Abstract

The development of electrochemical sensors has attracted great interest due to these sensors’ high sensitivity and selectivity. Here, we present the general concept and the classification of biosensors, their advantages and drawbacks, the main strategies in electrochemical biosensor technology and the materials used in electrochemical sensors, such as electrodes and supporting substrates, materials for improved sensitivity and selectivity, materials for bioreceptor immobilization, and biological recognition elements. Various nanomaterials, such as carbon-based materials (carbon nanotubes, graphene, carbon nanoparticles), inorganic and organic nanoparticles (magnetic and metal nanoparticles, nanosized clays), conductive and insulating polymers (nanosized and nanostructured polymers, molecularly imprinted polymers), and hybrid materials, etc., have been successfully applied for the enhancement of the electroanalytical performance of biosensors and for the immobilization of biorecognition elements. Among these, due to their unique physiochemical features, carbon-based materials, such as carbon nanotubes and graphenes, have received special attention in recent years, and examples of surface functionalization using various types of nanoparticles are presented. The future trends in sensor research activities and areas of development that are expected to have an impact in biosensor performance, like immobilization techniques, nanotechnology, miniaturization and multisensor array determinations, are also examined.

Keywords

  • Carbon-based nanomaterials
  • metal nanoparticles
  • magnetic nanoparticles
  • nanostructure
  • molecularly imprinted polymers

1. Introduction

The development of electrochemical sensors has attracted a great deal of interest due to their high sensitivity and selectivity, and they are being increasingly used in many fields, such as analytical chemistry, industrial process monitoring and control, clinical diagnostics, environmental monitoring and security, and food safety. By adding an enzyme at the electrode surface, they can also gain high specificity: thus the biosensor research area was born.

Biosensors (analytical devices coupling a transducer with a biorecognition element) are attractive for pharmaceutical and biomedical analysis due to their sensitivity (ng/mL or less) and high selectivity, and sometimes their specificity, high benefit/cost ratio, simple use and rapidity of data collection [1].

The major advantages of biosensors over traditional analytical methods, which will certainly lead in the near future to their even more pronounced use in the biomedical field, are: the fact that analyte detection can very often be made without prior separation; the short response times that make possible the real-time monitoring of biological and manufacturing processes; their ease of use, allowing in-field or point-of-care measurements; the flexibility and simplicity of preparation; the possibility of mass production and low production costs; and the possibility of miniaturization and automatization. Miniaturization is of great importance because biological samples are available in small amounts, and tissue damage must be minimized in cases of in-vivo monitoring. Therefore, the use of biosensors as components of modern medical devices has improved their portability, functionality and reliability for point-of-care analysis and real-time diagnosis [2].

However, many amperometric biosensors described in the recent literature still display a few drawbacks compared to other analytical methods. The most difficult problems to overcome for biosensors with biomedical applications are: the reduced stability, the electrochemical interferences, and the lack of or low response reproducibility. Removed from their natural environment, most biocomponents tend to rapidly lose their activity and thus limit the lifetime of the sensor. Modifying electrode surfaces in a way that exclusively favours one single electrochemical process is a difficult task and sample matrices in the biomedical field are very complex. Last but not least, in the case of in-vivo measurements biocompatibility and biofouling are critical issues [3].

The selectivity of the biosensor for the target analyte is mainly determined by the biorecognition element, whilst the sensitivity of the biosensor is greatly influenced by the transducer.

New sensing elements like genetically transformed cells and genetically engineered receptor molecules can improve affinity and specificity; therefore, genetic engineering and mass production of the molecular recognition may ultimately dictate the success of biosensor technologies. Enzyme variants that are specific for individual analytes have already been obtained by genetic modification, and novel gene fusions will lead to more sensitive biosensors [4].

As for the electrochemical transducer, important advances have been recently made thanks to the introduction of new platforms for biosensor design, such as nanotechnological materials and nanostructured architectures (i.e., nanoparticles, carbon nanotubes and nanofibres, graphenes, nanostructured surfaces, etc.), which have improved the sensitivity of the assembly [5].

Advertisement

2. Strategies of electrochemical biosensor technology

The biosensor development and construction strategy includes five features: 1) the detected or measured parameter and the matrix, 2) the working principle of the transducer, 3) the chemical/biochemical model, 4) the field of application and 5) the technology and materials for sensor fabrication [6].

Materials generally used for electrochemical sensors are classified as: (1) materials for the electrode and supporting substrate, (2) materials for improving electroanalytical performances, (3) materials for the immobilization of biological recognition elements and (4) biological elements; the last two are applicable for electrochemical biosensors [5].

Materials used for the electrode and supporting substrate are usually conductive materials exhibiting low currents in an electrolyte solution, free of any electroactive species, over a relatively wide potential window. Among the most frequently used materials for the electrode and supporting substrate, the following may be mentioned: metals (mercury, platinum, gold, silver and stainless steel), metal oxides (indium tin oxide, ITO) carbon-based materials (glassy carbon, graphite, carbon black and carbon fibre), new hybrid materials, and organic electroconductive polymers or salts. Boron-doped diamond is a special case of material, because bare diamond is a non-conducting, allotropic form of carbon. In the last decade, in order to avoid mercury-based electrodes (which are highly toxic), different metallic films (Au, Hg, Bi, Ag, Pb, Sn, Sb, Co, Ga, Se) that are relatively simple to achieve and easy to reproduce were developed. Some strategies to develop two- and three-dimensional nanostructured electrodes with larger surfaces (mesoporous silicates, metal oxides, polymers or carbons) were also recently reported [7]. The major challenge in the field of electrochemical sensors, consisting in the improvement of their electroanalytical performances, mainly in terms of sensitivity and selectivity, was addressed by the development of nanotechnology (especially nanoparticles, carbon nanotubes and graphenes) and nanostructured architectures, and this will be detailed in section 3.

The stable immobilization of a bioelement on an electrode surface, with complete retention of its biological activity and good diffusion properties for substrates, is a crucial problem for the commercial development of biosensors. A successful matrix should stably immobilize or integrate biomolecules at the transducer’s surface, it should efficiently maintain their functionality, while providing accessibility for the target analyte, and it should also ensure an intimate contact with the transducer surface [8]. The immobilization matrix may function purely as a support or may also be involved in the mechanism of signal transduction mediation. The selection of an appropriate immobilization method depends on the nature of the biological element, the type of transducer used, the physicochemical properties of the analyte and the biosensor’s operating conditions [9]. The choice of materials for the immobilization of biological recognition elements strongly depends on the employed procedure. The procedures that may be employed are as follows: 1) Entrapment behind a permeable membrane as a thin film covering the electrochemical detector (dialysis membrane) [10]; 2) Entrapment within a polymeric matrix (polyacrylonitrile, agar gel, polyurethane or poly(vinyl-alcohol) membranes, sol gels or redox hydrogels with redox centres such as [Os(bpy)2Cl]+, conducting polymers [11]; 3) Entrapment within self-assembled monolayers (SAMs) [12] or bilayer lipid membranes (BLMs) [13]; 4) Covalent bonding on membranes or surfaces activated by means of bifunctional groups or spacers (glutaraldehyde, carbodiimide, SAMs or multilayers, avidin-biotin, silanization) [14]; 5) Bulk modification of the electrode material (graphite epoxy resin or carbon paste modified with polyoxometallates, clays and double layered hydroxides, zeolites, functionalized silica, sol-gel-derived inorganic and hybrid materials, nanomaterials, sparingly soluble or insoluble inorganic salts, molecular and macrocyclic ligands, complex compounds, organic polymers, surfactants and lipids) [7].

The biorecognition elements used for biosensor development are classified into biological and artificial (biomimetic) receptors. They are selected depending on the application and the desired performance criteria, based on the following interaction principles: enzyme/substrate, antigen/antibody, mRNA/DNA, PNA/DNA or RNA, microorganism/substrate or toxic molecules, cell receptor/hormone and (semi)synthetic receptor/target molecule [15]. Initially, three types of biorecognition elements were identified [16]: 1) Enzyme (mono or multienzyme), the most common and best developed recognition system, including antibodies, natural receptors and nucleic acids; 2) Whole cells (microorganisms, such as bacteria, fungi, eukaryotic cells or yeast), or cell organelles or particles (mitochondria, cell walls); 3) Tissue (plant or animal tissue slice). More recently, the range of biochemical receptors was extended with biomimetic recognition elements (aptamers, ribosymes and molecularly imprinted polymers) [17]. The biological recognition elements are based either on catalytic or bioaffinity features, whilst biomimetic ones are based on biocomplexing and bioaffinity.

The biocatalytic recognition sensors are based on a reaction catalysed by macromolecules, either in their original biological environment, in their isolated form, or as a product of semisynthesis. Thus, a continuous consumption of substrate(s) is achieved by the immobilized biocatalyst incorporated into the sensor, and transient or steady-state responses are monitored by the integrated detector. In the construction of biosensors, enzymes are the most commonly used biological elements, despite their rather high extraction, isolation and purification costs, as they rapidly and cleanly form selective bonds with the substrate. Among the enzymes commercially available the most commonly used are the oxidases, such as glucose oxidase (GOx) and horseradish peroxidase (HRP), beta-lactamase, urease, tyrosinase, and acetylcholine esterase or choline oxidase.

The biocomplexing or bioaffinity sensors are based on antibodies, natural biological chemoreceptors, nucleic acids, aptamers or molecularly imprinted polymers (MIPs). These biosensors provide selective interactions of the analyte with a given ligand to form a thermodynamically stable complex. The analyte interacts with macromolecules or organized molecular assemblies that have either been isolated from their original biological environment or engineered. Thus, equilibrium is usually reached and there is no further consumption of the analyte by the immobilized biocomplexing agent. These equilibrium responses are either monitored by the integrated detector or by using a complementary biocatalytic reaction. The most developed examples of biosensors using biocomplexing receptors are based on immunochemical reactions, i.e., the binding of an antigen (Ag) to a specific antibody (Ab), and will be discussed in detail in the chapter concerning immunosensors.

The biological chemoreceptors [18] are used for drug and neurotransmitter detection, and they are based on an affinity reaction. Using nucleic acids as a biocomponent, one can detect the interaction between the synthetic drug and the genetic material and also the hybridization processes by working with single-strand DNA [19, 20].

The use of microorganisms and of whole vegetal or animal tissues is complex and very often their selectivity is limited [21].

Advertisement

3. Nanomaterials involved in biosensor construction

Advances in nanotechnology have led to the recent development of many nanomaterials, including carbon nanomaterials, magnetic and metallic nanoparticles [22]. Nanomaterials, generally defined as materials with feature sizes smaller than 100 nm, have a remarkable impact on various fields of application due to their properties (enhanced electrical conductivity, tensile strength and chemical reactivity), which are imparted by their increased surface area per unit weight. Nanomaterials have already been applied in electronics, foods, cosmetics, electronics, drug development and sensing devices. The use of nanomaterials, especially nanoparticles and nanostructured films, offers advantageous properties that can be exploited in order to maximize the interactions with specific bioelements, to maintain their activity, minimize structural changes, and enhance the catalytic step. In the biosensor field, the analytical exploitation of such protein-nanomaterial interactions are an emerging trend that spans many disciplines [23]. Nanomaterials based on metals, semiconductors and organic compounds contribute to the enhancement of optical, electrical, chemical and magnetic properties that are relevant in the case of sensing devices, and due to these facts they have been frequently used in this research field [24, 25].

Nanomaterials are mainly used for electrode construction or modification and as biomolecule tracers. Nanoparticles (NPs) are very stable (compared to enzyme labels), offering high sensitivity (thousands of atoms can be released from one nanoparticle) and a wide variety are available on the market. Nanoparticles are used nowadays as electrochemical labels or as vehicles containing several hundreds or thousands of electroactive labels, pushing detection limits down to several hundreds of biomolecules [26].

Various nanomaterials have been successfully applied for the immobilization of bioelements, such as metallic nanoparticles, carbon or metallic nanotubes, nanosilver-coated magnetic beads, magnetic particles, functionalized conductive polymers, etc.

3.1. Carbon-based materials

Different types of carbon nanostructures are becoming more and more popular due to their specific structures, properties and the possibility of being used for many applications. In fact, a wide variety of carbon-based materials are available, such as nanoparticles, nanodiamonds, nano-onions, peapods, nanofibers, nanorings, fullerenes and nanotubes, which have been extensively used in analytical applications. The basic structure of fullerenes and nanotubes consists of a layer of sp2-bonded carbon atoms. This configuration, which resembles that of graphene, is responsible for their good electrical conductivity and their ability to form charge-transfer complexes when in contact with electron donor groups [27]. This configuration is responsible for the development of strong van der Waals forces that significantly hamper the solubility and dispersion of carbon-based nanoparticles. In order to avoid these problems, different pretreatment methods have been proposed [28, 29, 30], such as the addition of polar groups (oxygen, hydroxyl, phenyl and polyvinylpyrrolidone). The surface defects could also affect the stability, the mechanical and electrical properties of carbon nanostructures [31, 32, 33, 34].

The carbon nanomaterials (Figure 1) cover a broad range of structures, beginning with zero-dimensional structures (fullerenes, diamond clusters), continuing with one-dimensional (nanotubes), two-dimensional (graphene), and three-dimensional structures (nanocrystalline diamond, fullerite).

Figure 1.

Main carbon entities at nanoscale level

Carbon nanomaterials, including fullerenes, graphene and carbon nanotubes, have many technological advantages such as facile modification by functional groups, high carrier capacity, incorporating both hydrophilic and hydrophobic substances, and high chemical stability.

3.1.1. Carbon nanotubes

Within the family of nanomaterials, carbon nanotubes (CNTs), described for the first time by Iijima [35], are arousing growing interest in relation to several modern devices, such as transducers, for the improvement of biosensing systems, mainly due to their unique structural, electronic and chemical properties, which include large surface area, high electrical conductivity, and excellent properties of electron transfer reactions. Furthermore, CNTs have a unique tubular structure, good biocompatibility and modifiable sidewall, making them ideal candidates for the construction of sensors with high performances [36].

Electrochemical nanobiosensors have been increasingly used in various types of nanomaterials, mainly CNTs, as electrochemical transducers, in order to improve them from the perspectives of automation, miniaturization and multiplexed analysis, but also to improve the analytical performances of such devices [37, 38]. For example, the detection of diseases at an early stage is one of the goals in developing and improving CNT biosensors, focusing on their sensitivity, fast response and small sample volume, in order to provide new strategies for the detection of specific biomarkers at low concentrations in complex sample media (e.g., serum) [39].

Many properties and the quality of CNTs are directly influenced by the way in which the graphene sheets are wrapped around [40], and they are affected by the operating conditions in their fabrication process. The production methods generally require transition metal nanoparticles (i.e., Fe, Co and Ni) to function as growth catalysts. When the catalysts are in contact with a gaseous carbon or a hydrocarbon source, the resulting carbon is deposited on the particle surface and the CNT grows rapidly at the surface of the catalyst. The diameter of the CNTs, their quantity and quality can be controlled by varying some reaction parameters (e.g., temperature, metal concentration, gas type and pressure). The diameter of the NPs defines the nature of the CNTs, whether single-walled (SWCNTs) with diameters in the range 0.4-3 nm, or a group of concentric CNTs sharing a common axis constituting double- and multiwalled (MWCNTs) with diameters in the range 1.4-100 nm [36].

In the case of CNT structures, the reactivity of atoms situated in the plane is different from those at the edges. For a number of biologically important compounds, the electrochemistry of atoms located at the edge-plane sites of the CNT is comparable to that of different planes of graphite [41], and the metal and non-metal impurities contained determine some of the electrochemical catalytic properties and limitations of CNTs [42, 43, 44]. It is also worth mentioning that the electrode surface modification with CNTs leads to a changed mass transport regime from semi-infinite diffusion to effective thin-layer diffusion in a porous layer, and the increased current may frequently be erroneously assigned to electrocatalysis [45, 46, 47].

Currently, SWCNTs and MWCNTs are synthesized by high-temperature processes (e.g., electric arc discharge and laser ablation, which are physical methods to convert carbon into NTs) and chemical vapour deposition (CVD), which is performed at lower temperatures and atmospheric pressure.

Usually, the nanostructured materials obtained using CNTs were morphologically characterized using transmission electron microscopy (TEM) and scanning electron microscopy (SEM). Figure 2 presents the SEM images obtained for SWCNT [48] and MWCNT [49].

Figure 2.

(A) SEM image of SWCNT composite film [48]; (B) FE-SEM images of MWCNT in high magnification [49]

Many papers published in recent years report CNT biosensors based on different oxidases applied for the detection of glucose [50, 51, 52, 53, 54], H2O2 [55, 56, 57], L-lactate [58, 59, 60], acetaminophen [61, 62, 63, 64, 65], etc. The construction and optimization of new biosensors with HRP immobilized onto the transducer surface, either screen-printed electrode (SPE) or glassy carbon electrode (GCE), by entrapment into polymer films (polyethylenimine (PEI) and polypyrrole (Ppy)) doped with SWCNT and MWCNT, was described [61, 66].

The obtained configurations were used to monitor the signal produced by the electrochemical reduction of the enzymatically generated electroactive N-acetylbenzoquinoneimine (the oxidized form of acetaminophen) in the presence of hydrogen peroxide using amperometry, cyclic voltammetry (CV) and differential pulse voltammetry (DPV) in synthetic and real samples, with the best limit of detection (LOD) obtained for acetaminophen of 1.36 μM [61].

3.1.2. Graphene

Graphene is another material that offers great prospects for the future of analytical chemistry, consisting of a one-atom-thick sheet of sp2 hybridized carbon atom composed of six member rings that provide a surface area that is nearly twice as large as that of SWCNT [43]. This material presents high mechanical strength, high elasticity and high thermal conductivity [43], but also presents a major disadvantage: its poor dispersion in aqueous medium. Stable dispersions of graphene sheets have been achieved using amphiphilic polymers, alkylamines and hydrophilic carboxyl groups, among others, as dispersing agents [67]. The functionalization of graphene is considered an important route for improving its dispersion. Therefore, different authors have tried producing graphene oxide, a graphite derivative with hydroxyl, epoxy and carboxyl groups covalently attached to its layers, with a better dispersion in some solvents being obtained.

Graphene can be interlinked with CNT for the fabrication of high-performance transparent electrodes, and the resulting films present features comparable to indium tin oxide [68]. The electrochemical activity of crystalline graphene is different from that of reduced graphene oxide flakes [69]. A wide variety of methods are available for graphene fabrication, the first being published in 2004 [70], and consist in the mechanical exfoliation (repeated peeling) of small patches of highly ordered pyrolytic graphite. Many other fabrication methods have been developed, including unzipping MWCNT to form graphene ribbons [71, 72], substrate-independent methods using micromoulding inside a capillary [73], and spray deposition of graphene oxide-hydrazine dispersions [74].

SEM images of a graphene film deposited on GCE revealed the typical crumpled and wrinkled graphene sheet structure on the rough surface of the film (Figure 3 A). The TEM image presented in Figure 3B shows the wrinkled graphene sheet with no aggregation, indicating that the functionalized graphene sheets were well dispersed in ethanol, forming stable suspensions [75].

Figure 3.

(A) SEM image of graphene-film-modified GCE. (B) TEM image of graphene in ethanol [75]

With regard to aptasensor development, graphene looks like it could have the necessary requirements to implement next-generation and high-performance aptasensors. Graphene naturally adsorbs the unfolded aptamer through the π–π stacking interaction between the purine and pyrimidine bases and the graphene plane, and hinders the adsorption of the folded aptamer [76, 77].

The currently developed strategies for graphene functionalization with aptamers include covalent and non-covalent approaches. The covalent approach often relies on the oxidation or the reduction of graphene oxide that presents carbonyl and carboxyl functional groups [78]. The density of carbonyl and carboxyl functional groups can be increased through chemical oxidation, which is widely used for the chemical modification of the inert graphitic structure [79]. Because this procedure affects the properties of graphene, the covalent modification of graphitic surface through a free radical addition reaction is used, without a preoxidation step and with less impact on the electronic properties [80]. Besides the immobilization methods mentioned above, amide [81] and phosphoramidate reactions [82] were employed towards the fabrication of graphene-based biosensors. Some graphene derivatives, such as graphene oxide, were also used as transducers in electrochemical biosensor development [83].

Graphene can also mediate electron transfer in view of the high-density edge-plane-like defects present within its nanostructure. In the absence of a target, aptamers immobilized on the gold electrode adsorb graphene nanosheets due to the strong π–π interaction that accelerates the electron transfer ratio between the electroactive species and the electrode surface. In the presence of a target, the binding reaction inhibits the adsorption of graphene and blocks the electron transfer. A label-free electrochemical aptasensor was constructed by taking advantage of the ultra-fast electron transfer ratio of graphene [84]. This detection strategy was further improved by nuclease for interferon gamma, yielding an LOD of 0.065 pM [85]. Other authors [86] employed graphene oxide nanoplatelets (GONPs) as electroactive labels for thrombin detection.

A simple label-free electrochemical immunosensor was constructed by modifying a graphite-based SPE with graphene oxide after functionalization with N-hydroxysuccinimide in the presence of 1-ethyl-3-(3-dimethyl aminopropyl) carbodiimide hydrochloride, Figure 4.

Figure 4.

Steps involved in the antiacetaminophen Ab-based immunosensor development; GO=Graphene oxide; NHS=N-hydroxysuccinimide; EDAC=1-ethyl-3-(3-dimethyl aminopropyl) carbodiimide hydrochloride; Ab anti APAP=antibody antiacetaminophen; APAP=acetaminophen; SWV=square wave voltammetry [87]

The above-described immunosensor was applied with good results for the determination of acetaminophen in synthetic and real samples by using square wave voltammetry (SWV) with an LOD of 0.17 μM [87].

3.1.3. Carbon nanoparticle-based electrochemical sensor

A sensitive and selective electrochemical sensor fabricated via the drop-casting of a suspension of carbon nanoparticles (CNPs) onto a GCE was investigated in simultaneous determination of acetaminophen and tramadol in pharmaceuticals and human plasma. CV and DPV studies of acetaminophen and tramadol were carried out at the modified and bare GCE. The results of the electrochemical investigations showed that CNPs enhanced the electroactive surface area and caused a remarkable increase in the peak currents, due to the diffusion within the porous layer. Enhanced sensitivity and a considerable decrease in the anodic overpotential leading to negative shifts in peak potentials was obtained with the modified GCE. By using DPV, the sensor showed good sensitivity and selectivity: LODs for acetaminophen and tramadol were 0.05 and 1 μM, respectively [88].

3.2. Inorganic and organic nanoparticles

Nanoparticles of major environmental, pharmaceutical and biomedical importance include magnetic nanoparticles, quantum dots, metal nanoparticles, silica nanoparticles and polymeric types with intrinsic properties contributing to their use in specific applications. Typically, nanoparticles are classified into: magnetic nanoparticles (possessing a core of Fe3O4) frequently used in drug delivery and lately in immunosensing; semiconductor/QDs (based on CdS, CdSe) used in bio- and immunosensing; metal nanoparticles (Ag, Au, Pd) used in biosensing and drug delivery; polymeric nanoparticles (polystyrene) mainly used in drug and gene delivery; hybrid nanoparticles (carbon or Fe3O4 covered with different materials like metal oxides, polymers, amino acids, etc.) [89].

3.2.1. Magnetic nanoparticles

In recent years, considerable efforts were made to develop magnetic nanoparticles (MNPs), due to their inherent advantages (magnetism, nanosize) and low cost of production. MNPs with 10-20 nm diameters exhibit their best performance, due to supermagnetism, which makes them especially suitable for a fast response. Due to their inherent dimensions, they possess a large surface area and high mass transfer capacity [90].

MNPs can be integrated into the transducer materials and/or be dispersed in the sample followed by their attraction by an external magnetic field onto the active detection surface of the (bio)sensor. Biorecognition molecules like enzymes, antibodies or oligonucleotides immobilized onto magnetic particles can be easily trapped by magnets and retained close to or on an electrode surface [90, 91]. Since the properties of MNPs depend strongly on their dimensions, their synthesis and preparation have to be designed in order to obtain particles with adequate size-dependent physicochemical properties. MNPs possessing adequate physicochemistry and tailored surface properties have been synthesized under precise conditions and applied for sensors, biosensors and other detection systems [92, 93].

Magnetic particles consist of a magnetic core surrounded by a non-magnetic shell. Iron oxides, such as magnetite (Fe3O4) or maghemite (γ-Fe2O3), are often used as core materials instead of iron due to their higher stability [94]. For biological applications, magnetic particles generally consist of nanosized superparamagnetic iron oxide distributed through a matrix, mainly composed of silica [95] or an organic polymer [96]. Silica-based magnetic beads can be prepared in different shapes and sizes with different degrees of porosity, which are much more attractive for the immobilization of biomolecules due to their chemical and mechanical stability and their resistance to bacterial attack [97]. Many types of magnetic particles, functionalized with different groups like carboxyl, amino, hydroxyl or epoxy, are currently available for different applications.

The majority of MNP-based systems use inorganic nanocrystals as magnetic cores. The composition of these inorganic nanocrystal ranges from metals and alloys to metal oxides. Among the metal oxide compounds, superparamagnetic iron oxide nanoparticles, including magnetite (Fe3O4) and maghemite (γ-Fe2O3), are the most popular [98]. The surfaces of the particles can be modified with organic polymer or inorganic compounds (e.g., gold, silica, alumina), suitable for further functionalization by the attachment of various bioactive molecules [99]. The functionalization of MNPs with bioelements exhibits specificity, due to both the bioreceptor and the magnetism.

MNPs amplify the electrochemical signal, improving the sensitivity of electrochemical devices [90], through their contact with the electrode surface, transport of a redox-active species to the electrode surface, and formation of a thin film on the electrode surface. The detection for MNP-based electrochemical biosensors can be performed by potentiometry [100, 101], amperometry [102, 103], electrochemiluminescence [104, 105], electrochemical impedance spectroscopy [106, 107] and especially voltammetry [108, 109, 110, 111, 112, 113, 114, 115].

Protein G-MNP-based label-free electrochemical immunosensor was elaborated for the ultrasensitive and specific detection of acetaminophen using a carbon-based SPE as immobilization platform (Figure 5). An LOD of 1.76 μM was calculated based on the calibration data. The performances of the optimized immunoassay were tested using two pharmaceutical products containing acetaminophen with excellent recoveries [116].

Figure 5.

Protein G-MNP-based label-free electrochemical immunosensor development [116]

Protein G-functionalized magnetic beads and graphite-based SPEs were also used to develop an HRP-labelled sandwich type immunosensor for Mucin 1 (MUC1) electrochemical detection with LOD of 1.34 ppb [117].

Fe3O4 magnetic dipolar attraction and its large surface area/volume ratio may lead to aggregation in clusters when exposed to biological solutions. Functionalization can overcome this problem and also enhance biocompatibility. Many strategies were used for the functionalization of MNPs, such as core-shell Au-Fe3O4 [109], core-shell Au-Fe3O4@SiO2 [102], core-shell Fe3O4@SiO2 [112], Au-Fe3O4composite NPs [108], Fe3O4@SiO2/MWCNTs [103], Fe3O4 anchored on reduced graphene oxide [113] and Fe3O4@ Au-MWCNT-chitosan [114]. Among all these, core-shell Fe3O4@SiO2 contributes to the stabilization of MNPs in solution and enhances the binding of ligands at the surface of MNPs. Electrode surface modified with core-shell Fe3O4@SiO2 shows good electrical conductivity and more electroactive interaction sites, providing enhanced mass transport and easier accessibility to the active sites, thus increasing the analytical signal and sensitivity [90]. A comparison between the analytical performances of some recently developed electrochemical sensors and biosensors based on MNPs is presented in Table 1.

Detection method MNPs type Analyte LOD Reference
Voltammetry Au-Fe3O4 Organochloride pesticides 56 pg/mL [108]
Voltammetry Core-shell Au-Fe3O4 Carcinoembryonic Ag 10 ng/mL [109]
Voltammetry Fe3O4 Au NP Clenbuterol 220 ng/mL [110]
Voltammetry Fe3O4 Au NP H2O2 20 μM [111]
Voltammetry Core-shell Fe3O4@SiO2 Metronidazole 18 nM [112]
Voltammetry Fe3O4-RGO Cr(III) - [113]
Voltammetry Fe3O4@Au-MWCNT-chitosan Streptomycin 1.5 nM [114]
Voltammetry Core-shell Fe3O4@SiO2/MWCNT Uric acid 0.13 μM [115]
Potentiometry MaMB Dynabeads Protein G Zearalenone 7 ng/mL [100]
Potentiometry Core-shell-Fe3O4 Glucose 0.5 μM [101]
Amperommetry Core-shell Au-Fe3O4 @SiO2 Glucose 0.01 mM [102]
Amperommetry Fe3O4@SiO2-MWCNT Glucose 800 nM [118]
EIS Fe3O4-COOH/MNP OchratoxinA 10 pg/mL [106]
EIS Fe@Au NP-2-aminoethanethiol
- graphene NP
DNA 2 fM [107]
Detection method MNPs type Analyte LOD Reference

Table 1.

Selected examples of sensors and biosensors based on magnetic nanoparticles

3.2.2. Metal nanoparticles

Metal nanoparticles and CNTs, have been employed to immobilize biomolecules for the construction of sol-gel biosensors, due to their large surface area, high catalytic activity and good biocompatibility. Moreover, the excellent conductivity of gold nanoparticles (AuNPs) provides the potential to construct amperometric biosensors based on redox enzymes (as sensing elements) and nanoparticle arrays (as conductive matrices where enzyme molecules are immobilized) [118, 119]. Metal nanoparticles, such as Pd and Pt, were deposited on SWCNTs and MWCNTs in order to increase the electrocatalytic activity of CNTs. An improved conductivity and electrocatalytic behaviour was obtained by combining sol-gel materials with CNTs, the electron transfer between the redox enzyme and the electrode being facilitated [120].

Gold nanomaterials have unusual optical and electronic properties, high stability and biological compatibility. Aptamer-conjugated gold nanomaterials provide powerful platforms to facilitate targeted recognition, detection and therapy applications, due to their controllable morphology, size dispersion and easy surface functionalization [121].

Many applications of nanoparticles, such as metal NPs (gold, silver) or semiconductor dots in biosensing, have been recently reported [122, 123]. A DNA array detection system based on oligonucleotide targets labelled with AuNPs was developed [124], and AuNPs or silver-enhanced colloidal gold were used as labels for electrochemical detection of DNA hybridization [125, 126, 127]. These NP-based electrochemical biosensors employed the stripping voltammetric technique as the read-out principle due to the preconcentration step resulting in ultra-trace-level LODs.

AuNPs are irreversibly immobilized on the transducer, the electrode surface being difficult to regenerate. To overcome this problem, Chen et al. [128] have developed a novel electrochemical system for the sensitive detection of glucose (LOD of 1 mM) as well as single-nucleotide polymorphism, which has the AuNPs directly in the supporting electrolyte. This one-pot detection system can be operated and regenerated very easily, since all the components are integrated in the electrolyte of the gold colloid and the unmodified electrode can be reused.

The sensing mechanism of potentiometric biosensors using NPs is based on a sandwich assay (Figure 6), the target being captured by the immobilized primary receptor and then attached to the NP-labelled secondary receptor. Potentiometry is finally used to detect the dissolution products of NPs after a release step with H2O2 [123].

Figure 6.

Potentiometric biosensors using nanoparticles as labels [123]

Miniaturized ion-selective electrodes (ISEs) designed to detect ions in microvolume samples, with a concentration of fM, were reported [129, 130], and an LOD of 12.5 pmol/50μL was obtained for IgG using an ISE based on silver-enhanced AuNPs [131]. Metal NPs can also be used as a catalyst for the reduction of metal ion, via the enzymatically generated reducing agent, and the biometallization processes can be monitored by ISEs in real time [132]. Thereby, an ISE based on AuNPs was developed and applied for highly sensitive detection of glucose [132, 133], while the NADH potentiometric detection was successfully obtained by monitoring the NADH-stimulated catalytic reduction of Cu2+ in the presence of AuNPs (which have a crucial role in improving the sensitivity) with a Cu2+-ISE [134]. Wang et al. [135] reported a highly sensitive potentiometric biosensor for detecting DNA hybridization based on the depletion of Ag+ ions induced by the biocatalytic reaction of alkaline phosphatase, recorded in the presence of AuNPs. The sensor has been used to detect DNA and 16S rRNA of Escherichia coli with LOD of 50 fM, using only 4 μL of sample solution.

Two simple electrochemical aptasensors were developed for the detection of Mucin 1 (MUC1) tumour marker based on its specific recognition by thiolated aptamers immobilized on Au NP-modified graphite and gold SPEs. The quantitative detection of MUC1 protein was electrochemically achieved by electrochemical impedance spectroscopy (EIS) and DPV. The estimated detection limits for MUC1 were 3.6 ng/mL at AuNP modified graphite SPE by EIS and 0.95 ng/mL at AuNP modified gold SPE by DPV [136].

3.2.3. Nanosized clays

Nanosized clays can serve as matrices for electroactive species because they are usually able to incorporate ions by an ion exchange process. Moreover, adsorption of proteins on clay mineral surfaces plays a very important role in the development of biosensors. By modifying the electrode surface with clays, the heterogeneous electron transfer process between the protein and the electrode surface is facilitated.

The entrapment of biomolecules in the clay matrix represents an inexpensive, fast and easy method for the elaboration of enzyme electrodes. The procedure consisting in the adsorption of an enzyme/clay aqueous colloid mixture onto the electrode surface offered the possibility of entrapping a large and known amount of enzyme into the clay film, and could also be adapted for interdigitated microelectrodes [137].

Mesoporous silica nanoparticles can be easily synthesized by using surfactant templating methods [138]. Silica nanoparticles have become an important platform in biomedical applications and bioanalysis because of their high density, which facilitates separation through centrifugation during their synthesis, and their modification and detection steps [139]. The silica surface also provides a biocompatible and versatile substrate for biomolecule immobilization, required for further utilization in biosystems and for drug delivery [140]. Following typical DNA silica surface conjugate chemistry, the aptamer immobilization on silica nanoparticles has been well developed for various applications [141].

Biosensors based on clay materials can play an important role as efficient tools in the field of forensic analysis considering their well-known advantages of sensitivity, selectivity, simple functioning and affordability [142]. Thereby, a bienzyme biosensor based on acid phosphatase (AcP) and polyphenol oxidase (PPO) simultaneously entrapped in anionic clays was constructed and applied for the specific amperometric determination of As(V). The biosensor relied on the successive hydrolysis of phenyl phosphate to phenol catalysed by AcP and the oxidation of phenol to o-quinone catalysed by PPO. The electrochemical reduction of the formed o-quinone was accomplished at a detection potential of -0.2 V vs Ag/AgCl with a detection limit of 0.15 ng/mL [143]. DNA-functionalized, layer-by-layer assembled SWCNT hybrids with clay nanoparticles were proposed for the detection of As(III) achieving an LOD of 0.05 mg L-1 [144]. Another simple amperometric biosensor for cyanide involved the immobilization of PPO into ion exchanger clay based on Zn-Al double-layered hydroxides (LOD of 0.1 nM). The low LOD was a consequence of the anion accumulation in the ion exchanger clay which also contained the immobilized enzyme [145].

3.2.4. Nanosized and nanostructured polymeric films

One of the most important aspects of biosensor development is the bioelement immobilization platform, which needs to ensure high stability and maintain its activity as long as possible. Several types of conductive or insulating polymers were used to obtain a nano/micropatterned surface with applications in the biosensor field [146]. Electropolymerization in the presence of a template, removed after the polymer formation, is an efficient route for the synthesis of 2D and 3D conductive polymers (CPs). Among the conductive polymers, polypyrrole and polyaniline [147, 148] were extensively studied because of their electrochemical properties, aqueous compatibility and their ability to form nanostructures like nanowires, nanotubes or microcontainers [149].

Hydrogels are cross-linked hydrophilic polymer structures that can hold large amounts of water or biological fluids in their pores. As one of the newest classes of polymer-based systems, target-responsive hydrogels have found numerous biomedical and pharmaceutical applications [121, 150]. Polyacrylamide/acrylate is a copolymer of acrylamide and acrylic acid, and acrydite-modified oligonucleotides can form hydrogels by the introduction of proper complementary DNA as a cross-linker. He et al. [151] reported an aptamer-based reversible DNA induced hydrogel system for molecular recognition and separation.

Different types of hydrogel nanoparticles were used together with various biocomponents, such as enzymes, proteins and nucleic acids, during the development of electrochemical biosensing systems, in order to increase their sensing performances [152].

CPs and in particular functionalized CPs are useful for biosensor construction as a platform to immobilize the primary Ab on the electrode surface. Poly o-aminobenzoic acid (PABA) can be considered a carboxyl-functionalized aniline CP, capable of self-doping, and is also a soluble derivative of polyaniline. Its carboxyl group can be exploited as a linker, either for PABA immobilization onto the electrode, or for covalent bond formation with biomolecules such as proteins and antibodies [153].

An electrochemical aptamer-based biosensing assay for MUC1 protein detection by the modification of the graphite-based SPE’s surface using a functionalized conductive polymer (PABA), and by using methylene blue (MB) as an electrochemical indicator was developed. The recognition of immunoreactions and aptamer binding events was assessed by monitoring the interfacial electron transfer resistance with EIS, while CV and DPV were employed for sensitive indirect quantification of MUC1 via MB with a detection limit of 0.62 ppb [154]. In order to establish the optimal configuration of HRP-based biosensors able to detect acetaminophen in various matrices, another conductive polymer, polypyrrole, was deposited on GCE and SPE by several methods. Chronoamperometric studies performed proved the interaction between the immobilized enzyme and the electroactive species of paracetamol, NAPQI, enzymatically generated in the presence of H2O2 [155].

The combination of nanosphere lithography, electrodeposition and click chemistry was applied for the immobilization of ferrocene at GCE. The nanostructuration of the surface was obtained by adsorption of 100 and 900 nm diameter polystyrene nanospheres followed either by electropolymerization of N-(10-azidodecyl) pyrrole or by electrografting of 4-azidobenzenediazonium. This approach offered potential applications such as engineering of porous materials and was used for HRP immobilization. The nanostructured biosensor was applied for acetaminophen detection in the presence of hydrogen peroxide, and showed better performances in terms of sensitivity and LOD in comparison with a non-structured electrode [156]. Another template based on nanostructured polymeric film was developed combining latex nanosphere lithography with electropolymerization of N-substituted pyrrole monomer in order to immobilize both tyrosinase and GOx and to obtain a biosensor with higher sensing performances [148].

3.2.5. Molecularly imprinted polymers

MIPs can be successfully used in applications relying on selective molecular binding as they are fully artificial macromolecular structures with biomimetic properties, imitating receptor-ligand, Ab-Ag, or enzyme-substrate biorecognition [157].

MIPs possess several advantages over their biological counterparts, including low cost, ease of preparation, long term stability and shelf life, inherent reusability, high mechanical strength, and remarkable chemical and thermal stability under various experimental conditions, allowing the very specific binding of the analyte in “harsh” chemical media [158, 159].

The rational design of MIPs is still considered a challenge due to the numerous experimental variables to be addressed, and it has been previously reviewed in detail by several groups of authors [157, 158, 160, 161] along with their most important analytical applications.

MIPs prepared by bulk polymerization have proven to be poorly compatible with transducers in terms of efficient immobilization strategies, mass transfer and rebinding kinetics, even if they are ground before use. Therefore, several aspects related to the in-situ polymerization, and more specifically to the electrosynthesis of MIPs and hybrid technologies involving MIPs in the fabrication of biosensors, as well as their most recent advances in addressing two important challenges for MIP-based electrochemical biosensors, the determination of biomacromolecules and enantiospecific analysis, will be very briefly presented.

Electroactive organic functional monomers can polymerize to form conducting, semiconducting or insulating polymers. Polymerization may be performed chemically (using an oxidizer as external reagent) or electrochemically (by electropolymerization). Electropolymerization is the more convenient approach, in terms of simplicity and time-effectiveness, for electrochemical biosensor fabrication as it leads to a tight binding of an MIP film onto the surface of the transducer. This procedure ensures both the synthesis and the surface immobilization of the biomimetic layer, along with the possibility of controlling polymer nucleation and growth, film thickness and morphology by the proper selection of the electropolymerization parameters. The reproducible preparation of thin MIP layers is of the utmost importance in sensing applications, because it reduces response time by significantly shortening diffusional path lengths. Solvent swelling and the inclusion of the supporting electrolyte’s ions tunes the rigidity and porosity of the MIP film. Electropolymerization may be performed under galvanostatic, potentiometric or most frequently potentiodynamic conditions, in both non-aqueous and aqueous solutions. The first two techniques are able to generate well-ordered, charged polymeric layers doped with counter ions for charge neutralization, whereas the latter generates a more tangled, charged or neutral polymeric matrix due to the continuous influx and release of solvated ions during its growth upon film charging and discharging [157].

Using electroinsulating MIP films in the fabrication of electrochemical sensors carries a major disadvantage due to the high resistance to charge transfer and the lack of direct path for electron conduction from the recognition sites to the electrode transducer, suffering from an incomplete template removal upon their overoxidation. Using electronically conducting polymers for MIP synthesis avoids these undesirable effects.

Some of the most common electroinsulating layers employed in MIP technology are acrylic- and vinylic-based polymers, but the most frequently reported electrosynthesized insulating MIPs are polyphenylenediamine, polyphenol, polyaminophenol and polythiophenol. Due to their low conductivity, sensors based on such MIPs usually use optical, piezoelectric or impedimetric signal transduction. However, as long as they are electrodeposited in a thickness that does not insulate the underlying conductive electrode surface (<10 nm) [162], a successful voltamperometric signal transduction may also be performed [163].

In the case of electrosynthesis of conducting MIP layers for biosensing, electroactive functional monomers such as pyrrole, thiophene, aniline, 3-aminophenylboronic acid and porphyrin derivatives are most often reported. Analyte determination using such conducting layers may be performed by any available signal transduction procedure, such as optical, piezoelectric, voltammetric, potentiometric, chronoamperometric, impedimetric or conductometric procedures. Poly(pyrrole), poly(thiophene), poly(aniline) and their derivatives are amongst the most widely studied conducting polymers [157, 160], due to their high conductivities in their oxidized form and their ability to reversibly switch between conducting and insulating states by doping and undoping [160]. As mentioned earlier, in the non-covalent molecular imprinting, a wide variety of templates are amenable to be used with elevated imprinting efficiencies, ranging from small analytes to macromolecules (proteins, nucleotides). Additional promising approaches include implementing surface MIP for cells (yeast), bacteria, virus or blood [164, 165, 166]. In the case of using small organic molecules as templates, MIP synthesis is well established, however the imprinting of much larger molecules is still considered to be a challenge. Larger templates are less rigid, which does not ensure the formation of well-defined binding cavities during macromolecular imprinting, and, furthermore, the secondary and tertiary structures of biomacromolecules (i.e., proteins) may be also affected under the imprinting conditions. The large size hinders the protein both in reaching and leaving the artificial binding sites, and their molecular complexity increases the chances of non-specific binding with the MIP, leading to poor selectivity and cross-reactivity. More importantly, proteins are often incompatible with the organic solvents used in MIPs synthesis. The use of an aqueous solution for macromolecular imprinting greatly restricts the choice of reactive functional monomers and cross-linkers. Additionally, water could compete for and potentially disrupt any hydrogen bonds between the template and the functional monomers. The protein as template in macromolecular imprinting may be introduced by three different approaches: it may be dissolved into the prepolymerization mixture (bulk imprinting), grafted to the surface of the transducer followed by a partial inclusion into the bulk polymer with the resulting binding sites situated on the MIP surface (surface imprinting), or only a small epitope part of the protein (epitope imprinting) may be used as a template, resulting in an MIP that is able to recognize the whole target protein. Surface and epitope imprinting enhance the imprinting effect for biomolecules, leading to higher affinity and capacity for the template, as well as to reduced non-specific binding; they improve protein binding kinetics, overcoming difficulties to mass transfer and protein removal in imprinted matrices. While surface imprinting is an efficient technique, the conditions used to accomplish the imprinting process must be carefully considered. The conformational stability of proteins is sensitive to their surrounding conditions, including temperature, pH, ion concentration, surfactant and its concentration, the ratio of template to functional monomers, the types of functional monomers used, and initiator type [167]. MIPs generated by epitope imprinting manifest minimal non-specific binding and improved affinity. Furthermore, it is less costly working with short peptides as templates, and more importantly, because of the template’s higher stability, organic solvents may also be used during the imprinting process. The detection of the bound non-electroactive target macromolecule is performed indirectly, in the presence or absence of a redox probe, by various signal transduction principles, such as voltammetry, piezomicrogravimetry, impedance spectroscopy, etc.

The high selectivity of such biomimetic polymers is also demonstrated by their efficient use for the molecular imprinting of optically active templates. This special application of MIP-based chiral electrochemical sensing has not been fully exploited, even though there is a great necessity in the pharmaceutical industry for developing fast and cost-effective methods of chiral analysis applicable from the early stages of drug development. The vast majority of MIP-based electrochemical sensors were developed for non-chiral analysis. Only around 20 articles report MIP-based sensors designed for the chiral analysis of different small-molecules, but not pharmaceutically active compounds, like amino acids [168, 169, 170] or monosaccharides [171]. Moreover, enantiospecific signal transduction was only reported in the case of amino acids. Nevertheless, very recently an MIP-based electrochemical sensor has been reported for the simultaneous enantiospecific recognition of several β-blocker enantiomers [163].

In spite of the electroinsulating properties of the R(+)-atenolol imprinted acrylate-based MIP, voltamperometric signal transduction was successfully performed, due to the resulting very thin (less than 4 nm) and porous MIP layer. The sensor exhibited distinctive enantiospecific oxidation peaks towards the R-antipodes of four β-blocker representatives and additional oxidation peaks common to both enantiomers of each studied β-blocker, thus allowing the simultaneous analysis of all of their enantiomers in a single determination.

3.3. Hybrid materials

Metal NPs decorating CNTs or graphene sheets, carbon-coated magnetic NPs, nanoparticles included in carbon paste electrodes, MWCNT-alumina-coated silica, and insulating MIPs combined with various nanostructures are examples of some hybrid nanomaterials used for the enhancement of biosensor sensitivity.

3.3.1. CNT-based hybrid materials

Electrodes modified by hybrid materials, consisting of CNTs and metal nanoparticles, have been developed for use as fuel cell catalysts and biosensors [172]. For example, platinum nanoparticles (PtNPs) catalyse the electrochemical oxidation of H2O2, which is generated by the enzymatic reaction, while MWCNTs can be employed to modify the working electrode for the enlargement of the electroactive surface area [173]. A highly sensitive Pt-CNT-glucose biosensor was developed using the incorporation of GOx on a Pt-CNT electrode but with poor enzyme stability after storage [174]. Electrodes modified by CNTs/PtNPs hybrids protected by dendrimers were first developed for methanol electrochemical oxidation [175], and then adapted for the detection of inorganic phosphorous pesticide through the inhibition of the acetylcholinesterase [172]. These electrochemical biosensors are based on a hybrid material, consisting of MWCNTs, PtNPs and poly(amidoamine) dendrimer (DEN) as a binder for enzyme immobilization (Figure 7).

Figure 7.

Biosensor preparation by loading MWCNT, PtNPs and enzyme through dendrimer binders [172]

MWCNTs increase the surface area and promote the electron transfer properties, PtNPs intensify the electron transferability in addition to catalytic performance, and dendrimers play a role in the uptake of the redox agent as well as the binder functionality.

Two new xanthine biosensors, based on GC modified with chitosan (CS), Co3O4NPs and MWCNTs were developed [176]. Two different enzyme systems were investigated involving monoenzyme xanthine oxidase (XO) immobilization and bienzyme xanthine oxidase (XO)/HRP co-immobilization. Xanthine amperometric determination was achieved by both electrooxidation and electroreduction of the enzymatically generated hydrogen peroxide, with LODs of 0.2 μM and 20 μM with anodic and cathodic processes, respectively. The proposed biosensors were applied for detecting xanthine in fish with average recoveries of 96.2±3.5 % with XO/Co3O4MWCNTs/CS/GCE and 95.4±2.1 % with XO/HRP/Co3O4MWCNTs/CS/GCE, respectively.

3.3.2. Graphene-based hybrid materials

Graphene can be decorated with AuNPs to enhance the signal transduction or to increase the grafting area for biocomponent immobilization. AuNPs are the most commonly used metal nanoparticles in graphene-based biosensing systems, mainly due to their ability to form strong covalent bonds with the thiol groups [26].

An electrochemical aptasensor using graphene-AuNPs composite obtained by the reduction of tetrachloroauric acid with sodium citrate in a graphene water suspension was reported [109]. Direct electrodeposition technique of AuNPs was also applied to graphene-based electrochemical transducers yielding a high density of nanoparticles that provide an ultra-large specific surface area available for aptamer immobilization [177].

A graphene-CNTs-incorporating zinc oxide NPs hybrid composite (GR-CNT-ZnO) was synthesized and applied to fabricate an enzyme-based glucose biosensor (Figure 8).

Figure 8.

The preparation of GR-CNT-ZnO composite [178]

By using CV, authors have concluded that the modified film in the presence of oxygen has high electrocatalytic ability in glucose detection with a wide linear detection range and a LOD of 4.5 μM of glucose. The sensor was successfully applied for human serum samples, and the results were in good agreement with those determined using a standard photometric method [178].

3.3.3. Molecularly imprinted polymer-based hybrid materials

Hybrid materials combining insulating MIPs with electronically conducting polymers tend to correct the drawbacks of the former and feature networks of molecular wires connecting recognitions sites to the electrode surface [179]. Molecularly imprinted nanoparticles, nanospheres, nanoshells or nanofibres obtained by various synthesis procedures (precipitation or emulsion/suspension polymerization, mini/emulsion precipitation, etc.), along with various organic/inorganic hybrid materials [165, 180, 181], metal nanoparticles, such as Ag [182] and Au [183] or carbon nanomaterials [173], may be combined with electrosynthesized MIP films, leading to composite MIP/nanomaterial biosensing systems [160, 161] with larger specific surface area, higher sensibility and specificity, and higher biocompatibility for trace electrochemical analysis of small analytes and biomacromolecules.

Advertisement

4. Perspectives and future development

The selection of materials and fabrication techniques is crucial for adequate sensor function and the performance of a biosensor often ultimately depends upon these, rather than upon other factors. Consequently, future developments in biosensor design will inevitably focus upon the technology of new materials, especially the new copolymers that promise to solve the biocompatibility problem and offer the prospect of more widespread use of biosensors in clinical monitoring. The future trends in sensor development that are expected to have an impact on biosensor performances mainly concern immobilization techniques, nanotechnology, miniaturization and design of multisensor arrays.

References

  1. 1. Cristea C, Hârceagă V, Săndulescu R. Electrochemical sensor and biosensors. In: Moretto LM, Kalcher K, editors. Environmental Analysis by Electrochemical Sensors and Biosensors. Vol. 1 Fundamentals. 1st ed. New York: Springer; 2014. p. 155-165.
  2. 2. Morrison DWG, Dokmeci MR, Demirci U, Khademhosseini A. Clinical applications of micro- and nanoscale biosensors. In: Gonsalves KE, Halberstadt CR, Laurencin CT, Nair LS, editors. Biomedical Nanostructures. 1st ed. New York: John Wiley & Sons Inc.; 2008. p. 1-10.
  3. 3. Castillo J, Gaspar S, Leth S, Niculescu M, Mortari A, Botidean I et al. Biosensors for life quality design, development and applications. Sensor. Actuat. B-Chem. 2004;102(2):179-194.
  4. 4. Campas M, Prieto-Simon B, Marty JL. A review of the use of genetically engineered enzymes in electrochemical biosensors. Semin. Cell. Dev. Biol. 2009;20(1):3-9.
  5. 5. Farré M, Kantiani L, Pérez S, Barceló D. Sensors and biosensors in support of EU directives. TrAC Trends Anal. Chem. 2009;28(2):170-185.
  6. 6. Zhang S, Wright G, Yang Y. Materials and techniques for electrochemical biosensor design and construction. Biosens. Bioelectron. 2000;15(5-6):273-282.
  7. 7. Walcarius A, Etienne M, Herzog G, Urbanova V, Vila N. Electrode materials (bulk materials and modification). In: Moretto LM, Kalcher K, editors. Environmental Analysis by Electrochemical Sensors and Biosensors. Vol. 1 Fundamentals. 1st ed. New York: Springer; 2014. p. 403-495.
  8. 8. Xu Z, Chen X, Dong S. Electrochemical biosensors based on advanced bioimmobilization matrices. TrAC Trends Anal. Chem. 2006;25(9):899-908.
  9. 9. Mello LD, Kubota LT. Review of the use of biosensors as analytical tools in the food and drink industries. Food Chem. 2002;77(2):237-256.
  10. 10. Lojou E, Bianco P. Membrane electrodes for protein and enzyme electrochemistry. Electroanal. 2004;16(13-14):1113-1121.
  11. 11. Teles FRR, Fonseca LP. Applications of polymers for biomolecule immobilization in electrochemical biosensors. Mat. Sci. Eng. C Biomin. 2008;28(8):1530-1543.
  12. 12. Samanta D, Sarkar A. Immobilization of bio-macromolecules on self-assembled monolayers: Methods and sensor application. Chem. Soc. Rev. 2011;40(5):2567-2592.
  13. 13. Tien HT. Bilayer lipid membrane-based electrochemical biosensors. J. Clin. Lab. Anal. 2005;2(4):256-264.
  14. 14. Walcarius A. Electroanalytical applications of microporous zeolites and mesoporous(organo) silicas: Recent trends. Electroanal. 2008;20(7):711-738.
  15. 15. Chambers JP, Arulanandam BP, Matta LL, Weis A, Valdes JJ. Biosensor recognition elements. Curr. Issues. Mol. Biol. 2008;10(1-2):1-12.
  16. 16. Thevenot DR, Toth K, Durst RA, Wilson GS. Electrochemical biosensors: Recommended definitions and classification. Biosens. Bioelectron. 16;1-2(121-131)
  17. 17. Bard AJ, Inzelt G, Scholz F, editors. Electrochemical Dictionary. 2nd ed. Berlin: Springer; 2012.
  18. 18. Mascini M, Minunni M, Guilbault GG, Carter R. Methods in biotechnology. In: Rogers KR, Mulchandani A, editors. Affinity Biosensors: Techniques and Protocols. 1st ed. Totowa, NJ: Humana Press Inc.; 1998.
  19. 19. Palecek E, Fojta M. DNA hybridization and damage. Anal. Chem. 2001;73(75A-83A).
  20. 20. Rauf S, Gooding JJ, Akhtar K, Ghauri MA, Rahman M, Anwar MA, et al. Electrochemical approach of anticancer drugs-DNA interaction. J. Pharm. Biomed. Anal. 2005;37(2):205-217.
  21. 21. Mulchandani A, Rogers K. Enzyme and microbial biosensors: Techniques and protocol. Totowa, NJ. 1st ed. Totowa, NJ: Humama Press; 1998.
  22. 22. Yamashita T, Yamashita K, Naeshi H, Yoshikawa T, Yoshioka Y, Tsunoda S et al. Carbon nanomaterials: Efficacy and safety for nanomedicine. Materials. 2012;5:350-363.
  23. 23. Bhakta SA, Evans E, Benavidez TE, Garcia CD. Protein adsorption onto nanomaterials for the development of biosensors and analytical devices: A review. Anal. Chim. Acta. Forthcoming. DOI: Doi:10.1016/j.aca.2014.10.031
  24. 24. Chan CPY, Mak WC, Cheung KY, Sin KK, Yu CM, Rainer TH et al. Evidence-based point-of-care diagnostics: Current status and emerging technologies. Annu. Rev. Anal. Chem. 2013;6:191-211.
  25. 25. Carregal-Romero S, Caballero-Díaz E, Beqa L, Abdelmonem AM, Ochs M, Hühn D et al. Multiplexed sensing and imaging with colloidal nano- and microparticles. Annu. Rev. Anal. Chem. 2013;6:53-81.
  26. 26. Pumera M, Sanchez S, Ichinose I, Tang J. Electrochemical nanobiosensors. Sensor. Actuat. B-Chem. 2007;123(2):1195-1205.
  27. 27. Barnes TM, van de Lagemaat J, Levi D, Rumbles G, Coutts TG, Weeks CL et al. Optical characterization of highly conductive single-wall carbon-nanotube transparent electrodes. Phys. Rev. B. 2007;75(23):23541001-22354110.
  28. 28. Zhao J, Chen X, Xie JRH. Optical properties and photonic devices of doped carbon nanotubes. Anal. Chim. Acta. 2006;568:161-170.
  29. 29. Wooten M, Gorski W. Facilitation of NADH electrooxidation at treated carbon nanotubes. Anal. Chem. 2010;82(4):1299-1304.
  30. 30. Bose S, Khare RA, Moldenaers P. Assessing the strengths and weaknesses of various types of pretreatment of carbon nanotubes on the properties of polymer carbon nanotubes composites: A critical review. Polymer. 2010;51:975-993.
  31. 31. Wang HZ, Huang ZP, Cai QJ, Kulkarni K, Chen CL, Carnahan D et al. Reversible transformation of hydrophobicity and hydrophilicity of aligned carbon nanotube arrays and buckypapers by dry processes. Carbon. 2010;48(3):868-875.
  32. 32. Soetedjo H, Mora MF, Garcia CD. Optical properties of single-wall carbon nanotube films deposited on Si/SiO2 wafers. Thin Solid Films. 2010;518(14):3954-3959.
  33. 33. Qureshi A, Kang WP, Davidson JL, Gurbuz Y. Review on carbon-derived, solid-state, micro and nano sensors for electrochemical sensing applications. Diam. Relat. Mater. 2009;18(12):1401-1420.
  34. 34. Masheter AT, Abiman P, Wildgoose GG, Wong E, Xiao L, Rees NV et al. Investigating the reactive sites and the anomalously large changes in surface pKa values of chemically modified carbon nanotubes of different morphologies. Mater. Chem. 2007;17(25):2616-2626.
  35. 35. Iijima S. Helical microtubules of graphitic carbon. Nature. 1991;354:56-58.
  36. 36. Hu C, Hu S. Carbon nanotube-based electrochemical sensors: Principles and applications in biomedical systems. Journal of Sensors. 2009;2009:187615. DOI:10.1155/2009/187615.
  37. 37. Sadik OA, Mwilu SK, Aluoch A. Smart electrochemical biosensors: From advanced materials to ultrasensitive devices. Electrochim. Acta. 2010;55(14):4287-4296.
  38. 38. Erdem A. Nanomaterials based sensor development towards electrochemical sensing of bioreactions. In: Vaseashta A, Braman E, Susmann P, editors. Technological innovation in sensing and detection of chemical, biological, radiological, nuclear threats and ecological terrorism. NATO Science for Peace and Security Series A: Chemistry and Biology. 1st ed. New York: Springer; 2012. p. 165-169.
  39. 39. Giljohann DA, Mirkin CA. Drivers of biodiagnostic development. Nature. 2009;462:461-464.
  40. 40. Belin T, Epron F. Characterization methods of carbon nanotubes: A review. Mat. Sci. Eng. B-Solid. 2005;119:105-118.
  41. 41. Banks CE, Davies TJ, Wildgoose GG, Compton RG. Electrocatalysis at graphite and carbon nanotube modified electrodes: Edge-plane sites and tube ends are the reactive sites. Chem. Commun. 2005;7(21):829-841.
  42. 42. Banks CE, Crossley A, Salter C, Wilkins SJ, Compton RG. Carbon nanotubes contain metal impurities which are responsible for the “electrocatalysis” seen at some nanotube-modified electrodes. Angew. Chem. Int. Ed. 2006;45(16):2533-2537.
  43. 43. Pumera M, Ambrosi A, Bonanni A, Chng ELK, Poh HL. Graphene for electrochemical sensing and biosensing. TrAC Trends Anal. Chem. 2010;29(9):954-965.
  44. 44. Ambrosi A, Pumera M. Nanographite impurities dominate electrochemistry of carbon nanotubes. Chem.-A Eur. J. 2010;16(36):10946-10949.
  45. 45. Streeter I, Wildgoose GG, Shao LD, Compton RG. Cyclic voltammetry on electrode surfaces covered with porous layers: An analysis of electron transfer kinetics at single-walled carbon nanotube modified electrodes. Sensor. Actuat. B-Chem. 2008;133:462-466.
  46. 46. Sims MJ, Rees NV, Dickinson EJF, Compton RG. Effects of thin-layer diffusion in the electrochemical detection of nicotine on basal plane pyrolytic graphite (BPPG) electrodes modified with layers of multi-walled carbon nanotubes (MWCNT-BPPG). Sensor. Actuat. B-Chem. 2010;144(1):153-158.
  47. 47. Henstridge MG, Dickinson EJF, Aslanoglu M, Batchelor-McAuley C, Compton RG. Voltammetric selectivity conferred by the modification of electrodes using conductive porous layers or films: The oxidation of dopamine on glassy carbon electrodes modified with multiwalled carbon nanotubes. Sensor. Actuat. B-Chem. 2010;145(1):417-427.
  48. 48. Li W, Jung H, Hoa ND, Kim D, Hong S-K, Kim H. Nanocomposite of cobalt oxide nanocrystals and single-walled carbon nanotubes for a gas sensor application. Sensor. Actuat. B-Chem. 2010;150(1):160-166.
  49. 49. Ghadimi H, Tehrani RMA, Mohamed Ali AS, Mohamed N, Ab Ghani S. Sensitive voltammetric determination of paracetamol by poly (4-vinylpyridine)/multiwalled carbon nanotubes modified glassy carbon electrode. Anal. Chim. Acta. 2013;765:70-76.
  50. 50. Gao Q, Guo Y, Liu J, Yuan X, Qi H, Zhang C. A biosensor prepared by co-entrapment of a glucose oxidase and a carbon nanotube within an electrochemically deposited redox polymer multilayer. Bioelectrochemistry. 2011;81(2):109-113.
  51. 51. Salimi A, Compton RG, Hallaj R. Glucose biosensor prepared by glucose oxidase encapsulated sol–gel and carbon nanotube modified basal plane pyrolytic graphite electrode. Anal. Biochem. 2004;333(1):49-56.
  52. 52. Sato N, Okuma H. Development of single-wall carbon nanotubes modified screen-printed electrode using a ferrocene-modified cationic surfactant for amperometric glucose biosensor applications. Sensor. Actuat. B-Chem. 2008;129(1):188-194.
  53. 53. Rahman MM, Umar A, Sawada K. Development of amperometric glucose biosensor based on glucose oxidase co-immobilized with multi-walled carbon nanotubes at low potential. Sensor. Actuat. B-Chem. 2009;137(1):327-333.
  54. 54. Deng C, Chen J, Chen X, Xiao C, Nie L, Yao S. Direct electrochemistry of glucose oxidase and biosensing for glucose based on boron-doped carbon nanotubes modified electrode. Biosens. Bioelectron. 2008;23(8):1272-1277.
  55. 55. Li S, Zhu X, Zhang W, Xie G, Feng W. Hydrogen peroxide biosensor based on gold nanoparticles/thionine/gold nanoparticles/multi-walled carbon nanotubes – chitosan composite film-modified electrode. Appl. Surf. Sci. 2012;258(7):2802-2807.
  56. 56. Liu Y, Yuan R, Chai Y, Tang D, Dai J, Zhong X. Direct electrochemistry of horseradish peroxidase immobilized on gold colloid/cysteine/Nafion-modified platinum disk electrode. Sensor. Actuat. B-Chem. 2006;115(1):109-115.
  57. 57. Zhang Y, Yuan R, Chai Y, Xiang Y, Hong C, Ran X. An amperometric hydrogen peroxide biosensor based on the immobilization of HRP on multi-walled carbon nanotubes/electro-copolymerized nano-Pt-poly (neutral red) composite membrane. Biochem. Eng. J. 2010;51(3):102-109.
  58. 58. Huang J, Li J, Yang Y, Wang X, Wu B, Anzai J-I et al. Development of an amperometric L-lactate biosensor based on L-lactate oxidase immobilized through silica sol-gel film on multi-walled carbon nanotubes/platinum nanoparticle modified glassy carbon electrode. Mater. Sci. Eng. C. 2008;28(7):1070-1075.
  59. 59. Wang YT, Yu L, Wang J, Lou L, Du WJ, Zhu ZQ et al. A novel L-lactate sensor based on enzyme electrode modified with ZnO nanoparticles and multiwall carbon nanotubes. J. Electroanal. Chem. 2011;661(1):8-12.
  60. 60. Goran JM, Lyon JL, Stevenson KJ. Amperometric detection of L-lactate using nitrogen-doped carbon nanotubes modified with lactate oxidase. Anal. Chem. 2011;83(21):8123-8129.
  61. 61. Tertiş M, Florea A, Săndulescu R, Cristea C. Carbon based electrodes modified with horseradish peroxidase immobilized in conducting polymers for acetaminophen analysis. Sensors Basel. 2013;13:4841-4854. DOI:10.3390/s130404841
  62. 62. Baghayeri M, Namadchian M. Fabrication of a nanostructured luteolin biosensor for simultaneous determination of levodopa in the presence of acetaminophen and tyramine: Application to the analysis of some real samples. Electrochim. Acta. 2013;108(1):22-31.
  63. 63. Amiri-Aref M, Raoof JB, Ojani R. A highly sensitive electrochemical sensor for simultaneous voltammetric determination of noradrenaline, acetaminophen, xanthine and caffeine based on a flavonoid nanostructured modified glassy carbon electrode. Sensor. Actuat. B-Chem. 2014;192:634-641.
  64. 64. Babaei A, Afrasiabi M, Babazadeh M. A glassy carbon electrode modified with multiwalled carbon nanotube/chitosan composite as a new sensor for simultaneous determination of acetaminophen and mefenamic acid in pharmaceutical preparations and biological samples. Electroanal. 2010;22(15):1743-1749.
  65. 65. Shahrokhian S, Asadian E. Simultaneous voltammetric determination of ascorbic acid, acetaminophen and isoniazid using thionine immobilized multi-walled carbon nanotube modified carbon paste electrode. Electrochim. Acta. 2010;55(3):666-672.
  66. 66. Tertiş M, Florea A, Feier B, Marian IO, Silaghi-Dumitrescu L, Cristea A et al. Electrochemical impedance studies on single and multi-wall carbon nanotubes-polymer nanocomposites for biosensors development. J. Nanosci. Nanotechnolog. 2015;15(5):3383-3393.
  67. 67. Bustos-Ramírez K, Martínez-Hernández AL, Martínez-Barrera G, de Icaza M, Castaño VM, Velasco-Santos C. Covalently bonded chitosan on graphene oxide via redox reaction. Materials. 2013;6:911-926.
  68. 68. Guoqing X, Hwang W, Kim N, Cho SM, Chae H. A graphene sheet exfoliated with microwave irradiation and interlinked by carbon nanotubes for high-performance transparent flexible electrodes. Nanotechnol. 2010;21:405201.
  69. 69. Lim CX, Hoh HY, Ang PK, Loh KP. Direct voltammetric detection of DNA and pH sensing on epitaxial graphene: An insight into the role of oxygenated defects. Anal. Chem. 2010;82(17):7387-7393.
  70. 70. Novoselov KS, Geim AK, Morozov SV, Jiang D, Zhang Y, Dubonos SV et al. Electric field effect in atomically thin carbon films. Science. 2004;306(5696):666-669.
  71. 71. Bai J, Huang Y. Fabrication and electrical properties of graphene nanoribbons. Mat. Sci. Eng. R. 2010;70(3-6):341-353.
  72. 72. Kim WS, Moon SY, Bang SY, Choi BG, Ham H, Sekino T et al. Fabrication of graphene layers from multiwalled carbon nanotubes using high DC pulse. Appl. Phys. Lett. 2009;95:083103-183103.
  73. 73. He Q, Sudibya HG, Yin Z, Wu S, Li H, Boey F et al. Centimeter-long and large-scale micropatterns of reduced graphene oxide films: Fabrication and sensing applications. ACS Nano. 2010;4(6):3201-3208.
  74. 74. Pham VH, Cuong TV, Hur SH, Shin EW, Kim JS, Chung JS et al. Fast and simple fabrication of a large transparent chemically-converted graphene film by spray-coating. Carbon. 2010;48:1945-1951.
  75. 75. Kang X, Wang J, Wu H, Liu J, Aksay IA, Lin Y. A graphene-based electrochemical sensor for sensitive detection of paracetamol. Talanta. 2010;81:754-759.
  76. 76. Liu YX, Dong XC, Chen P. Biological and chemical sensors based on graphene materials. Chem. Soc. Rev. 2012;41(6):2283-2307.
  77. 77. Park JW, Lee SJ, Choi EJ, Kim J, Song JY, Gu MB. An ultra-sensitive detection of a whole virus using dual aptamers developed by immobilization-free screening. Biosens. Bioelectron. 2014;51:324-329.
  78. 78. Biju V. Chemical modifications and bioconjugate reactions of nanomaterials for sensing, imaging, drug delivery and therapy. Chem. Soc. Rev. 2014;43:744-764.
  79. 79. Maiti UN, Lee WJ, Lee JM, Oh Y, Kim JY, Kim E et al. 25th anniversary article: Chemically modified/doped carbon nanotubes & graphene for optimized nanostructures & nanodevices. Adv. Mater. 2014;26(1):40-67.
  80. 80. Bekyarova E, Itkis ME, Ramesh P, Berger C, Sprinkle M, de Heer WA et al. Chemical modification of epitaxial graphene: Spontaneous grafting of aryl groups. J. Am. Chem. Soc. 2009;131(4):1336-1337.
  81. 81. Liu S, Xing XR, Yu JH, Lian WJ, Li J, Cui M et al. A novel label-free electrochemical aptasensor based on graphene–polyaniline composite film for dopamine determination. Biosens. Bioelectron. 2012;36(1):186-191.
  82. 82. Yan M, Sun GQ, Liu F, Lu JJ, Yu JH, Song XR. An aptasensor for sensitive detection of human breast cancer cells by using porous GO/Au composites and porous PtFe alloy as effective sensing platform and signal amplification labels. Anal. Chim. Acta. 2013;798:33-39.
  83. 83. Zhao J, Chen GF, Zhu L, Li GX. Graphene quantum dots-based platform for the fabrication of electrochemical biosensors. Electrochem. Commun. 2011;13(1):31-33.
  84. 84. Wang L, Xu M, Han L, Zhou M, Zhu CZ, Dong SJ. Graphene enhanced electron transfer at aptamer modified electrode and its application in biosensing. Anal. Chem. 2012;84(17):7301-7307.
  85. 85. Yan GP, Wang YH, He XX, Wang KM, Liu JQ, Du YD. A highly sensitive label-free electrochemical aptasensor for interferon-gamma detection based on graphene controlled assembly and nuclease cleavage-assisted target recycling amplification. Biosens. Bioelectron. 2013;44:57-63.
  86. 86. Loo AH, Bonanni A, Pumera M. Thrombin aptasensing with inherently electroactive graphene oxide nanoplatelets as labels. Nanoscale. 2013;5:4758-4762.
  87. 87. Tertis M, Hosu O, Fritea L, Farcau C, Cernat A, Sandulescu R et al. A novel label-free immunosensor based on activated graphene oxide for acetaminophen detection. Electroanal. 2015; 27(3):638-647. DOI: 10.1002/elan.201400583
  88. 88. Ghorbani-Bidkorbeh F, Shahrokhian S, Mohammadi A, Dinarvand R. Simultaneous voltammetric determination of tramadol and acetaminophen using carbon nanoparticles modified glassy carbon electrode. Electrochim. Acta. 2010;55(8):2752-2759.
  89. 89. Weingart J, Vabbilisetty P, Sun XL. Membrane mimetic surface functionalization of nanoparticles: Methods and applications. Adv. Colloid. Interfac. Sci. 2013;197-198:68-84.
  90. 90. Rocha-Santos TAP. Sensors and biosensors based on magnetic nanoparticles. TrAC Trends Anal. Chem. 2014;62:28-36.
  91. 91. Rusling JF, Sotzing G, Papadimitrakopoulos F. Designing nanomaterials-enhanced electrochemical immunosensors for cancer biomarker proteins. Bioelectrochemistry. 2009;76(1-2):189-194.
  92. 92. Aguilar-Arteaga K, Rodriguez JA, Barrado E. Magnetic solids in analytical chemistry: A review. Anal. Chim. Acta. 2010;674(2):157-165.
  93. 93. Xu Y, Wang E. Electrochemical biosensors based on magnetic micro/nanoparticles. Electrochim. Acta. 2012;84:62-73.
  94. 94. Nomura A, Shin S, Mehdi OO, Kauffmann JM. Preparation, characterization, and application of an enzyme-immobilized magnetic microreactor for flow injection Analysis. Anal. Chem. 2004;76(18):5498-5502.
  95. 95. Zhang XF, Dong XL, Huang H, Lv B, Zhu KG, Lei JP et al. Synthesis, structure and magnetic properties of SiO2-coated Fe nanocapsules. Mater. Sci. Eng. A. 2007;454-455:211-215.
  96. 96. Gruttner C, Rudershausen S, Teller J.J. Improved properties of magnetic particles by combination of different polymer materials as particle matrix. J. Magn. Magn. Mater. 2001;225(1-2):1-7.
  97. 97. Deere J, Magner E, Wall JG, Hodnett BK. Adsorption and activity of cytochrome c on mesoporous silicates. Chem. Commun. 2001;5:465-466.
  98. 98. Yang L, Zhang X, Ye M, Jiang J, Yang R, Fu T. Aptamer-conjugated nanomaterials and their applications. Adv. Drug Deliv. Rev. 2011;63(14-15):1361-1370.
  99. 99. Berry CC, Curtis ASG. Topical review: Functionalisation of magnetic nanoparticles for applications in biomedicine. J. Phys. D: Appl. Phys. 2003;36(13):R198–R206.
  100. 100. Hervas M, Lopez MA, Escarpa A. Simplified calibration and analysis on screen-printed disposable platforms for electrochemical magnetic bead-based inmunosensing of zearalenone in baby food samples. Biosens. Bioelectron. 2010;25(7):1755-1760.
  101. 101. Yang Z, Zhang C, Zhang J, Bai W. Potentiometric glucose biosensor based core-shell Fe3O4-enzyme-polypyrrole nanoparticles. Biosens. Bioelectron. 2014;51:268-273.
  102. 102. Chen X, Zhu J, Chen Z, Xu C, Wang Y, Yao C. A novel bienzyme glucose biosensor based on three layers Au-Fe3O4@SiO2 magnetic nanocomposite. Sensor. Actuat. B-Chem. 2011;159(1):220-228.
  103. 103. Baby TT, Ramaprabhu R. SiO2 coated Fe3O4 magnetic nanoparticle dispersed multiwalled carbon nanotubes based amperometric glucose biosensor. Talanta. 2010;80(5):2016-2022.
  104. 104. Zhou H, Gan N, Li T, Cao Y, Zeng S, Zheng L et al. The sandwich-type electroluminescence immunosensor for α-fetoprotein based on enrichment by Fe3O4-Au magnetic nano probes and signal amplification by CdS-Au composite nanoparticles labeled anti-AFP. Anal. Chim. Acta. 2012;746:107-113.
  105. 105. Li J, Xu Q, Wei X, Hao Z. Electrogenerated chemiluminescence immunosensor for Bacillus thuringiensis Cry1Ac based on Fe3O4@Au nanoparticles. J. Agric. Food Chem. 2013;61(7):1435-1440.
  106. 106. Zamfir L-G, Geana I, Bourigua S, Rotariu L, Bala C, Errachid A, et al. Highly sensitive label-free immunosensor for ochratoxin A based on functionalized magnetic nanoparticles and EIS/SPR detection. Sensor. Actuat. B-Chem. 2011;159(1):178-184.
  107. 107. Yola ML, Eren T, Atar N. A novel and sensitive electrochemical DNA biosensor based on Fe@Au nanoparticles decorated graphene oxide. Electrochim. Acta. 2014;125:38-47.
  108. 108. Gan N, Yang X, Xie D, Wu Y, Wen W. A disposable organophosphorus pesticides enzyme biosensor based on magnetic composite nano-particles modified screen printed carbon electrode. Sensors. 2010;10(1):625-638.
  109. 109. Li J, Gao H, Chen Z, Wei X, Yang CF. An electrochemical immunosensor for carcinoembryonic antigen enhanced by self-assembled nanogold coatings on magnetic particles. Anal. Chim. Acta. 2010;665(1):98-104.
  110. 110. Yang X, Wu F, Chen D-Z, Lin H-W. An electrochemical immunosensor for rapid determination of clenbuterol by using magnetic nanocomposites to modify screen printed carbon electrode based on competitive immunoassay mode. Sensor. Actuat. B-Chem. 2014;192:529-535.
  111. 111. Xin Y, Fu-Bing X, Hong-Wei L, Feng W, Di-Zhao C, Zhao-Yang W. A novel H2O2 biosensor based on Fe3O4-Au magnetic nanoparticles coated horseradish peroxidase and grapheme sheets-Nafion film modified screen-printed carbon electrode. Electrochim. Acta. 2013;109:750-755.
  112. 112. Chen D, Deng J, Liang J, Xie J, Hue C, Huang K. A core-shell molecularly imprinted polymer grafted onto a magnetic glassy carbon electrode as a selective sensor for the determination of metronidazole. Sensor. Actuat. B-Chem. 2013;183:594-600.
  113. 113. Prakash A, Chandra S, Bahadur D. Structural, magnetic, and textural properties of iron oxide-reduced graphene oxide hybrids and their use for the electrochemical detection of chromium. Carbon. 2012;50(11):4209-4212.
  114. 114. Hu Y, Zang Z, Zhang H, Luo L, Yao S. Selective and sensitive molecularly imprinted sol-gel film-based electrochemical sensor combining mecaptoacetic acid modified PbS nanoparticles with Fe3O4@Au-multi-walled carbon nanotubes-chitosan. J. Solid State Electrochem. 2012;16(3):857–867.
  115. 115. Arvand M, Hassannezhad M. Magnetic core-shell Fe3O4@SiO2/MWCNT nanocomposite modified carbon paste electrode for amplified electrochemical sensing of uric acid. Mater. Sci. Eng. C. 2014;36:161-167.
  116. 116. Hosu O, Tertis M, Cristea C, Sandulescu R. Protein G magnetic beads based immunosensor for sensitive detection of acetaminophen. Farmacia. 2015;63(1):143-146.
  117. 117. Taleat Z, Cristea C, Marrazza G, Sandulescu R. Electrochemical sandwich immunoassay for the ultrasensitive detection of human MUC1 cancer biomarker. Int. J. Electrochem. 2013; Article ID 740265, http://dx.doi.org/10.1155/2013/740265
  118. 118. Jia J, Wang B, Wu A, Cheng G, Li Z, Dong S. A method to construct a third-generation horseradish peroxidase biosensor: Self-assembling gold nanoparticles to three-dimensional sol-gel network. Anal. Chem. 2002;74:2217-2223.
  119. 119. Di J, Shen C, Peng S, Tu Y, Li S. A one-step method to construct a third-generation biosensor based on horseradish peroxidase and gold nanoparticles embedded in silica sol–gel network on gold modified electrode. Anal. Chim. Acta. 2005;553:196-200.
  120. 120. Xu Z, Chen X, Dong S. Electrochemical biosensors based on advanced bioimmobilization matrices. TrAC Trends Anal. Chem. 2006;25(9):899-908.
  121. 121. Yang L, Zhang X, Ye M, Jiang L, Yang R, Fu T et al. Aptamer-conjugated nanomaterials and their applications. Adv. Drug Deliv. Rev. 2011;63(14-15):1361-1370.
  122. 122. Merkoçi A. Nanoparticles-based strategies for DNA, protein and cell sensors. Biosens. Bioelectron. 2010;26(4):1164-1177.
  123. 123. Yin T, Qin W. Applications of nanomaterials in potentiometric sensors. TrAC Trends Anal. Chem. 2013;51:79-86.
  124. 124. Taton TA, Mirkin CA, Letsinger RL. Scanometric DNA array detection with nanoparticle probes. Science. 2000;289(5485):1757-1760.
  125. 125. Wang J, Polsky R, Xu D. Silver-enhanced colloidal gold electrochemical stripping detection of DNA hybridization. Langmuir. 2001;17(19):5739-5741.
  126. 126. Wang J, Xu D, Kawde A-N, Polsky R. Metal nanoparticle-based electrochemical stripping potentiometric detection of DNA hybridization. Anal. Chem. 2001;73(22):5576-5581.
  127. 127. Wang J, Liu G, Merkoçi A. Electrochemical coding technology for simultaneous detection of multiple DNA targets. J. Am. Chem. Soc. 2003;125(11):3214-3215.
  128. 128. Chen G, Tong H, Gao T, Chen Y, Li G. Direct application of gold nanoparticles to one-pot electrochemical biosensors. Anal. Chim. Acta. 2014;849:1-6.
  129. 129. Malon A, Vigassy T, Bakker E, Pretsch E. Potentiometry at trace levels in confined samples: Ion-selective electrodes with subfemtomole detection limits. J. Am. Chem. Soc. 2006;128(25):8154-8155.
  130. 130. Rubinova N, Chumbimuni-Torres K, Bakker E. Solid-contact potentiometric polymer membrane microelectrodes for the detection of silver ions at the femtomole level. Sensor. Actuat. B-Chem. 2007;121(1):135-141.
  131. 131. Chumbimuni-Torres K, Dai Z, Rubinova N, Xiang Y, Pretsch E, Wang J et al. Potentiometric biosensing of proteins with ultrasensitive ion-selective microelectrodes and nanoparticle labels. J. Am. Chem. Soc. 2006;128(42):13676-13677.
  132. 132. Willner I, Baron R, Willner B. Growing metal nanoparticles by enzymes. Adv. Mater. 2006;18:1109-1120.
  133. 133. Ngeontae W, Janrungroatsakul W, Maneewattanapinyo P, Ekgasit S, Aeungmaitrepirom W, Tuntulani TW. Novel potentiometric approach in glucose biosensor using silver nanoparticles as redox marker. Sensor. Actuat. B-Chem. 2009;137(1):320-326.
  134. 134. Chumbimuni-Torres KY, Wang J. Nanoparticle-induced potentiometric biosensing of NADH at copper ion-selective electrodes. Analyst. 2009;134(8):1614-1617.
  135. 135. Wu J, Chumbimuni-Torres KY, Galik M, Thammakhet C, Haake DA, Wang J. Potentiometric detection of DNA hybridization using enzyme-induced metallization and a silver ion selective electrode. Anal. Chem. 2009;81(24):10007-10012.
  136. 136. Florea A, Taleat Z, Cristea C, Mazloum-Ardakani M, Săndulescu R. Label free MUC1 aptasensors based on electrodeposition of gold nanoparticles on screen printed electrodes. Electrochem. Comm. 2013;33:127-130.
  137. 137. Mousty C. Sensors and biosensors based on clay-modified electrodes-new trends. Appl. Clay Sci. 2004;27:159-177.
  138. 138. Direnzo F, Cambon H, Dutartre R. A 28-year-old synthesis of micelle-templated mesoporous silica. Micropor. Mater. 1997;10:283-286.
  139. 139. Qhobosheane M, Santra S, Zhang P, Tan WH. Biochemically functionalized silica nanoparticles. Analyst. 2001;126:1274-1278.
  140. 140. Yan L, Chen X. Nanomaterials for drug delivery. In: Tjong SC, editor. Nanocrystalline Materials. 2nd ed. London: Elsevier; 2014. p. 221-268.
  141. 141. Cheng SH, Lee CH, Yang CS, Tseng FG, Mou CY, Lo LW. Mesoporous silica nanoparticles functionalized with an oxygen-sensing probe for cell photodynamic therapy: Potential cancer theranostics. J. Mater. Chem. 2009;19(9):1252-1257.
  142. 142. Yáñez-Sedeño P, Agüí L, Villalonga R, Pingarrón JM. Biosensors in forensic analysis. A review. Anal. Chim. Acta. 2014;823:1-19.
  143. 143. Cosnier S, Mousty C, Cui XQ, Yang X, Dong S. Specific determination of As(V) by an acid phosphatase polyphenol oxidase biosensor. Anal. Chem. 2006;78(14):4985-4989.
  144. 144. Liu YX, Wei WZ. Layer-by-layer assembled DNA functionalized single-walled carbon nanotube hybrids for arsenic(III) detection. Electrochem. Comm. 2008;10(6):872-875.
  145. 145. Shan D, Mousty C, Cosnier S. Subnanomolar cyanide detection at polyphenol oxidase/clay biosensors. Anal. Chem. 2004;76(1):178-183.
  146. 146. Santos L, Martin P, Ghilane L, Lacaze P-C, Randriamahazaka H, Abrantes LM et al. Electrosynthesis of well-organized nanoporous poly(3,4-ethylenedioxythiophene) by nanosphere lithography. Electrochem. Commun. 2010;12(7):872-875.
  147. 147. Cernat A, Griveau S, Richard C, Bedioui, Săndulescu R. Horseradish peroxidase nanopatterned electrodes by click chemistry: Application to the electrochemical detection of paracetamol. Electroanal. 2013;25(6):1369-1372.
  148. 148. Cernat A, Le Goff A, Holzinger M, Sandulescu R, Cosnier S. Micro- to nanostructured poly(pyrrole-nitrilotriacetic acid) films via nanosphere templates: Applications to 3D enzyme attachment by affinity interactions. Anal. Bioanal. Chem. 2014;406(4):1141-1147.
  149. 149. Stetter JR, Penrose WR, Yao S. Sensors, chemical sensors, electrochemical sensors, and ECSJ. Electrochem. Soc. 2003;150(2):S11-S16.
  150. 150. Ravi Kumar MNV, Kumar N, Kashyap N. Hydrogels for pharmaceutical and biomedical applications. Crit. Rev. Ther. Drug Carrier Syst. 2005;22(2):107-150.
  151. 151. He X, Wei B, Mi Y. Aptamer based reversible DNA induced hydrogel system for molecular recognition and separation. Chem. Commun. 2010;46(34):6308-6310. DOI: DOI:10.1039/c0cc01392g.
  152. 152. Hasanzadeh M, Shadjou N, de la Guardia M. Electrochemical biosensing using hydrogel nanoparticles. TrAC Trends Anal. Chem. 2014;62:11-19.
  153. 153. Sriwichai S, Baba A, Phanichphant S, Shinbo K, Kato K, Kaneko F. Electrochemically controlled surface plasmon resonance immunosensor for the detection of human immunoglobulin G on poly(3-aminobenzoic acid) ultrathin films. Sensor. Actuator. B-Chem. 2010;147(1):322-329.
  154. 154. Taleat Z, Cristea C, Marrazza G, Mazloum-Ardakani M, Săndulescu R. Electrochemical immunoassay based on aptamer–protein interaction and functionalized polymer for cancer biomarker detection. J. Electroanal. Chem. 2014;717-718:119-124.
  155. 155. Cernat A, Tertis M, Griveau S, Bedioui F, Săndulescu R. New modified electrodes with HRP immobilized in polymeric films for paracetamol analysis. Farmacia. 2012;1:1-12.
  156. 156. Cernat A, Griveau S, Martin P. Lacroix JC, Farcau C, Săndulescu R et al. Electrografted nanostructured platforms for click chemistry. Electrochem. Comm. 2012;23:141-144.
  157. 157. Sharma PS, Pietrzyk-Le A, D’Souza F, Kutner W. Electrochemically synthesized polymers in molecular imprinting for chemical sensing. Anal. Bioanal. Chem. 2012;402:3177-3204.
  158. 158. Yan H, Row KH. Characteristic and synthetic approach of molecularly imprinted polymer. Int. J. Mol. Sci. 2006;7:155-178.
  159. 159. Moreira FTC, Dutra RAF, Noronha JPC, Sales MGF. Electrochemical biosensor based on biomimetic material for myoglobin detection. Electrochim. Acta. 2013;107:481-487.
  160. 160. Malitesta C, Mazzotta E, Picca RA, Poma A, Chianella I, Piletsky SA. MIP sensors – the electrochemical approach. Anal. Bioanal. Chem. 2012;402:1827-1846.
  161. 161. Irshad M, Iqbal N, Mujahid A, Afzal A, Hussain T, Sharif A et al. Molecularly imprinted nanomaterials for sensor applications. Nanomaterials. 2013;3:615-637.
  162. 162. Herlem G, Gharbi T. Electrodeposition of insulating thin film polymers from aliphatic monomers as transducers for biosensor applications. In: Serra PA, editor. Biosensors – Emerging Materials and Applications. 1st edition. InTech; 2011. DOI: 10.5772/17584
  163. 163. Iacob BC, Bodoki E, Florea A, Bodoki AE, Oprean R. Simultaneous enantiospecific recognition of several β-blocker enantiomers using molecularly imprinted polymer-based electrochemical sensor. Anal. Chem. Forthcoming. DOI: 10.1021/ac504036m
  164. 164. Ge Y, Turner APF. Too large to fit? Recent developments in macromolecular imprinting. Trends Biotechnol. 2008;26:218-224.
  165. 165. Takeuchi T, Hishiya T. Molecular imprinting of proteins emerging as a tool for protein recognition. Org. Biomol. Chem. 2008;6:2459-2467.
  166. 166. Hussain M, Wackerlig J, Lieberzeit PA. Biomimetic strategies for sensing biological species. Biosensors. 2013;3:89-107.
  167. 167. Ding XC, Heiden PA. Recent developments in molecularly imprinted nanoparticles by surface imprinting techniques. Macromol. Mat. Engineer. 2014;299:268-289.
  168. 168. Chen Y, Chen L, Bi R, Xu L, Liu Y. A potentiometric chiral sensor for L-phenylalanine based on crosslinked polymethylacrylic acid-polycarbazole hybrid molecularly imprinted polymer. Anal. Chim. Acta. 2012;754:83-90.
  169. 169. Prasad BB, Pandey I, Srivastava A, Kumar D, Tiwari MP. Multiwalled carbon nanotubes-based pencil graphite electrode modified with an electrosynthesized molecularly imprinted nanofilm for electrochemical sensing of methionine enantiomers. Sensor. Actuat. B-Chem. 2013;176:863-874.
  170. 170. Prasad BB, Pandey I. Metal incorporated molecularly imprinted polymer-based electrochemical sensor for enantio-selective analysis of pyroglutamic acid isomers. Sensor. Actuat. B-Chem. 2013;186:407-416.
  171. 171. Granot E, Tel-Vered R, Lioubashevski O, Willner I. Stereoselective and enantioselective electrochemical sensing of monosaccharides using imprinted boronic acid-functionalized polyphenol films. Adv. Funct. Mater. 2008;18:478-484.
  172. 172. Siriviriyanun A, Imae T, Nagatani N. Electrochemical biosensors for biocontaminant detection consisting of carbon nanotubes, platinum nanoparticles, dendrimers, and enzymes. Anal. Biochem. 2013;443(2):169-171.
  173. 173. Ahmed MU, Hossain MM, Tamiya E. Electrochemical biosensors for medical and food applications. Electroanal. 2008;20:616-626.
  174. 174. Liu G, Riechers SL, Mellen MC, Lin Y. Sensitive electrochemical detection of enzymatically generated thiocholine at carbon nanotube modified glassy carbon electrode. Electrochem. Commun. 2005;7(11):1163-1169.
  175. 175. Siriviriyanun A, Imae T. Advantages of electrodes with dendrimer-protected platinum nanoparticles and carbon nanotubes for electrochemical methanol oxidation. Phys. Chem. Chem. Phys. 2013;15:4921-4929.
  176. 176. Dalkiran B, Kacar C, Erden PE, Kilic E. Amperometric xanthine biosensors based onchitosan-Co3O4-multiwall carbon nanotube modified glassy carbon electrode. Sensor. Actuat. B-Chem. 2014;200:83-91.
  177. 177. Jiang BY, Wang M, Chen Y, Xie JQ, Xiang Y. Highly sensitive electrochemical detection of cocaine on graphene/AuNP modified electrode via catalytic redox-recycling amplification. Biosens. Bioelectron. 2012;32(1):305-308.
  178. 178. Hwa K-Y, Subramani B. Synthesis of zinc oxide nanoparticles on graphene-carbon nanotube hybrid for glucose biosensor applications. Biosens. Bioelectron. 2014;62:127-133.
  179. 179. Lakshmi D, Bossi A, Whitcombe MJ, Chianella I, Fowler SA, Subrahmanyam S et al. Electrochemical sensor for catechol and dopamine based on a catalytic molecularly imprinted polymer-conducting polymer hybrid recognition element. Anal. Chem. 2009;81:3576-3584.
  180. 180. Lei R, Guo C, Xiong H, Dong C, Zhang X, Wang S. A novel electrochemical sensor for β2-agonists with high sensitivity and selectivity based on surface molecularly imprinted sol-gel doped with antimony-doped tin oxide. Electroanal. 2014;26:1004-1012.
  181. 181. Fatoni A, Numnuam A, Kanatharana P, Limbut W, Thavarungkul P. A novel molecularly imprinted chitosan–acrylamide, graphene, ferrocene composite cryogel biosensor used to detect microalbumin. Analyst. 2014;139:6160-6167.
  182. 182. Du D, Chen S, Cai J, Tao Y, Tu H, Zhang A. Recognition of dimethoate carried by bi-layer electrodeposition of silver nanoparticles and imprinted poly-o-phenylenediamine. Electrochim. Acta. 2008;53:6589-6595.
  183. 183. Kan X, Liu T, Zhou H, Li C, Fang B. Molecular imprinting polymer electrosensor based on gold nanoparticles for theophylline recognition and determination. Microchim. Acta. 2010;171:423-429.

Written By

Robert Săndulescu, Mihaela Tertiş, Cecilia Cristea and Ede Bodoki

Submitted: 29 January 2015 Reviewed: 11 March 2015 Published: 24 September 2015