Open access peer-reviewed chapter

Study of Metrological Properties of Voltammetric Electrodes in the Time Domain

Written By

Krzysztof Suchocki

Submitted: 05 October 2016 Reviewed: 17 February 2017 Published: 07 June 2017

DOI: 10.5772/67944

From the Edited Volume

Applications of the Voltammetry

Edited by Margarita Stoytcheva and Roumen Zlatev

Chapter metrics overview

1,376 Chapter Downloads

View Full Metrics

Abstract

Metrological properties of voltammetric electrodes, in the situation where on their surface an electrochemical reaction of oxidizing/reduction takes place, were analyzed in this chapter. The properties of electrodes on which a reaction controlled by ion transport process takes place were taken into consideration. Also, it was analyzed how the electrode’s shape and the voltage polarizing the electrode influence this electrode’s metrological properties. The result of the analysis conducted is that in case of a reaction controlled by charge exchange process, such a voltammetric electrode functions like a converter type 0. Its metrological properties in the time domain are defined solely by sensitivity. However, if on the surface of the electrode there is a reaction controlled by ion transport process, the electrode will function like a converter type I. Its metrological properties in the time domain are defined by the sensitivity and time constant. Numeric simulations were conducted in order to determine the influence of the electrode’s shape and the polarizing voltage on metrological properties of the electrode. The results show that both the sensitivity and the time constant of the electrode can be influenced by choice of an electrode’s shape and the shape of the polarizing voltage.

Keywords

  • voltammetric electrodes
  • voltammetric measurements
  • DC voltammetry
  • AC voltammetry
  • metrologic properties
  • time constant

1. Introduction

Voltammetric measurements are one of the most frequently conducted measurements in order to determine ion concentration in water [17]. Their commonness is connected most of all to its simplicity and relatively high accuracy. There are many different types of voltammetric methods [715]. These methods differ from each other mainly in the voltage shape polarizing the voltammetric electrode, and in result, they also differ in accuracy of measurements conducted. These methods are successfully applied in electrochemical measurements in stationary conditions when the marked ion concentration in the volume of analyzed solution is constant in time. At the same time, much more voltammetric measurements are conducted in situ, where the concentration of marked ions can change during the marking process [7, 10]. Some questions raise concerning the accuracy of conducted measurements, metrological properties of voltammetric electrodes, and methods of their improvement [1215]. Hence, some work is undertaken in order to define metrological properties of voltammetric electrodes and the influence of the electrode’s shape and the shape of polarizing voltage on these electrodes.

Advertisement

2. Metrological properties of voltammetric electrodes

Generally, metrological properties of voltammetric electrodes as measuring converters can be divided into static and dynamic ones.

Static properties are the characteristic of voltammetric electrodes which are in the steady state, i.e., in the state in which the concentration of marked ions does not change in the volume of the analyzed solution nor on the surface of the electrode. Dynamic properties are the characteristics of the electrode in the transient state, when these concentrations change while the voltammetric measurements are being conducted.

In order to simplify the analysis of the voltammetric electrode metrological properties following assumptions have been accepted:

  • the input signal is the marked ion concentration Ci0(t) in the analyzed electrolyte volume,

  • the output signal is the current ii(t) of the electrochemical reaction on the voltammetric electrode’s surface,

  • the time of charge exchange between the ions in the analyzed electrolyte and the voltammetric electrode equals 0.

Generally, the electrochemical reaction of oxidizing/reduction on the voltammetric electrode’s surface is divided into several stages. The first stage is about delivery of depolarizer’s ions from the volume of the electrolyte into the vicinity of the electrode’s surface. The second stage of the electrochemical reaction is to transport, to or from, an electron or electrons through the depolarizer’s ion. The third stage is to transport away the products from the reaction to the volume of analyzed solution. Stage four of the reaction is when the products of the electrochemical reaction can still react with other ions in the electrolyte after being transported away.

Advertisement

3. Electrochemical reaction controlled by a process of charge exchange on the surface of the voltammetric electrode

When on the surface of the voltammetric electrode an electrochemical reaction controlled by a process of charge exchange takes place, then the marked ion concentration in oxidizing/reduction form on the electrode surface equals the concentration of the same ion form in the analyzed solution volume [8]:

Ci,ox,0(t)=Ci,ox0(t),E1
Ci,red,0(t)=Ci,red0(t).E2

The value of the output signal of a voltammetric electrode, which is the current on this electrode, is defined by the Butler‐Volmer equation [8]:

ii(t)=ziFA[ki,ox(t)Ci,ox,0(t)ki,red(t)Ci,red,0(t)]=ii,ox(t)ii,red(t),E3

where the values of the reaction rate coefficients ki,ox and ki,red are defined by the following relations [8]:

ki,ox(t)=k0exp{αziFRT[Epol(t)E0]},E4
ki,red(t)=k0exp{(1α)ziFRT[Epol(t)E0]}.E5

Keeping in mind that an oxidizing/reduction reaction may take place on the surface of the voltammetric electrode, the relation (3) may be denoted as follows:

ii,ox(t)=ziFAki,ox(t)Ci,ox,0(t),E6
ii,red(t)=ziFAki,red(t)Ci,red,0(t).E7

It is immediately clear that the voltammetric electrode functions exactly like a converter type 0. Hence, its metrological properties are defined solely by the sensitivity coefficient denoted as follows:

Si(t)=Δii(t)ΔCi0(t)=ziFAki(t),E8

by taking into consideration relations (6) and (7) we get:

Si,ox(t)=ziFAki,ox(t),E9
Si,red(t)=ziFAki,red(t).E10

It results from the above analysis presented that metrological properties of a voltammetric electrode are described solely by sensitivity. This parameter is characteristic to an electrode both in the steady state and in the transient state. In such electrochemical reactions, the electrode does not present any delays or dynamic errors. Its sensitivity is determined by parameters defining marked ions, area of the electrode, electrochemical reaction rate on the surface of the electrode, and the voltage polarizing the electrode.

Advertisement

4. Electrochemical reaction controlled by a process of ion transport to the surface of the voltammetric electrode

In the case when an electrochemical reaction controlled by a process of ion transport takes place on the surface of the electrode, its flux to the surface is defined by this relation [8]:

Ni(t)=DiCi(t)+uiziFCi(t)U(t)+Vi(t)Ci(t).E11

Distribution of ion concentration Di=0 in the solution volume as a function of time t is defined by the flux divergence. Hence [8]:

Ci(t)t=Ni(t).E12

Keeping in mind the relation (11) and assuming that Di=0 we get as a result:

Ci(t)t=Di2Ci(t)+ziuiFCi(t)U(t)+ziuiFCi(t)2U(t)+Vi(t)Ci(t)+Vi(t)Ci(t)E13

In real terms, voltammetric measurements are conducted with stationary electrodes in presence of excess of concentrated basic electrolyte, which allows to simplify the relation (13) to:

Ci(t)t=Di2Ci(t).E14

It is clear that in such a case the ion transport to or from the surface of a voltammetric electrode is determined solely by the diffusion.

4.1. Metrological properties for a general case

In general cases, without making any assumptions about reversibility or irreversibility of electrochemical reactions happening on the surface of the voltammetric electrode, one may present the relation (14) using the finite difference method

Ci,0(t)t=Di[Ci0(t)Ci,0(t)δi2(t)],E15

Which, after transformation, leads to:

[δi2(t)Di]Ci,0(t)t+Ci,0(t)=Ci0(t)E16

Keeping in mind the relation (3) and transforming it we get:

Ci,0(t)=ii(t)ziFAki(t),E17
Ci,0(t)t=1ziFAki(t)[ii(t)tziFACi,0(t)ii(t)t].E18

Substituting the relation (16) with relations (17) and (18) we get:

[δi2(t)Di]ii(t)t+[1δi2(t)Di1ki(t)ki(t)t]ii(t)=ziFAki(t)Ci0(t).E19

Keeping in mind all the assumptions, we can transform the above relation to:

[δi2(t)Di]{1[δi2(t)Di]1ki(t)ki(t)t}ii(t)t+ii(t)=ziFAki(t){1[δi2(t)Di]1ki(t)ki(t)t}Ci0(t).E20

It results from the presented analysis that in this case the voltammetric electrode functions like converter type I. Its metrological properties are defined by sensitivity and the time constant. Static properties of the electrode are defined by sensitivity, and its dynamic properties are characterized by sensitivity and the time constant. Values of these parameters define the following relations:

Si,ox(t)=ziFAki,ox(t){1[δi,ox2(t)Di,ox]1ki,ox(t)ki,ox(t)t},E21
NT,i,ox(t)=[δi,ox2(t)Di,ox]{1[δi,ox2(t)Di,ox]1ki,ox(t)ki,ox(t)t},E22
Si,red(t)=ziFAki,red(t){1[δi,red2(t)Di,red]1ki,red(t)ki,red(t)t},E23
NT,i,red(t)=[δi,red2(t)Di,red]{1[δi,red2(t)Di,red]1ki,red(t)ki,red(t)t}.E24

In the case of this type of electrochemical reactions, an electrode will present a dynamic error whose values are determined above all by the time constant and the nature of changes in the marked ion concentration. One can see in the presented relations that the parameters defining metrological properties of the voltammetric electrode are determined by the parameters defining marked ions, areas of the electrodes, a thickness of the diffusion layer, a rate of the electrochemical reaction happening on the surface of the electrode, and thereby voltage polarizing the electrode.

4.2. The influence of the voltammetric electrode’s shape on its metrological properties in the time domain

Not only flat electrodes are being used in voltammetric measurements, but spherical and cylindrical as well. That is why we also analyzed the influence of the voltammetric electrode on its metrological properties.

4.2.1. Spherical voltammetric electrode

In the case of spherical voltammetric electrode, the relation of ion transport to/from the electrode’s surface give in relation (14) appears as follows:

Ci,0(t)t=Di{1r2r[r2Ci,0(t)r]+1r2sinθθ[sinθCi,0(t)θ]+1r2sin2θ2Ci,0(t)ϕ2}E25

Assuming that the marked ion concentration on the surface of the voltammetric electrode is not determined by θ and ϕ, so

Ci,0(t)θ=0,E26
2Ci,0(t)θ2=0,E27

and

Ci,0(t)ϕ=0,E28
2Ci,0(t)ϕ2=0,E29

relation of the ion transport given in relation (25) is simplified as follows:

Ci,0(t)t=Di{1r2r[r2Ci,0(t)r]},E30

and it results in:

Ci,0(t)t=2DirCi,0(t)r+Di2Ci,0(t)r2.E31

We can present the relation using the finite difference method:

Ci,0(t)t=2Dir[Ci0(t)Ci,0(t)δi(t)]+Di[Ci0(t)Ci,0(t)δi2(t)],E32

which after transformation leads to the relation:

Ci,0(t)t+[2Diδi(t)+rDirδi2(t)]Ci,0(t)=[2Diδi(t)+rDirδi2(t)]Ci0(t),E33

and it results in:

rδi2(t)2Diδi(t)+rDiCi,0(t)t+Ci,0(t)=Ci0(t).E34

Substituting the relation (34) with relations (17) and (18) we get:

ii(t)t+[2Diδi(t)+rDirδi2(t)1ki(t)ki(t)t]ii(t)=ziFAki(t)[2Diδi(t)+rDirδi2(t)]Ci0(t),E35

which can be denoted as:

1[2Diδi(t)+rDirδi2(t)1ki(t)ki(t)t]ii(t)t+ii(t)=ziFAki(t)[2Diδi(t)+rDirδi2(t)][2Diδi(t)+rDirδi2(t)1ki(t)ki(t)t]Ci0(t).E36

The above relation proves that the spherical voltammetric electrode functions as converter type I, both for the oxidizing reaction and for the reduction reaction. Its metrological properties are defined by sensitivity and the time constant. Static properties of the electrode are defined by sensitivity, and its dynamic properties are defined by sensitivity and the time constant. Values of these parameters are defined by the following relations:

Si,ox(t)=ziFAki,ox(t)[2Di,oxδi,ox(t)+rDi,oxrδi,ox2(t)][2Di,oxδi,ox(t)+rDi,oxrδi,ox2(t)1ki,ox(t)ki,ox(t)t],E37
NT,i,ox(t)=1[2Di,oxδi,ox(t)+rDi,oxrδi,ox2(t)1ki,ox(t)ki,ox(t)t],E38
Si,red(t)=ziFAki,red(t)[2Di,redδi,red(t)+rDi,redrδi,red2(t)][2Di,redδi,red(t)+rDi,redrδi,red2(t)1ki,red(t)ki,red(t)t],E39
NT,i,red(t)=1[2Di,redδi,red(t)+rDi,redrδi,red2(t)1ki,red(t)ki,red(t)t].E40

Presented relations show that a change of the shape of voltammetric electrode does not cause a change of the measuring converter’s type. A spherical electrode functions as converter type I. This electrode, in this type of electrochemical reactions, will produce a dynamic error whose value is determined above all by the time constant and by the nature of changes of the marked ion concentration. The relations also show that the parameters defining metrological properties of this voltammetric electrode are determined by the parameters defining marked ions, an electrode radius, a thickness of the diffusion layer, a rate of electrochemical reaction taking place on the surface of the electrode, and thereby by the voltage polarizing the electrode.

4.2.2. Cylindrical voltammetric electrode

The relation of ion transport to/from the surface of the cylindrical voltammetric electrode shown by relation (14) is:

Ci,0(t)t=Di{1ρρ[ρCi,0(t)ρ]+1ρ22Ci,0(t)ϕ2+2Ci,0(t)z2}.E41

Assuming that the concentration of marked ions on the surface of the cylindrical electrode is determined by ϕ, i.e.,:

Ci,0(t)ϕ=0,E42

we get:

Ci,0(t)t=DiρCi,0(t)ρ+Di2Ci,0(t)ρ2+Di2Ci,0(t)z2.E43

Defining this relation with finite difference method we get:

Ci,0(t)t=Diρ[Ci0(t)Ci,0(t)δi(t)]+Di[Ci0(t)Ci,0(t)δi2(t)]+Di[Ci0(t)Ci,0(t)δi2(t)],E44

which finally leads to:

Ci,0(t)t=[Diρδi(t)+2Diδi2(t)]Ci,0(t)=[Diρδi(t)+2Diδi2(t)]Ci0(t),E45

and it results in:

ρδi2(t)2ρDi+δi(t)DiCi,0(t)t+Ci,0(t)=Ci0(t).E46

Substituting relation (46) with relations (17) and (18) we get:

ρδi2(t)2ρDi+δi(t)Diii(t)t+[2ρDi+δi(t)Diρδi2(t)1ki(t)ki(t)t]ii(t)=ziFAki(t)[2ρDi+δi(t)Diρδi2(t)]Ci0(t).E47

Keeping in mind the assumptions, the above relation may be transformed as follows:

[ρδi2(t)2ρDi+δi(t)Di][2ρDi+δi(t)Diρδi2(t)1ki(t)ki(t)t]ii(t)t+ii(t)=ziFAki(t)[2ρDi+δi(t)Diρδi2(t)][2ρDi+δi(t)Diρδi2(t)1ki(t)ki(t)t]Ci0(t).E48

In the above relation, it is clear that a cylindrical voltammetric electrode functions as a measuring converter type I both for the oxidizing reaction and for the reduction reaction. Its metrological properties are defined by sensititvity and the time constant. Static properties of the electrode are defined by sensitivity, and its dynamic properties are characterized by sensitivity and the time constant. Values of the parameters are described as follows:

Si,ox(t)=ziFAki,ox(t)[2ρDi,ox+δi,ox(t)Di,oxρδi,ox2(t)][2ρDi,ox+δi,ox(t)Di,oxρδi,ox2(t)1ki,ox(t)ki,ox(t)t],E49
NT,i,ox(t)=[ρδi,ox2(t)2ρDi,ox+δi,ox(t)Di,ox][2ρDi,ox+δi,ox(t)Di,oxρδi,ox2(t)1ki,ox(t)ki,ox(t)t],E50
Si,red(t)=ziFAki,red(t)[2ρDi,red+δi,red(t)Di,redρδi,red2(t)][2ρDi,red+δi,red(t)Di,redρδi,red2(t)1ki,red(t)ki,red(t)t],E51
NT,i,red(t)=[ρδi,red2(t)2ρDi,red+δi,red(t)Di,red][2ρDi,red+δi,red(t)Di,redρδi,red2(t)1ki,red(t)ki,red(t)t].E52

In presented relations, one may notice that the change of the shape of a voltammetric electrode does not change its type as a measuring converter. A cylindrical electrode functions also as a converter type I. This electrode in case of this type of electrochemical reaction will create a dynamic error whose value is determined above all by the time constant and the nature of changes in marked ion concentration. In presented relations, one may notice that parameters defining metrological properties of the voltammetric electrode are determined by parameters of marked ions, an electrode radius, a thickness of the diffusion layer, a rate of the electrochemical reaction taking place on the surface of the electrode, and thereby by voltage polarizing the electrode.

4.3. The influence if polarizing voltage on metrological properties of a voltammetric electrode

In order to enforce a certain course of an electrochemical reaction on the surface of a voltammetric electrode, it should be polarized with a proper voltage. In measuring practice, there are different voltammetric methods applied. Most frequently used method is direct current voltammetry. In this method, an electrode used is polarized by voltage with a value changing linearly. The advantage of such a solution is the simplicity of the measuring system. At the same time, its disadvantage is relatively low accuracy. It is a solution connected with relatively great influence of capacitive current.

In order to eliminate the influence of the volume of the double layer on accuracy of voltammetric markings, there are different types of alternating currents voltammetry applied. In such cases, the voltage polarizing a voltammetric electrode has two components: variable and static. A static component is identical with one in the direct current voltammetry, and a variable component may be for example sinusoidal voltage, square wave voltage, or triangle wave voltage.

4.3.1. The influence of polarizing voltage with a value changing linearly

In the method of direct current voltammetry, a voltammetric electrode used for measuring is polarized with a voltage described as follows:

Epol(t)=E0±SUt.E53

In such a case, coefficients of the rate of an electrochemical reaction which are defined by relations (4) and (5) will take on a form:

ki,ox(t)=k0exp{αziFRT[E0+SUtE0]},E54
ki,red(t)=k0exp{(1α)ziFRT[E0SUtE0]},E55

and their derivatives are, respectively:

ki,ox(t)t=αziFRTSUki,ox(t),E56
ki,red(t)t=(1α)ziFRTSUki,red(t).E57

Substituting relations (21) and (22) with relations (56) and (23) and relation (24) with relation (57) we get, respectively:

Si,ox(t)=ziFAki,ox(t){1+[δi,ox2(t)Di,ox]αziFRTSU},E58
NT,i,ox(t)=[δi,ox2(t)Di,ox]{1+[δi,ox2(t)Di,ox]αziFRTSU},E59

and:

Si,red(t)=ziFAki,red(t){1+[δi,red2(t)Di,red](1α)ziFRTSU},E60
NT,i,red(t)=[δi,red2(t)Di,red]{1+[δi,red2(t)Di,red](1α)ziFRTSU}.E61

In relations above, one can see that a flat voltammetric electrode polarized with a voltage as in relation (53) functions as a converter type I. Parameters defining its metrological properties are determined by the rate of changes in polarizing voltage. Hence, their values can be alternated by an appropriate choice of the rate of changes in this voltage.

4.3.2. The influence of polarizing voltage with a sinusoidal variable component

In the method of alternating current sinusoidal voltammetry, the voltammetric electrode used for measuring is polarized with a voltage described this way:

Epol(t)=E0±SUt±Umsinωt.E62

In this case, coefficients of the rate of an electrochemical reaction described by the relations (4) and (5) are denoted:

ki,ox(t)=k0exp{αziFRT[E0+SUt+UmsinωtE0]},E63
ki,red(t)=k0exp{(1α)ziFRT[E0SUtUmsinωtE0]},E64

and their derivatives are, respectively:

ki,ox(t)t=αziFRT(SU+ωUmcosωt)ki,ox(t),E65
ki,red(t)t=(1α)ziFRT(SU+ωUmcosωt)ki,red(t).E66

Because the variable component of polarizing voltage causes changes of ion concentration in the volume of the analyzed solution, also on the surface of the electrode, accordingly to the relation:

[ΔCi,0(t)]t=t[Ci0(t)sinωt]=Ci0(t)tsinωt+ωcosωtCi0(t),E67

then the relation (14) denoted on the surface of the voltammetric electrode polarized by voltage with a static and sinusoidal variable component is:

[Ci,0(t)+ΔCi,0(t)]t=Ci,0(t)t+[ΔCi,0(t)]t=Di2Ci(t)+Ci0(t)tsinωt+ωcosωtCi0(t).E68

Denoting this relation with the use of finite difference method we have:

Ci,0(t)t=Di[Ci0(t)Ci,0(t)δi2(t)]+Ci0(t)tsinωt+ωcosωtCi0(t),E69

which results in:

[δi2(t)Di]Ci,0(t)t+Ci,0(t)=[δi2(t)Di][Diδi2(t)+ωcosωt]{Ci0(t)+Ci0(t)tsinωt[Diδi2(t)+ωcosωt]}.E70

Substituting the relation (70) with relations (17) and (18) we get:

[δi2(t)Di]ii(t)t+[11ki(t)ki(t)tδi2(t)Di]ii(t)={ziFAki(t)[1+δi2(t)Diωcosωt]}{Ci0(t)+ziFAki(t)δi2(t)DiCi0(t)tsinωtziFAki(t)[1+δi2(t)Diωcosωt]}.E71

Keeping in mind the assumptions taken, we can transform the above relation into:

[δi2(t)Di][11ki(t)ki(t)tδi2(t)Di]ii(t)t+ii(t)=ziFAki(t)[1+δi2(t)Diωcosωt][11ki(t)ki(t)tδi2(t)Di]{Ci0(t)+ziFAki(t)δi2(t)DiCi0(t)tsinωtziFAki(t)[1+δi2(t)Diωcosωt]}.E72

It is clear from the above relation that both for the oxidizing reaction and for the reduction reaction, a flat voltammetric electrode polarized by the voltage denoted as in relation (62) functions as a converter type I with properties defined as follows:

Si,ox(t)=ziFAki,ox(t)[1+δi,ox2(t)Di,oxωcosωt][1+αziFRT(SU+ωUmcosωt)δi,ox2(t)Di,ox],E73
NT,i,ox(t)=δi,ox2(t)Di,ox[1+αziFRT(SU+ωUmcosωt)δi,ox2(t)Di,ox],E74
Si,red(t)=ziFAki,red(t)[1+δi,red2(t)Di,redωcosωt][1+(1α)ziFRT(SU+ωUmcosωt)δi,red2(t)Di,red],E75
NT,i,red(t)=δi,red2(t)Di,red[1+(1α)ziFRT(SU+ωUmcosωt)δi,red2(t)Di,red].E76

It is clear that both the sensitivity of the electrode as well as its time constant are determined by the rate of changes of the static component of polarizing voltage and by the amplitude and sinusoidal frequency of the variable component. Hence, the parameters describing metrological properties of the electrode can be influenced by an appropriate choice of polarizing voltage parameters.

4.3.3. The influence of the triangle waveform variable component

In the method of triangular waveform AC voltammetry, a voltammetric electrode used for measuring is polarized by voltage denoted as follows:

Epol(t)=E0±SUt±Ut(t),E77

with a variable component of the polarizing voltage which can be denoted as:

Ut(t)={2AtπT2tT22A(πt)πT2tT.E78

Triangular bipolar waveform component Ut(t) can be denoted with an expansion in the Fourier series:

Ut(t)=8Aπ2n=0(1)nsin[(2n+1)ωt](2n+1)2=8Aπ2(sinωt19sin3ωt+125sin5ωt).E79

In this case coefficients of the rate of electrochemical reactions denoted by the relations (4) and (5) are as follows:

ki,ox(t)=k0exp{αziFRT[E0+SUt+8Aπ2(sinωt19sin3ωt+125sin5ωt)E0]},E80
ki,red(t)=k0exp{(1α)ziFRT[E0SUt8Aπ2(sinωt19sin3ωt+125sin5ωt)E0]},E81

and their derivatives are, respectively:

ki,ox(t)t=αziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]ki,ox(t),E82
ki,red(t)t=(1α)ziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]ki,red(t).E83

Because the variable component of the polarizing voltage causes changes in the ion concentration of the analyzed solution, also on the surface of the electrode, accordingly to the relation:

[ΔCi,0(t)]t=t[8Aπ2Ci0(t)n=0(1)nsin[(2n+1)ωt](2n+1)2],E84

which leads to:

[ΔCi,0(t)]t=8Aπ2Ci0(t)tn=0(1)nsin[(2n+1)ωt](2n+1)2+8Aπ2Ci0(t)n=0(1)nωcos[(2n+1)ωt](2n+1),E85

then the relation (14) denoted for the surface of the voltammetric electrode polarized by the voltage with a static component and a triangular waveform variable component is:

[ Ci,0(t)+ΔCi,0(t) ]t=Di2Ci(t)+8Aπ2 [ (sinωt19sin3ωt+125sin5ωt)Ci0(t)t +ωCi0(t)(cosωt13cos3ωt+15cos5ωt) ].E86

Denoting this relation by using finite difference method, we get:

Ci,0(t)t=Di[ Ci0(t)Ci,0(t)δi2(t) ]+8Aπ2 [ (sinωt19sin3ωt+125sin5ωt)Ci0(t)t +ωCi0(t)(cosωt13cos3ωt+15cos5ωt) ],E87

which results in:

[δi2(t)Di]Ci,0(t)t+Ci,0(t)=[δi2(t)Di][Diδi2(t)+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]{Ci0(t)+8Aπ2Ci0(t)t(sinωt19sin3ωt+125sin5ωt)[Diδi2(t)+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]}.E88

Substituting the relation (88) with relations (17) and (18), we get the relation:

[δi2(t)Di]ii(t)t+[11ki(t)ki(t)tδi2(t)Di]ii(t)={ziFAki(t)[1+δi2(t)Di8ωAπ2(cosωt13cos3ωt+15cos5ωt)]}{Ci0(t)+ziFAki(t)δi2(t)Di8Aπ2Ci0(t)t(sinωt19sin3ωt+125sin5ωt)ziFAki(t)[1+δi2(t)Di8ωAπ2(cosωt13cos3ωt+15cos5ωt)]}.E89

Keeping in mind the assuptions taken, the above relation may be transformed into:

[δi2(t)Di]{1+δi2(t)DiαziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]}ii(t)t+ii(t)=ziFAki(t)[1+δi2(t)Di8ωAπ2(cosωt+cos3ωt+cos5ωt+)]{1+δi2(t)DiαziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]}{Ci0(t)+ziFAki(t)δi2(t)Di8Aπ2Ci0(t)t(sinωt19sin3ωt+125sin5ωt)ziFAki(t)[1+δi2(t)Di8ωAπ2(cosωt13cos3ωt+15cos5ωt)]}.E90

It is clear from the above relation that both for the oxidizing reaction and for the reduction reaction, a flat voltammetric electrode polarized by voltage as in relation (77) functions as a converter type I with properties defined as follows:

Si,ox(t)=ziFAki,ox(t)[1+δi,ox2(t)Di,ox8ωAπ2(cosωt+cos3ωt+cos5ωt+)]{1+δi,ox2(t)Di,oxαziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]},E91
NT,i,ox(t)=[δi,ox2(t)Di,ox]{1+δi,ox2(t)Di,oxαziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]},E92
Si,red(t)=ziFAki,red(t)[1+δi,red2(t)Di,red8ωAπ2(cosωt+cos3ωt+cos5ωt+)]{1+δi,red2(t)Di,red(1α)ziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]},E93
NT,i,red(t)=[δi,red2(t)Di,red]{1+δi,red2(t)Di,red(1α)ziFRT[SU+8ωAπ2(cosωt13cos3ωt+15cos5ωt)]}.E94

It is clear that the sensitivity of an electrode and its time constant are determined by the rate of changes of the static component in the polarizing voltage and by the amplitude and frequency of the fundamental component and individual harmonic components of the triangle waveform polarizing voltage.

Hence, the parameters describing metrological properties of an electrode can be influenced by an appropriate choice of polarizing voltage parameters.

Advertisement

5. Numerical simulations and discussion

Numeric simulations were conducted in order to determine how the electrode’s shape and the shape of the voltage influencing the voltammetric electrode metrological properties.

The influence of the voltammetric electrode’s shape on its metrological properties was analyzed assuming that the electrode is flat, spherical, and cylindrical and polarized only by linearly increasing voltage.

Also, the influence of the shape of the voltage polarizing the electrode on its metrological properties was analyzed assuming that the electrode is flat and polarized by the linearly increasing voltage, linearly increasing voltage with sinusoidal variable component and linearly increasing voltage with a triangular waveform variable component. It was assumed in the simulations that there is an oxidizing reaction of marked ions on the surface of the electrode and the values defining the marked ions, voltammetric electrode’s shapes and polarizing voltages are: zi = 1, α = 0.5, E0 = 0.25…0.75 V; Di = 10−5…10−7 cm/s, A = 1 mm2, r = 0.1…1 mm, ρ = 0.1…1 mm, SU= 5…100 mV/s, Um= 5…100 mV, f = 1…100 Hz.

Numerical simulations prove that the shape of the electrode and its geometrical dimensions influence its sensitivity. When the electrode functions like a converter type 0, the geometrical values influence the sensitivity solely through the surface area of the electrode. However, if the electrode functions like a converter type I, the electrode’s radius influences greatly the sensitivity. The sensitivity of the electrode is the highest when the electrode functions like a converter type 0, which is shown in Figure 1.

Figure 1.

Influence of the shape on the sensitivity of an electrode; E0=0.50 V; Di=106cm2/s,  k0=108 cm/s , A=1 mm2, r=0.5 mm, ρ=0.5 mm, SU=0.025 mV/s; The type of an electrode: (1) converter type 0; converter type I, (2) cylindrical, (3) spherical, (4) flat.

The highest sensitivity is the characteristic of a cylindrical electrode when there are determined: rate of changes in the voltage polarizing an electrode, a rate of oxidizing reaction, an ion diffusion coefficient, and equal geometrical dimensions. It was also proved that spherical and cylindrical electrodes sensitivity is determined by their radius. Reducing geometrical dimensions of a flat voltammetric electrode leads to reduction of its sensitivity. And reducing the radius of spherical and cylindrical electrode leads to an increase of their sensitivity.

Numeric simulation results show that the shape of the electrode influences its time constant. The lowest time constant is a characteristic of a cylindrical electrode when there are determined: rate of changes in the voltage polarizing an electrode, an ion diffusion coefficient, and equal geometrical dimensions of electrodes, which is shown in Figure 2.

Figure 2.

Influence of the shape on the time constant of an electrode; Di=107cm2/s, r=0.5 mm, ρ=0.5 mm, SU=0.010 mV/s; The type of an electrode: (1) flat, (2) spherical, (3) cylindrical.

The results of calculations show that the shape of the voltage polarizing a flat voltammetric electrode influences its metrological properties. It has been proved that a flat electrode polarized by voltage with a triangular waveform variable component has the highest sensitivity when there are determined: an ion diffusion coefficient, steady rate of the reaction, equal parameters of the polarizing voltage, which is shown in Figure 3.

Figure 3.

Influence of the shape of polarization voltage on the sensitivity of an electrode; Di=107cm2/s, SU=0.010 mV/s; Um=5 mV, f=1 Hz; The type of polarization voltage: (1) direct voltage, (2) direct voltage with a sinusoidal component, (3) direct voltage with a triangular component.

The sensitivity of an electrode polarized with such a type of voltage is determined by the number of harmonics of the electrode’s triangular waveform variable component. Including in calculations further, harmonics of this component increases the sensitivity of the electrode. The results of such calculations show that the sensitivity of a flat voltammetric electrode is also determined by the frequency of a polarizing voltage variable component. The increase of a frequency of sinusoidal and triangular waveform component voltage leads to the increase of a voltammetric electrode’s sensitivity. The lower the rate of change of polarizing voltage static component the greater the increase of sensitivity. The amplitude of the polarizing voltage variable component influences the sensitivity of the electrode as well. The increase of this component’s amplitude leads to the reduction of the electrode’s sensitivity.

It was also proved that the time constant of a flat voltammetric electrode is determined by the shape of polarizing voltage. Results of these calculations revealed that the electrode polarized by voltage only with a static component has the highest time constant. The increase of the rate of changes in this voltage leads to the decrease of time constant of the electrode. A flat voltammetric electrode has the lowest time constant for voltage with a sinusoidal variable component when there are determined: rate of changes of the polarizing voltage static component and amplitude and frequency of the variable component, which is shown in Figure 4.

Figure 4.

Influence of the shape on the time constant of a flat electrode; Di=107cm2/s, r=0.5 mm, ρ=0.5 mm, SU=0.010 mV/s; Um=5 mV, f=1 Hz. The type of polarization voltage: (1) direct voltage, (2) direct voltage with a sinusoidal component, (3) direct voltage with a triangular component.

The increase of the variable component amplitude leads to the decrease of the voltammetric electrode time constant. Also an increase of a polarizing voltage variable component frequency leads to the decrease of the electrode’s time constant. In the case when a flat electrode is polarized by voltage with a triangular waveform variable component including this electrode’s individual harmonics leads to the decrease of the time constant. The electrode’s time constant also decreases when the rate of gain of the electrode polarizing voltage static component increases.

Advertisement

6. Conclusions

Analyses conducted show that depending on the concentration of marked ion solution, the voltammetric electrode may function like a converter type 0 or type I. In the first case, its metrological properties in the time domain are determined solely by sensitivity, in the latter case by sensitivity and the time constant. Values of both parameters are determined by the type of marked ions in the analyzed solution, by the shape of the electrode and the shape of the polarizing voltage. Results of simulations conducted show that in the case of marking low ion concentrations, the cylindrical electrode has the best metrological properties. It is characterized by the highest sensitivity and the lowest time constant. It was also proved that the most beneficial metrological properties are the ones of a voltammetric electrode polarized by voltage with a static component and a sinusoidal variable component. The choice of the amplitude and the frequency of polarized voltage variable component may influence both sensitivity and the time constant of a voltammetric electrode used in markings. As a result of such activities metrological properties of a used measuring electrode may be shaped in an optional way. The ion concentration may be measured with a high sensitivity or low time constant. It will allow to match closer metrological properties of a voltammetric electrode to the expected nature of changes in marked ion concentration and as a consequence it leads to the decrease of static and dynamic errors of electrochemical markings conducted.

Advertisement

Major symbols

ASurface area of electrode
Ci0(t)Bulk concentration of ions i at time t
Ci,0(t)Surface concentration of ions i at the electrode surface at time t
Ci(x)tGradient of concentration of ions i at time t at voltammetric electrode
Ci(x)Concentration gradient of ions i at distance x
Ci(x)Concentration gradient of ions i at distance x
Ci,ox,0(t)Concentration of oxidation ions i at the electrode surface at time t
Ci,red,0(t)Concentration of reduction ions i at the electrode surface at time t
DiDiffusion coefficient of ions i
E0Standard potential of an electrode
E0Initial potential
FThe Faraday constant
i(t)Current
iox(t)Anodic oxidation current
ired(t)Cathodic reduction current
ki,ox(t)Heterogeneous rate constant for oxidation
ki,red(t)Heterogeneous rate constant for reduction
k0Rate constant for a heterogeneous reaction
Ni(t)Flux gradient of ions i
uiMobility of ions i
ziCharge on ions i in signed units of electronic charge
SULinear potential scan rate
tTime
TAbsolute temperature
UmAmplitude of a sinusoidal component
Ut(t)Amplitude of a triangle component
UPotential gradient
ViVelocity of ions i
αTransfer coefficient
ωAngular frequency of rotation
δi(t)Diffusion layer thickness for ions i at the electrode surface at time t
δi,ox(t)Diffusion layer thickness for oxidation ions i at the electrode surface at time t
δi,red(t)Diffusion layer thickness for reduction ions i at the electrode surface at time t

References

  1. 1. Pierre MB, Jacques B. Polarography, voltammetry and tensammetry: tools for day‐to‐day analysis in the industrial laboratory. Analyst. 1989; 114(12):1531–1544. DOI : 10.1039/AN9891401531
  2. 2. Robert IM, Alan MB, Terence JC, Robert WC, Roger WK, Michael GNO, Bruce RC On‐line and off‐line voltammetric methods for the determination of nickel in zinc plant electrolyte. Analyst. 1994; 119(5):1057–1061. DOI: 10.1039/AN994190105
  3. 3. Alan MB. Past, present and future contributions of microelectrodes to analytical studies employing voltammetric detection. A review: Analyst. 1994, 119(11):1R–21R. DOI: 10.1039/AN994190001R
  4. 4. Williams G, D’Silva C. Field‐based heavy metal analyser for the simultaneous determination of multiple cations on‐site. Analyst. 1994; 119(11):2337–2341. DOI: 10.1039/AN9941902337
  5. 5. Damien WMA. Tutorial review. Voltammetric determination of trace metals and organics after accumulation at modified electrodes. Analyst. 1994; 119(9):1953–1966. DOI: 10.1039/AN9941901953
  6. 6. Roy MH. Understanding Our Environment, an Introduction to Environmental, Chemistry and Pollution. 3rd ed. Royal Society of Chemistry; 1999. p. 485. ISBN 0‐85404‐584‐8. DOI:10.1039/9781847552235-00071.
  7. 7. Allen JB, Larry RF. Electrochemical Methods: Fundamentals and Applications. 2nd ed. New York: John Wiley & Sons Inc.; 2001. p. 864. ISBN: 978‐0‐471‐04372‐0
  8. 8. Patrycja C, Wojciech W. Sensor arrays for liquid sensing – electronic tongue systems. Analyst. 2007; 132(10):963–978
  9. 9. Wassana Y, Kitiya H, Cynthia LW, Daiwon C, Thanapon S, Mychailo BT, Marvin GW, Glen EF, Shane RA, Charles T. Direct detection of Pb in urine and Cd, Pb, Cu, and Ag in natural waters using electrochemical sensors immobilized with DMSA functionalized magnetic nanoparticles. Analyst. 2008; 133(3):348–355. DOI: 10.1039/B711199A
  10. 10. Niina JR, Brian HH, William RH. Electrochemical biosensors. Chemical Society Reviews. 2010; 39(5):1747–1763. DOI: 10.1039/B714449K
  11. 11. Jonathan PM, Rashid OK, Craig EB. New directions in screen printed electroanalytical sensors: an overview of recent developments. Analyst. 2011; 136(6):1067–1076. DOI: 10.1039/C0AN00894J
  12. 12. Rongsheng C, Yong L Kaifu H, Paul KC. Microelectrode arrays based on carbon nanomaterials: emerging electrochemical sensors for biological and environmental applications. Royal Society of Chemistry Advances. 2013; 3(41):18698–18715. DOI: 10.1039/C3RA43033B
  13. 13. Jonathan PM, Edward PR, Craig EB. Screen‐printed back‐to‐back electroanalytical sensors. Analyst. 2014;139(21):5339–5349. DOI: 10.1039/C4AN01501K
  14. 14. de la MGM, Rosario MPM, Lorenzo CB, Eduardo PG. Applicability of the bismuth bulk rotating disk electrode for heavy metal monitoring in undisturbed environmental and biological samples: determination of Zn(II) in rainwater, tap water and urine. Analytical Methods. 2014; 6(21):8668–8674. DOI: 10.1039/C4AY01626B
  15. 15. Abdel‐Nasser K, Nadeem B, Muhammad S. Graphite pencil electrodes as electrochemical sensors for environmental analysis: a review of features, developments, and applications. Royal Society of Chemistry Adavances. 2016; 6(94):91325–91340. DOI: 10.1039/C6RA17466C

Written By

Krzysztof Suchocki

Submitted: 05 October 2016 Reviewed: 17 February 2017 Published: 07 June 2017