processing steps, we will only work with the largest object. Fig. 6 demonstrates the difference between a 3-D model of the whole vessel system (Fig. 6A), and the largest continuous object only (Fig. 6B).
The volumetric image (Fig. 6) is useful for visualization. However, it is not desirable for topological investigation. A shape simplification is needed that would preserve the tree connectivity and geometric conditions. In this study, we used the algorithm introduced by Lee et al. (1994). This algorithm is based on parallel 3-D binary image thinning, which is suitable for large data sets and produces a 1-voxel-thin 3-D skeleton. The topology analysis of the skeleton involves identifying the node and terminal points of the tree. The points are located based on their 26-connectivity to neighboring voxels. Based on the properties of the skeletonization algorithm, we can find nodes of valence 3 (i.e., nodes at which three microvascular segments join). Existing nodes of valence 4 are decomposed into two nodes of valence 3. Fig. 7 shows the resulting skeleton with located node points. In this sample, 149 nodes were detected. Using the number of branching points with a known valence, the numerical density of the microvessels
Currently, we are able to segment a vessel tree formed by a corrosive cast. In the example analyzed in this study, continuous regions of the volumetric model were counted to detect the number of individual objects within the volume. The volume of each individual object was estimated. Using skeletonization, the number of branching nodes was counted, and the number of vascular segments between the nodes was estimated, as was the length of each segment. Knowing the vessel length and volume, it is also possible to compute the average diameter of the vessel. However, there are still challenges for future work, such as estimating the vessel diameter in each voxel of the skeleton using a distance transform. With this information, the diameter distribution with respect to the vessel length or the vessel volume could be acquired. With a known diameter, the surface area of the vessels can also be computed. Tracking the skeleton leading its binary graph construction determines the spatial distribution of the branching nodes from the proximal vascular segments to the periphery of the vascular tree.
The approach presented in this chapter is based on several assumptions that deserve to be discussed, as they also represent the limitations of this study. We assume, for our computation, that the vessels are in the form of generalized cylinders, which means that the cross-section orthogonal to the medial axis is a circle. The voxels in our data set were cubical; however, a data set with unequal voxel edges may be resampled into cubical voxels.
An important question is the choice of the threshold value. The segmented vessel diameter is partially linearly dependent on the chosen threshold value. Setting an excessively low threshold causes too many artifacts to appear and objects to merge into each other; contrariwise, an excessively high threshold makes vessels very thin, such that small vessels disappear. This effect implies that the threshold value influences the absolute diameter of the vessels and biases statistical markers. However, the construction of the skeleton and the topological analysis do not appear to be affected by the selection of the threshold.
Let us summarize the advantages and disadvantages of our automatic image processing approach. Compared with stereological methods currently in use (see the next chapter), automatic image processing requires virtually no user interaction. This approach is able to compute the distributions of the volumes, surfaces and lengths of vessels of given diameters with a high precision. Moreover, the stereological methods may be applied to our volumetric model to achieve results comparable to those of stereological measurements performed by a human, but with no interaction at all. A disadvantage is the necessity of the choice of the threshold used for the segmentation, due to the sensitivity of certain results to the threshold settings. As stated before, the threshold value does not greatly influence the topology of the vessel tree but strongly correlates with the diameters of the vessels. A proposed solution to this issue is to compute the skeleton with a higher threshold and the diameters with a lower one.
7. Quantitative micro-CT, histology and stereology
Current micro-CT devices are typically bundled with sophisticated software packages that offer a number of automated quantification procedures. However, correlating the micro-CT results with quantitative histology favors the use of unbiased stereological methods, which are highly standardized and widely accepted in biomedical microscopy research (Howard and Reed, 1998; Mouton, 2002). This chapter illustrates the stereological assessment of micro-CT scans of bone scaffolds and microvascular corrosion casts, including the quantification of the volume fraction (
In bone tissue samples, the micro-CT resolution is currently capable of providing images that can be used for both analysis of bone vascular canals, and counting individual osteocyte lacunae. Quantification of bone microporosities is used for testing their effect on the viscoelastic properties of bone tissue (Tonar et al., 2011). The microporosity has at least two functional levels, the vascular porosity (related to the vascular canals; the order of magnitude is 10–1000 μm) and the lacunar-canalicular porosity (surrounding the osteocytes; the order of magnitude is 0.1–10 μm). Mechanical experiments clearly demonstrated that the hierarchical organization of bone architecture is crucial and that bone structural heterogeneity varies with the scale of magnification. Whereas Fig. 8 demonstrates two-dimensional sections of compact bone produced by histology and micro-CT, Fig. 9 shows the results of 3-D micro-CT reconstruction of cancellous bone.
The vascular corrosion casts described in section 4 and used in section 5 also can be assessed with spatial stereological methods. Image data acquired by micro-CT are demonstrated in Fig. 10.
After manually tracing the microvessel profiles within a series of consecutive two-dimensional micro-CT sections (software Ellipse, ViDiTo, Košice, Slovak Republic), a three-dimensional system of oriented lines can be acquired (Fig. 11).
Next, the orientation of each skeletonized vessel can be described using a spherical coordinate system (Fig. 12). Each blood vessel segment is described as a vector connecting the center of the coordinate system with the surface of the sphere. This vector is described by its length and a combination of azimuth ranging between [0, 2π] and elevation [0, π/2].
Next, the combinations of vessel lengths and their 3-D orientation can be assessed using various 2-D plots (Fig. 13). These plots are very useful when assessing the directionality and anisotropy of the vascular segments. Should the anisotropy be quantified, several methods are available, such as the ellipsoidal anisotropy, fractional anisotropy, or a chi-square method comparing the observed length densities of lines with the discrete uniform distribution of an isotropic line system (Kochová et al., 2011).
Using the skeletons of the microvessels, their lengths
The volume fraction occupied by the microvascular corrosion cast easily can be estimated using the Cavalieri principle (Howard & Reed, 2005), as shown in equation 4,
where
To simulate the histological measurement of microvessel density
The surfaces of the microvascular casts can also be estimated using stereological methods. However, several of these methods require isotropic uniform random sections or vertical uniform random sections. Randomized orientation of the sections cannot always be guaranteed in micro-CT, as the sample is typically oriented with its long axis perpendicular to the X-ray beam (Fig. 1). The section plane is often arbitrary and cannot be regarded as random. A suitable solution without randomizing the cutting plane is using an isotropic virtual test probe named fakir (Larsen et al., 1998; Kubínová & Janáček, 1998; Kubínová & Janáček, 2001); see Fig. 15. The ratio between the surface area
8. Conclusion
This chapter reviews the current approaches in micro-CT imaging and the quantitative evaluation of the resulting image data sets. Both automatic image processing methods and interactive stereological methods are applied for the quantification of the volume fraction, surface density, length density, numerical density, orientation and anisotropy. Micro-CT imaging of bone tissue, tissue-engineered bone scaffolds, and microvascular corrosion casts is quantified using unbiased methods that are already acknowledged in quantitative histology.
Acknowledgement
This work was supported by the European Regional Development Fund (ERDF) project “NTIS - New Technologies for Information Society”, European Centre of Excellence, CZ.1.05/1.1.00/02.0090. The micro-CT technique was developed within the CENTEM project, reg. no. CZ.1.05/2.1.00/03.0088, which is cofunded from the ERDF within the OP RDI program of the Ministry of Education, Youth and Sports. The corrosion casting was funded by the Charles University in Prague, Project No. SVV 264808, and by the Internal Grant Agency of the Ministry of Health of the Czech Republic under Project No. IGA MZ ČR 13326. The quantification of vascular trees was funded by the Grant Agency of the Czech Republic, Project No. 106/09/0740. The bone scaffold research was funded by the Grant Agency of the Czech Republic, Project No. P304/10/1307, and by the The Grant Agency of the Charles University, Project GAUK No. 96610.
References
- 1.
Badea C. T. Fubara B. Hedlund L. W. Johnson G. A. 2005 D micro-CT of the mouse heart. Mol. Imaging,4 110 116 1535-3508 - 2.
Bag S. . Schambach S. J. . Boll H. . Schilling L. . Groden C. . Brochmann M. A. 2010 Aktueller Stand der Mikro-CT in der experimentellen Kleintierbildgebung. Fortschr. Roentgestr.,182 390 403 1438-9029 - 3.
Bartel H. Lametschwandtner A. 2000 Intussusceptive microvascular growth in the lung of larval Xenopus laevis Daudin: a light microscope, transmission electron microscope and SEM study of microvascular corrosion casts. Anat. Embryol. (Berl).,202 55 65 0340-2061 - 4.
Bentley M. D. Ortiz M. C. Ritman E. L. Romero C. 2002 The use of Microcomputed tomography to study microvasculature in small rodents. Am. J. Physiol.-Reg. Integ. Comp. Physiol.,282 1267 1279 0363-6119 - 5.
Boll H. Nittka S. Doyon F. Neumaier M. Marx A. Kramer M. Groden C. Brockmann M. A. 2011 Micro-CT based experimental liver imaging using a nanoparticulate contrast agent: a longitudinal study in mice. PLoS One,6 e25692.1932-6203 - 6.
Burghardt A. J. Link T. M. Majumar S. 2011 High-resolution computed tomography for clinical paging of bone microstructure. Clin. Orthop. Relat. Res.,469 2179 2193 0000-9921 X. - 7.
Cancedda R. Cedola A. Giuliani A. Komlev V. Lagomarsino S. Mastrogiacomo M. Peyrin F. Rustichelli F. 2007 Bulk and interface investigations of scaffolds and tissue-engineered bones by X-ray microtomography and X-ray microdiffraction. Biomaterials,28 2505 2524 0142-9612 - 8.
Dorsey S. M. Lin-Gibson S. Simon C. G. Jr 2009 X-ray microcomputed tomography for the measurement of cell adhesion and proliferation in polymer scaffolds. Biomaterials,30 2967 2974 0142-9612 - 9.
Fraser G. M. Milkovich S. Goldman D. Ellis C. G. 2012 Mapping 3-D functional capillary geometry in rat skeletal muscle in vivo. Am. J. Physiol. Heart. Circ. Physiol.,302 H654 H664 0363-6135 - 10.
Fujisaka M. Ohtani O. Watanabe Y. 1996 Distribution of lymphatics in human palatine tonsils: a study by enzyme-histochemistry and scanning electron microscopy of lymphatic corrosion casts. Arch. Histol. Cytol.,59 273 280 0914-9465 - 11.
Gadžijev E. M. Ravnik D. 1996 Atlas of Applied Internal Liver Anatomy. Springer,978-3-21182-793-2 Wien. - 12.
Giuvărăşteanu I. 2007 Scanning electron microscopy of vascular corrosion casts-standard method for studying microvessels. Rom. J. Morphol. Embryol.,48 257 261 0377-5038 - 13.
Gössl M. Rosol M. Malyar N. M. Fitzpatrick L. A. Beighley P. E. Zamir M. Ritman E. L. 2003 Functional anatomy and hemodynamic characteristics of vasa vasorum in the walls of porcine coronary arteries. Anat Rec A Discov Mol Cell Evol Biol.,272 526 537 1552-4884 - 14.
Gundersen H. J. Jensen E. B. 1987 The efficiency of systematic sampling in stereology and its prediction. J. Microsc.,147 229 263 0022-2720 - 15.
Hajnis K. 1988 The capacity of the sphenoid sinus. Anat. Anz.,167 23 28 0003-2786 - 16.
Havránková J. Skoda V. Holusa R. 1989 The use of the Dentacryl rapid (Spofa) resin for preparation of the rat tracheobronchial casts. Z. Versuchstierkd.,32 97 100 0044-3697 - 17.
Ho S. T. Hutmacher D. W. 2006 A comparison of micro CT with other techniques used in the characterization of scaffolds. Biomaterials27 1362 1376 0142-9612 - 18.
Hojo T. 1993 Scanning electron microscopy of styrene-methylethylketone casts of the airway and the arterial system of the lung. Scanning Microsc.,7 287 293 0891-7035 - 19.
Howard C. V. Reed M. G. 2005 Unbiased Stereology: Three Dimensional Measurement in Microscopy, (2nd edition), Garland Science/BIOS Scientific,978-1-85996-089-9 New York. - 20.
Jorgensen S. M. Demirkaya O. Ritman E. L. 1998 Three-dimensional imaging of vasculature and parenchyma in intact rodent organs with X-ray micro-CT. Am. J. Physiol.,275 H1103 H1114 0002-9513 - 21.
Kachlik D. Baca V. 2006 Macroscopic and microscopic intermesenteric communications. Biomed. Pap. Med. Fac. Univ. Palacky Olomouc Czech Repub.,150 121 124 1213-8118 - 22.
Kachlík D. Hoch J. 2008 The blood supply of the large intestine. Karolinum,68 71 978-8-02461-397-0 Prague. - 23.
Kochová P. Cimrman R. Janáček J. Witter K. Tonar Z. 2011 How to asses, visualize and compare the anisotropy of linear structures reconstructed from optical sections--a study based on histopathological quantification of human brain microvessels. J. Theor. Biol.,286 67 78 0022-5193 - 24.
Kubínová L. Janáček J. 1998 Estimating surface area by the isotropic fakir method from thick slices cut in an arbitrary direction. J Microsc.,191 201 211 1365-2818 - 25.
Kubínová L. Janáček J. 2001 Confocal microscopy and stereology: estimating volume, number, surface area and length by virtual test probes applied to three-dimensional images. Microsc. Res. Techn.53 425 435 1097-0029 - 26.
Lametschwandtner A. Minnich B. Kachlik D. Setina M. Stingl J. 2004 Three-dimensional arrangement of the vasa vasorum in explanted segments of the aged human great saphenous vein: scanning electron microscopy and three-dimensional morphometry of vascular corrosion casts. Anat. Rec. A Discov. Mol. Cell. Evol. Biol.,281 1372 1382 1552-4884 - 27.
Lametschwandtner A. Minnich B. Stöttinger B. Krautgartner W. D. 2005 Analysis of microvascular trees by means of scanning electron microscopy of vascular casts and 3D-morphometry. Ital. J. Anat. Embryol.,110 87 95 1122-6714 - 28.
Larsen J. O. Gundersen H. J. Nielsen J. 1998 Global spatial sampling with isotropic virtual planes: estimators of length density and total length in thick, arbitrarily orientated sections. J. Microsc.,191 238 248 1365-2818 - 29.
Lee T. C. Kashyap R. L. Chu C. N. 1994 Building skeleton models via 3-D medial surface/axis thinning algorithms.Graph. Model. Image Process.,56 6 462 478 1077-3169 - 30.
Lokkegaard A. Nyengaard J. R. West M. J. 2001 Stereological estimates of number and length of capillaries in subdivisions of the human hippocampal region. Hippocampus,11 726 740 1050-9631 - 31.
Ma X. Tian J. Yang X. Qin C. 2011 Molecular paging in tumor angiogenesis and relevant drug research. Int. J. of Biomedical Imaging,2011 1687-4196 - 32.
Marques-Sampaio B. P. Pereira-Sampaio M. A. Henry R. W. Favorito L. A. Sampaio F. J. 2007 Dog kidney: anatomical relationships between intrarenal arteries and kidney collecting system. Anat Rec. (Hoboken),290 1017 1022 1932-8486 - 33.
Marsen M. Paget C. Yu X. Henkelman R. M. 2006 Estimating perfusion using micro-CT to locate microspheres. Phys. Med. Biol.,51 N9 16 0031-9155 - 34.
Mather M. L. Morgan S. P. Crowe J. A. 2007 Meeting the needs of monitoring in tissue engineering. Regen. Med.,2 145 160 1746-0751 - 35.
Mather M. L. Morgan S. P. White L. J. Tai H. Kockenberger W. Howdle S. M. 2008 Image-based characterization of foamed polymeric tissue scaffolds. Biomed. Mater.,3 015011 pp),0955-7717 - 36.
Minnich B. Bartel H. Lametschwandtner A. 2002 How a highly complex three-dimensional network of blood vessels regresses: the gill blood vascular system of tadpoles of Xenopus during metamorphosis. A SEM study on microvascular corrosion casts. Microvasc Res.,64 425 437 0026-2862 - 37.
Missbach-Guentner J. Hunia J. Alves F. 2011 Tumor blood vessel visualization. Int. J. Dev. Biol.,55 535 546 1696-3547 - 38.
Morris D. E. Mather M. L. Crowe J. A. 2009 Generation and simulated imaging of pseudo-scaffolds to aid characterisation by X-ray micro CT. Biomaterials,30 4233 4246 0142-9612 - 39.
Mouton P. R. 2002 Principles and Practices of Unbiased Stereology. An Introduction for Bioscientists, The Johns Hopkins University Press,0-80186-797-5 - 40.
Müller R. 1994 Non-invasive bone biopsy: a new method to analyse and display the 3D structure of trabecular bone. Phys. Med. Biol.,39 145 164 0031-9155 - 41.
Müller R. Hildebrand T. Hauselmann H. J. Rüegsegger P. 1996 Resolution dependency of microstructural properties of cancellous bone based on 3d-microct. Technol. Health Care,4 113 119 0928-7329 - 42.
Müller R. van Campenhout H. van Damme B. van der Perre G. Dequeker J. Hildebrand T. Rüegsegger P. 1998 Morphometric analysis of human bone biopsies: a quantitative structural comparison of histological sections and micro-computed tomography. Bone,23 59 66 8756-3282 - 43.
Müller R. 2009 Hierarchical microimaging of bone structure and function. Nat. Rev. Rheumatol.,5 373 381 1759-4790 - 44.
Paluzzi A. Belli A. Bain P. Viva L. 2007 Brain ‘imaging’ in the Renaissance. J. R. Soc. Med.,100 540 543 0141-0768 - 45.
Peyrin F. 2011 Evaluation of bone scaffolds by micro-CT. Osteoporos. Int.,20 2043 2048 0093-7941 X. - 46.
Prosecká E. Rampichová M. Vojtová L. Tvrdík D. Melčáková Š. Juhasová J. Plencner M. Jakubová R. Jančář J. Nečas A. Kochová P. Klepáček K. Tonar Z. Amler Z. 2011 Optimized conditions for mesenchymal stem cells to differentiate into osteoblasts on a collagen/hydroxyapatite matrix. J. Biomed. Mater. Res. A,99A 307 315 1552-4965 - 47.
Razavi H. Dusch M. N. Zarafshar S. Y. Taylor C. A. Feinstein J. A. 2012 A method for quantitative characterization of growth in 3-D structure of rat pulmonary arteries. Microvasc. Res.,83 146 153 0026-2862 - 48.
Renghini C. Komlev V. Fiori F. Verné E. Baino F. Vitale-Brovarone C. 2009 Micro-CT studies on 3-D bioactive glass-ceramic scaffolds for bone regeneration. Acta Biomater.,5 1328 1337 1742-7061 - 49.
Ritman E. L. 2011 Current status of developments and application of micro-CT. Annu. Rev. Biomed. Eng.,13 531 552 1523-9829 - 50.
Roth H. Neubrand T. Mayer T. 2010 Improved inspection of miniaturised interconnections by digital X-ray inspection and computed tomography, Proceedings of Electronics Packaging Technology Conference (EPTC),978-1-42448-560-4 Singapore, December 2010. - 51.
Rüegsegger P. 2001 Imaging of bone structure, In: Bone Mechanics Handbook, 2nd ed., Cowin S.C. (Ed.),9 CRC Press,978-0-84939-117-0 Boca Raton. - 52.
Savai R. Langheinrich A. C. Schermuly R. T. Pullamsetti S. S. Dumitrascu R. Traupe H. Rau W. S. Seeger W. Grimminger F. Banat G. A. 2009 Evaluation of angiogenesis using micro-computed tomography in a xenograft mouse model of lung cancer. Neoplasia,11 48 56 1522-8002 - 53.
Schambach S. J. Bag S. Schilling L. Groden Ch. Brockmann M. A. 2010 Application of micro-CT in small animal paging. Methods,50 2 13 1046-2023 - 54.
Schladitz K. 2011 Quantitative micro-Ct. J. Microscopy,243 111 117 1365-2818 - 55.
Sonka M. Hlavac V. Boyle R. 1998 Image processing, Analysis, and Machine Vision. PWS Publishing,123 134 053495393 - 56.
Sterio D. C. 1984 The unbiased estimation of number and sizes of arbitrary particles using the disector. J. Microsc.,134 127 136 1365-2818 - 57.
Tjia J. S. Moghe P. V. 1998 Analysis of 3-D microstructure of porous poly(lactide-glycolide) matrices using confocal microscopy. J. Biomed. Mater. Res.,43 291 299 1552-4965 - 58.
Tonar Z. Khadang I. Fiala P. Nedorost L. Kochová P. 2011 Quantification of compact bone microporosities in the basal and alveolar portions of the human mandible using osteocyte lacunar density and area fraction of vascular canals. Ann Anat.,193 211 219 0940-9602 - 59.
Villa C. Erratico S. Razini P. Fiori F. Rustichelli F. Torrente Y. Belicchi M. 2010 Stem cell tracking by nanotechnologies. Int J. Mol. Sci.,11 1070 1081 1422-0067 - 60.
Zagorchev L. Oses P. Zhuang Z. Moodie K. Mulligan-Kehoe M. J. Simons M. Couffinhal T. 2010 Micro computed tomography for vascular exploration. J. Angiogen. Res.,2 7 18 e-2040-2384