Abstract

Kjeldahl method using concentrated boric acid is a common practice in many laboratories. A thorough study of the titration with hydrochloric acid of ammonia trapped in a solution of boric acid is made in an attempt to explain the fundamentals of a widely applied standard method. A new potentiometric method for the determination of the end point in the Kjeldahl titrimetric finish is proposed based on the linearization of the titration curve of the ammonia‐boric acid system. The method is strictly based on mole and charge balances, and no approximations are made in deriving the equations. The proposed method has proved very accurate when applied to synthetic titration curves and data. Some problems, however, are experienced in the practice, because the behavior of the experimental system studied is far from the expected one on the basis of the theoretical model. However, a slight modification of the devised method has been applied to the experimental titration of ammonia with hydrochloric acid, in boric acid as trapping solution, getting good results.

Keywords

  • Kjeldahl method
  • titrimetric finish
  • ammonia‐boric acid system

1. Introduction

Nearly 130 years ago, on March 7, 1883, at a meeting of the Danish Chemical Society, Johan Gustav Christoffer Thorsager Kjeldahl (Head of Chemistry Department of the Carlsberg Foundation Laboratory of the Danish Brewing Carlsberg Company) introduced a method known later under the eponym “Kjeldahl Method” that basically is still in use. Within the same year, the method was published in a German journal Zeitschrift für Analytische Chemie [1] and written in French and Danish languages in communications from the Carlsberg Laboratory.

The Danish brewer Mr. J.C. Jacobsen had a great respect for Pasteur and his work for the French wine industry, and for this reason, French extensive summaries of the Carlsberg papers were also published. As an extended summary of the Kjeldahl paper appeared in Chemical News in August [2], the method was quickly taken up [3]. The Analyst first gave details of the method in 1885 [4, 5], although the method had been briefly mentioned by Blyth [6] though giving the Kjeldahl name incorrectly as Vijeldahl. A surprisingly short period went by between the publication of the Kjeldahl method and the appearance of new publications concerning the topic, both in Europe and America. None of the analytical methods has been as widely chosen, in so short a time [7], as the “Kjeldahl Method” for the estimation of nitrogen.

The Kjeldahl method was originally designed for the brewing industry as an aid in following protein changes in grain during germination. It was Berzelius, who suggested the use of the word “protein” in 1838 in a letter to Mulder because it was derived from the Greek word meaning “to be in the first place” [8]. The Kjeldahl method lacks of analytical selectivity because does not distinguish between protein‐based nitrogen from non‐protein nitrogen (NPN). Adulteration incidents (i.e., adulteration of protein‐based foods with melanine and related nonprotein compounds) exploiting this analytical vulnerability have been recently detected and are new examples of a problem that dates back to before the Kjeldahl method was introduced [9]. The presence of NPN compounds in foods (amino acids, ammonia, urea, trimethylamine oxide) overestimates their true protein content as derived from the current nitrogen determination methods. Separation of NPN from true protein nitrogen may be carried out by addition of a protein precipitating agent such as trichloroacetic acid or perchloric acid.

The protein content in a foodstuff is estimated by multiplying the nitrogen content by a nitrogen-to-protein conversion factor, usually set at 6.25, which assumes the nitrogen content of proteins to be 16%. However, pure proteins differ in terms of their nitrogen contents because of differences in their amino acids composition, ranging from 13.4 to 19.3%. So, different multiplying factors are suitable for samples of different kind.

The protein content determines the market value of major agricultural commodities. In addition, the quantitative analysis of protein content is necessary for quality control, being also a prerequisite for accurate food labeling. In recent years, soy protein products have increased the interest of consumers, especially in Western cultures, due to the high‐quality protein of soy foods together with their associated health benefits [10] (according to FDA [11], 25 g of soy protein per day may improve cardiovascular health).

Table 1 [1232] gives the chemical methods, most commonly used for protein determination. Some of the most significant methods (Dumas, Kjeldahl, and biuret assay) date from the late 1800s. Since the nineteenth century, many other analytical methodologies have been developed to determine the total protein in the field of biochemistry, biology and proteomics, but most of them to address research needs and not necessarily to determine the purity and/or adulteration of food products.

Technique/eponymReferences
Dumas method[13]
Nessler reagent[14, 15]
Biuret method[1618]
Berthelot’s method (Alkali‐phenol reagent)[19, 20]
Kjeldahl method[1, 2]
Folin‐Ciocalteu[21]
Dye binding[22]
Lowry method[2326]
Direct alkaline distillation[27]
NIR (Near‐infrared reflectance)[27]
Modified Berthelot reaction[20]
Modified Lowry method[25, 28]
Bradford method (Coomassie blue dye‐binding method)[29, 30]
BCA (Bicinchoninic acid method)[31]
3‐(4‐carboxy benzyl)quinoline‐2‐carboxaldehyde[32]

Table 1.

Methods for food protein analysis: some key references [12].

Though there are several experimental approaches to evaluate the nitrogen content in different kinds of samples (Dumas combustion method, NIR methods), the Kjeldahl method still remains as the reference method, being really the “golden standard” for validating other quantifying methodologies in the biopharmaceutical and food industries [9, 33]. The Kjeldahl method is applied in official methods [34] to determine the nitrogen content measurement in foods as well as in many other samples, pharmaceutical, agricultural, food products, biological sediments and surface and waste waters. The diversity of papers dealing with Kjeldahl method [12] is attributable to the immense usefulness of the method, to its need for modifications for applications to various types of organic and inorganic compounds, and to the search for catalysts to provide such modifications and to accelerate the digestion [35, 36].

W. Johannsen (1857–1927), a pharmacist, wrote the Kjeldahl obituary, first published in German [37] and then translated to French [38] and English [39]. Johannsen, one of the founders of the science of genetics, was in his beginnings an assistant in the chemistry department at the Carlsberg Laboratory under the chemist Johan Kjeldahl and is well known as coined the term gene in 1909. Kjeldahl was elected to membership in the scientific academies of Denmark and Christiania and received an honorary doctorate from the University of Copenhagen. Kjeldahl (like Nessler) has been verbalized, an honor not usually accorded to a chemist [40]. All chemists understand what it means when it is said that a substance was kjeldahled [5] or that one kjeldahlizes a sample [41]. Kjeldahl was predecessor to S.P.L. Sörensen as Head of the Carlsberg Laboratory in Copenhagen, who introduced the notation of pH [41, 42].

Advertisement

2. The Kjeldahl’s three steps

Titration analysis is one of the oldest analytical methods, and as a matter of fact, it plays an important role studies [4346] in various analytical fields as well as in routine analysis [4750]. Generally, the quantity of tested components of a sample is determined by adding to the measured sample an exactly known quantity of the standard titration solution with which the desired constituent reacts in a definite, known proportion. The content of the components is calculated according [45, 46] to the concentration of the standard solution, the consumed volume, the measuring relationship between chemical reactions and the weight of the tested substances, etc. If the tested substance cannot chemically react with the titrant directly, indirect ways of other reactions can be applied to measure its content.

Although some analytical techniques allow the direct determination of species without sample treatment, it is usually necessary to dissolve the sample prior analysis. The analytical adage: “once the problem is dissolved, the problem is solved” denotes clearly this fact. Most analyses are preferentially performed on solution samples, a more homogeneous analysis sample representative of the bulk properties of a large solid sample being thus obtained, improving precision and accuracy [51]. Digestion may be defined as the process in which a complex substance is decomposed into volatile gases and simple salts that are soluble in dilute acid solution [52]. Wet decomposition or acid digestion involves the use of mineral acid, alone or in combination with other acids and oxidizing agents to affect the dissolution of a sample [53].

The three steps [1, 54] of Kjeldahl method includes:

  1. Wet acidic (conc. H2SO4) digestion (mineralization) of nitrogen‐containing sample, in a long‐necked digestion flask, causing its conversion into NH4+ ions. The Kjeldahl flask, which he constructed in 1888 to simplify the method, is still in use today. Traditionally, Kjeldahl flasks with a capacity for 500–800 mL and gas or electric heating have been used for the digestion. The digest must contain residual H2SO4 to retain the NH3 as NH4+. Water is added manually or automatically to prevent the digest from solidifying, which also may cause bumping, and to avoid mixing concentrated alkali with concentrated acid [55] during the distillation.

  2. Transformation of NH4+ ions into NH3(neutralization with alkali) and distillation of the NH3. The flask is heated after the addition of water and alkali to the digested sample, in order to distill a volume of distillate and collect NH3 in the acidic distillation receiver. Direct steam distillation drastically decreases the time needed for distillation [55]. Experimental conditions including rate of steam flow, vigor of distillation and volume of solution must be correctly balanced to ensure efficient removal of ammonia without any carry‐over of alkaline spray.

  3. Titration of the solution from the distillation receiver. The ammonia from the distillation is frequently collected in an excess of standard acid and determined by a back titration with standard alkali solution. A more common practice is the use of boric acid for trapping ammonia. The titration in this case should be carried out as soon as possible after the distillation is complete, ensuring that the temperature of the distillate does not exceed 25°C. Under these conditions, losses of ammonia are avoided [56] (EN ISO 5983‐1, 2009).

Complete details [12] of the three Kjeldahl steps are too lengthy to include here. A series of factors such as sample origin, homogeneity, stability, laboratory skillfulness, sample handling procedures, and composition are of critical importance, as are the size of the test portion taken for analysis, and concentration of the titrant used in the Kjeldahl analysis [57]. A relatively large analytical sample (1–2 g) was used in the original method, requiring large amounts of acid.

Green alternatives exert considerable pressure to ensure the safe disposal of mercury (when used as catalyst) and, especially, to minimize acid usage [5860]. The use of Micro‐Kjeldahl methods is common practice in order to reduce the amount of acid fumes and also require less acid and catalyst mixture.

The introduction of aluminum blocks (at Tecator, now FOSS) in the early 1970 [61] made possible to improve the speed and accuracy of the digestion procedure, thus saving space, chemicals and energy [60]. The digestion system has since been improved, and the distillation step has been speeded up by the use of the semi‐automated systems now available. Block digestions followed by steam distillation are named as rapid Kjeldahl. As a result of technical innovations, there are also available fully automated protein analysis systems that are based on the classical Kjeldahl procedure, for example, the Kjeltec series of Foss Tecator manufactures.

Advertisement

3. The ammonia determination: titrimetric finish

Transformation of NH4+ into NH3 caused by addition of NaOH (pH growth) into Pregl‐Parnas‐Wagner results [12] from equation

[NH3][NH4+]=10pHpK1,aE1

where pK1,a = −log K1,a; K1,a refers to reaction NH4+ = H+ + NH3, pK1,a = 9.35 at 20°C.

The distillation titration method is a standard procedure used by most laboratories to measure ammonium nitrogen in the total Kjeldahl nitrogen digests of various kinds of agricultural and environmental samples [62]. Ammonia may be collected into a solution of a strong acid (HCl or H2SO4). After distillation, the excess of standard acid may be iodometrically determined with starch as indicator, as was done by Kjeldahl, according to consecutive reactions IO3 + 5I+ 6H+= 3 I2+ 3H2O; I2 +2 S2O32− = S4O62−+ 2 I, though this method is seldom used. The excess of standard acid is instead usually titrated with a strong base. Two standard solutions are then needed: (1) titrant (NaOH solution) and (2) the strong acid in the receiver.

Titration of ammonia absorbed in H3BO3 solution was proposed first by Winkler [63] which notes “Boric acid is indeed such a week acid, that its solution does not cause a noticeable colour change of certain indicators. Ammonia is, however, completely fixed by it, provided that a suitable excess of acid is present.” Boric acid is also commonly used to trap ammonia because only one standardized solution (e.g., HCl or H2SO4) is needed (as titrant).

Mixtures of methyl red and tetrabromophenol blue or methyl red and methylene blue (2:1) (Tashiro indicator) or methyl red alone were suggested as indicators; the latter changes its color at about pH 5.2. After addition of HCl, the Tashiro indicator changes its color from green to violet. Alternatively, a pH titration with end point (Veq) at a preset pH value of 5.0 is also done. A mixed indicator (bromocresol green and methyl red) has been recently used by Beljkaš et al. [64] to locate the end point.

Advertisement

4. Base or acid as titrant: the boric acid trapping choice

The ammonia can be distilled into an excess of standard strong acid (HCl or H2SO4), and the excess determined by back titration with a strong standard base, since the solution at the equivalence point contains (NH4)2SO4, or NH4Cl, which hydrolyses,

pH=logKaCA=pKa+pCA2=9.36+1.302=5.33E2

(i.e., 0.05 M of ammonium chloride at the end point), methyl red (transition range 4.4–6.2) being used as an indicator [65].

A mistake appears in the (excellent) Bradstreet monograph [66], p. 152, who wrote “in the case of back titration of a distillate, the equivalence point will occur at pH 7, since this is the point of neutralization of a strong acid by a strong base. Any indicator, therefore, changing colour at or close to pH 7 is suitable.” The millimoles of NH3 in the sample are equal to the total millimoles of HCl added minus millimoles of base used in back titration. This method has two disadvantages: i) the amount of NH3 is obtained as a difference; and ii) two standard solutions are required.

Preferably, the distilled ammonia can be absorbed into a solution of boric acid [63] or other weak acid [67]. Boric acid is sufficiently acid to react with ammonia and prevent loss by volatilization, but it is apparently too weak an acid to interfere with the titration of ammonium borate with diluted hydrochloric acid. When boric acid is used instead of standard acid as the recipient of the distillate, the use of a weak standard alkaline solution is avoided, which suppose a distinct advantage. The alkaline solution is vulnerable to the absorption of atmospheric carbon dioxide with subsequent changes in molarity. Neither the amount nor the concentration (about 4%) of boric acid in the receiving bottle has to be precise.

Then, in the Winkler modification, the NH3is caught in an unmeasured excess of boric acid

NH3+H3BO3=NH4++H2BO3E3

The borate formed is determined by titration with standard HCl, one mole of HCl being required for each mole of NH3

H2BO3+H+=H3BO3E4

The solution at the equivalence point contains H3BO3 and NH4Cl, a mixture of two weak acids,

pH=log(Ka1CA1+Ka2CA2)E5

so that an indicator transiting in the acid region (pH 5–6) is satisfactory [65].

Note that the acidity constants of boric acid and ammonium ion are very similar. The equilibrium constant of Eq. (3) is given by

Keq=[NH4+][H2BO3][NH3][H3BO3]=K1,H3BO3K1,NH4+=109.24109.35=1.288E6

Though this equilibrium constant is low, the fraction of ammonia converted into ammonium ion increases with increasing the difference of concentrations between boric acid and ammonia. As we are studying the trapping process (without titrant added), the volume may take as constant and the mass balances are given by

CH3BO3= [H3BO3]+[H2BO3]E7
CNH3= [NH3]+[NH4+]E8

The following relationship is satisfied in the trapping solution (if solution is not very diluted, the contribution of dissociation of water being in those cases is negligible).

[NH4+]=[H2BO3]E9

and then, by combining Eqs. (6) to (9), we get

Keq=[NH4+]2(CNH3[NH4+])(CH3BO3[NH4+])E10

which on rearrangement gives the second degree equation

(11Keq)[NH4+]2(CNH3+CH3BO3)[NH4+]+CNH3CH3BO3=0E11

which may be solved for given concentrations of boric acid and ammonia. Once the value of [NH4+] is known, that is, the fraction of ammonium ion present

f1,α=[NH4+]CNH3E12

the pH values may be calculated from Eqs. (2) and (12)

[H+]=K1,α[NH4+][NH3]=K1,αf1,αf0,α=K1,αf1,α1f1,αE13

Hydrochloric acid titrates then the borate ion from the above reaction (Eq. (4)) as well as the ammonia, which is not converted into ammonium ion in Eq. (3), that is, H2BO3 + H+ = H3BO3; NH3 + H+ = NH4+ (the sum of both being equivalent to the ammonia distilled).

The trapping process may be illustrated by means of semilogarithmic diagrams (Figure 1). Figure 1 shows the fraction of ammonium ion (equivalent to the borate ion), at a given p[NH3] for varying concentrations of trapping boric acid. In the absence of other equilibria than those considered previously, this diagram would help to known the concentration of ammonia from the pH initial of the sample to be titrated, that is, a pH of about 7 for a 2% boric acid trapping solutions corresponds to a p[NH3] of about 2. However, in numerous cases, things are not an easy business as may appear at first sight and apparently some complications arise in the boric acid‐ammonia system associated when high boric acid concentrations are present. When significant quantities of ammonia vapor are delivered into 4% aqueous boric acid containing methyl red‐methylene blue indicator, it is necessary to dilute with distilled water in order to the expected change in color (purple to green) may occur. Only after considerable dilution had been effected could the ammonia [68] be satisfactorily titrated with hydrochloric acid. Boric acid forms [68, 69] polymeric borate species (triborate, tetraborate, pentaborate) in concentrated solutions whose acidity constants are greater than the K1,b of boric acid by a factor between 210 and 440, behaving thus boric acid as a stronger acid than in when it is diluted.

Figure 1.

Semilogarithmic diagram molar fraction of ammonium ion versus minus logarithm of the total ammonia concentration (p[NH3]), at varying percentage of boric acid as trapping solution.

As far as we know, the anomaly behavior of boric acid was revealed first by Prideaux [70] one century ago when studying the titration curve of boric acid, and then by Thygesen [71] through conductivity measurements. In spite of the fact that Kjeldahl titration is one of the titrimetric methods more applied worldwide, scarce mention of this fact has been made in the literature. However, some problems posses the location of the end point in Kjeldahl titration for the indicators recommended, for example, bromophenol blue, methyl red, methyl orange, and congo red, as we have previously indicated.

This fact is another argument in favor to carry out a potentiometric study of the ammonia‐boric acid system with the aim of to devise a straight‐line linearization of titration curve in order to accurately locate the end point of the titration curve. The theoretical background for the new method is outlined in the following sections.

Advertisement

5. Titration curves in the ammonia‐boric hydrochloric acid titration system

A thorough study of the titration of ammonia trapped in a solution of boric acid is made in this section based on a mathematical approach with equations derived for the titration curves involved. The calculations presented are based on charge and concentration balances, and expressions for equilibrium constants related to acid‐base equilibria. This approach can be perceived as the clear confirmation of the statement [72], ascribed to J.C. Maxwell that “a good theory is the best practical tool.

Let us consider the titration of ammonia trapped in boric acid solution with hydrochloric acid as titrant. The initial concentration of ammonia and boric acid is denoted by CNH3 and CH3BO3, respectively, and the initial volume by V0, CHCl denotes the concentration of strong acid and V the volume added. The electroneutrality rule for the solution is

[H+]+[NH4+]=[OH]+[Cl]+[H2BO3]E14

Eq. (14) should also include concentrations of ions originating from the neutral salt added to adjust ionic strength (when necessary). As the salts used for this purpose usually potassium nitrate and chloride are completely dissociated, the concentration of the anion and cation is the same and cancel out in the equation. Also, the contribution of the second and third dissociation steps of boric acid is considered negligible at the working pH range.

The law of mass action holds for the reactions

NH4+NH3+H+K1,ac=[NH3][H+][NH4+]E15
H3BO3H2BO3+H+K1,bc=[H2BO3][H+][H3BO3]E16

The constants K1i (i = a, b) are concentration constants, useful when the pH‐meter is calibrated in term of hydrogen ion concentrations at an ionic strength fixed. If this not the case, mixed constants (at an ionic strength fixed), where the activity of the hydrogen and hydroxide ions (denoted by (H+) and (OH)) are used in conjunction with concentrations of all other species. The activity of the hydrogen ion is related to concentration by the expression

(H+)=[H+]γHE17

where γH is the activity coefficient of the ion. Then

K1,aB=[NH3](H+)[NH4+]=K1,aCγH+K1,bB=[H2BO3](H+)[H3BO3]=K1,bCγH+E18

The total concentrations of the ammonia and boric acid during the titration can be expressed by their respective mass balances

[H3BO3]+[H2BO3]=V0CH3BO3V0+VE19
[NH4+]+[NH3]=V0CNH3V0+VE20

The concentration of the chloride ion can be expressed by the following equation

[Cl]=VCHClV0+VE21

The molar fractions of ammonia and borate ions are given by

f1,a=[NH4+][NH3]+[NH4+]=[NH4+][NH3]1+[NH4+][NH3]=[H+]K1,ac1+[H+]K1,ac=10pK1,acpH1+10pK1,acpHE22
f0,b=[H2BO3][H3BO3]+[H2BO3]=11+[H3BO3][H2BO3]=11+[H+]K1,bc=11+10pK1,bcpHE23

and the ammonia and borate concentrations expressed in term of their respective molar fractions may be substituted into Eq. (14), giving

[H+]+f1,aV0CNH3V0+V=[OH]+VCHClV0+V+fo,bV0CH3BO3V0+VE24

and then we get

Δ=[H+][OH]=VCHCl+fo,bV0CH3BO3f1,aV0CNH3V0+VE25

From the Eq. (25), we can, after a simple mathematical manipulation, obtain

V=V0f1,aCNH3fo,bCH3BO3+ΔCHClΔE26

thus obtaining a relatively simple, closed‐form expression for the titration curve valid at any moment of the titration. It relates the volume of added titrant to the hydrogen ion concentration (p[H]) of the solution. Then, by substituting a series of values of [H+], we calculate the corresponding values of V.

Advertisement

6. Graphical method for the determination of the equivalence point

Numerical methods based on the mathematical modeling of the titration curve may be applied for the determination of the end point instead of approximate methods. Gran’s linearization known since 1950 [7375] is one of the examples. In this chapter, a new method for the determination of the end point is proposed based on the linearization of the titration curve of the ammonia‐boric acid system. The method is inspired in a previous method described first by Schwartz [76] and also explained with great detail in two relatively recent analytical chemistry textbooks [77, 78].

Eq. (25) gives

(V0+V)Δ=V·CHCl+fo,bV0CH3BO3f1,aV0CNH3E27

and dividing the left and right member through CHCl we have

(V0+V)ΔCHCl=V+f0,bV0CH3BO3CHClf1,aVeqE28

The following equation should be valid at the equivalence point

V0CNH3=VeqCHClE29

and then, by combining Eqs. (31) and (30), we get

V(V0+V)Δf0,bV0CH3BO3CHCl=f1,aVeqE30

The left member of Eq. (32) is denoted by V'

V´=V(V0+V)Δf0,bV0CH3BO3CHClE31

and taking into account that Eq. (22) is equivalent to

f1,a=[H+][H+]+K1,aE32

we get

V´=[H+][H+]+K1,aE33

By simple manipulation, Eq. (35) gives

V´([H+]+K1,a)=[H+]VeqV´K1,a=[H+]VeqV´[H+]E34

By multiplying the left and right members of Eq. (36) through 1/(K1,a [H+]) we finally have

V´[H+]=VeqK1,aV´K1,aE35

Eq. (35) gives a straight line when V'/[H] is plotted against V’. The plot has a slope of −1/K1,a and intersects the V’ axis at the point Veq. By use of Eq. (35) the equivalence point can be located with considerable accuracy. Eq. (35) may be considered as a variant (an extension) of the method of Schwartz [76] for the determination of the equivalence point in the titration of a weak base with a strong acid.

The method, which we propose, is based strictly on mole and charge balance equations. No approximations are made in deriving Eq. (35). Both the dissociation of water as well as the dilution effects is precisely accounted for. The only restrictions of the method are, on the one hand, the accuracy with which the pH‐meter and electrodes can be calibrated, and on the other hand, that linear regression entails several fundamental assumptions that are not always satisfied with data obtained. One, for example, is that the values of x are free from error but those of y are drawn from a population having normally distributed errors [79, 80]. When the precision of experimental measurements is very high, no special problems, however, are to be expected.

Note that the use of the complete Eq. (31) which gives the modified volume function V’ presupposes knowledge of the pK1,b, necessary to obtain the f0,b value (Eq. (23)). In any case, the pK1,b value may be extracted from appropriate tables, or from the analytical bibliography [81], though it is not necessary to be known exactly. A trial value of pK1,b may be assumed and values for Veq and K1,a calculated by least squares method on Eq. (35). The procedure is repeated for other assumed values of pK1,b, and the best value is taken as that for which minimizes sy/x. the standard deviation of the corresponding regression line. The sy/x values can be easily got using linear regression (method of the least squares), in EXCEL, with the function LINEST.

Taking into account Eqs. (3) and (4), the computation of ionic strength may be made as a first approximation, when V < Veq, as

I=12(2CNH3V0CHClVV0+V+[OH])E36

and when V > Veq as

I=12(CNH3V0V0+V+[H+])E37

As the activity coefficient γH will change only to a very small extension during a titration, log γH may be considered constant. A large ratio of sample volume to titrant volume (i.e., 10:1) minimizes errors introduced by variations in activity coefficients due to dilution. The value of most constants has been determined at an ionic strength of I = 0.1. It is thus practical to adjust the ionic strength to this value. However, addition of a basal electrolyte into the solution and the titrant to keep the ionic strength approximately constant and rather high is not usually practiced [82] in the titrations involved with Kjeldahl method.

γH ≈ 1 is assumed for simplicity of considerations when the titrations are carried out at low ionic strength. The success of the system of pH standardization depends on the validity of putting–log (H+) equal to pH. As [H+] normally makes only a small contribution, no sensible error is introduced if (H+) is used in its place: in some exceptional cases, when [H+] is relatively more important, the quantity may be estimated with sufficient exactness from Eq. (17); on the acid side of pH (when [OH] is negligible).

If we represent the pH as A ± ε, where ε is the uncertainty, then we have [H+] = 10−(A±ε) = 10‐A times 10±ε, where 10±ε = 1 for ε = 0 [77]. The resulting uncertainty in [H+] is now a multiplicative factor, and therefore a relative uncertainty, which applies regardless of the value of pH. A relatively small error in the pH of 0.01, 0.02, or 0.05 corresponds to an uncertainty in [H+] of about 2, 5 or 12%, respectively. Consequently, Eq. (35) plot is characterized by extreme sensitivity to small changes in pH. This problem is compounded by possible activity effects and by the requirement to use a precise value of pK1,b.

Differential (approximate) methods for the determination of end points of titration are based on the presumption that the end point of a titration is the inflection point of the titration curve [83, 84], where the absolute value of the first derivative reaches a maximum (titration of a weak acid with a strong base) or a minimum (titration of a weak base with a strong acid) and the second derivative changes of sign. A number of points very closely spaced and preferably of high precision are needed [83, 84] in order that the method is successfully applied. Only the points in the vicinity of the inflection are used for the calculation. This may result in increasing errors as titration data are least accurate right near the end point [85], because buffering is minimal and electrode response is sluggish. The local pH fluctuations in the inflection region are mostly due [77] to insufficiently rapid mixing of the titrant and the sample, and localized pH sampling by the glass electrode.

Two advantages may be ascribed to the method devised in this chapter, based on the Eq. (35). Fewer titration points need to be taken than with conventional methods, and measurements need not be made close to the equivalence point since this point may be obtained by extra(inter)polation. Therefore, problems related to incompleteness of reaction or instability of measurements close to the end point might be thus avoided. Note that, when very weak acids or bases are titrated, the approximations assumed by Gran when deriving their equations are no longer valid.

Harris [85] have comment in reference to the linearization of titration curve by Gran [74]: “The beauty of a Gran plot is that it enables us to use data taken before the end point to find the end point.” This sentence is undoubtedly applied to linear extrapolation methods based [7678, 86] on an improvement in the la methodology proposed by Gran, as it is the case in this chapter.

Advertisement

7. Approximation expressions derived from the complete equation

Eq. (31) may be expressed as

V´=V(V0+V)ΔCHCl+f0,bV0CH3BO3CHClE38

Which is in the form

V´=V+U+WE39
U=(V0+V)ΔCHClE40
W=f0,bV0CH3BO3CHClE41

Figure 2 depicts the contributions of the different terms in Eq. (39) to the value of V'. It is valid for a titration of 150 mL of 0.0096 M ammonia solution and 0.2156 M trapping acid boric solution, titrated with HCl 0.12 M (pK1,a = 9.27; pK1,b = 9.12; pKwc = 13.80). It is seen that before the equivalence point, the U term may be considered as negligible whereas beyond the equivalence point it is the W term, which may be neglected.

Figure 2.

Contributions of the different terms in Eq. (38) to the value of V'.

Thus, before the equivalence point we approximately get

V´=V+f0,bV0CH3BO3CHClE42

Note that, the mole fraction of borate varies from about 0.043 at the beginning of the titration to about 0 at the equivalence point (Figure 3). Despite the fact that initial volume V0 and concentrations of HCl and H3BO3 are constants, the contribution of the second term of the right hand of Eq. (42) is null only when approximate at the equivalence point. There is no way to use a conventional Gran equation method based on the quotient V/[H] in the ammonia‐boric acid titration case.

Figure 3.

Mole fraction of ammonium (f1,a) and borate (f0,b) ions as a function of V.

Beyond the equivalence point we have instead

V´=V(V0+V)ΔCHClE43

When V > Veq, [OH] is negligible, and

Δ=[H+][OH][H+]E44

and f1,a, the mole fraction of ammonium ion is virtually equal to the unity, and then by combining Eqs. (43), (44) and (35) we get

V=V(V0+V)[H+]CHCl=VeqE45

and then we obtain the Gran expression for the titration of a weak base with a strong acid, beyond the equivalence point

(V0+V)[H+]=VCHClVeqCHClE46

By plotting the left hand of Eq. (46) against V, a straight line is obtained with slope CHCl, intersecting the V axis at the point Veq.

An alternative route to obtain Eq. (46) is from Eq. (27). At V > Veq, f0,b ≈ 0, f1,a≈1, Δ≈[H+], and then,

(V0+V)[H+]VCHClV0CNH3E47

which is combined with Eq. (29), gives finally Eq. (46).

Advertisement

8. Titration error

From Eq. (26), we may calculate the fraction titrated, T, as

T=CHClVCNH3V0=CHClCNH3(f1,aCNH3f0,bCH3BO3+ΔCHClΔ)=f1,af0,bCH3BO3CNH3+Δ1ΔCHClE48

In the vicinity of the equivalence point, the mole fraction of borate ion, f0,b, is close to zero, and the titration error, ΔT, may be approximated as

ΔT=[T1]end=[f1,a+Δ1ΔCHCl1]endE49
Advertisement

9. Checking the proposed linearization method with synthetic data

In order to verify the goodness of the proposed linearization method, based on the use of Eq. (35), a series of theoretical data with variations of 0.05 units of pH, has been generated with aid of Eq. (26), using the analogous conditions that previously published by Cruz [68], V0= 150 mL; CNH3=0.0096 M; CH3BO3=0.2156 M; CHCl=0.12 M; pK1,a=9.27; pK1,b= 9.12.

Two sets of data [V, pH] (N = 24 and N = 11) were selected from the theoretical data in the region prior to the equivalence point (Veq=12 mL), ranging from approximately 0.6 to 11.16 mL for the first series, and from 0.6 to 10.8 mL for the second (Table 2). Note that, the points selected for study are far from the region of the equivalence point: Veq ± 1 mL. A set of data [V, pH] (N = 11) was also selected after the equivalence point, between about 13 and 23.5 mL (Table 3).

V0=150 mLVeq=12.000 mL
C(HCl)=0.12 MpK1,a=9.270
pKa(NH4+)=9.27
pKa(H3BO3)=9.12a1=‐1.864E + 092.237E + 10=a0
C(H3BO3)=0.2156 Ms(a1)=1.971E + 072.327E + 08=s(a0)
pKw=13.8R2=0.99906.386E + 06=s(y/x)
VV(2)V(2)/[H]pHΔf0,bV′(2)v′(2)/[H]
0.621160.623.49E + 077.75‐8.73 E‐074.09 E‐0211.656.55E + 08
1.813681.819.07E + 077.70‐7.74 E‐073.66 E‐0211.685.85E + 08
2.885452.891.29E + 087.65‐6.86 E‐073.28 E‐0211.725.24E + 08
3.847853.851.53E + 087.60‐6.06 E‐072.93 E‐0211.754.68E + 08
4.711364.711.67E + 087.55‐5.34 E‐072.62 E‐0211.774.18E + 08
5.485605.491.74E + 087.50‐4.70 E‐072.34 E‐0211.803.73E + 08
6.800616.801.71E + 087.40‐3.58 E‐071.87 E‐0211.842.97E + 08
7.854207.851.57E + 087.30‐2.66 E‐071.49 E‐0211.872.37E + 08
8.696878.701.38E + 087.20‐1.88 E‐071.19 E‐0211.901.89E + 08
9.653559.651.08E + 087.05‐8.87 E‐088.44 E‐0311.921.34E + 08
10.8192410.826.08E + 076.758.87 E‐084.25 E‐0311.966.73E + 07

Table 2.

Proposed method applied to theoretical data before the equivalence point and rounded to the second decimal figure in volume (penultimate and last column).

V0=150 mLVeq=11.999 mL
C(HCl)=0.12 MC(HCl)=0.120 M
pKa(NH4+)=9.27
pKa(H3BO3)=9.12a1=0.120−1.440=a0
C(H3BO3)=0.2156 Ms(a1)=4.471 E‐058.219 E‐04=s(a0)
pKw=13.8R2=0.99994.778 E‐04=s(y/x)
V(V0+V) [H]V(3)(V(3)+V0) [H]V(2)(V(2)+V0) [H]pH
13.079220.12953842913.0790.13013.080.133.10
14.161810.26017893714.1620.26014.160.262.80
15.079630.36956725715.0800.37015.080.372.65
15.896250.46755915915.8960.46815.900.472.55
16.935860.59231078316.9360.59216.940.592.45
18.263320.75160464218.2630.75218.260.752.35
19.060890.84731159819.0610.84719.060.852.30
19.964820.95578242119.9650.95619.960.962.25
20.990621.07887787520.9911.07920.991.082.20
22.156421.21877411822.1561.21922.161.222.15
23.483551.3780288223.4841.37823.481.382.10

Table 3.

Gran method applied to theoretical data after the equivalence point and rounded to the second decimal figure in volume (antepenultimate and penultimate column).

The proposed linearization method has been applied to data corresponding to V < Veq (basic zone), rounding the volumes to 5, 3 and 2 decimal places, respectively, proving that the method works well in all cases, obtaining correct values for both Veq and pK1,a. The same rounding procedure has been applied to the data corresponding to V > Veq (acid zone), but applying in this latter case the Gran method based on Eq. (46), which also leads to the correct values of Veq and CHCl. However, a pronounced curvature is observed when the conventional Gran function (V/[H+]) is applied to the data obtained in the basic region (V < Veq), which makes it impossible to apply this method to ammonia‐boric acid system to values away from pHeq.

From the data obtained and using linear regression (method of the least squares), in EXCEL, with the function LINEST, the volume at the equivalence point is obtained as

Veq=a0a1pK1,a=log(1a1)E50

to data corresponding to V < Veq, and

Veq=a0a1CHCl=a1E51

to data corresponding to V > Veq. The data corresponding to the statistic parameters of the straight line using the least squares method appear squared in Tables 2 and 3. The parameters obtained using linear regression by the function LINEST in EXCEL are: intercept, slope and coefficient of determination (correlation coefficient squared): a0, a1 y R2, respectively; and standard deviations of the intercept, slope and regression: s(a0), s(a1) y s(y/x), respectively.

Figure 4 (top) shows the titration curve corresponding to the series of eleven pairs of data (left side) and the application of the conventional Gran method (right side). Figure 4 (middle) shows the linearization method proposed in this chapter (left side) and the Gran method for data beyond the equivalence point (right side). Finally, at the bottom of Figure 4, a comparative study of the three methods is shown together. Gran and modified Schwartz (proposed method) on the same scale, showing the superiority of the latter transforming the whole curve in a straight line. The scale corresponding to the application of the Gran method in acid medium (V > Veq) is shown on the secondary axis.

Figure 4.

Top: ammonia titration curve (pH = f(V)) in boric acid (LEFT) and Gran method V < Veq (right). Middle left: modified Schwartz method. Middle right: Gran method V > Veq. Bottom: comparative study.

On the other hand, the proposed method, based on the use of Eq. (35), has been applied to a series of 19 data pairs (compiled in the legend of Figure 5) in the first instance, assuming unknown the value of pK1,b and varying its in 0.01 units of pK in the vicinity of pK1,b true. The standard deviation of the corresponding regression lines (sy/x) obtained as a function of the assumed pK1,b value is shown in Figure 5 (top). The standard deviation of the regression is minimal when the value of pK1,b agrees on the true value (or differs in less than 0.01 unit). The values obtained for the equivalence volume and for pK1,a (ammonium ion), as a function of the assumed value of pK1,b, are plotted in the central part of Figure 5.

Figure 5.

Top: Standard deviation of the regression as a function of pK1,b supposed. Middle: equivalence volume and pK1,a values as a function of pK1,b supposed. Bottom: plots for a series of pK1,b values selected. Data [V; pH]: [0.62; 7.75] [1.81; 7.70] [2.89; 7.65] [3.85;7.60] [4.71;7.55] [5.49;7.50] [6.18;7.45] [6.80;7.40] [7.36; 7.35] [7.85;7.30] [8.30; 7.25] [9.05; 7.15] [9.65; 7.05] [10.13; 6.95][10.52; 6.85][10.68; 6.80] [10.82; 6.75][11.06; 6.65][11.53; 6.35].

Two conclusions can be drawn from the study:

  1. The proposed method is sufficiently robust regarding to variations in the supposed value for pK1,b. Values of pK1,b between 9.09 and 9.16 give rise to an error in the determination of the equivalence volume < 0.1%.

  2. The proposed method does not work to estimate, simultaneously, the value of the acidity constant of the ammonium ion, as shown in the central part of Figure 5 by plotting pK1,a obtained (secondary axis) versus the supposed value of pK1,b. Small variations in the supposed value of pK1,b lead to large variations in the value obtained for pK1,a. Moreover, in case of pK1,a is not applicable a minimization criterion because a minimum or maximum value is not reached when pK1,b agrees on the true value. When pK1,b < 9.11, the slope of the line changes from negative to positive, so Eq. (50b) is not applicable, since the value of the constant obtained is negative, which has no physical meaning.

Advertisement

10. Checking the proposed linearization method with experimental data

Once tested that our proposed method works well, the theory devised has been applied to the experimental data recently reported by Cruz [68] in the Journal of Chemical Education in a study about the determination of ammonia with HCl using concentrated (4% w/v) and diluted (1.3% w/v) boric acid to reproduce the final‐Kjeldahl titration when two different volumes of ammonia distillate are collected (data [V, pH] appear as supporting information).

Table 4 shows the data [V, pH] corresponding to the titration curves of ammonia in solutions of H3BO3 at 4% w/v (0.6469 M) and 1.33% w/v (0.2156 M). The shape of titration curves and the application of the conventional Gran method in the region prior to equivalence point are shown in Figure 6. A curvilinear shape is obtained with both solutions (being the curve most flattened when [H3BO3]=0.6469 M), so that, in its original philosophy, the Gran method is not applicable.

Figure 6.

Ammonia titration curve (pH = f(V)) in boric acid (top) and conventional Gran method (bottom).

V0=50 mL; C(HCl) = 0.12 MV0=150 mL; C(HCl) = 0.12 M
C(H3BO3)=0.6469 M (4% w/v)C(H3BO3)=0.2146 M (1.33% w/v)
VpHVpHVpHVpHVpHVpH
0.006.4511.354.9511.914.120.007.3811.375.8012.014.32
2.416.3111.384.9411.934.021.317.2911.485.7012.044.15
3.486.2311.434.8911.963.912.217.2311.545.6512.074.00
4.606.1411.494.8411.993.793.207.1611.595.5912.103.88
5.386.0811.524.8212.013.674.287.0711.625.5512.133.79
6.395.9911.554.7912.043.575.217.0011.655.5112.163.71
7.425.8811.584.7612.073.466.226.9011.695.4712.213.58
8.405.7611.634.6912.103.387.266.8011.715.4312.273.48
9.375.6111.664.6612.133.318.146.6911.745.3712.333.40
10.385.3811.694.6712.183.189.026.5511.785.3212.413.30
10.565.3211.724.5812.243.089.506.4711.805.2512.503.23
10.925.1811.754.5312.303.0010.006.3511.835.1812.623.14
11.045.1311.774.4812.352.9310.406.2411.865.0912.743.07
11.155.0711.804.4212.442.8410.786.1111.894.9912.882.99
11.215.0411.834.3610.926.0511.924.8713.062.92
11.275.0111.854.2911.115.9611.954.7213.292.84
11.324.9711.884.2111.225.8911.984.52

Table 4.

Data [V, pH] corresponding to the titration curves of ammonia in solutions of H3BO3 at 4% w/v (0.6469 M) and 1.33% w/v (0.2156 M) [68].

In order to apply our proposed linearization method, data with VHCl < 11 mL are selected (region prior to Veq, which is around 12 mL). Tables 5 and 6 show the calculations performed when applying the proposed linearization method for the optimum pK1,b (value shown in the box) in the studies of the ammonia solutions diluted (0.2156 M) and concentrated (0.6469 M), respectively.

V0=150 mLV(eq)=12.024 mL
C(HCl)=0.12 MpK(*)=7.179
pKa(H3BO3)=9.117
C(H3BO3)=0.2156 Ma1=‐1.510E + 071.816E + 08=a0
pKw=13.8s(a1)=1.353E + 051.189E + 06=s(a0)
R2=0.99901.015E + 06=s(y/x)
VV/[H]pHΔf0,bV'V'/(H)
1.312.554E + 077.29‐2.577 E‐071.468 E‐025.265E + 001.027E + 08
2.213.753E + 077.23‐2.103 E‐071.281 E‐025.661E + 009.614E + 07
3.204.625E + 077.16‐1.599 E‐071.092 E‐026.143E + 008.880E + 07
4.285.029E + 077.07‐1.011 E‐078.894 E‐036.677E + 007.845E + 07
5.215.210E + 077.00‐5.849 E‐087.580 E‐037.253E + 007.253E + 07
6.224.941E + 076.900.000E + 006.031 E‐037.845E + 006.232E + 07
7.264.581E + 076.805.849 E‐084.796 E‐038.553E + 005.396E + 07
8.143.987E + 076.691.265 E‐073.727 E‐039.144E + 004.479E + 07
9.023.200E + 076.552.256 E‐072.703 E‐039.748E + 003.459E + 07
9.502.804E + 076.472.921 E‐072.249 E‐031.011E + 012.982E + 07
10.002.239E + 076.354.112 E‐071.707 E‐031.046E + 012.342E + 07
10.401.807E + 076.245.479 E‐071.326 E‐031.076E + 011.869E + 07
10.781.389E + 076.117.558 E‐079.830 E‐041.104E + 011.423E + 07
10.921.225E + 076.058.735 E‐078.563 E‐041.115E + 011.251E + 07

Table 5.

Linearization method applied to the titration data lower than 11 mL (Veq ≈ 12 mL): diluted boric acid series (1.33%, 0.2156 M).

V0=50 mLV(eq)=11.827 mL
C(HCl)=0.12 M
pKa(H3BO3)=7.509
C(H3BO3)=0.6469 Ma1=5.662E + 06‐6.697E + 07=a0
pKw=13.8s(a1)=2.662E + 043.828E + 05=s(a0)
R2=0.99981.749E + 05=s(y/x)
VV/[H]pHΔf0,bV'V'/(H)
2.414.921E + 066.314.574 E‐075.948 E‐021.844E + 013.765E + 07
3.485.910E + 066.235.619 E‐074.997 E‐021.695E + 012.878E + 07
4.606.350E + 066.147.026 E‐074.100 E‐021.565E + 012.161E + 07
5.386.468E + 066.088.127 E‐073.590 E‐021.506E + 011.810E + 07
6.396.245E + 065.991.008 E‐062.938 E‐021.431E + 011.398E + 07
7.425.629E + 065.881.306 E‐062.296 E‐021.361E + 011.032E + 07
8.404.834E + 065.761.729 E‐061.751 E‐021.312E + 017.549E + 06
9.373.817E + 065.612.448 E‐061.246 E‐021.273E + 015.185E + 06
10.382.490E + 065.384.165 E‐067.375 E‐031.237E + 012.966E + 06
10.562.206E + 065.324.783 E‐066.430 E‐031.229E + 012.568E + 06
10.921.653E + 065.186.605 E‐064.666 E‐031.217E + 011.843E + 06

Table 6.

Linearization method applied to the titration data lower than 11 mL (Veq ≈ 12 mL): concentrated boric acid series (4% w/v, 0.6469 M).

The most diluted solution ([H3BO3]=0.2156 M) is firstly studied. Figure 7 (top) shows the plots obtained for different supposed values of pK1,b, in the range from 9.04 to 9.24. The best straight line corresponds to a value of pK1,b around 9.12, as shown in the bottom left side of Figure 7, which plots the standard deviation of the regression obtained for each line versus the supposed value of pK1,b, which reaches a minimum when pK1,b equals to 9.12. The supposed pK1,b is then varied around 9.12 in 0.001 unit (Figure 7, bottom right), which minimizes the standard deviation of the regression to 9.117.

Figure 7.

Top: proposed linearization method as a function of pK1,b supposed. Data [V, pH] are shown in Table 4. Bottom: standard deviation of the regression as a function of pK1,b supposed.

In the study of [H3BO3]=0.6469 M, the results obtained indicate that is necessary to explore the values for supposed pK1,b in a wider range toward lower values of pK, consistent with the fact that the boric acid strength increases with its concentration [68, 70, 71]. It should be noted that the pK1,b value used with the proposed method is an adjustment parameter, and in fact apparent, without a specific physical meaning. From Figure 8 (top) can be observed that when the apparent pK1,b reaches a value around 7.5, the representation of V′[H] versus V’ has the form of a straight line, so that this parameter is varied in 0.001 units around this value, being verified that the standard deviation of the regression is minimum (Figure 8 bottom) when it reaches the value of 7.509, value corresponding to a Veq of 11.827 mL.

Figure 8.

Top: proposed linearization method in the titration with [H3BO3] = 0.6469 and V0 = 50 mL when V < Veq. Bottom: equivalence volume and s(y/x) values as a function of pK1,b supposed; [H3BO3]=0.6469 M.

Figure 9 shows the application of the proposed method for the ammonia solutions diluted (0.2156 M) and concentrated (0.6469 M) using the optimum value for pK1,b in each case (9.117 and 7.509, respectively). Note how the response of the diluted sample in boric acid, in a relation to 1:3, is more sensitive (greater slope) than that of the sample more concentrated in boric acid.

Figure 9.

Left: proposed linearization method in the titration with [H3BO3] = 0.2156 and V0= 150 mL. Right: proposed linearization method in the titration with [H3BO3] = 0.6469 and V0 = 50 mL.

An empirical parameter, Γ, can be defined which take into account the variation in the value of pK1,b optimum when the boric acid concentration increases, reflecting the increase in acidity, motivated by the appearance of polynuclear species. The complexity of such treatment far exceeds the objective of this chapter.

Γ=[K1,b]c[K1,b]d=10([K1,b]c[K1,b]d)=10(9.1177.509)=40.55E52

The determination of the various polyborate species in solution has proved to be difficult because they appear at fairly high concentrations of boric acid and involve the addition of no more than one OH ion per B(OH)3 group presents [87]. The pK values of triborate (3B(OH)3=B3O3(OH)4 + H+ + 2H2O) and pentaborate (5B(OH)3=B5O6(OH)4 + H+ + 5H2O) are of the order of 6.60 and 6.90, respectively [68].

The titration with H3BO3 0.2156 M results in a wider pH range and a steeper slope, contrary to that predicted by the theoretical models of the titration curves. This is for the concurrent presence of other polymer species more abundant in H3BO3 0.6469 M, which disappear progressively as the dilution increases. The pH value at the equivalence point is variable depending on the initial concentration of boric acid, which can lead to systematic errors when using colored indicators in the detection of the end point.

Figure 10 (top) shows the conventional Gran method (I) to the data compiled in Table 4. Despite the precision of the data, the application of this method has difficulties in the appreciation of the equivalence volume. Cruz [68] applied the Gran method (II) to the titration data between 11 and 12 mL, which could be questioned from the methodological point of view. The results obtained by Cruz are shown in Figure 10 (bottom).

Figure 10.

Top: Gran I method. Bottom: Gran II method (conventional) for 11 < V < 12 mL (as applied by Cruz (2013)).

Table 7 summarizes a comparative study of the results obtained in the calculation of the equivalence point using the proposed method described in this chapter and those obtained by Cruz [68]. The differences range from 0.15% for [H3BO3]=0.2156 M, and −1.6% for [H3BO3]=0.6469 M. The values of the standard deviation of the volume at the equivalence point are equal to 0.036 and 0.017 volume units for the proposed method for [H3BO3] = 0.2156 M and 0.6469 M, respectively, and 0.3 and 0.15%, respectively, in terms of relative standard deviation. If the covariance is not taken into account, these values would increase in absolute terms to 0.133 and 0.102 units of volume, respectively, corresponding to 1.11 and 0.86%, respectively (an increase of the relative error between four and six times more). So, the covariance of measurements can be important as the variances and both contribute significantly to the total analytical error.

[H3BO3]pH rangeNMethodpK1,bV(eq)s[V(eq)]s[V(eq)]*
0.21561.31–10.9214Proposed9.11712.0240.0360.133
0.61642.41–10.9211Proposed7.50911.7940.0170.102
0.215611.04–11.9323Gran II12.006
0.616411.11–11.9217Gran II11.987

Table 7.

Comparison results obtained in the calculation of the equivalence point using the proposed method and the Gran method [68].

VpHVpHVpHVpH
0.106.6091.206.5062.806.3404.815.922
0.196.5941.426.4893.006.3025.005.847
0.306.5901.516.4833.206.2785.205.769
0.416.5771.606.4743.406.2435.405.687
0.506.5651.716.4663.606.2105.605.537
0.596.5621.816.4443.806.1715.805.303
0.706.5492.006.4294.006.1316.004.907
0.806.5442.196.4064.206.0956.204.275
0.926.5342.406.3834.406.0386.403.787
1.106.5222.606.3584.605.9856.603.574

Table 8.

Data [V, pH] corresponding to the titration curves of ammonia in solutions of H3BO3 (0.1941 M) with HCl (0.077 M).

Finally, the proposed method has been applied to experimental data carried out in our laboratory reproducing the conditions reported by Cruz [68] in the case of diluted boric acid.

10.1. Reagents

All reagents were analytical grade unless otherwise specified:

  • Boric acid (H3BO3) M = 61.83 g/mol (Merck > 99.5%), Ammonia (30%) M = 17.03 g/mol. (Panreac); Trishydroxymethyl aminomethane (TRIS; C4H11NO3) M = 121.14 g/mol.

  • Buffers pH = 3, 5 (Riedel‐de Häen), 4.01, 7 (Crison).

  • Phial of HCl 1 M (Riedel‐de Häen).

10.2. Instruments

Analytical balance (Metler AE200) (4 decimals), Granatario (Metler PJ 400) (2 decimals), pH‐meter Crison GLP 21, with a combined Ag/AgCl glass electrode were used. The pH‐meter was calibrated using pH buffers 3, 4.01, 5 and 7, using a two‐point calibration method. Burette of 5 mL (Brand) (± 0.01 at 20ºC) was used.

10.3. Experimental

About 5 mL of ammonia solution (0.12 M) was pipetted into a 50 mL volumetric flask and made up to the mark with a boric acid solution (4% w/v). The contents of this solution were transferred to a 200 mL Erlenmeyer flask containing 100 mL of distilled water. Then, the ammonium borate solutions were titrated potentiometrically with HCl (0.077 M) (previously standardized with TRIS) using the glass pH electrode.

Table 8 shows the data [V, pH] corresponding to the titration curve of ammonia in solution of H3BO3 (0.1941 M). The shape of the valuation curve is shown in the upper part of Figure 11. All data prior to the equivalence point, before the jump in the titration curve, have been taken to apply the calculus by the proposed method, which leads to an equivalence volume equal to 6.17 mL. The equivalence volume obtained by the first derivative method is equal to 6.20 mL (although not many points are available in this case).

Figure 11.

Top left: ammonia titration curve (pH = f(V)) in boric acid (0.1941 M) with HCl (0.077 M). Top right: conventional Gran method. Middle: proposed linearization method as a function of pK1,b supposed. Bottom: s(y/x) values as a function of pK1,b supposed.

If the first points of the titration in which the solution is not sufficiently buffered are neglected, a slightly lower volume (6.14 mL) is obtained. An analogous result is obtained by using only the last 9 points prior to the equivalence point. However, the great contribution of this method is the complete linearization.

Figure 11 (middle) shows the plots obtained for different supposed values of pK1,b, in the range from 8.20 to 9.24. The best straight line corresponds to a value of pK1,b around 8.279, as shown in the bottom of Figures 11 and 12.

Figure 12.

Proposed linearization method in the titration of ammonia in [H3BO3] = 0.1941 with HCl (0.077 M).

11. Conclusions

The knowledge on the mathematical approach involved in the titration curves is very useful to check the validity of the procedure. A thoroughly theoretical study of the boric acid version of the titrimetric finish of the important Kjeldahl method has been carried out in this chapter.

In order to locate the equivalence point of the ammonia‐boric acid titration, an extension of the method of Schwartz has been devised. Though the method is non‐approximate, it requires the knowledge of the acidity constant of the boric acid, which may be calculated by a trial and error procedure (minimization of the standard deviation of the regression line). Unlike the differential method, the method proposed makes use of all the experimental data, so no preliminary knowledge concerning the end point is needed. The method has proved very accurate when applied to synthetic titration curves and data, and in order to check its utility, it has been applied to the experimental data recently reported by Cruz [68]. In addition, titration data have been obtained in the laboratory and processed consequently, with good results.

The study of experimental ammonia‐boric acid systems titrated with hydrochloric acid allows us to extract interesting conclusions. First, the behavior of experimental systems subject of study is far from the expected on the basis of the theoretical model. This difference will be greater as the concentration of boric acid used as trapping agent for ammonia increases.

Second, and in spite of the above indicated, the proposed method works well so that it allows the straight line model fits properly the experimental data and lead to a reliable equivalence end point value by using a minimization criteria as indicated above. However, the parameter varied is empirical, without any physical significance, given the complexity of the systems with polynuclear borate species being present at high boric acid concentration, which tend to disappear by diluting the solution. It would be interesting on this respect to assay another weak acid other than boric acid as trapping agent, thus avoiding the concurrent equilibria of polyborate polymer species.

References

  1. 1. Kjeldahl J. Neue methode zur Bestimmung des Stickstoffs in organischen Körpern. Z. Analytical Chemistry. 1883;22:366–382
  2. 2. Kjeldahl J. New method for the determination of nitrogen. Chemistry News. 1883;48(1240):101–102
  3. 3. Sella A. Classic kit: Kjeldahl flask. The Royal Society of Chemistry (http://www.rsc.org/): Chemistry World; 2008
  4. 4. Analyst Ed. On Kjeldahl method for determination of nitrogen. Analyst. 1885;10(6):127–128
  5. 5. Burns DT. Kjeldahl, the man, the method and the Carlsberg Laboratory. Analytical Proceedings. 1984;21(6):210–214
  6. 6. Blyth AW. A method of determining organic nitrogen in liquids. Analyst. 1884;9(7):115–116
  7. 7. Kebler JF. Notes on the estimation of nitrogen in nitrates by Kjeldahl’s method, and an index to the literature on the estimation of nitrogen. Journal of Analaytical and Pure Chemistry. 1891;5:257–278
  8. 8. Zelitch I. Hubert Bradford Vickery 1893–1978, Biographical Memoir. Washington, DC: National Academy of Sciences; 1985
  9. 9. Möller J. Protein analysis revisited. In Focus. 2010;34(2):22–23
  10. 10. Jung S, Rickert DA, Deak NA, Aldin ED, Recknor J, Johnson LA, Murphy PA. Comparison of Kjeldahl and Dumas methods for determining protein contents of soybean products. Journal of the American Oil Chemists’ Society. 2003;80(12):1169–1173
  11. 11. FDA. Food Labeling Health Claims: Soy Protein and Coronary Heart Diseases. Silver Spring, Md: Food and Drug Administration; Final Rule, Fed. Reg. 1999;64:57700–57733
  12. 12. Sáez‐Plaza P, Navas MJ, Wybraniec S, Michałowski T, García Asuero A. An overview of the Kjeldahl method of nitrogen determination. Part II. sample preparation, working scale, instrumental finish, and quality control. Critical Reviews in Analytical Chemistry. 2013;43:224–272
  13. 13. Dumas J‐B, Lettre de M and Dumas a M. Gay‐Lussac, Sur les Procédés de l’Analyse Organique. Annales de Chimie et de Physique. 1831;2(47):198–215
  14. 14. Rammelsberg C. Ueber einige Salze der Untersehwfelsäure und ihre Verbindungen mit Ammoniak. Annual Review of Physical Chemistry. (Pogg. Ann.) 1843;58:295–298
  15. 15. Nessler J. Verhalten des Jodquecksilbers zu Ammoniak, und eine neue Reaktion auf Ammoniak. Chemistry Central Journal. 1856;27:529–541; On the behavior of iodide of mercury to ammonia, and a new reaction for ammonia. Journal für praktische Chemie. 1856;14(445–453):463–466
  16. 16. Wiedemann G. Ueber ein neues Zersetzungsproduct des Hanstofts. Pharmaceutical Chemistry Journal. (Pogg. Ann.) 1849;74:67–84
  17. 17. Riegler E. Eine kolorimetrische Bestimmungsmethode des Eiweisses. Analytical and Bioanalytical Chemistry. 1914;53:242–254
  18. 18. Gornall AG, Bardawill CJ, David MM. Determination of serum proteins by means of biuret reagent. The Journal of Biological Chemistry. 1949;177(2):751–766
  19. 19. Berthelot MPE. Violet D’aniline. Repert. Chim. Appl. 1859;1:284
  20. 20. Searle PL. The berthelot or indophenol reaction and its use in the analytical chemistry of nitrogen. Analyst. 1984;109(5):549–568
  21. 21. Folin O, Ciocalteu J. On tyrosine and tryptophan determinates in proteins. The Journal of Biological Chemistry. 1927;73(2):627–650
  22. 22. Fraenkel‐Conrat H, Cooper M. The use of dye for the determination of acid and basic groups in proteins. The Journal of Biological Chemistry. 1944;154(1):239–246
  23. 23. Wu H. A new colorimeric method for the determination of plasma proteins. The Journal of Biological Chemistry. 1922;51(1):33–39
  24. 24. Lowry OH, Rosebrough NJ, Farr AL, Randall RJ. Protein measurement with the folin phenol reagent. The Journal of Biological Chemistry. 1951;193(1):265–275
  25. 25. Peterson GL. Review of the folin phenol protein quantitation of lowry, Rosebrough, Farr and Randall. Analytical Biochemistry. 1979;100(2):201–220
  26. 26. Legler G, Muller‐Platz CM, Mentges‐Hettkamp M, Pflieger G, Julicj E. On the chemical basis of the lowry protein determination. Analytical Biochemistry. 1985;150(2):278–287
  27. 27. Owusu‐Apenten RK. Kjeldahl method, Quantitative Aminoacid Analysis and Combustion Analysis. In Food Protein Analysis. Quantitative Effects on Processing; New York‐Basel: Marcel Dekker; 2002. Chapter1, pp. 1–45
  28. 28. Chandrarajan J. Klein L. Lowry assay of dilute protein solutions containing high concentrations of Triton X‐100. Analytical Biochemistry. 1975;69(2):632–636
  29. 29. Bradford MM. A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein–dye binding. Analytical Biochemistry. 1976;72(1–2):248–254
  30. 30. Compton SJ, Jones CG. Mechanism of dye response and interference in the bradford protein assay. Analytical Biochemistry. 1985;151(2):369–374
  31. 31. Smith PK, Krohn RI, Hermanson GT, Malla AK, Gartner FH, Provenzano MD, Fujimoto EK, Goecke NM, Olson BJ, Klenk DC. Measurement of protein using bicinchoninic acid. Analytical Biochemistry. 1985;150(1):76–85; Erratum, 1987;163(1):279
  32. 32. You WW, Haugland RP, Ran DK, Haugland RP. 3‐(4‐Carboxy Benzyl) Quinoline‐2‐Carboxaldehyde, a reagent with broad dynamic range for the assay of proteins and lipoproteins in solution. Analytical Biochemistry. 1997;244(2):277–282
  33. 33. Möller J. Kjeldahl‐still going strong. In Focus. 2009;33(1):14–16
  34. 34. Official Methods of Analysis of AOAC International 18th Ed. (Revision 4). Gaithersburg, MD: AOAC; 2011
  35. 35. McKenzie HA. The Kjeldahl determination of nitrogen: Retrospect and prospect. Trends in Analytical Chemistry. 1994;13(4):138–144
  36. 36. Stephen WI. Determination of nitrogen in organic compounds in the years before Kjeldahl’s method. Analytical Proceedings. 1984;21:215–220
  37. 37. Johannsen W. Johan Kjeldahl 1849–1900. Obituary. Ber. Dtsch. Chem. Ges. 1900;33(3):3881–3888
  38. 38. Johannsen W. Johan Kjeldahl 1849–1900. CR Trav. Lab. Carlsb. (Edition Française). 1903;5:I‐VIII
  39. 39. Holter H, Møller KM. The Carlsberg Laboratory 1876/1976. Rhodos, Denmark: The Carlsberg Foundation; 1976
  40. 40. Veibel S. John Kjeldahl (1849–1900). Journal of Chemical Education. 1949;26(9):459–461
  41. 41. Hansen EH. The current status of analytical chemistry in Denmark. A personnel view. Analytical Proceedings. 1993;30(1):3–9
  42. 42. Rideal EK. Soren Peter Lauritz Sörensen. Journal of the Chemical Society. 1940;554–561
  43. 43. Asuero AG, Michalowski T. Comprehensive formulation of titration curves for complex acid‐base systems and its analytical implications. Critical Reviews in Analytical Chemistry. 2011;41:151–187
  44. 44. Michalowski T, Asuero AG. New approaches in modeling the carbonate alkalinity. Critical Reviews in Analytical Chemistry. 2012;42(3):220–244
  45. 45. Breuer PL, Sutchiffe CA, Meakin RL. Cyanide measurement by silver nitrate titration: Comparison of rhodanine and potentiometric end points. Hydrometallurgy. 2011;106:135–140
  46. 46. Felber H, Rezzonico S, Máriásy M. Titrimetry at a metrological level. Metrologia. 2003;40:249–254
  47. 47. Thordarson P. Determining association constants from titration experiments in supramolecular chemistry. Chemical Society Reviews. 2011;40:1305–1323
  48. 48. Michalowski T, Asuero AG, Ponikrar‐Svet M, Toporek M, Pietrzyk A, Rymanowski M. Liebig‐Denigès method of cyanide determination: A comparative study of two approaches. Journal of Solution Chemistry. 2012;41:1224–1239
  49. 49. Michalowski T, Toporek M, Michalowska‐Kaczmarcyk AM, Asuero AG. New trends in studies on electrolytic systems. Electrochimica Acta. 2013;109:519–531
  50. 50. Winkler‐Oswatistsch R, Eigen M. The art of titration. From classical end points to modern differential and dynamic analysis. Angewandte Chemie International Edition. 1979;18:20–49
  51. 51. Sneddon J, Hardaway C, Bobbadi KK, Reddy AK. Sample preparation of solid samples for metal determination by atomic spectroscopy–An overview and selected recent applications. Applied Spectroscopy Reviews. 2006;41(4):1–14
  52. 52. Brayton SV. Acid Digestions using the Hach Digesdahl Digestion Apparatus: Sample Preparation for Protein and Elemental Analysis. Hach Company, Loveland, CO: Tech. Info. Ser. Booklet; No 14; 1992
  53. 53. Flores EMM, Barin JS, Mesko MF, Knapp G. Sample preparation tehniques based on combustion reactions in closed vessels–A brief overview and recent applications. Spectrochimica Acta Part B. 2007;62(9):1051–1064
  54. 54. Egli H. Kjeldahl Guide, BUCHI. CH‐9230 Flawill, Switzerland: Büchi Labortechnik AG; 2008
  55. 55. Thiex NJ, Manson H, Anderson S, Persson JA. Determination of crude protein in animal feed, forage, grain, and oilseeds by using block digestion with a copper catalyst and steam distillation into boric acid: Collaborative study. Journal of AOAC INTERNATIONAL Int. 2002;85(2):309–317
  56. 56. EN ISO 5983–1. Animal feeding stuffs–Determination of nitrogen content and calculation of crude protein content–Part 1: Kjeldahl method. Genève: International Organization for Standardization; 2009
  57. 57. Lynch JM, Barbano DM. Kjeldahl nitrogen analysis as a reference method for protein determination in dairy products. Journal of AOAC INTERNATIONAL. 1999;82(6):1389–1398
  58. 58. AMC. Methods for the destruction of organic matter. Analyst. 1960;85(9):643–656
  59. 59. Greenfield H, Southgate DAT. Food composition data. Production, management and use. 2nd ed. Rome: FAO; Review of Methods of Analysis; 2003. Chapter 7; pp. 97–148
  60. 60. Persson JA, Wennerholm M, O’Halloran S. Handbook for Kjeldahl digestión. Hilleroed: Denmark FOSS: DK‐3400; 2008
  61. 61. Thiex NJ, Manson H. New AOAC standard for crude protein determinations in animal feed, grain, and oilseed using Tecator Kjeltec equipment, In Focus. 2002;26(2):10–12
  62. 62. Saha UK, Sonon L, Kissel DE. Comparison of conductimetric and colorimetric methods with distillation titration method of analyzing ammonium‐nitrogen in total Kjeldahl digests. Communications in Soil Science and Plant Analysis. 2012;43(18):2323–2341
  63. 63. Winkler JW. Beitrag zur titrimetrischen bestimmung des ammoniaks. Angewandte Chemie International Edition. 1913;26(31):231–232
  64. 64. Beljkaš B, Matić AJ, Milovanović I, Jovanov P, Mišan A, Šaric L. Rapid method for determination of protein content in cereals and oilseeds: Validation, measurement uncertainty and comparison with the Kjeldahl method. Accreditation and Quality Assurance. 2010;15(10):555–561
  65. 65. Blaedel WJ, Meloche VW. Elementary quantitative analysis: Theory and practice. Evanston, IL: Row, Petterson and Co; 1957
  66. 66. Bradstreet RB. The Kjeldahl Method for Organic Nitrogen. London: Academic Press; 1965
  67. 67. Laitinen HA, Harris WE. Chemical Analysis. 2nd ed. New York: McGraw‐Hill; 1975. p. 115
  68. 68. Cruz G. Boric acid in Kjeldahl analysis. Journal of Chemical Education. 2013;90:1645–1648
  69. 69. Ingri N. Equilibrium studies of polyanions. On the first equilibrium steps in the acidification of B(OH)4. An application of the self medium method. Acta Chemica Scandinavica. 1963;17:573–580
  70. 70. Prideaux BR. Note on the neutralization curve of boric acid. Transactions of the Faraday Society. 1915;11:76–78
  71. 71. Thygesen JE. Über die Selbstkomplexbildung der Borsäure. Zeitschrift für anorganische und allgemeine Chemie. 1938;237(1):101–112
  72. 72. Kantowitz BH, Roediger H, Elmes DG. Experimental Psicology. 9thed. Belment, CA: Wadsworth; 2009
  73. 73. Gran G. Determination of the equivalent point in potentiometric titrations. Acta Chemica Scandinavica. 1950;4:559–577
  74. 74. Gran G. Determination of the equivalence point in potentiometric titrations. Part II. Analyst. 1952;77:661–671
  75. 75. Gran G. Equivalence volumes in potentiometric titrations. Analytica Chimica Acta 1988;206:111–123
  76. 76. Schwartz LM. Advances in acid‐base Gran plot methodology. Journal of Chemical Education. 1987;64(11):947–950
  77. 77. de Levie R. Principles of Quantitative Chemical Analysis. New York: McGraw‐Hill; 1997
  78. 78. de Levie R. Aqueous Acid‐Base Equilibria and Titrations. Oxford: Oxford University Press, 1999
  79. 79. Asuero AG, Gonzalez G. Fitting straight lines with replicated observations by linear regression. III weighting data. Critical Reviews in Analytical Chemistry. 2007;37(3):143–172
  80. 80. Asuero AG, Bueno JM. Fitting straight lines with replicated observations by linear regression. IV. Transforming data. Critical Reviews in Analytical Chemistry. 2011;41(1):36–69
  81. 81. Ivaska A and Wanninen E. Potentiometric titration of weak acids. Analytical Letters. 1973;6:961–967
  82. 82. Michalowski T, Asuero AG, Wybraniec S. The titration in the kjeldahl method of nitrogen determination: Base or acid as titrant. Journal of Chemical Education. 2013;90(2):191–197
  83. 83. Asuero AG. Buffer capacity of a polyprotic acid: First derivative of the buffer capacity and pKa values of single and overlapping equilibria. Critical Reviews in Analytical Chemistry. 2007;37(4):269–301
  84. 84. Asuero AG. Inflexion points of the titration curve for two step overlapping equilibria. Microchemical Journal. 1993;47:386–398
  85. 85. Harris DC. Quantitative Chemical Analysis. 8th ed. New York: W.H. Freeman and Company; 2010
  86. 86. Schwartz LM. Uncertainty of a Titration equivalence point. A graphical method using spreadsheets to predict values and detect systematic errors. Journal of Chemical Education. 1992;69(11):879–883
  87. 87. Baes CF, Mesmer RE. The Hydrolysis of Cations. Malabar, FL: Krieger; 1976. pp. 104–111

Written By

Julia Martín, Lucía Fernández Sarria and Agustín G. Asuero

Submitted: 02 December 2016 Reviewed: 27 March 2017 Published: 27 September 2017